The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
Device Story
Capillus laser domes (82, 202, 272 Pro, 272 OfficePro, 302, 312, 352) are Low-Level Laser Therapy (LLLT) devices for home or salon use. Devices consist of helmet-style domes containing 82 to 352 red diode lasers (650 nm). Operated by consumers; portable models use rechargeable Li-Ion batteries; OfficePro model uses direct DC power. Devices emit audible tones (2 short beeps start, 1 long beep end) to signal therapy session status. Intended to treat Androgenetic Alopecia and promote hair growth. Users wear the device passively; hands-free operation. No physician intervention required for OTC use. Benefit is non-invasive hair growth stimulation.
Clinical Evidence
Bench testing only. Compliance with IEC-60825-2007-03 (Class 3R laser) and IEC 61959 (charger) verified. Usability study (n=55) conducted to support OTC clearance: 94.6% pass rate for self-selection and 90.9% pass rate for user instruction questionnaire, exceeding the 80% success threshold required by FDA.
Technological Characteristics
Helmet-style LLLT device; 82-352 red diode lasers (650nm); Class 3R laser (AEL 5mW max). Materials: outer helmet, protective inner. Power: rechargeable Li-Ion battery or DC charger. Connectivity: none. Standards: IEC-60825-2007-03, IEC 61959. Audible tone signaling for therapy start/end.
Indications for Use
Indicated for males with Norwood-Hamilton Classifications IIa to V and females with Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal patterns of Androgenetic Alopecia; Fitzpatrick Skin Types I to IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
K143199 — Capillus 272 Pro · Capillus, LLC · Jan 29, 2015
K151516 — Capillus82 · Capillus, LLC · Aug 21, 2015
K150613 — LaserCap LC Pro and LC Elite · Transdermal Cap, Inc. · Jul 22, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 31, 2017
Capillus, LLC Patricia Schnoor Director, Ouality & Regulatory Affairs 1715 NW 82nd Avenue Miami. Florida 33126
Re: K163170
Trade/Device Name: Capillus 82, Capillus 202, Capillus 272 Pro, 272 Office Pro, Capillus 302, Capillus 312, and Capillus 352 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: November 4, 2016 Received: November 14, 2016
Dear Patricia Schnoor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/1/Picture/7 description: The image shows the signature of Jennifer R. Stevenson. The text "Sincerely" is written above the name. The name is written in a larger font size than the word "Sincerely". The letters "FDA" are faintly visible in the background.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K163170
Device Name
Devillus Laser Domes: Capillus202, Capillus272 Pro, Capillus272 Office Pro, Capillus302, Capillus312, and Capillus352
#### Indications for Use (Describe)
Indications for Use (Describe)
The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## Capillus, LLC
| Date Prepared: | November 04, 2016 | |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------|--|
| Submit t er's Cont act Inform at i on: | | |
| Name: | Patricia Schnoor | |
| Address: | Capillus LLC<br>1715 NW 202nd Avenue<br>Miami, FL 33126 | |
| Establishment<br>Registration# | 3010123655 | |
| Telephone:<br>Facsimile: | (786) 888-1874<br>(305) 418-7581 | |
| Name of Device and Name / Address of Sponsor: | | |
| Trade Name: | Capillus82, Capillus202, Capillus272 Pro, 272 OfficePro,<br>Capillus302, Capillus312, and Capillus352 | |
| Common or Usual Name: | Lamp, non-heating, for promotion of hair growth | |
| Classification Name: | Infrared lamp per 21 CFR 890.5500 | |
| Classification Code: | OAP (Laser, Comb, Hair) | |
| Sponsor Contact Information: | Patricia Schnoor<br>Capillus LLC<br>1715 NW 82nd Avenue<br>Miami, FL 33126 | |
| Telephone: | (786) 888-1874 | |
## Predicate Devices:
| Device Trade Name | Manufacturer | 510(k) |
|-------------------|-------------------------|----------------|
| Hairmax Lasercomb | Lexington International | 103368, 142573 |
| iGrow II | Apira | 141567 |
## Intended Use / Indications for Use:
The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood -Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
{4}------------------------------------------------
### Technological Characteristics
The Capillus272 Pro and Capillus272 OfficePro both consist of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner, and configured for portable use with rechargeable battery and adapter (Capillus272 Pro) and salon/spa use (Capillus272 OfficePro). The Capillus82, Capillus202, Capillus302, Capillus312, and Capillus352 are exactly the same as the Capillus272 Pro with the exception of the number of diodes which are respective of model (e.g. 82 diodes in Capillus82).
As stated in prior submissions, the devices emit an audible tone at the beginning and end of a therapy session, indicating that therapy has begun (2 short beeps) or ended (one long beep). The portable systems are powered by rechargeable Li-Ion battery cells assembled into a proprietary battery pack. Both the battery are fully compliant to recognized, international standards. The OfficePro is directly powered by the DC charger, which is fully compliant to recognized, international standards.
### Performance Data:
Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criteria were met by the devices. The Capillus202, Capillus272 Pro, 272 OfficePro, Capillus302, Capillus312, and Capillus352 conform to the standard IEC-60825-2007-03. This IEC standard is a recognized and accepted standard by the FDA. The guidance document for this accepted standard is found in the Federal register, July 26, 2001 (volume 66, Number 144) [page 39049 39050]. This report validates for the Capillus82, Capillus202, Capillus272 Pro, 272 OfficePro, Capillus312, and Capillus352 the laser class of 3R which establishes the AEL (accessible emission limits) as 5 milliWatts maximum. The charger conforms to IEC 61959.
The data presented in this submission is restricted to self-selection and usability to include comprehension of user instructions, warnings, and precautions. The results of a presubmission request (Q160385) verified the objectives that must be fulfilled for the Capillus family of devices to be classified for over-the-counter (OTC) sale. One of the responses for the Q-submission indicated that a minimum of an 80% success rate must be achieved for the intended use of OTC to be granted. The testing was administered to 55 adult subjects of any age (>18), educational background, race, disease status present or pre-education about the testing to be performed. This was designed to provide for the broadest test criteria without any bias being imposed, thereby assessing the real-world capability of the average adult, "retail customer" who might wish to purchase a device without the benefit of a physician or other qualified health care provide counsel as is the case with prescription devices. The study was conducted in two (2) parts specifically intended to demonstrate that the device is appropriate for consumer use by lay persons (see data and report in Section 13). The first part addressed self-selection (box labeling), while the second part addressed the content of the user manual including warnings and precautions. The Capillus usability test demonstrated a pass rate of 94.6% (self-selection) and 90.9% pass rate for user instruction questionnaire, satisfying the FDA's requirements for clearance (sale) as Over-the-Counter Intended Use.
Just as for Capillus272 Pro (K143199) and reference devices, there are no reported adverse events for this technology.
{5}------------------------------------------------
### Substantial Equivalence
The requirements set for substantial equivalence in the traditional sense, do not have applicability in this process as the purpose of this submission is expansion of indications for use to include OTC sale capability. However, The Hairmax Lasercomb and iGrow Hair Growth System (predicate devices) have been cleared with the Intended Use of Over-the-Counter, and the Capillus devices can demonstrate the minimum pass-rate for usability by lay persons. The Capillus devices utilize the same technology and share the same risk profile as the predicates. The Capillus family of devices referenced in this application are the same technology used by all LLLT devices cleared under device code OAP. The Capillus272 Pro, Capillus272 OfficePro, Capillus202, Capillus202, Capillus302, Capillus312 and Capillus352 are as safe and effective as the predicate devices, as well as other reference devices in its class.
| Capillus Devices | iGrow | OAP (General) |
|---------------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------------------|
| K143199; K150578; K151516; K153618 | K122248; K140931 | N/A |
| LLLT Device Type | LLLT Device Type | LLLT Device Type |
| Prescription Use | Prescription Use | Prescription and OTC |
| Intended Use - Androgenic Alopecia | Intended Use - Androgenic Alopecia | Intended Use -Androgenic Alopecia |
| Contain Laser Diodes- Class 3R | Contain Laser Diodes- Class 3R and LEDs | Contain Laser Diodes- Class 3R; LEDs, or a<br>combination of both |
| Helmet/Cap Design | Helmet/Cap Design | Helmet/Cap, comb, or flat Design |
| 650nm | 650nm | 650nm |
| Marketing clearance for Females | Marketing clearance for Males and Females | Marketing clearance for both genders |
| Passive Use-Hands Free | Passive Use-Hands Free | Passive Use-Hands-Free, or comb |
| 272 Laser Diodes | 21 Laser Diodes; 30 LEDs | Various configurations of lasers, LEDs or both |
| OAP Classification | OAP Classification | OAP Classification |
| Classification Name -Infrared Lamp | Classification Name -Infrared Lamp | Classification Name - Infrared Lamp |
| Common Usage Name - Lamp, Non-Heating | Common Usage Name - Lamp, Non-Heating | Common Usage Name - Lamp, Non- Heating |
| General & Plastic Surgery Committee | General & Plastic Surgery Committee | General & Plastic Surgery Committee |
| Fitzpatrick Skin Phototypes - I- IV | Fitzpatrick Skin Phototypes - I- IV | Fitzpatrick Skin Phototypes - I -IV |
| Ludwig-Savin I-II (females)<br>Norwood Hamilton IIA-V (males) | Ludwig-Savin I-II (females)<br>Norwood Hamilton IIA-V (males) | Ludwig-Savin I-II (females); Norwood<br>Hamilton IIA-V (males); or both |
| Efficacy Rates - High Compared to Placebo | Efficacy Rates -34-41% | Various; trial results are all significant<br>increase with respect to placebo. |
| Treatment- 17 weeks, every other day<br>(indefinite) | Treatment- 17 weeks, every other day<br>(indefinite) | Treatment - ~16 weeks, every other day<br>(indefinite) |
| Device Class II | Device Class II | Device Class II |
For these reasons, the Capillus family of laser domes satisfy the FDA's substantial equivalence requirements with respect to risk profile, intended use, and technological and design characteristics. Additionally, no new safety or efficacy concerns are raised due to the minor differences present between devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.