Ranfac Fat Aspiration Transfer Syringe (FATS) Procedure Pack, Ranfac Fat Aspiration Transfer Syringe (FATS) Containment Vessel
K150156 · Ranfac, Corp. · MUU · Oct 19, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150156
Device Name
Ranfac Fat Aspiration Transfer Syringe (FATS) Procedure Pack, Ranfac Fat Aspiration Transfer Syringe (FATS) Containment Vessel
Applicant
Ranfac, Corp.
Product Code
MUU · General, Plastic Surgery
Decision Date
Oct 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ranfac FATS Procedure Pack is used in medical procedures involving the harvesting and transferring of autologous adipose tissue. The FATS Procedure Pack is for concentrating adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the concentration of harvested adipose is desired: Neurosurgery, gastrointestinal surgery, urological surgery, plastic & reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, arthroscopic surgery.
Device Story
Device facilitates sterile processing and transfer of harvested autologous adipose tissue. System includes centrifugation syringes, transfer syringes, luer fittings, and a reusable containment vessel. Principle of operation: density-gradient separation via centrifugation. User aspirates fat into syringe; removes plunger/handle to shorten profile; places in generic centrifuge (1250 g-force for 4 minutes). Concentrated adipose tissue is transferred to injection syringe via luer-lock fittings. Used in clinical/surgical settings by healthcare providers. Benefits patient by providing concentrated autologous tissue for surgical applications.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including fluid transfer performance, centrifugation testing, and simulated use. Biocompatibility testing performed per ANSI/AAMI/ISO 10993-1:2009.
Technological Characteristics
Components: polypropylene syringe barrel, plunger with elastomer seal, luer fittings. Centrifugation: 1250 g-force for 4 minutes. Sterilization: Ethylene-Oxide (EtO). Conformance: ISO 7886-1:1993 (sterile hypodermic syringes). Single-use disposables.
Indications for Use
Indicated for patients undergoing medical procedures requiring the harvesting and concentration of autologous adipose tissue across various surgical specialties including neurosurgery, gastrointestinal, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, laparoscopic, and arthroscopic surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized depiction of a human figure in profile, with three overlapping faces representing the department's focus on health and human well-being. The logo is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 19, 2015
Ranfac Corporation Mr. Christopher P. Whelan Senior Vice President - Quality 30 Doherty Avenue Avon, Massachusetts 02322
Re: K150156
Trade/Device Name: Ranfac Fat Aspiration Transfer Syringe (FATS) Procedure Pack Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: September 11, 2015 Received: September 14, 2015
Dear Mr. Whelan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and
{2}------------------------------------------------
809]); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150156
### Device Name
Ranfac Fat Aspiration Transfer Syringe (FATS) Procedure Pack
### Indications for Use (Describe)
The Ranfac FATS Procedure Pack is used in medical procedures involving the harvesting of autologous adipose tissue. The FATS Procedure Pack is for concentrating adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the concentration of harvested adipose is desired: Neurosurgery, gastrointestinal surgery, urological surgery, plastic & reconstructive surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, arthroscopic surgery.
Type of Use (Select one or both, as applicable):
| <span> <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 12px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (Page 1 of 4)
The contents of this 510(k) summary have been provided in conformance with 21 CFR §807.92 "Content and Format of a 510(k) Summary".
#### Submitter/Sponsor 1.
Ranfac Corp. 30 Doherty Avenue Avon MA 02322 FDA Registration Number 1211566 Telephone Number/Fax: 508-588-4400 ext. 106/508-584-8588 Contact Person: Christopher P. Whelan Date Prepared: December 30, 2014
### 2. Device Name
Trade Name: Ranfac Fat Aspiration Transfer Syringe (FATS) Procedure Pack Common or Usual Name: Fat Transfer Syringe Procedure Pack Classification Name: Suction Lipoplasty System 21 CFR §878.5040, Procode MUU Classification: Class II
#### 3. Predicate Device:
| Trade Name | 510(k) | Company |
|------------------------|---------|----------------------|
| Harvest AdiPrep System | K121005 | Harvest Technologies |
### 4. Device Description
The Ranfac Fat Aspirate Transfer System (FATS) is designed to allow for sterile processing and transfer of the collected adipose tissue. The device consists of a procedure pack including the following components: two sterile centrifugation syringes, two each 30 mL and 20 mL sterile syringes for adipose tissue aspirate transfer, and other luer fittings as accessories to be used to transfer into and from the centrifugation syringe. Additionally, a non-sterile, reusable centrifugation syringe Containment Vessel is provided as a separate accessory which allows the centrifugation syringe to fit within a commercially available centrifuge.
### K150156/S001 RAI #2
{5}------------------------------------------------
### 510(k) Summary (Page 2 of 4)
#### 5. Indications For Use
The Ranfac FATS Procedure Pack is used in medical procedures involving the harvesting and transferring of autologous adipose tissue. The FATS Procedure Pack is for concentrating adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the concentration of harvested adipose is desired: Neurosurgery, gastrointestinal surgery, urological surgery, plastic & reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, arthroscopic surgery.
#### Comparison of the Technological Characteristics With the Predicate Devices: 6.
As compared with the predicate device, and as shown below, the Ranfac FATS device has the same indications for use, and has similar technological and operational characteristics when compared with the predicate device.
| | Ranfac FATS (This Submission) | Harvest AdiPrep (K121005) |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Ranfac FATS Procedure Pack is used in<br>medical procedures involving the harvesting<br>and transferring of autologous adipose tissue.<br>The FATS Procedure Pack is for concentrating<br>adipose tissue harvested with a legally<br>marketed lipoplasty system. The device is<br>intended for use in the following surgical<br>specialties when the concentration of harvested<br>adipose is desired:<br><br>Neurosurgery, gastrointestinal surgery,<br>urological surgery, plastic & reconstructive<br>surgery, general surgery, orthopedic surgery,<br>gynecological surgery, thoracic surgery,<br>laparoscopic surgery, arthroscopic surgery. | The AdiPrep Adipose Transfer System is used in<br>medical procedures involving the harvesting and<br>transferring of autologous adipose tissue. The<br>AdiPrep system is for concentrating adipose<br>tissue harvested with a legally marketed<br>lipoplasty system. The AdiPrep Adipose<br>Transfer System is intended for use in the<br>following surgical specialties when the<br>concentration of harvested adipose is desired:<br><br>Neurosurgery, gastrointestinal surgery,<br>urological surgery, plastic & reconstructive<br>surgery, general surgery, orthopedic surgery,<br>gynecological surgery, thoracic surgery,<br>laparoscopic surgery, arthroscopic surgery. |
| Components | Syringe with removable plunger, aspiration &<br>fat injection syringes and accessory caps and<br>luer fittings. | Syringe with removable plunger, aspiration & fat<br>injection syringes, accessory caps and luer<br>fittings, skin puncture needles, and oil extraction<br>syringe & needle. |
Table 5.1 Comparison of the Proposed Ranfac FATS System to the Predicate Harvest AdiPrep Adipose Transfer System
### K150156/S001 RAI #2
{6}------------------------------------------------
# 510(k) Summary (Page 3 of 4)
| | Ranfac FATS (This Submission) | Harvest AdiPrep (K121005) |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Centrifugation<br>Syringe | Modified standard hypodermic syringe with a<br>polypropylene barrel with printed graduations,<br>and a plunger composed of a polypropylene body<br>supporting an elastomer seal (i.e., plunger tip).<br>The plunger is removed prior to centrifugation to<br>allow the syringe to fit within the centrifuge.<br>The FATS centrifugation syringe has a second<br>gasket that serves as a barrier to help maintain<br>vacuum pressure within the syringe during<br>centrifugation and minimizes exposure of the<br>syringe contents with air. | Modified standard hypodermic syringe with a<br>Cyclo Olefin Polymer (COP) barrel with printed<br>graduations, and a plunger composed of a<br>polypropylene or ABS body supporting a<br>silicone seal (i.e., plunger tip). The plunger is<br>removed prior to centrifugation to allow the<br>syringe to fit within the centrifuge.<br>The AdiPrep centrifugation syringe has a lipid<br>barrier which provides isolation of oils and lipids<br>from adipose tissue and minimizes exposure of<br>the concentrated adipose tissue with air. |
| | The centrifugation syringe has a male luer-lock<br>which is attached to a female-female luer lock to<br>transfer the concentrated fat to an injection<br>syringe. | The centrifugation syringe has a male luer-lock<br>which is attached to a female-female luer lock to<br>transfer the concentrated fat to an injection<br>syringe. |
| Fill volumes | Volume = 5ml to 25ml | Volume = 5ml to 25ml |
| Principle of<br>Operation | Fat is aspirated into syringe (i.e., centrifugation<br>syringe) that has a removable handle. Shortened<br>profile of syringe after removing the handle<br>allows the syringe to fit in a standard centrifuge.<br>Fat and supernatant material are separated via<br>density-gradient separation. | Fat is aspirated into syringe (i.e., centrifugation<br>syringe) that has a removable handle. Shortened<br>profile of syringe after removing the handle<br>allows the syringe to fit in a standard<br>centrifuge. Fat and supernatant material are<br>separated via density-gradient separation. |
| Centrifugation<br>Protocol | 1250 g-force 4 minutes | 1250 g-force 4 minutes |
| Centrifuge | Generic centrifuge that has appropriate sized<br>bucket. | Dedicated standard swinging bucket centrifuge |
| Sterilization Method<br>of Procedure Pack | Ethylene-Oxide Gas (EtO) | Ethylene-Oxide Gas (EtO) |
| How Used | Single use only | Single use only |
# Table 5.1 Comparison of the Proposed Ranfac FATS System to the Predicate Harvest AdiPrep Adipose Transfer System
#### 7. Performance Data
Design verification tests were performed based on the risk analysis and product requirements, and the results of these tests demonstrate that the Ranfac FATS Procedure Pack is adequately designed for the intended use indicated. Design verification tests included fluid transfer into and from the process disposable, centrifugation tests, data to support conformance with applicable requirements of ISO 7886-1:1993 and simulated use.
{7}------------------------------------------------
### 510(k) Summary (Page 4 of 4)
Biocompatibility testing have also been performed in compliance with ANSVAAMI/ISO 10993-1: 2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and consistent with FDA's Blue Book Guidance G95 "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing. Results of this testing demonstrate that the materials used in the assembly of the FATS device are suitable for their intended use.
#### 8. Clinical Data
Not applicable
#### 9. Conclusion
Based on the similarities in indications for use, materials, design, principles of function, biocompatibility and sterilization between the Ranfac Fat Aspirate Transfer Syringe Procedure Pack, subject of this premarket notification and the predicate devices, the proposed subject device has been shown to be substantially equivalent to the predicate devices in accordance with Section 510(k) of the Federal Food, Drug and Cosmetic Act.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.