VHT-200 Wound Treatment System

K212121 · Vaporox, Inc. · KPJ · Mar 23, 2023 · General, Plastic Surgery

Device Facts

Record IDK212121
Device NameVHT-200 Wound Treatment System
ApplicantVaporox, Inc.
Product CodeKPJ · General, Plastic Surgery
Decision DateMar 23, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of VHT 200 wound treatment system is to provide humidified oxygen to open, acute or chronic wounds as an adjunct therapy in wound management and treatment. The VHT-200 wound treatment system is intended for the following kinds of wounds: - skin ulcerations due to diabetes, venous stasis, and post-surgical infections, and gangrenous lesions - decubitus ulcers - amputations/infected stumps - skin grafts - burns - frostbite

Device Story

VHT-200 is a prescription-only topical oxygen therapy system for clinical office use. It delivers humidified oxygen to open, acute, or chronic wounds to support wound management. Unlike predicate systems requiring external oxygen sources, the VHT-200 features an integrated oxygen concentrator. The system utilizes a treatment chamber bag to enclose the wound site. Medical providers operate the device, manually adjusting oxygen flow rates and treatment frequency based on professional assessment of the patient's clinical needs. By providing a controlled environment of oxygen and moisture, the device serves as an adjunct therapy to promote healing in various skin injuries and ulcerations.

Clinical Evidence

Bench testing only. Evaluations included software validation, biocompatibility, cleaning validation, distribution testing, and compliance with IEC 60601-1-2, IEC 80601-1-2, and IEC 80601-2-69 standards.

Technological Characteristics

Topical oxygen chamber system; integrated oxygen concentrator; humidified oxygen delivery; treatment chamber bag form factor; clinical office use; software-controlled operation; compliant with IEC 60601-1-2, IEC 80601-1-2, and IEC 80601-2-69.

Indications for Use

Indicated for patients with open, acute, or chronic wounds including diabetic, venous stasis, or post-surgical ulcers, gangrenous lesions, decubitus ulcers, amputations/infected stumps, skin grafts, burns, and frostbite. Prescription use only.

Regulatory Classification

Identification

A topical oxygen chamber for extremities is a device that is intended to surround a patient's limb and apply humidified oxygen topically at a pressure slightly greater than atmospheric pressure to aid healing of chronic skin ulcers such as bedsores.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Topical Oxygen Chamber for Extremities.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 23, 2023 Vaporox, Inc. % Allyson Mullen Hyman, Phelps, & McNamara, P.C. 700 Thirteenth Street NW. Suite 1200 Washington, District of Columbia 20005 Re: K212121 Trade/Device Name: VHT-200 Wound Treatment System Regulation Number: 21 CFR 878.5650 Regulation Name: Topical Oxygen Chamber For Extremities Regulatory Class: Class II Product Code: KPJ Dated: July 6, 2021 Received: July 7, 2021 Dear Allyson Mullen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Julie A. Morabito -5 Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212121 #### Device Name VHT-200 Wound Treatment System Indications for Use (Describe) The intended use of VHT 200 wound treatment system is to provide humidified oxygen to open, acute or chronic wounds as an adjunct therapy in wound management and treatment. The VHT-200 wound treatment system is intended for the following kinds of wounds: - · skin ulcerations due to diabetes, venous stasis, and post-surgical infections, and gangrenous lesions - decubitus ulcers - · amputations/infected stumps - · skin grafts - burns - · frostbite | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary VHT-200 Wound Treatment System August 20, 2021 | Company: | Vaporox, Inc.<br>7012 S. Revere Pkwy, Suite 100<br>Centennial, CO 80112 | |--------------------|-------------------------------------------------------------------------| | Company Contact: | Alan Sage CEO, Vaporox<br>(303) 558-5145 | | Trade Name: | VHT-200 Wound Treatment System | | Common Name: | Topical oxygen chamber for extremities | | Classification: | Class II | | Regulation Number: | 21 CFR 878.5650 | | Product Code: | KPJ | | Predicate Device: | Hyper-Box Topical Wound Oxygen System (K080966 | ### Device Description: The VHT-200 Wound Treatment System is a prescription medical device. This system is designed for most topical skin injuries that can benefit from the properties of oxygen and moisture therapy treatments. All the benefits of oxygen and moisture modalities are self-contained into one system and applied without the need of switching applications. The VHT-200 is made for clinical office use. The VHT-200 wound treatment system is designed for the medical provider to add oxygen flow rate and frequency of the treatments based on experience and professional assessment of the individual patient's medical need. ### Indications for Use: VHT-200 is intended to provide humidified oxygen to open, acute and chronic wounds as an adjunct therapy in wound management and treatment. The VHT-200 System is intended for the following kinds of wounds: - . Skin ulcerations due to diabetes, venous stasis, and post-surgical infections - . Gangrenous lesions - . Decubitus ulcers - Amputations / infected stumps - Skin grafts - . Burns - Frostbite {4}------------------------------------------------ ### K212121 # Substantial Equivalence: The subject device is substantially equivalent to the predicate Hyper-Box Topical Wound Oxygen System (K080966). The subject device is similar in indications, technological characteristics, and materials to the predicate device. The subject device includes an integrated oxygen concentrator versus the predicate device's use of an external oxygen supply and uses a treatment chamber bag rather than the predicate's sleeve and boot design for the treatment environment. However, neither of these differences in technology raise different questions of safety or effectiveness and performance testing on the subject device has verified it is appropriate for its intended use. # Performance Testing: The VHT-200 has been evaluated in various conditions and determined to be safe and effective. Testing included software validation, IEC 60601-1-2, IEC 80601-1-2, IEC 80601-2-69, biocompatibility evaluation, performance testing, cleaning validation, and distribution testing. # Conclusion: The VHT-200 has equivalent indications for use, similarities in design, and equivalent performance to the predicate device. Therefore, it can be concluded that the subject device is as safe, as effective, and performs at least as safely and effectively as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...