K022028 · Vascular One, Inc. · KPJ · Dec 17, 2002 · General, Plastic Surgery
Device Facts
Record ID
K022028
Device Name
VX-400 TOPICAL HYPERBARIC OXYGEN CHAMBER
Applicant
Vascular One, Inc.
Product Code
KPJ · General, Plastic Surgery
Decision Date
Dec 17, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For Treatment of open acute or chronic wounds, e.g., Skin ulcerations due to diabetes, venous stasis, post surgical infections and gangrenous lesions Decubitus ulcers Amputations/infected stumps Skin grafts Burns Frostbite
Device Story
VX-400 Topical Hyperbaric Oxygen Chamber; portable system for limb wound healing. Inputs: 50-psi oxygen source; AC power. Components: rigid plastic chamber, control unit, hand-held controller, silicone sleeve, power supply. Operation: clinician programs therapy duration/pressure via hand controller; system delivers oxygen to limb at prescribed pressure. Output: controlled topical hyperbaric oxygen environment. Used in home or clinical settings. Clinician sets parameters; patient operates on/off switch. LCD display provides status. Benefits: facilitates wound healing through controlled oxygen delivery.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including environmental (temperature/humidity), mechanical (vibration/impact), and electrical safety testing conducted per FDA 1993 Draft Reviewer Guidance.
Technological Characteristics
Rigid plastic shell (two heat-sealed half-panels); silicone sleeve; 50-psi oxygen source; pressure sensor; flowmeter; hand-held controller with LCD display; software-driven control system; AC power supply.
Indications for Use
Indicated for patients with open acute or chronic wounds, including diabetic skin ulcerations, venous stasis, post-surgical infections, gangrenous lesions, decubitus ulcers, amputations/infected stumps, skin grafts, burns, and frostbite.
Regulatory Classification
Identification
A topical oxygen chamber for extremities is a device that is intended to surround a patient's limb and apply humidified oxygen topically at a pressure slightly greater than atmospheric pressure to aid healing of chronic skin ulcers such as bedsores.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Topical Oxygen Chamber for Extremities.” See § 878.1(e) for the availability of this guidance document.
Topox Topical Sacral Hyperbaric Oxygen Chamber, Topox, Inc. (K920948)
Topical Hyperbaric Oxygen Extremity Chamber, Advanced Hyperbaric Technologies, Inc. (Grandfathered)
Submission Summary (Full Text)
{1}------------------------------------------------
KO22028
# DEC 1 7 2002
Attachment 3
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 5l0(k) number is:________________________________________________________________________________________________________________________________________________
### Submitter's Identification: 1.
Vascular One. Inc. 8711 E. Pinnacle Peak Road Suite 306 Scottsdale, AZ 85255-3517
Contact: Mr. Edward Lisinski, President
Date Summary Prepared: June 20, 2002
### 2. Name of the Device:
VX-400 Topical Hyperbaric Oxygen Chamber
### న. Predicate Device Information:
- O2 Boot, GWR Medical, LLP, Chadds Ford, PA, K#971507 .
- Topox Topical Sacral Hyperbaric Oxygen Chamber, Topox, Inc. North Attleboro, . MA. K#920948
- The Topical Hyperbaric Oxygen Extremity Chamber, Advanced Hyberbaric ● Technologies, Inc., Farmingdale, NJ. (Grandfathered, Pre-Amendment Status).
### 4. Device Description:
The VX-400 Topical Hyperbaric Oxygen Chamber is a small, portable system that provides oxygen, at prescribed pressure for a prescribed duration, to a patient's limb. It is used at home or a clinical setting to help speed wound healing. The system consists of five (5) major components:
- 1. The Blue/Clear Plastic Chamber
- 2. The Control Unit
- 3. The Hand Held Controller
- 4. The Silicone Sleeve
- 5. Power Cube (Power Supply)
{2}------------------------------------------------
The VX-400 is a rectangular, yet rounded (10 x 24 x 16) rigid plastic shell constructed of two plastic half-panels, to which epoxy is applied and then heat-sealed together along the edges to form one chamber.
The chamber is a rigid plastic sleeve into which the patient places their limbs. There are two (2) hose barbs on the chamber, one for oxygen tubing that fills the chamber, and one for tubing to a pressure sensor. The controller consists of a plastic box with two (2) pieces of attached clear flexible tubing; these connect to the two (2) barbs on the chamber. The controller has a hose connected from the 50-psi oxygen source, a connector for a "power cube" that plugs into an AC wall outlet, and a cable that connects to the hand controller. The controller also incorporates a flowmeter. The hand-held controller can be programmed by the clinician to the prescribed time and pressure therapy, and then set to allow the patient to only turn the unit on and then off; it has a small 2 line LCD alphanumeric display and several push buttons.
### ട. Intended Use:
# For Treatment of open acute or chronic wounds, e.g.,
- Skin ulcerations due to diabetes, venous stasis, post surgical infections and . gangrenous lesions
- Decubitus ulcers .
- Amputations/infected stumps .
- Skin grafts ●
- . Burns
- Frostbite .
### 6. Comparison to Predicate Devices:
Both the VX-400 Topical Hyperbaric Oxygen and the Advanced Hyperbaric Technologies Topical Hyperbaric Oxygen Extremity Chamber are constructed of enclosed, rigid plastic to allow for pressures of up to 50 mmHg of the extremities. The VX-400 device is constructed of materials similar to the Advance Hyperbaric device and is used in an identical manner for the treatment of chronic skin and pressure sores of the extremities. The GWR O2 Boot device employs a disposable sleeve, and is flexible in construction. All of the devices are constructed of transparent materials to allow the wounds to be viewed and monitored during set-up and treatment. The VX-400 device and the predicates use the hospital wall oxygen or tank oxygen supplies. The prescribed oxygen pressures are maintained and controlled at all times. The VX-400 device is software driven. The GWR device is not.
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:
Performance testing to confirm the operations by setting several treatment combinations, and checking time and pressure (clock) was conducted in order to show that the device meets product specifications.
{3}------------------------------------------------
Testing information demonstrating safety and effectiveness of the VX-400 Topical Hyperbaric Qxygen Chamber in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.
A Contract Testing Laboratory conducted the following testing:
a. Operation at 5 +5°C and at +40°C, 90- 95% relative humidity (to assure that the device operates normally in the environment of use)
b. Storage at -20°C and at +60°C, 90-95% relative humidity (storage resistance)
c. Sinusoidal vibration
d. Leakage current and dielectric withstand of power cube (electrical safety)
e. Thermal Safety
Vascular One, Inc. conducted Impact (drop) resistance testing.
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazard. It was the contract testing laboratory and Vascular One, Inc.'s conclusion that the VX-400 device samples(s) tested met all relevant requirements of the aforementioned test.
### 8. Discussion of Clinical Tests Performed:
Not applicable
### ഗ് Conclusions:
The subject device, The Vascular One VX-400 Topical Hyperbaric Oxygen Chamber, has the same intended use and similar characteristics as the predicate devices. Moreover, bench tenting contained in our submission and non-clinical testing supplied demonstrates that there are no differences in their technological characteristics, thereby not raising any new questions of safety or effectiveness. Thus, the VX-400 Topical Hyperbaric Oxygen Chamber is substantially equivalent to the predicate devices.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services.
# Public Health Service .
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vascular One, Inc. Susan D. Goldstein-Falk c/o MDI Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K022028
Trade/Device Name: VX-400 Topical Hyperbaric Oxygen Chamber Regulation Number: 878.5650 Regulation Name: Topical oxygen chamber for extremities Regulatory Class: Class III Product Code: KPJ Dated: September 18, 2002 Received: September 23, 2002
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the
{5}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Attachment #2
Page __ of
# 510(k) Number (if known): K#022028
Device Name VX-400 Topical Hyperbaric Oxygen Chamber
Indications For Use:
For Treatment of open acute or chronic wounds, e.g.,
- Skin ulcerations due to diabetes, venous stasis, post surgical infections . and gangrenous lesions
- Decubitus ulcers t
- Amputations/infected stumps 아
- Skin grafts
- Burns
- Frostbite
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
neral, Restorative and Neurological Devices
Number K022028
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.