Smart Kit Basic, Smart Kit Pro

K202443 · Bsl Co., Ltd. · MUU · Mar 11, 2021 · General, Plastic Surgery

Device Facts

Record IDK202443
Device NameSmart Kit Basic, Smart Kit Pro
ApplicantBsl Co., Ltd.
Product CodeMUU · General, Plastic Surgery
Decision DateMar 11, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Smart Kit is disposable and is indicated for autologous fat transfer. The autologous fat transfer is also technically called as autologous fat grafting or reinjection of concentrated adipose-tissue into the recipient body areas. The Smart kit can be used for concentration and transfer of homogenous adipose tissues with legally marketed lipoplasty system when the concentrated and homogenous fat tissue is re-injected into the recipient body areas of the same patient autologously. This device is intended for autologous fat transfer. The Smart Kit is intended for use in the following surgical specialties when the concentration of harvested adipose tissue is desired. Neurosurgery / Gastrointestinal Surgery / Urological Surgery / General surgery / Orthopedic Surgery / Gynecological Surgery / Thoracic Surgery / Laparoscopic Surgery / Arthroscopic Surgery / Plastic and reconstructive surgery

Device Story

Device processes lipoaspirate for autologous fat transfer; removes bloody impurities, tumescent solution, and free oil. Input: lipoaspirate harvested via cannula and manual suction. Operation: centrifugation in specialized syringe; piston-based filtration separates adipose tissue from fluids/oils; Adinizer mesh (400-4000 um) further refines tissue homogeneity. Output: concentrated, homogeneous adipose tissue for reinjection. Used in clinical settings by trained physicians. Benefits: provides sterile, closed-system processing to prepare autologous fat for grafting, improving tissue quality for transfer.

Clinical Evidence

Bench testing only. Performance testing included appearance, dimensional measurement, water tightness, centrifugation leakage, cell viability (>85%), pH, heavy metals, and UV-visible absorbance. Biocompatibility testing performed per ISO 10993-1 (hemocompatibility, cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity).

Technological Characteristics

Preassembled, single-use, sterile (EtO) system. Components: filter piston syringe (10/20/60mL), Adinizer mesh filter (400-4000 µm), syringe handle, luer-to-luer connector, manual suction unit. Principle: gravity/centrifugal separation and mechanical filtration. Materials: biocompatible per ISO 10993. Shelf life: 3 years per ISO 11607.

Indications for Use

Indicated for autologous fat transfer (grafting/reinjection of concentrated adipose tissue) in patients undergoing surgical procedures in neurosurgery, gastrointestinal, urological, general, orthopedic, gynecological, thoracic, laparoscopic, arthroscopic, and plastic/reconstructive surgery.

Regulatory Classification

Identification

A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.

Special Controls

*Classification.* Class II (special controls). Consensus standards and labeling restrictions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. March 11, 2021 BSL Co., Ltd % James Monroe President & CEO Global RQC Med Device Solutions, LLC. 319 Shilling Dr. Somerset, New Jersey 08873 Re: K202443 Trade/Device Name: Smart Kit Basic, Smart Kit Pro Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: February 3, 2021 Received: February 5, 2021 Dear James Monroe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202443 Device Name Smart Kit Indications for Use (Describe) The Smart Kit is disposable and is indicated for autologous fat transfer is also technically called as autologous fat grafting or reinjection of concentrated adipose-tissue into the recipient body areas. The Smart kit can be used for concentration and transfer of homogenous adipose tissues with legally marketed lipoplasty system when the concentrated and homogenous fat tissue is re-injected into the recipient body areas of the same patient autologously. This device is intended for autologous fat transfer. The Smart Kit is intended for use in the following surgical specialties when the concentration of harvested adipose tissue is desired. Neurosurgery / Gastrointestinal Surgery / General surgery / Orthopedic Surgery / Gynecological Surgery / Thoracic Surgery / Laparoscopic Surgery / Arthroscopic Surgery / Plastic and reconstructive surgery | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Date of Preparation: | February 10, 2021 | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Jun Seok Lee, President<br>BSL Co., Ltd<br>6-13, Chilsan-ro 237beon-gil<br>Gimhae-si, Gyeongsangnam-do, Korea | | Phone: | +82-55-328-9235 | | Contact Person: | James W. Monroe<br>President & CEO<br>Global RQC Med Device Solutions,<br>LLC<br>319 Shilling Dr.<br>Somerset, NJ 08873 USA<br>Phone: (908) 809-0081<br>Fax: (908) 540-2084<br>Email: jmonre@globalrqc.com | | Trade/Device Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Predicate Device: | Smart Kit<br>Suction lipoplasty system<br>System, Suction, Lipoplasty<br>21 CFR Part 878.5040<br>General & Plastic Surgery<br>Class II<br>MUU<br>Dermapose Refresh (K193363, clearance received April<br>14,2020) | {4}------------------------------------------------ # Device Description: The fat harvested by liposuction, called as lipoaspirates, contains fat tissues, bloody impurities and tumescent solution. The harvested lipoaspirates are not appropriate for autologous fat transfer without removal of the bloody impurities and tumescent solution containing anesthetic agent, anti-coagulant agent, antibiotics and etc. In addition, the harvested fat tissues are not homogeneous and contain bigger fibrotic tissues. Smart kit is designed and developed for harvest of a variety of sizes of homogeneous adipose tissues in a closed sterile way for autologous fat transfer surgical technique. Using a separate cannula, Manual handheld suction unit, and plunger, aspirate the adipose tissue or fill the syringe with adipose tissue and centrifuge with the cap in front of the syringe. Centrifuged adipose tissue is divided into blood and fluid layer, adipose tissue layer, and free oil layer. Those are located at the bottom of the piston. After centrifugation, the plunger is pushed unlocked to the piston and free oil is drained to the outside of the syringe as the free oil moves to the top of the piston through the hole at the front of the piston. Open the cap in front of the syringe, lock the plunger completely and push it, the blood and fluid layers at the bottom of the adipose tissue will be drained and only the adipose tissue will remain inside the syringe. Depending on the syringe capacity, there are three types of piston: 10ml, 20ml, 60ml. The built-in filter adapter consists of a filter for separating the fiber and adipose tissue to facilitate movement through needles of lumpy adipose tissues and a body for receiving it. Connect the syringe containing the adipose tissue and the empty syringe to the connection parts at both ends of the built-in filter adapter and push the plunger of the syringe containing the adipose tissue to allow the adipose tissue to pass through the filter. Fiber and adipose tissues larger than the pore size is filtered, and adipose tissue smaller than the pore size of the filter is passed through to separate adipose tissue. ## Smart Syringe: Filter piston syringe is equipped with a piston built-in filter. After the filter piston syringe is filled with aspirated adipose tissue, filter piston syringe capped with a syringe CAP, centrifuged and divided into blood and fluid layers, adipose tissue layer, and free oil layer due to differences in specific gravity. During the centrifugation process, the weight valve in the piston operates to open the oilpassage, and the free oil passes through the filter of the piston. The adipose tissue that is larger than the pore of the filter can not pass through, so it stays between the front of the filter piston and the blood & fluid layer. The Adipose tissue and free oil are separated. After the centrifugation is completed, push the piston built-in filter {5}------------------------------------------------ through the filter piston plunger to remove blood and fluid layer, leaving only adipose tissue in the device. #### Smart Lock A separate handheld suction unit that is reusable. ## Syringe Connector Luer to Luer connector between two syringes for fat transfer, that is disposable # Syringe Handle Assist in fat processing, disposable # Adinizer Concentrates adipose-tissue passes through (4000-400 um) mesh outlet into the reconstructing area. Smart Kit (System) - The device comes is available either as a) Smart Kit Basic, b) Smart Kit Pro - 1. Smart Kit Basic: - a. Piston - b. Plunger - Syringe CAP C. - d. Syringe Handle - Syringe Connector e. - Smart Syringe f. 2. Smart Kit Pro: - g. Piston - h. Plunger - 1. Syringe CAP - Syringe Handle i. - k. Syringe Connector - Smart Syringe l. - m. Adinizer (Mesh Filter) Smart kit is recommended to be used by the well-trained physicians for liposuction and fat transfer technique. This device is sterile and for single-use only. {6}------------------------------------------------ ## Intended Use/Indications for Use: The Smart Kit is disposable and is indicated for autologous fat transfer. The autologous fat transfer is also technically called as autologous fat grafting or reinjection of concentrated adipose-tissue into the recipient body areas. The Smart kit can be used for concentration and transfer of homogenous adipose tissues with legally marketed lipoplasty system when the concentrated and homogenous fat tissue is re-injected into the recipient body areas of the same patient autologously. This device is intended for autologous fat transfer. The Smart Kit is intended for use in the following surgical specialties when the concentration of harvested adipose tissue is desired. Neurosurgery / Gastrointestinal Surgery / Urological Surgery / General surgery / Orthopedic Surgery / Gynecological Surgery / Thoracic Surgery / Laparoscopic/ Arthroscopic Surgery/Plastic and reconstructive surgery. #### Summary of Technology Characteristics: Smart Kit is substantially equivalent to the predicate device. A comparison of technological characteristics is provided in the table below. Both the subject device and the predicate device provide a method of harvesting, concentrating, and transferring of autologous adipose tissue. In addition, both the subject and predicate devices are preassembled systems which are provided sterile (via ethylene oxide) and intended for single use. | Attribute | Subject<br>Device<br>Smart Kit | Predicate<br>DermaposeTM Refresh<br>(K193363) | Same/Different | |--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------| | Device<br>Classification | System, Suction,<br>Lipoplasty | System, Suction,<br>Lipoplasty | | | Regulation<br>Number | 21CFR<br>878.5040 | 21 CFR 878.5040 | | | Product Code | MUU | MUU | | | Intended Use | The Smart Kit is disposable<br>and is indicated for autologous<br>fat transfer. The autologous fat<br>transfer is also technically<br>called as autologous fat<br>grafting or reinjection of<br>concentrated adipose-tissue<br>into the recipient body areas.<br>The Smart kit can be used for<br>concentration and transfer of<br>homogenous adipose tissues<br>with legally marketed<br>lipoplasty system when the<br>concentrated and homogenous<br>fat tissue is re-injected into the<br>recipient body areas of the<br>same patient autologously.<br><br>This device is intended for<br>autologous fat transfer. The | The DermaposeTM Refresh is a<br>sterile medical device intended<br>for the closed-loop processing<br>of lipoaspirate tissue in medical<br>procedures involving the<br>harvesting, concentrating and<br>transferring of autologous<br>adipose tissue harvested with a<br>legally marketed lipoplasty<br>system. The device is intended<br>for use in the following surgical<br>specialties when the transfer of<br>harvested adipose tissue is<br>desired: orthopedic surgery,<br>arthroscopic surgery,<br>neurosurgery, gastrointestinal<br>and affiliated organ surgery,<br>urological surgery, general | | | Attribute | Subject<br>Device<br>Smart Kit | Predicate<br>Dermapose™ Refresh<br>(K193363) | Same/Different | | | Smart Kit is intended for use in<br>the following surgical<br>specialties when the<br>concentration of harvested<br>adipose tissue is desired.<br><br>Neurosurgery /<br>Gastrointestinal Surgery /<br>Urological Surgery / General<br>surgery / Orthopedic Surgery /<br>Gynecological Surgery /<br>Thoracic Surgery /<br>Laparoscopic Surgery /<br>Arthroscopic Surgery / Plastic<br>and reconstructive surgery | surgery, gynecological surgery,<br>thoracic surgery, laparoscopic<br>surgery, and plastic and<br>reconstructive surgery when<br>aesthetic body contouring is<br>desired. Only legally marketed<br>accessory items, such as<br>syringes, should be used with<br>the system. If harvested fat is to<br>be reimplanted, the harvested<br>fat is only to be used without<br>any additional manipulation | | | Technological<br>Characteristics<br>(Components) | Smart Kit is composed of 4<br>major parts: 1. Harvester<br>(Piston, plunger, syringe cap,<br>luer to luer syringe connector,<br>filter piston syringe(Smart<br>syringe) need to remove piston<br>plunger,2. Builtin filter<br>adaptor(Adinizers), 3. Syringe<br>handle 4. Manual handheld<br>suction unit (Smart Lock)<br><br>Smart Kit Variations:<br><br>1. Smart Kit Basic(SKT-<br>010,SKT-020, SKT-060,<br>SKT-011, SKT-021,<br>SKT-061)<br>a. Smart Lock<br>(PLK-002,<br>PLK-003)<br>b. Smart<br>Syringe<br>(SKT-SS-<br>100)<br><br>2. Smart Kit Pro(SKT-<br>14060,SKT-24060, SKT-<br>12440, SKT-22440, SKT-<br>12460, SKT-22460)<br>a. Adinizer<br>(SKT- AN-<br>100, SKT-<br>AN-200,<br>SKT- AN-<br>400, SKT-<br>AN-600,<br>SKT- AN-<br>1200, SKT-<br>AN-<br>2400,SKT-<br>AN-4000,<br>SKT-AN-<br>4060, SKT-<br>AN-4040, SKT- | 50 mL polycarbonate<br>harvesting and<br>processing syringe unit<br>with vacuum lock and 30<br>mL output syringe<br>containing a two- stage<br>sieve filter on the inletTwo 30 mL<br>polypropylene<br>luer-lock syringesThree luer capsLuer-to-luer adapter | | | Attribute | Subject<br>Device<br>Smart Kit | Predicate<br>Dermapose™ Refresh<br>(K193363) | Same/Different | | | AN-2440, SKT-<br>AN-101, SKT-<br>AN-201, SKT-<br>AN-401, SKT-<br>AN-601, SKT-<br>AN-1201, SKT-<br>AN-2401, SKT-<br>AN-4001, SKT-<br>AN-2460, SKT-<br>AN-1240, SKT-<br>AN-6020, SKT-<br>AN-24201,<br>SKT-AN-<br>40201)<br>b. Smart Syringe<br>(SKT-SS-100) | | | | Device Size | • BD syringe 10cc (NB<br>No.0086)<br>• BD syringe 20cc (NB<br>No.0050)<br>• BD syringe 60cc (NB<br>No.0050) | 50 ml syringe | | | User<br>Environment | Physician | Physician | | | Method of<br>Harvesting | Cannula | Cannula | | | Construction | Preassembled | Preassembled | | | Device<br>Performs<br>Harvestin<br>g | Syringe generated vacuum-<br>lock | Syringe generated vacuum-<br>lock | | | Filter/Mesh | Concentrated adipose-tissue<br>passes through (4000-400 µm)<br>mesh outlet into the<br>reconstructing area. | 1000 and 800 micron mesh at<br>outlet - collected adipose<br>tissue passes though both | | | Sterilization<br>Method | Ethylene-Oxide Gas (EtO) | Ethylene-Oxide Gas (EtO) | | | Sterility<br>Assurance<br>Level | SAL 10-6<br>Shelf life: 3 years | SAL 10-6<br>Shelf life: 3 years | | | Disposable or<br>reusable | Single Use, Disposable | Single Use, Disposable | | | Biocompatibility | Biocompatible | Biocompatibility | | {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ ## Performance Testing: Smart Kit was tested in accordance with established protocols and met the acceptance criteria for all tests performed. Performance testing include: | Test Item | Test Requirement | |------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Appearance | There should be no visible scratches, damage, or<br>foreign material | | Measurement | When measuring with Vernier calipers, according to<br>the dimension of "shape and structure" part, the<br>stated should be within $\pm$ 5% | | Water Tightness | When tested according to the test method, there<br>should be no leakage | | Centrifugation<br>Compatibility<br>Test<br>(Leakage) | When tested according to the test method, it should<br>be aspirated to the maximum capacity without<br>leakage | | Cell Viability Test | The finally filtered adipose tissues for autologous fat<br>transfer should maintain at least over 85 % of the cell<br>viability. | | Pore Size | When measuring with Vernier calipers, according to<br>the dimension of 'shape and structure' part, the stated<br>should be within $\pm$ 15 %. | | pH | Difference of pH $\leq$ 1.5 | | Potassium permanganate<br>reducing substances | Difference of the consumption $\leq$ 2.0 mL | | Evaporating residue | Difference of residues $\leq$ 1.0mg | | Heavy Metals | The test solution should not be darker than blank<br>solution. | | Ultraviolet-visible | Difference of the maximum absorbance at (250~350)<br>nm $\leq$ 0.1 | # Sterilization and Shelf Life Smart Kit was verified for validation for 3 years (36 months) by testing physicochemical changes and packaging of the materials and the product before, middle, and after acceleration aging according to ISO 11607. Sterilization was conducted on the Smart Kit Fat Separator with Ethylene oxide gas. {10}------------------------------------------------ Sterilization is performed with the cycle validation on the report in accordance with EN ISO11135:2014 Sterilization of medical devices validation and routine control of sterilization by Ethylene oxide gas. The sterilizing cycle has been validated to ensure our products to be sterilized to an S.A.L. of at least 106 by EN ISO11135:2014 by the method of over-kill. #### Biocompatibility Biocompatibility testing for Smart Kit was performed in accordance with the requirements of ISO 10993-1:2009 for external communicating devices having contact < 24 hours. In all instances, the test articles were found to be biocompatible. The following tests were conducted: - . Hemocompatibility, per ISO 10993-4:2017 - Cytotoxicity, per ISO-10993-5: 2009 - Sensitization, Intracutaneous Reactivity, per ISO 10993-10: 2010 - Acute Systemic Toxicity, Pyrogen per ISO 10993-11:2018 ● #### Statement of Substantial Equivalence Based on the information contained in this submission, it is concluded that the Smart Kit is substantially equivalent to the predicate device, Dermapose Refresh. Smart Kit has the same indications for use, as well as similar technological characteristics and principles of operation as its predicate device. Thus, Smart Kit is substantially equivalent to the Dermapose Refresh (K193363).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...