Radiant Renewal Skincare Wand (HD-15, HD-15A)

K232863 · Shenzhen Kaiyan Medical Equipment Co., Ltd. · OHS · Dec 13, 2023 · General, Plastic Surgery

Device Facts

Record IDK232863
Device NameRadiant Renewal Skincare Wand (HD-15, HD-15A)
ApplicantShenzhen Kaiyan Medical Equipment Co., Ltd.
Product CodeOHS · General, Plastic Surgery
Decision DateDec 13, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Radiant Renewal Skincare Wand (Model: HD-15) is intended for treating wrinkles on the face and is indicated for over-the-counter use. The Radiant Renewal Skincare Wand (Model: HD-15A) is intended for facial and neck stimulation and is indicated for over-the-counter cosmetic use, the red light is intended for treating wrinkles on the face and is indicated for over-the-counter use.

Device Story

Hand-held, battery-powered skincare wand; models HD-15 and HD-15A. Inputs: user-selected modes via single button. Operation: delivers transcutaneous electrical currents for stimulation and 630nm red light for wrinkle treatment. Output: electrical stimulation and red light emission. Used in home environment by consumers. Device features 12-minute automatic shutdown; 3-minute treatment intervals with vibration feedback. Healthcare providers do not operate; output is cosmetic. Benefits: facial/neck stimulation and wrinkle reduction. Device includes charging cable and carrying case.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, 60601-2-10, 60601-2-57), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), battery safety (IEC 62133-2), biocompatibility (ISO 10993-5, -10, -23), software validation, and usability testing (IEC 62366-1, 60601-1-6).

Technological Characteristics

Materials: PC and Aluminum. Energy: 3.7V 500mAh Lithium-Ion rechargeable battery. Output: Pulsed biphasic electrical current and 630nm red light LED. Form factor: Hand-held. Connectivity: None. Sterilization: Not sterile. Software: Microprocessor-controlled with automatic shut-off and overload protection.

Indications for Use

Indicated for over-the-counter cosmetic use for facial and neck stimulation (Model HD-15A) and treatment of facial wrinkles (Models HD-15 and HD-15A).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 13, 2023 Shenzhen Kaiyan Medical Equipment Co., Ltd Alain Dijkstra Official Correspondent Building#3 and Building#5, 40th of Fuxin Street Huaide Community Fuyong Town, Baoan District Shenzhen, 518103 China Re: K232863 Trade/Device Name: Radiant Renewal Skincare Wand (HD-15, HD-15A) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, NFO Dated: August 23, 2023 Received: September 15, 2023 Dear Alain Dijkstra: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jianting Wang Digitally Digitally signed by Jianting Date: 2023.12.13 13:54:39 -2 -05'00' For Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices {2}------------------------------------------------ OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232863 Device Name Radiant Renewal Skincare Wand (HD-15, HD-15A) Indications for Use (Describe) The Radiant Renewal Skincare Wand (Model: HD-15) is intended for treating wrinkles on the face and is indicated for over-the-counter use. The Radiant Renewal Skincare Wand (Model: HD-15A) is intended for facial and neck stimulation and is indicated for over-the-counter cosmetic use, the red light is intended for treating wrinkles on the face and is indicated for over thecounter use. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary of K232863 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. #### 1. Submitter's Information Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Establishment Registration Number: 3011644607 Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China Contact Person (including title): Alain Dijkstra (Manager) Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: requlation@kaiyanmedical.com #### Application Correspondent: Contact Person: Alain Dijkstra Company: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China Tel: +86 755 82129361 Fax: +86 755 25024651 Email: regulation@kaiyanmedical.com #### 2. Subject Device Information: Trade Name: Radiant Renewal Skincare Wand, model: HD-15, HD-15A Classification Name: Light based OTC for winkle reduction; Transcutaneous Electrical, Aesthetic Purposes Review Panel: General & Plastic Surgery Product Code: OHS, NFO Regulation Number:21 CFR 878.4810, 21 CFR 882.5890 Regulation Class: II #### 3. Predicate Device Information Predicate Device 1 (K171821) Sponsor: Heat In A Click Trade Name: 2 Face / Face Evolution (Model: 2 Face / Face Evolution) Classification Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes; Light Based Over the Counter Wrinkle Reduction, over-the counter powered light based laser for acne Review Panel: General & Plastic Surgery Product Code: NFO, OHS, OLP Regulation Number: 21 CFR 882.5890, 21 CFR 878.4810 Requlation Class: II ## Predicate Device 2 (K212342) Sponsor: ZIIP, Inc. Trade Name: ZIIP+ Device {5}------------------------------------------------ Classification Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief. Review Panel: Neurology Product Code: NFO Regulation Number: 21 CFR 882.5890 Regulation Class: II #### Predicate Device 3 (K163470) Sponsor: ShenZhen Siken 3D Technology Development Co.,Ltd. Trade Name: Galvanic Spa Classification Name: stimulator, transcutaneous electrical, aesthetic purposes Review Panel: Neurology Product Code: NFO Regulation Number: 21 CFR 882.5890 Regulation Class: II ### Predicate Device 4 (K202055) Sponsor: Heat In A Click LLC Trade Name: Looper (Model: ZX-579S) Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over The Counter Wrinkle Reduction (OHS) Review Panel: General & Plastic Surgery Product Code: OLP, OHS Regulation Number: 21 CFR 878.4810 Regulation Class: II #### 4. Device Description The Radiant Renewal Skincare Wand (Model: HD-15, HD-15A) is a hand-held, battery-powered device used to stimulate the face and neck superficially through application of transcutaneous electrical currents and treat wrinkles by emitting 630 nanometers of red light. The device is powered by a Lithium-lon rechargeable battery, and it has a charging cable, carrying case and instruction manual. There are two models of the device, with only differences in operating modes. Model HD-15 only has a red-light wrinkle treatment effect, while Model HD-15A has two functional modes: one is red light wrinkle treatment, and the other is micro current stimulation of the face and neck. Both of them have one button to turn on/off the device. For model HD-15A, this button also can switch the mode. The device will automatically shut down after 12 minutes of operation. The recommended treatment time is 3 minutes. After every three minutes of treatment, the device will vibrate the time. If you need to continue treatment, simply turn on the device again. #### 5. Intended Use / Indications for Use The Radiant Renewal Skincare Wand (Model: HD-15) is intended for treating wrinkles on the face and is indicated for over-the-counter use. The Radiant Renewal Skincare Wand (Model: HD-15A) is intended for facial and neck stimulation and is indicated for over-the-counter cosmetic use, the red light is intended for treating wrinkles on the face and is indicated for over-the-counter use. #### 6. Comparison to predicate devices Compare with the predicate devices, the subject device is very similar in design principle, intended use, {6}------------------------------------------------ indications for use, functions, material and the applicable standards. The differences between the subject device and predicate devices do not raise new questions of safety or effectiveness. | Elements of<br>Comparison | Subject<br>device | Predicate<br>device<br>(K171821) | Predicate<br>1 device<br>(K212342) | Predicate<br>2 device<br>(K163470) | Predicate<br>3 device<br>(K202055) | 4 Remark | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------| | Manufacturer | Shenzhen<br>Kaiyan<br>Medical<br>Equipment<br>Co., Ltd | Heat In A<br>Click | ZIIP Inc. | ShenZhen Siken<br>3D Technology<br>Development<br>Co., Ltd. | Heat In A Click<br>LLC | | | 510<br>Number | (K) K232863 | K171821 | K212342 | K163470 | K202055 | | | Device Name | Radiant<br>Renewal<br>Skincare<br>Wand | 2 Face / Face<br>Evolution<br>(Model:<br>Face / Face<br>Evolution) | Micro-current<br>Wrinkle<br>Reduction<br>Facial Service | Galvanic Spa,<br>Model:<br>SKB-<br>1405 | Looper | | | OTC/Rx | OTC | OTC | OTC | OTC | OTC | Same | | Regulation<br>Class | Class II | Class II | Class II | Class II | Class II | Same | | Product Code | NFO, OHS | NFO, OHS,<br>OLP | NFO | NFO | OHS, OLP | Same | | Mechanism of<br>Action | Microcurrent<br>electrical<br>therapy to<br>deliver skin<br>stimulation;<br>Red light<br>treatment for<br>facial<br>wrinkles. | Microcurrent<br>electrical<br>therapy to<br>deliver skin<br>stimulation;<br>Red light<br>treatment for<br>facial wrinkles;<br>Bule light<br>treatment for<br>mild to<br>moderate<br>inflammatory<br>acne. | Microcurrent<br>electrical<br>therapy to<br>deliver skin<br>stimulation | Microcurrent<br>electrical therapy<br>to deliver skin<br>stimulation | Blue light therapy<br>for mild to<br>moderate<br>inflammatory<br>acne;<br>Red light<br>treatment for<br>facial wrinkles. | Same | | Regulation<br>Number | 21 CFR<br>882.5890,<br>21 CFR<br>878.4810 | 21 CFR<br>882.5890,<br>21 CFR 878.4810 | 21 CFR<br>882.5890 | 21 CFR<br>882.5890 | 21 CFR<br>878.4810 | Same | | Indications for<br>Use / Intended<br>use | The Radiant<br>Renewal<br>Skincare<br>Wand<br>(Model: HD-<br>15) is<br>intended for<br>treating<br>wrinkles on<br>the face and | 2 Face / Face<br>Evolution is a<br>hand-held<br>device for<br>over-the<br>counter<br>aesthetic<br>purposes.<br>(1) The EMS<br>mode is<br>indicated<br>for | The ZIIP+<br>Device is<br>intended for<br>facial and neck<br>stimulation and<br>is indicated for<br>over-the-<br>counter use. | The Galvanic<br>Spa (Model:<br>SKB-1405) is<br>intended for<br>facial stimulation<br>and is indicated<br>for over-the<br>counter cosmetic<br>use. The<br>anatomical site | Looper (Model:<br>ZX-579S) is a<br>hand-held device<br>for over-the<br>counter aesthetic<br>purposes. The<br>Photon mode red<br>light is indicated<br>for the use in<br>treating wrinkles<br>on the face, and | Same | | Elements of Comparison | Subject device | Predicate device (K171821) | Predicate 1 device (K212342) | Predicate 2 device (K163470) | Predicate 3 device (K202055) | 4 Remark | | | is indicated for over-the- counter use. The Radiant Renewal Skincare Wand (Model: HD-15A) is intended for facial and neck stimulation and is indicated for over-the- counter cosmetic use, the red light is intended for treating wrinkles on the face and is indicated for over-the- counter use. | facial stimulation; (2) The Photon mode: The red light is intended for the treatment of periorbital wrinkles and the blue light is intended for the treatment of the mild to moderate inflammatory acne. | | the Galvanic Spa is the face. | the blue light is indicated for the treatment of the mild to moderate inflammatory acne. | | | Material/ Biocompatibility | PC and Aluminum | ABS Plastic & Stainless Steel | Biocompatible materials typically used in medical devices and identical to predicate ZIIP device | Stainless steel | ABS Plastic & Stainless Steel | & Different , note 1 | | Power Source | Lithium battery: 3.7V, 500mAh, 1.85Wh | DC 2200mAh | 3.7V Internal Rechargeable Lithium Ion Battery | 3.7V Li-battery | DC 3.7V 1000mA Li battery | Different , note 2 | | Sterility | Not applicable – this device is not sold sterile | Not applicable – this device is not sold sterile | Not applicable – this device is not sold sterile | Not applicable – this device is not sold sterile | Not applicable – this device is not sold sterile | Same | | Human Factors | Hand-held device | Hand-held device | Hand-held device | Hand-held device | Hand-held device | Same | | Electrical Safety | Compliant with | Compliant with | Compliant with IEC 60601-1 | Compliant with IEC 60601-1 | Compliant with IEC 60601-1 | Same | | Elements of<br>Comparison | Subject<br>device | Predicate<br>device 1<br>(K171821) | Predicate<br>device<br>(K212342) | Predicate<br>2 device<br>(K163470) | Predicate<br>3 device<br>(K202055) | 4 Remark | | | 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-11,<br>IEC 62471,<br>IEC 60601-2-10,<br>IEC 60601-2-57 | 60601-1,<br>IEC 60601-1-2,<br>IEC 62471,<br>IEC 60601-2-57 | IEC 60601-1-2,<br>IEC 60601-1-11 | IEC 60601-1-2,<br>IEC 60601-2-10 | 60601-1-2,<br>60601-1-11,<br>62471,<br>60601-2-57 | IEC | | Type of<br>Energy | LED,<br>Electrical<br>current | LED,<br>Electrical<br>current | Electrical<br>current | Electrical<br>current | LED | Same | | Charging<br>Method | External wall<br>adaptor | External wall<br>adaptor | External wall<br>adaptor | External wall<br>adaptor | No publicity | Same | | Detailed Output Characteristics | | | | | | | | Power<br>Source(s) | -- | -- | -- | -- | -- | -- | | a) Method of<br>Line Current<br>Isolation | One<br>rechargeable<br>Lithium-Ion<br>Battery and<br>External<br>Charger<br>Isolation | One<br>rechargeable<br>Lithium-Ion<br>Battery and<br>External<br>Charger<br>Isolation | One<br>rechargeable<br>Lithium-Ion<br>Battery and<br>External<br>Charger<br>Isolation | One<br>rechargeable<br>Lithium-Ion<br>Battery and<br>External Charger<br>Isolation | Type BF | Same | | b) Patient<br>Leakage<br>Current | -- | -- | External<br>Charger<br>Included | -- | No publicity | -- | | - Normal<br>Condition | ≤100μΑ | No publicity | 46 μΑ | No publicity | No publicity | Different<br>, note 2 | | - Fault<br>Condition | ≤500μΑ | No publicity | 46 μΑ | No publicity | No publicity | Different<br>, note 2 | | Number of<br>Output<br>Modules | 1 | 2 | 1 | 1 | Not applicable | Same | | Number of<br>Output<br>Channels | 1 | 2 | 1 | 1 | Not applicable | Same | | a)<br>Synchronous<br>or Alternating | N/A – 1<br>Output<br>Channel | No publicity | N/A – 1 Output<br>Channel | Alternating | Not applicable | Same | | b) Method of<br>Channel<br>Isolation | N/A – 1<br>Output<br>Channel | Type BF | N/A - 1 Output<br>Channel | No publicity | Not applicable | Same | | Regulated<br>Current or<br>Regulated<br>Voltage | Both | Both | Both | Both | Not applicable | Same | | Software/ | Yes | Yes | Yes | Yes | Yes | Same | | Elements<br>of Comparison | Subject<br>device | Predicate<br>device<br>(K171821) | Predicate<br>device<br>(K212342) | Predicate<br>device<br>(K163470) | Predicate<br>device<br>(K202055) | Remark | | Firmware/<br>Microprocessor<br>Controlled | | | | | | | | Automatic<br>Overload Trip | Yes | No | Not required<br>because of<br>circuit design<br>(Current and<br>Voltage<br>Limited by<br>Circuit Design<br>and Firmware) | Yes | Not applicable | Same | | Automatic No-<br>Load Trip | No | Yes | Yes<br>(Reversion to<br>Fixed Voltage<br>Output) | Yes | Not applicable | Different,<br>note 2 | | Automatic<br>Shut Off | Yes | Yes | Yes | Yes | Yes | Same | | Patient<br>Override<br>Control | Yes | Yes | Yes | Yes | Yes | Same | | Indicator Display | | | | | | | | a)<br>On/Off<br>Status | Yes | Yes | Yes<br>(LED<br>Illumination on<br>Conduction) | Yes | Yes | Same | | b) Low Battery | Yes | Yes | Yes | No | Yes | Same | | c)<br>Voltage/Current<br>Level | Yes | Yes | Yes<br>(LED<br>Illumination on<br>Target Current<br>Levels) | Yes | Yes | Same | | Output Specifications | | | | | | | | Waveform | Pulsed<br>Biphasic | Pulsed<br>Biphasic,<br>Modulated<br>Square | Pulsed<br>Biphasic | Pulsed Biphasic | Not applicable | Same | | Shape | Rectangular | Rectangular | Modulated<br>Square Wave | Symmetric<br>biphasic,<br>rectangular | Not applicable | Same | | Maximum<br>Output Voltage | $164 mV @ 500Ω$<br>$540 mV @ 2kΩ$<br>$2.28V @ 10kΩ$ | $310mV @ 500Ω$<br>$1.16V @ 2kΩ$<br>$5.56V @ 10kΩ$ | $154 mV @ 500Ω$<br>$465 mV @ 2kΩ$<br>$2.2 V @ 10kΩ$ | $156mV @ 500Ω$<br>$0.78V @ 2kΩ$<br>$2.6V @ 10kΩ$ | Not applicable | Similar,<br>note 3 | | Maximum<br>Current<br>Density | $328 μA @ 500Ω$<br>$270 μA @ 2kΩ$ | $620μA @ 500Ω$<br>$580μA @ 2kΩ$ | $309 μA @ 500Ω$<br>$232 μA @ 2kΩ$ | $0.31mA @ 500Ω$<br>$0.39mA @ 2kΩ$<br>$0.26mA @ 10kΩ$ | Not applicable | Similar,<br>note 3 | | Elements of Comparison | Subject device | Predicate device (K171821) | Predicate 1 device (K212342) | Predicate 2 device (K163470) | Predicate 3 device (K202055) | 4 Remark | | | 228 μA @ 10kΩ | 10kΩ | 10kΩ | | | | | Frequency range | 10.5Hz | 8.333Hz | No publicity | 9.09 Hz±10 % | Not applicable | Similar, note 3 | | Pulse duration | 48ms±10% | 60ms | No publicity | 55 ms | Not applicable | Similar, note 3 | | Maximum Phase Charge | 15.74 μC@ 500Ω | 26.31μC @ 500Ω | 6.16 μC@ 500Ω | 17.1 μC @ 500Ω | Not applicable | Similar, note 3 | | Maximum Current Density | 0.10 mA/cm²@ 500Ω | 0.330mA/cm² @500Ω | 0.34 mA/cm²@ 500Ω | 0.274mA/cm²@50Ω (The Electrode Size: 1.13 cm²) | Not applicable | Similar, note 3 | | Maximum Power Density | 0.004 mW/cm²@ 500Ω | 4.34μW/cm² @500Ω | 3.44 W/ cm²@ 500Ω | 42.5μW/cm²@50Ω (The Electrode Size:1.13cm²) | Not applicable | Similar, note 3 | | ON time | Constant | Constant | Constant | Adjustable, due to different levels | Not applicable | Same | | OFF time | None | None | None | Adjustable, due to different levels | Not applicable | Same | | Burst Mode (i.e. pulse trains) | N/A - no burst mode | No publicity | N/A - no burst mode | No publicity | Not applicable | Same | | LED output Specification | | | | | | | | LED wavelength | Red: 630nm ± 10nm | Red: 630 ± 3nm, Blue: 415 ± 3nm | Not applicable | Not applicable | Red: 630 ± 10nm, Blue: 415 ± 10nm | Same | | Irradiances | 55mW/cm² ± 10% | Red light: 73.26mw/cm² ± 10% Blue: 64.10mw/cm² ± 10% | Not applicable | Not applicable | Red light: 55mw/cm² ± 10% Blue: 65mw/cm² | Similar, note 4 | | LED output Specification | |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...