Hair Growth System

K200929 · Xuzhou Kernel Medical Equipment Co., Ltd. · OAP · Jun 3, 2020 · Physical Medicine

Device Facts

Record IDK200929
Device NameHair Growth System
ApplicantXuzhou Kernel Medical Equipment Co., Ltd.
Product CodeOAP · Physical Medicine
Decision DateJun 3, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000D/KN-8000E) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV.

Device Story

Hands-free, portable, noninvasive low-level laser therapy (LLLT) device; utilizes red visible light diode lasers and/or LEDs (650nm/655nm) in a helmet form factor. Input: user-positioned head within the helmet. Operation: dot matrix laser arrangement targets hair follicles; automatic pause/resume function based on head position; audible voice prompts for treatment status (models KN-8000D/E). Power: internal rechargeable Li-ion battery or external AC adapter. Used in home/clinic settings by patients. Output: red light radiation to scalp. Clinical benefit: promotion of hair growth. Healthcare provider uses device output to support hair regrowth therapy.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including electrical safety (IEC 60601-1, IEC 60825-1, IEC 60601-2-57), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-5, ISO 10993-10), and usability (IEC 62366).

Technological Characteristics

LLLT helmet; red visible light diode lasers/LEDs (650-655nm); max output <5mW per diode; Class 3R laser classification. Powered by 7.2V Li-ion battery or 9V DC adapter. Patient-contacting materials biocompatible per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 62471, IEC 60601-2-57.

Indications for Use

Indicated for females with androgenic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood Hamilton IIa-V), Fitzpatrick Skin Phototypes I-IV.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 3, 2020 Xuzhou Kernel Medical Equipment Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China Re: K200929 Trade/Device Name: Hair Growth System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: March 15, 2020 Received: April 7, 2020 Dear Jet Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K200929 Device Name Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000D/KN-8000E) #### Indications for Use (Describe) The Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000E) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Date of the summary prepared: May 30, 2020 # 510(k) Summary This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a traditional 510(K) submission with no previous application. #### 1. Submitter's Information #### Sponsor - � Company Name: Xuzhou Kernel Medical Equipment Co., Ltd. - Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, � China - � Phone: +86 1831 685 8036 - Fax: +86 0516-87732208 13776587162 � - Contact Person (including title): Wang Jing (Management Representative) � - E-mail: wikernel@126.com � #### Application Correspondent: - Company Name: Guangzhou KEDA Biological Tech Co., Ltd. � - Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China � - � Contact Person: Jet Li - Tel: +86-18588874857 � - Email: med-jl@foxmail.com � #### Subject Device Information 2. - � Trade Name: Hair Growth System - Model: KN-8000B/KN-8000C/KN-8000D/KN-8000E � | Data<br>Model | LD Quantity | LED Quantity | |---------------|-------------|--------------| | KN-8000B | 204 | -- | | KN-8000C | 80 | 30 | | KN-8000D | 21 | 30 | | KN-8000E | 81 | -- | - Common Name: Lamp, non-heating, for promotion of hair growth � - Classification name: Infrared lamp � - � Review Panel: General & Plastic Surgery {4}------------------------------------------------ - Product Code: OAP (Laser, comb, hair) � - � Requlation Class: 2 - Regulation Number: 21 CFR 890.5500 � # 3. Predicate Device Information - � Predicate Devices for KN-8000B and KN-8000E | | Predicate I | Predicate II | |-------------------|----------------------|-----------------------------------------------------------------------------------------------------------| | Sponsor | DermaScalp LLC | Capillus LLC | | Device Name | DermaScalp Laser Cap | Capillus 82, Capillus 202, Capillus 272 Pro, 272 Office Pro, Capillus 302, Capillus 312, and Capillus 352 | | 510(k) Number | K173846 | K163170 | | Product Code | OAP | OAP | | Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | | Regulation Class | 2 | 2 | #### Predicate Devices for KN-8000C and KN-8000D � | | Predicate I | Predicate II | |-------------------|------------------------------|-----------------------------| | Sponsor | Freedom Laser Therapy, Inc. | Freedom Laser Therapy, Inc. | | Device Name | iRestore Hair Grow th System | iRestore Professional 282 | | 510(k) Number | K151662 | K183417 | | Product Code | OAP | OAP | | Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | | Regulation Class | 2 | 2 | #### 2. Device Description The Hair Growth System (Model: KN-8000B/KN-8000D/KN-8000E) is hands-free, portable, noninvasive, low-level laser device, which consists of red visible light diode lasers and/or LEDs, to produce red light operating at 650nm wavelength (maximum output power of each is 5mW). The laser sources are arranged in a dot matrix arrangement in the inner of the helmet, which could take into account of every hair follicle and promote rapid hair growth. The device will automatically pause therapy if the user's head is moved outside of the zone of radiation and will resume therapy when the correct head position is reestablished. The device could be powered by internal rechargeable li-on battery (Rated 7.2Vd.c. 2100mAh), and it can also be supplied by specified external adapter with rated Input 100-240Vac. 50/60Hz and rated Output 9Vdc 3A. For KN-8000D and KN-8000E, the device emits an audible voice prompt when the treatment is going to start/pause/end, and the device is going to shut down. {5}------------------------------------------------ # 3. Intended Use / Indications for Use The Hair Growth System (Model: KN-8000B/KN-8000C/KN-8000E) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV. # 4. Test Summary The Hair Growth System has been evaluated for its safety and performance by lab bench testing as following: - � Electrical safety and performance test according to IEC 60601-1, IEC 60825-1, IEC 62471 and IEC 60601-2-57 standard - � Electromagnetic compatibility test according to IEC 60601-1-2 standard - � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standard - � Usability test according to IEC 62366 - � Software verification and validation test according to the requirements of the FDA "Guidance for Premarket Submissions and for Software Contained in Medical Devices" #### 5. Comparison to predicate device and conclusion The technological characteristics, features, specifications, materials, and intended use of Hair Growth System, model: KN-8000B/KN-8000D/KN-8000E is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. {6}------------------------------------------------ | Substantial Equivalence Comparison Table for<br>KN-8000B/KN-8000E | | | | | | |-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--| | Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | | | 510(k) Number | TBD | K173846 | K163170 | -- | | | Device Name | Hair Grow th System<br>KN-8000B/ KN-8000E | DermaScalp Laser Cap | Capillus 82, Capillus 202, Capillus<br>272 Pro, 272 Office Pro, Capillus<br>302, Capillus 312, and Capillus 352 | -- | | | Product Code | OAP | OAP | OAP | SE | | | Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 | SE | | | Regulation Class | 2 | 2 | 2 | SE | | | LLLT Device Type | LLLT | LLLT | LLLT | SE | | | Prescription | OTC | OTC | OTC | SE | | | Intended Use | The Hair Grow th System (Model: KN-<br>8000B/KN-8000C/KN-8000D/KN-8000E) is<br>intended for the promotion of hair grow th in<br>females with androgenic alopecia w ho<br>have Ludwig-Savin Classifications I-II, and<br>in males with androgenetic alopecia w ho<br>have Norwood Hamilton Classifications Ila-<br>V; and both genders having Fitzpatrick<br>Classification of Skin Phototypes I-IV. | The family of DermaScalp Laser Cap<br>devices are indicated to treat<br>Androgenetic Alpopecia and promote<br>hair grow th in Males w ho have<br>Norw ood Hamilton Classifications of<br>Ila to V patterns of hair loss and to<br>treat Androgenetic Alopecia and<br>promote hair grow th in Females w ho<br>have Ludwig (Savin) Scale I-1 to I-4,<br>Il-1, Il-2, or frontal; both with<br>Fitzpatrick Skin Types I to IV. | The Capillus laser domes 82, 202, 272<br>Pro, 272 OfficePro, 302, 312, and 352,<br>are intended to treat Androgenetic<br>Alopecia and promote hair grow th in<br>males w ho have Norw ood -Hamilton<br>Classifications of Ila to V patterns of<br>hair loss and to treat Androgenetic<br>Alopecia and promote hair grow th in<br>females who have Ludwig (Savin) Scale<br>I-1 to I-4, II-1, II-2, or frontal; both with<br>Fitzpatrick Skin Types I to IV. | SE<br>Minor<br>difference | | | Waveform | Visible red laser | Visible red laser | Visible red laser | SE | | | Wavelength | 655nm ±5nm | 650nm | 650nm | SE<br>Note 1 | | | Amounts of laser diode | KN-8000B: 204<br>KN-8000E: 81 | 50, 80, 120, 148, 180, 202, 224, 272 | 82, 202, 272, 302, 312, 352 | SE<br>Note 2 | | | Laser radiation output | <5mW | <5mW | ≤5mW | SE<br>Note 1 | | | Classification according to<br>IEC60825-1 | Class 3R | Class 3R | Class 3R | SE | | | Substantial Equivalence Comparison Table for<br>KN-8000B/KN-8000E | | | | | | | Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | | | Treatment | Each Treatment: 25-35min,<br>Total Treatment: every other day | 17weeks, every other day (indefinite) | 17weeks, every other day (indefinite) | SE<br>Note 2 | | | Total laser irradiance | KN-8000B: ≤1,020mW<br>KN-8000E: ≤405mW | Diode 50: <250mW,<br>Diode 80: <400mW,<br>Diode 120: < 600mW,<br>Diode 148: <740mW,<br>Diode 180: <900mW,<br>Diode 202: <1,010mW,<br>Diode 224: <1,120mW,<br>Diode 272: <1,360mW | Diode 82: <410mW,<br>Diode 202: <1,010mW,<br>Diode 272: <1,360mW,<br>Diode 302: <1,510mW,<br>Diode 312: <1,560mW,<br>Diode 352: <1,760mW | SE<br>Note 2 | | | Appearance Design | Helmet cap | Cap | Cap | SE<br>Minor<br>difference | | | Safety and Performance<br>Feature | Complied with IEC 60601-1, IEC 60601-1-<br>2, IEC 60825-1 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1, IEC 60601-<br>1-11 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1 | SE<br>Note 3 | | | Biocompatibility | All patient contacting materials are<br>complied with ISO 10993-5, ISO 10993-10 | All patient contacting materials are<br>complied with ISO 10993-1, ISO<br>10993-5 | All patient contacting materials are<br>complied with ISO 10993-5, ISO 10993-<br>10 | SE<br>Note 3 | | {7}------------------------------------------------ | Substantial Equivalence Comparison Table for<br>KN-8000C/KN-8000D | | | | | |-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | | 510(k) Number | TBD | K151662 | K183417 | -- | | Device Name | Hair Grow th System | iRestore Hair Grow th System | iRestore Professional 282 | -- | | Substantial Equivalence Comparison Table for<br>KN-8000C/KN-8000D | | | | | | Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | | | KN-8000C/ KN-8000D | | | | | Product Code | OAP | OAP | OAP | SE | | Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 | SE | | Regulation Class | 2 | 2 | 2 | SE | | LLLT Device Type | LLLT | LLLT | LLLT | SE | | Prescription | OTC | OTC | OTC | SE | | Intended Use | The Hair Growth System (Model: KN-<br>8000B/KN-8000C/KN-8000D/KN-8000E) is<br>intended for the promotion of hair growth in<br>females with androgenic alopecia who have<br>Ludwig-Savin Classifications I-II, and in<br>males with androgenetic alopecia who have<br>Norwood Hamilton Classifications IIa-V; and<br>both genders having Fitzpatrick Classification<br>of Skin Phototypes I-IV. | The iRestore Hair Growth System is<br>indicated to promote hair growth in<br>females with androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I - II, males who<br>have Norwood-Hamilton<br>Classifications of IIa - V and for both,<br>Fitzpatrick Classification of Skin<br>Phototypes I to IV. | The iRestore Professional 282 is<br>indicated to promote hair growth in<br>females with androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I-II, males who have<br>Norwood-Hamilton Classifications of<br>IIa-V and for both, Fitzpatrick<br>Classification of Skin Phototypes I to<br>IV. | SE | | Waveform | Visible red laser and light | Visible red laser and light | Visible red laser and light | SE | | Wavelength | LD: 655nm±5nm<br>LED: 655nm±20nm | 655nm | 650nm±10nm | SE<br>Note 1 | | Amounts of LED | KN-8000C: 30<br>KN-8000D: 30 | 30 | 200 | SE<br>Note 2 | | Amounts of laser diode | KN-8000C: 80<br>KN-8000D: 21 | 21 | 82 | SE<br>Note 2 | | Classification according to<br>IEC60825-1 | Class 3R | Class 3R | Class 3R | SE | | Treatment time | Each Treatment: 25-35min<br>Total Treatment: every other day | Each Treatment: 25min<br>Total Treatment: every other day, on<br>non-consecutive days, for 16weeks | Each Treatment: 25min<br>Total Treatment: every other day, on<br>non-consecutive days, for 16weeks | SE<br>Note 2 | | Appearance Design | Helmet cap | Helmet | Helmet | SE | | Substantial Equivalence Comparison Table for<br>KN-8000C/KN-8000D | | | | | | Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Verdict | | Safety and Performance<br>Feature | Complied with IEC 60601-1, IEC 60601-1-2,<br>IEC 60825-1, IEC 62471, IEC 60601-2-57 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1, IEC 60601-<br>1-11 | Complied with IEC 60601-1, IEC<br>60601-1-2, IEC 60825-1, IEC 60601-<br>1-11 | SE<br>Note 3 | | Biocompatibility | All patient contacting materials are complied<br>with ISO 10993-5, ISO 10993-10 | All patient contacting materials are<br>complied with ISO 10993-5, ISO<br>10993-10 | All patient contacting materials are<br>complied with ISO 10993-5, ISO<br>10993-10 | SE | {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ # Comparison in Detail(s): ### Note 1: Although the Wavelength and Laser radiation output of subject device have a little difference to the predicate devices, these parameters are similar to the predicate devices, and these minor difference of wavelength will not affect the main function. So the differences will not raise any safety or effectiveness issue. ### Note 2: Although the Amounts of LED and/or laser diode/Treatment Time/Total Irradiance of subject device and predicate devices are a little difference, the energy and power parameters' range of subject device can be covered by predicate device's several models' range; they are very similar. So these parameters' differences will not raise any safety or effectiveness issue. ### Note 3: Although the accepted FDA recognized standards of subject device are a little different from the predicate devices, they all comply with IEC 60101-1/IEC 60601-1-2/IEC 60825, and the subject device also complies with IEC 62471 requirements, so these differences will not affect the critical functions or the normal use. #### Summary for clinical test 6. Clinical performance is not deemed necessary. # 7. Conclusion The subject device Hair Growth System (KN-8000B/KN-8000D/KN-8000E) has all features of the predicate devices for intended use. Thus, the subject device is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...