K153618 · Capillus, LLC · OAP · Jan 19, 2016 · Physical Medicine
Device Facts
Record ID
K153618
Device Name
Capillus202
Applicant
Capillus, LLC
Product Code
OAP · Physical Medicine
Decision Date
Jan 19, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K153618 · Jan 19, 2016
Capillus202
Capillus, LLC
Postmarket vigilance/surveillance data for predicate device (Capillus272 Pro)
The sponsor utilized postmarket surveillance data from the predicate device to support the safety profile of the modified device (Capillus202), noting no adverse events were reported in the real-world setting.
Postmarket Vigilance for Capillus272 Pro; Postmarket surveillance
Users of the predicate Capillus272 Pro device
Not applicable for this study
Adverse events
Indications for Use
The Capillus202 is indicated to promote hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications of I-II, and with Fitzpatrick Skin Phototypes I-IV.
Device Story
Capillus202 is a wearable, helmet-style low-level laser therapy (LLLT) device designed for home use. It utilizes 202 red visible light diode lasers (650 nm) to deliver radiant energy to the scalp. The device is powered by a rechargeable Nickel Metal Hydride battery pack. Operation is hands-free; a safety interlock automatically pauses therapy if the device is moved out of the radiation zone and resumes when repositioned. Audible tones signal the start (2 beeps) and end (one long beep) of a session. The device is intended for use for a maximum of 30 minutes every other day. By delivering light energy to the scalp, the device aims to promote hair growth in patients with androgenic alopecia. It serves as a lower-cost alternative to the predicate Capillus272 Pro, with a reduced total energy dose due to fewer diodes.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed that the device meets design specifications, including laser output power, safety interlock operation, and battery performance. Compliance with IEC 60825-1:2007 (laser safety) and IEC 61959 (charger safety) was verified. No new clinical data was provided; efficacy is supported by the established performance of the predicate Capillus272 Pro and reference devices.
Technological Characteristics
Helmet-style LLLT device; 202 red diode lasers (650 nm); Class 3R laser system (≤5mW per diode); rechargeable NiMH battery; safety interlock for head positioning; audible tone feedback; ANSI/AAMI/ISO 14971:2007/(R)2010 risk management; IEC 60825-1:2007 and IEC 61959 compliant.
Indications for Use
Indicated for females with androgenic alopecia, Ludwig-Savin Classifications I-II, and Fitzpatrick Skin Phototypes I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement. The profiles are arranged in a way that they appear connected. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2016
Capillus LLC Ms. Patricia Schnoor Manager, Quality & Regulatory Affairs 1715 Northwest 82nd Avenue Miami, Florida 33126
Re: K153618
Trade/Device Name: Capillus202 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: December 16, 2015 Received: December 18, 2015
Dear Schnoor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K153618
Device Name Capillus202
### Indications for Use (Describe)
The Capillus202 is indicated to promote hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications of I-II, and with Fitzpatrick Skin Phototypes I-IV.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Special 510(k) Summary
### Capillus, LLC
### Date Revised:
### January 13, 2016
### Submitter's Contact Information:
| Name: | Patricia Schnoor |
|--------------|---------------------------------------------------------|
| Address: | Capillus LLC<br>1715 NW 202nd Avenue<br>Miami, FL 33126 |
| Estab't Reg# | 3010123655 |
| Telephone: | (786) 888-1874 |
| Facsimile: | (305) 418-7581 |
## Name of Device and Name / Address of Sponsor:
| Trade Name: | Capillus202 |
|------------------------------|----------------------------------------------------------------------------|
| Common or Usual Name: | Lamp, non-heating, for promotion of hair growth |
| Classification Name: | Infrared lamp per 21 CFR 890.5500 |
| Classification Code: | OAP (Laser, Comb, Hair) |
| Sponsor Contact Information: | Patricia Schnoor<br>Capillus LLC<br>1715 NW 82nd Avenue<br>Miami, FL 33126 |
| Telephone: | (786) 888-1874 |
### Predicate Devices:
| Device Trade Name | Manufacturer | 510(k) |
|-------------------|----------------------|--------|
| Capillus272 Pro | Capillus LLC | 143199 |
| LaserCap LCPro | Transdermal Cap, Inc | 150613 |
### Reference Devices:
| Device Trade Name | Manufacturer |
|-------------------|-------------------------|
| Hairmax Lasercomb | Lexington International |
### Intended Use / Indications for Use:
The Capillus202 is indicated to promote hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications of I-II, and with Fitzpatrick Skin Phototypes I-IV.
{4}------------------------------------------------
### Technological Characteristics
The Capillus202 is exactly the same as the Capillus272 Pro in every way except the following:
#### The Capillus202 has fewer diodes - 202 compared to 272 for Capillus272 Pro 1.
The Capillus272 Pro consists of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner liner, and configured for portable use with rechargeable battery and adapter. The Capillus202 consists of 202 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner liner, and configured for portable use with rechargeable battery and adapter.
Just as for the previous version, the helmet system will automatically pause therapy if the subject's head is moved outside of the zone of radiation and will resume therapy when the correct head position is re-established. This is achieved by a safety interlock. At the beginning and of a therapy session, audible tones are delivered to the user, indicating that therapy has begun (2 beeps) or ended (one long beep). The system is powered by rechargeable Nickel Metal Hydride battery cells assembled into a proprietary battery pack. Both the battery pack and charger are fully compliant to recognized, international standards.
### Design Control Activities:
The Capillus design and development team followed ANSI/AAMI/ISO 14971:2007/(R)2010 Risk Management: Medical devices - Application of risk management to medical devices. Significant changes included assessing the effect a reduction of lasers would have on the end product, both for safety and for efficacy. All residual risks are found acceptable and the risk management file and summary are available for review. Based on the Capillus272 Pro, the verification and validation activities remain the same (e.g. output of each laser, operation of safety interface, and output of power pack), and were performed and assessed by designated personnel qualified to perform such activities. All methods, tests, and acceptance criteria are stipulated on the verification reports. All documents are available for review.
Safety ratings remained the same; there was no change to the individual lasers (≤ SmW); however, the "dose" (total delivered energy/cm2) is reduced number of lasers. This may affect efficacy; this observation is made based on the performance of other devices similar to Capillus that are currently cleared under device code OAP and utilize much fewer than 272 lasers (The closest equivalent in light sources is Transdermal with a total of 224 lasers).
Use instructions do not change. Indications for use and the dose schedule remain exactly the same as for Capillus272 Pro and as for other devices in the same category (product code OAP); which is for maximum 30 minutes, every other day.
The output of the laser diodes was verified to remain the same (each diode) as for the previous model. Design control activities were followed per 21 CFR 820.30:
{5}------------------------------------------------
- a) General: The Capillus LLC Quality Management System is compliant to the requirement of the quality system regulation and specifically to design controls as stipulated by 21 CFR 820.30.
- b) Design and development planning: During project planning activities, the risks were identified and assessed with respect to the proposed design change. The proposed design change was minimal and affected only the number of diodes in the PCB array. The power pack remains exactly the same. Review of the risk assessment from previous model revealed no significant changes or risks are identified for the new model Capillus202. Vigilance activities for the Capillus272 Pro indicated that there were no significant problems identified (no adverse events reported).
- c) Design input: Design requirements for the Capillus272 Pro are transferred to the Capillus202 and an input/output matrix completed as for the original project.
- d) Design output: Design requirements for the Capillus272 Pro are transferred to the Capillus202 and an input/output matrix completed as for the original project. Design outputs are verified per performance tests.
- e) Design review: Appropriate design reviews were conducted; in this case (due to minimal change), a technical evaluation, and a verification of inputs and outputs (physical values) were completed.
- Design verification: All verification activities to assess the impact of the change (reduction of 0 number of laser diodes) were completed. Laser diode output remains the same, but total dosage is reduced. Since the cap is intended to be used indefinitely (if treatment is stopped, results are lost see instructions for use), this will likely affect only the time required to witness the same result as achieved with the Capillus272 Pro. The Capillus202 is offered as a lower-cost alternative.
- g) Design validation: Feedback from hair restoration specialists and consumers indicate that a lower-cost alternative is desired. The slower efficacy is accepted, with the understanding that treatment is intended to be indefinite.
- h) Design transfer: The Capillus family of products are manufactured on-site in the factory in Miami. Design transfer was minimal; training requirement to the new PCB was minimal, and all other components (including overall build) remained the same.
- Design changes: The design of all models follows our established design control, change control, i) document control procedures.
- Design history: The design history file for the Capillus202 is available for review. i)
{6}------------------------------------------------
### Performance Data:
All verification and validation activities were performed by designated individuals. The results (available for review) demonstrate that the predetermined acceptance criteria were met.
Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criteria were met by the device.
The Capillus202 conforms to the standard IEC-602025-2007-03; This IEC standard is a recognized and accepted standard by the FDA. The guidance document for this accepted standard is found in the Federal register, July 26, 2001 (volume 66, Number 144) [page 39049 39050]. This report validates for the Capillus202 the laser class of 3R which establishes the AEL (accessible emission limits) as 5 milliWatts maximum. The charger conforms to IEC 61959.
The performance data available for review demonstrates that the Capillus202 has exactly the same laser wavelength, output power (per diode), output beam, energy type, laser field, and treatment area as the original Capillus272 Pro cleared under K143199. Total "dose" - delivered energy over time - (J/cm2) is logically reduced. Just as for Capillus272 Pro (K143199) and reference devices, there are no reported adverse events for this technology.
{7}------------------------------------------------
### Substantial Equivalence
The Capillus202 is the same technology used by Capillus272 Pro (K143199) and other reference devices; specifically. Transdermal Cap (K150613). The Capillus202 is as safe and (commensurate to number of diodes) effective as the Capillus272 Pro, as well as other reference devices in its class, such as the Hairmax Lasercomb.
| Capillus272 Pro | Capillus202 | LaserCap LCPro |
|------------------------------------------------------|------------------------------------------------------|------------------------------------------------------|
| K143199 | Candidate | K150613 |
| LLLT Device Type | LLLT Device Type | LLLT Device Type |
| Prescription Use | Prescription Use | Prescription Application |
| Intended Use - Androgenic Alopecia | Intended Use - Androgenic Alopecia | Intended Use -Androgenic Alopecia |
| Contain Laser Diodes- Class 3R | Contain Laser Diodes- Class 3R | Contain Laser Diodes-Class3R |
| Helmet/Cap Design | Helmet/Cap Design | Helmet/Cap Design |
| 650nm | 650nm | 650nm |
| Marketing clearance for Females | Marketing clearance for Females | Marketing clearance for both genders |
| Passive Use-Hands Free | Passive Use-Hands Free | Passive Use-Hands-Free |
| 272 Laser Diodes | 202 Laser Diodes | 224 Laser Diodes |
| OAP Classification | OAP Classification | OAP Classification |
| Classification Name -Infrared Lamp | Classification Name -Infrared Lamp | Classification Name - Infrared Lamp |
| Common Usage Name - Lamp. Non-Heating | Common Usage Name - Lamp, Non-Heating | Common Usage Name - Lamp, Non- Heating |
| General & Plastic Surgery Committee | General & Plastic Surgery Committee | General & Plastic Surgery Committee |
| Fitzpatrick Skin Phototypes - I-IV | Fitzpatrick Skin Phototypes - I-IV | Fitzpatrick Skin Phototypes - I-IV |
| Ludwig-Savin 1-11 (females) | Ludwig-Savin I-II (females) | Ludwig-Savin I-II (females) |
| Efficacy Rates - High Compared to Placebo | Efficacy Rates - High Compared to Placebo | NA |
| Treatment- 17 weeks, every other day<br>(indefinite) | Treatment- 17 weeks, every other day<br>(indefinite) | Treatment - 16 weeks, every otherday<br>(indefinite) |
| Device Class II | Device Class II | Device Class II |
Capillus LLC believes that with the exception of the configuration of the optical elements, the Capillus202 is the same device in form, function, safety, and efficacy as the predicate device(s). The Hairmax Lasercomb, offered as a reference, is proof of the functionality and acceptability of devices with fewer laser diodes cleared by the FDA in the category of OAP, both technically and clinically. Capillus LLC believes that the difference in the physical appearance, number of diodes, or in the method of delivering the radiant energy of the systems is of no consequence and does not affect the therapeutic value or the safety profile.
All compliant LLLT systems which use red light diode lasers are classified as class 3R laser systems by the IEC standard for allowable emission levels, which is a recognized standard by the FDA as well, and the adverse event profile is the same.
For these reasons, the Capillus202 satisfies the FDA's requirements (for device modification notification) with respect to intended use, and technological and design characteristics. Additionally, no new safety or efficacy concerns are
{8}------------------------------------------------
raised due to the minor difference present between devices.
Signed:
Patricia Schnoor
01.13.2016
Quality, Safety & Compliance Manager
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.