K202055 · Heat IN A Click, LLC · OLP · Apr 16, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202055
Device Name
Looper (Model: ZX-579S)
Applicant
Heat IN A Click, LLC
Product Code
OLP · General, Plastic Surgery
Decision Date
Apr 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Looper (Model: ZX-579S) is a hand-held device for over-the counter aesthetic purposes. The Photon mode red light is indicated for the use in treating wrinkles on the blue light is indicated for the treatment of the mild to moderate inflammatory acne.
Device Story
Hand-held, multifunctional beauty instrument for home use; features Sonic and Photon modes. Sonic mode provides facial skin vibration with 5 intensity levels. Photon mode emits red light (630nm) for facial wrinkle treatment and blue light (415nm) for mild to moderate inflammatory acne. Device consists of applicator and charging base; no external data connectivity. User operates via PHOTON and SONIC keys to select modes and adjust intensity. Therapeutic probe contacts skin directly. Automatic shut-off included. Benefits include non-invasive aesthetic treatment for wrinkles and acne.
Hand-held device; ABS plastic and stainless steel housing. Energy source: DC 3.7V 1000mA Li-battery. Light source: LED (Red 630±10nm, Blue 415±10nm). Irradiance: Red 55mW/cm2, Blue 48mW/cm2. Sonic vibration mode. No external connectivity. Type BF isolation. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 62133-2, IEC 62304.
Indications for Use
Indicated for individuals with facial wrinkles and mild to moderate inflammatory acne. For over-the-counter aesthetic use.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
April 16, 2021
Heat In A Click LLC % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, Guangdong 510006 China
Re: K202055
Trade/Device Name: Looper (Model: ZX-579S) Regulation Number: 21 CFR 878.4810 Regulation Name: over-the-counter powered light based laser for acne Regulatory Class: Class II Product Code: OLP, OHS Dated: March 9, 2021 Received: March 12, 2021
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202055
Device Name Looper (Model: ZX-579S)
Indications for Use (Describe)
Looper (Model: ZX-579S) is a hand-held device for over-the counter aesthetic purposes. The Photon mode red light is indicated for the use in treating wrinkles on the blue light is indicated for the treatment of the mild to moderate inflammatory acne.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of K202055
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
Company Name: Heat In A Click LLC Establishment Registration Number: 3008929787 Address: 1975 Tigertail Blvd Dania Florida 33034 USA Phone: 1(954)518-9777 Fax No.: 1(954)320-7984 Contact Person (including title): Guy Levi (CEO) E-mail: hagar@palmmassager.com
#### Factory:
510(k) Owner's Name: Hong Qiang Xing (Shenzhen) Electronics Limited Establishment Registration Number: Applying Address: 4F., Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen, Guangdong Province, PRC Phone: +86-755-26423615 Fax No.: +86-755-29915485 Contact Person (including title): Mr. Xu Jianhua (General Manager) E-mail: info@sunmas.com
#### Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
# 2. Subject Device Information
Common Name: Electrosurgical device for over-the-counter aesthetic use Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over The Counter Wrinkle Reduction (OHS) Trade Name: Looper (Model: ZX-579S)
{4}------------------------------------------------
Review Panel: Neurology; General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810, 878.4420 Regulation Class: II
## 3. Predicate Device Information
| Sponsor | Heat In A Click | Li-Tek Electronics<br>Technologies | AEGIS Regulatory, Inc. |
|-----------------------------|-----------------------------------------------------------|------------------------------------|------------------------|
| Device<br>Name and<br>Model | 2 Face / Face Evolution(Model:2<br>Face / Face Evolution) | LED Phototherapy Device | Elevare Plus |
| 510(k)<br>Number | K171821 | K162098 | K172909 |
| Product<br>Code | NFO, OHS, OLP | OLP, OHS | OHS |
| Regulation<br>Number | 882.5890 | 878.4810 | 878.4810 |
| Regulation<br>Class | II | II | II |
## 4. Device Description
Looper (Model: ZX-579S) is a hand-held device for over-the-counter aesthetic purposes, it's a multifunctional comprehensive beauty instrument. It's combined with the two kinds of operation functions: SONIC and PHOTON mode.
Sonic mode: The Sonic waves can provide vibrating to facial skin. The device provides 5 vibration intensities for users to choose from.
Photon mode: The red light is intended for use in treating wrinkles on the face, and the blue light is intended for the treatment of the mild to moderate inflammatory acne. It emits energy in the red spectrum for the treatment of facial wrinkles. The therapeutic probe is designed for direct contact with the face for treatment.
The Looper device consists of an applicator and charging base. An applicator is a hand-held unit used for treatment. The treatment surface is located at the applicator tip and comes in direct contact with the skin. The charging base charging for the applicator. And there is no external data connection capability (e.g., through USB, Ethernet, WiFi, Bluetooth, etc.) for the device.
There are two keys: the PHOTON key and the SONIC key. The PHOTON is to select red light and the Sonic is to select the SONIC mode and adjust the sonic. Both keys can turn on/off the device.
#### 5. Intended Use / Indications for Use
{5}------------------------------------------------
Looper (Model: ZX-579S) is a hand-held device for over-the counter aesthetic purposes. The Photon mode red light is indicated for the use in treating wrinkles on the blue light is indicated for the treatment of the mild to moderate inflammatory acne.
# 6. Test Summary
The Looper (Model: ZX-579S) has been evaluated the safety and performance by lab bench testing as following:
- . IEC 60601-1: 2005+A1: 2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- � IEC 60601-1-11 (Edition 2.0): 2015, Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- ◆ IEC 60601-1-2: 014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic Compatibility.
- ◆ IEC 60601-2-57 (First Edition): 2011 for use in conjunction with IEC 60601-1:2005, Medical Electrical Equipment - Part 2-57 : Particular requirements for the basic safety and essential performance of nonlaser light source equipment intended for therapeutic, monitoring and cosmetic/aesthetic use.
- ◆ IEC 62133-2 Edition 1.0 2017-02, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
- . IEC 60601-1-6: 2013 (Edition 3.1), Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- . IEC 62366-1: 2015 (First Edition), Medical devices - Application of usability engineering to medical devices.
- ◆ IEC 62304: Edition 1.1 2015-06, Medical device software, Software life- cycle processes.
- ◆ Self Selection and Usability
# 7. Comparison to predicate device and conclusion
Although there has some slight differences between with subject device and predicate device, the subject device Looper (model: ZX-579S) is Substantial Equivalence to all predicate devices. Because compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1<br>(Primary<br>Device) | Predicate<br>Device 2 | Predicate<br>Device 3 | Remark |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1<br>(Primary<br>Device) | Predicate<br>Device 2 | Predicate<br>Device 3 | Remark |
| Company | Heat In A Click<br>LLC | Heat In A Click | Li-Tek<br>Electronics<br>Technologies | AEGIS<br>Requlatory,<br>Inc. | -- |
| Trade Name | Looper | 2 Face / Face<br>Evolution | LED<br>Phototherapy<br>Device | Elevare Plus | -- |
| Classification<br>Name | Light Based<br>over the<br>Counter<br>Wrinkle<br>Reduction,<br>Over-the-<br>counter<br>powered light<br>based for acne | Transcutaneo<br>us Electrical,<br>Aesthetic<br>Purposes,<br>Light Based<br>Over The<br>Counter<br>Wrinkle<br>Reduction,<br>Over-The-<br>Counter<br>Powered Light<br>Based Laser<br>For Acne | Light Based<br>over the<br>Counter<br>Wrinkle<br>Reduction,<br>Over-the-<br>counter<br>powered light<br>based for acne | Light Based<br>Over-The-<br>Counter<br>Wrinkle<br>Reduction<br>Device | SE |
| 510(k)<br>Number | K202055 | K171821 | K162098 | K172909 | -- |
| Product Code | OHS, OLP | NFO, OHS,<br>OLP | OLP, OHS | OHS | SE |
| Intended Use /<br>Indications for<br>Use | Looper<br>(Model: ZX-<br>579S) is a<br>hand-held<br>device for<br>over-the<br>counter<br>aesthetic<br>purposes. The<br>Photon mode<br>red light is<br>indicated for<br>the use in<br>treating<br>wrinkles on<br>the face, and<br>the blue light<br>is indicated<br>for the<br>treatment of<br>the mild to<br>moderate<br>inflammatory<br>acne. | The subject<br>device is a<br>hand-held<br>device for<br>over-the<br>counter<br>aesthetic<br>purposes and<br>it's intended<br>use for:<br>(1) The EMS<br>mode is<br>indicated for<br>facial<br>stimulation;<br>(2) The<br>Photon mode:<br>The red light is<br>intended for<br>the treatment<br>of periorbital<br>wrinkles and<br>the blue light<br>is intended for<br>the treatment<br>of the mild to<br>moderate | The red light is<br>intended for<br>the treatment<br>of periorbital<br>wrinkles, and<br>the blue light<br>is intended for<br>the treatment<br>of the mild to<br>moderate<br>inflammatory<br>acne. | The Elevare<br>Plus is an<br>Over-the-<br>Counter (OTC)<br>device<br>intended for<br>the use in<br>treating<br>wrinkles on<br>the face. | SE |
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1<br>(Primary<br>Device) | Predicate<br>Device 2 | Predicate<br>Device 3 | Remark |
| | | inflammatory<br>acne. | | | |
| Anatomical<br>Sites | Entire Face | Entire Face | Entire Face | Entire Face | SE |
| Design | Hand-held<br>device | Hand-held<br>device | Hand-held<br>device | Hand-held<br>device | SE |
| Target<br>Population | Individuals<br>with wrinkles<br>on face,<br>Individuals<br>with mild to<br>moderate<br>inflammatory<br>acne | Individuals<br>with periorbital<br>wrinkles,<br>Individuals<br>with mild to<br>moderate<br>inflammatory<br>acne | Individuals<br>with periorbital<br>wrinkles,<br>Individuals<br>with mild to<br>moderate<br>inflammatory<br>acne | Individuals<br>with wrinkles<br>on face | SE |
| Environment<br>of Use | Home | Home | Home | Home | SE |
| Method of Line<br>current<br>Isolation | Type BF | Type BF | Type BF | Not publicly<br>available | SE |
| Main Unit<br>Weight | 230g | 200g | 150g | Not publicly<br>available | SE<br>Note 1 |
| Dimensions of<br>device | 234.5mm x<br>30mm x 46mm | 158mm x<br>56mm x 51.5m<br>m | 187mm x<br>65mm x 51<br>mm | Not publicly<br>available | SE<br>Note 1 |
| Housing<br>Materials of<br>main unit | ABS Plastic &<br>Stainless Steel | ABS Plastic &<br>Stainless Steel | ABS Plastic | Not publicly<br>available | SE |
| Power Source | DC 3.7V<br>1000mA Li<br>battery | DC 3.7V<br>2200mAh | DC 3.7V<br>1050mA Li<br>battery | 2 Li-Ion<br>rechargeable<br>batteries | SE |
| Software/Firm<br>ware/Micropro<br>cessor<br>Control? | Yes | Yes | Yes | Yes | SE |
| Automatic<br>Shut Off | Yes | Yes | Yes | Yes | SE |
| User Override<br>Control? | Yes | Yes | Yes | Yes | SE |
| Indicators | Indicates for<br>low battery,<br>PHOTON<br>mode<br>information,<br>SONIC mode<br>information,<br>intensity level,<br>Charging<br>Indicator | EMS Mode<br>intensity<br>Indicator<br>lights, Photon<br>Mode working<br>indicator<br>lights, Photon<br>Mode Indicator<br>lights,<br>Charging<br>Indicator | Not publicly<br>available | Not publicly<br>available | SE<br>Note 1 |
| Time Range | 3 minutes per | EMS Mode (5 | 3 minutes per | 3 minutes | SE |
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1<br>(Primary<br>Device) | Predicate<br>Device 2 | Predicate<br>Device 3 | Remark |
| (minutes) | target area; 2<br>treatments per<br>week for 6<br>weeks | minutes)<br>Photon Mode<br>(5~7 minutes) | target area; 2<br>treatments per<br>week for 6<br>weeks | daily, 5 days<br>per week for 8<br>weeks | |
| Wavelengths | Red:<br>630±10nm<br>Blue:<br>415±10nm | Red Light<br>(630nm±3nm<br>Wavelength)<br>Blue Light<br>(415nm±3nm<br>Wavelength) | Red: 630±3nm<br>Blue:<br>415±3nm | RED: 610,<br>630, 660,<br>850nm<br>(±5mm) | SE |
| The distance<br>between the<br>LEDs to<br>treatment<br>surface | Close to the<br>facial skin | 2-3 cm | 2-3 cm | Close to the<br>facial skin | SE |
| Irradiation<br>area | 12cm2±10% | 26cm2±5% | 30cm2±5% | Over 17cm² | SE<br>Note 2 |
| Irradiances | Red light:<br>55mW/cm2±10<br>%<br>Blue light:<br>48mW/cm2±10<br>% | Red light:<br>73.26mW/cm2<br>±10%<br>Blue light:<br>64.10mW/cm2<br>±10% | Red light:<br>80mW/cm2±10<br>%<br>Blue light:<br>65mW/cm2±10<br>% | 65mW/cm2<br>(±5mW) | SE<br>Note 2 |
| Irradiance<br>source | LED | LED | LED | LED | SE |
| Visible light<br>LEDs | Yes | Yes | Yes | Yes | SE |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | SE |
| Safety | IEC 60601-1,<br>IEC 60601-1-<br>11 and IEC<br>60601-2-57 | IEC 60601-1,<br>IEC 60601-2-<br>10 and IEC<br>60601-2-57 | IEC 60601-1<br>and IEC<br>60601-2-57 | IEC 60601-1,<br>IEC 60601-1-<br>11 | SE |
| Working<br>Environment | Temperature:<br>5~35°C<br>Humidity:<br>10~80%RH<br>Atmospheric<br>pressure:<br>700hPa~1060<br>hPa | Not publicly<br>available | Temperature:<br>5~40°C<br>Humidity:<br>10~80%RH<br>Atmospheric<br>pressure:<br>700hPa~1060<br>hPa | Not publicly<br>available | SE |
| Transportation<br>and Storage<br>Environment | Temperature: -<br>10~40°C<br>Humidity:<br>5~95%RH<br>Atmospheric<br>pressure:<br>600hPa~1060<br>hPa | Not publicly<br>available | Temperature: -<br>10~40°C<br>Humidity:<br>5~95%RH<br>Atmospheric<br>pressure:<br>600hPa~1060<br>hPa | Not publicly<br>available | SE |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### Comparison in Detail(s): Note 1:
Although the "Main Unit Weight", "Dimensions of device", "Power source" and "Indicator" of subject device is different from the predicate devices, they are all compliance with IEC 60601-1 requirement for the product and not affect the safety or effectiveness. So the differences of function specification will not raise any safety or effectiveness issue.
# Note 2:
Although the "Irradiation area" and "Irradiance" is a little different from the predicate devices, they are all compliance with IEC 60601-1-11 and IEC 60601-2-57 requirements. So the differences will not raise any safety or effectiveness issue.
# Note 3:
The Nature of body contact is skin contact. And the contact duration is less than 24 hours. The materials and manufacturing used for the subject device are identical to those of the device K171821, and the materials have been demonstrated to conform with the following biocompatibility standards:
· ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity · ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and delayedtype hypersensitivity
#### Final Conclusion:
Although there has some slight differences between with subject device and predicate device, the subject device Looper (Model: ZX-579S) is Substantial Equivalence to the predicate devices K171821, K162098 and K172909. Because compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. And the subject meets the same standards as the predicate device, so the differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
8. Date of the summary prepared: 2021-04-16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.