K121005 · Harvest Technologies, Corp. · MUU · Sep 21, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121005
Device Name
ADIPREP ADIPOSE TRANSFER SYSTEM
Applicant
Harvest Technologies, Corp.
Product Code
MUU · General, Plastic Surgery
Decision Date
Sep 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AdiPrep™ Adipose Transfer System is used in medical procedures involving the harvesting and transferring of autologous adipose tissue. The AdiPrep system is used for concentrating adipose tissue harvested with a legally marketed lipoplasty system. The AdiPrep™ Adipose Transfer System is intended for use in the following surgical specialties when the concentration of harvested adipose tissue is desired. • Neurosurgery • Gastrointestinal Surgery • Urological Surgery • Plastic and Reconstructive Surgery • General surgery • Orthopedic Surgery • Gynecological Surgery • Thoracic Surgery • Laparoscopic Surgery • Arthroscopic Surgery
Device Story
AdiPrep Adipose Transfer System is a sterile, single-use disposable process pack used with the SmartPReP2 centrifuge; designed for aspiration, harvesting, and concentration of autologous adipose tissue. System components include syringes with removable plungers, centrifuge tubes with filters, aspiration/injection cannulas, and skin puncture needles. Operated by clinicians in surgical settings; adipose tissue is harvested using a separate lipoplasty system, then processed in the AdiPrep pack via centrifugation (approx. 4 minutes at 1250 g-force) to concentrate the tissue. The resulting concentrated autologous adipose product is then transferred by the clinician for the intended surgical application. The device facilitates tissue concentration, potentially improving graft quality for reconstructive or surgical procedures.
Clinical Evidence
Bench testing only. Comparative study verified that the output of the AdiPrep system (nucleated cell count and viability) is substantially equivalent to the Viafill and Vortech predicate devices. Results indicated functional equivalence within the intended use.
Technological Characteristics
Components include medical-grade plastic syringes, centrifuge tubes with filters, and stainless steel cannulas. Sterilized via Ethylene-Oxide (EtO) gas. Operates via centrifugation (1250 g-force for 4 minutes) using the SmartPReP2 laboratory centrifuge. Designed for 5 mL to 25 mL fill volumes. Biocompatibility confirmed per ISO 10993.
Indications for Use
Indicated for patients undergoing medical procedures requiring harvesting and transfer of autologous adipose tissue; used for concentrating adipose tissue harvested via legally marketed lipoplasty systems across various surgical specialties including neurosurgery, gastrointestinal, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, laparoscopic, and arthroscopic surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}------------------------------------------------
K121005 page 1/4
# 510(k) SUMMARY
| 510(k) Summary of Safety and Effectiveness | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Summary Prepared: | September 18, 2012 |
| Applicant/Sponsor: | Harvest Technologies Corporation<br>40 Grissom Road, Suite 100<br>Plymouth, MA 02360 |
| Phone Number: | 508-732-7541 |
| Fax Number: | 508-732-0400 |
| Contact Person: | Susan Finneran - Director Regulatory Affairs |
| Device Trade Name: | Harvest AdiPrep™ Adipose Transfer System |
| Common name: | Fat Concentration System |
| Classification Name: | Lipoplasty System, Class II<br>(21 CFR 878.5040, Suction Lipoplasty System,<br>Product Code MUU) |
| Legally Marketed Devices<br>to Which Substantial<br>Equivalence Is Claimed: | K081848, Viafill System, Lipose Corporation and,<br>K100114, Vortech Adipose Transfer System, Biomet Biologics,<br>Inc. |
| Device Description: | The Harvest AdiPrep Adipose Transfer System consists of a<br>disposable process pack to be utilized with the SmartPReP2<br>centrifuge. The process pack is a sterile (EtO) single<br>use disposable pack used for the aspiration, harvesting, and<br>transferring of autologous adipose tissue. It is intended for the<br>concentration of aspirated adipose tissue for subsequent transfer<br>during the same procedure. |
| Indications for Use: | The AdiPrep™ Adipose Transfer System is used in medical<br>procedures involving the harvesting and transferring of autologous<br>adipose tissue. The AdiPrep system is used for concentrating<br>adipose tissue harvested with a legally marketed lipoplasty system.<br>The AdiPrep™ Adipose Transfer System is intended for use in the<br>following surgical specialties when the concentration of harvested<br>adipose tissue is desired.<br>• Neurosurgery<br>• Gastrointestinal Surgery<br>• Urological Surgery<br>• Plastic and Reconstructive Surgery<br>• General surgery |
- Orthopedic Surgery .
{2}------------------------------------------------
- . Gynecological Surgery
- Thoracic Surgery .
- Laparoscopic Surgery .
- Arthroscopic Surgery ●
#### Substantial Equivalence:
The device system has been cleared by the FDA by the 510(k) Premarket Notification process. This submission describes this new adipose concentration system as compared to the predicate devices.
### Technological Characteristics:
Based on the intended use, design, materials, and technological characteristics presented in this premarket notification as summarized in Table 1 below, the Harvest AdiPrep Adipose Transfer System has been shown to be substantially equivalent to the currently marketed predicate devices. Biocompatibility testing was conducted as per ISO 10993 to demonstrate that the materials used in the manufacture of the the AdiPrep Adipose Transfer System are suitable for the intended use.
Clinical Testing:
Clinical testing was performed to determine substantial equivalence to the predicate device. The testing included verification that the output of the AdiPrep Adipose Transfer System is substantially equivalent to the predicate Viafill System (Lipose Corp.) and the Vortech Adipose Transfer System (Biomet Biologics). The results indicated the devices were functional within their intended use. A summary of the results appears in Table 1 below.
{3}------------------------------------------------
K12.1005 Page 3/4
| | DEVICES | | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FEATURES | AdiPrep Adipose Transfer<br>System<br>(This submission) | Lipose Fat Transfer System,<br>K081848, (Predicate device) | Biomet Vortech™ Adipose Transfer<br>System (VATS), K100114, (Predicate<br>device) |
| Indications for<br>Use (IFU) | The AdiPrep™ Adipose<br>Transfer System is used in<br>medical procedures involving<br>the harvesting and transferring<br>of autologous adipose tissue.<br>The AdiPrep system is used for<br>concentrating adipose tissue<br>harvested with a legally<br>marketed lipoplasty system.<br>The AdiPrep™ Adipose<br>Transfer System is intended for<br>use in the following surgical<br>specialties when the<br>concentration of harvested<br>adipose tissue is desired.<br>Neurosurgery, Gastrointestinal<br>Surgery, Urological Surgery,<br>Plastic and Reconstructive<br>Surgery, General Surgery,<br>Orthopedic Surgery,<br>Gynecological Surgery,<br>Thoracic Surgery, Laparoscopic<br>Surgery, Arthroscopic Surgery | The Lipose Fat Transfer<br>System is intended to be used<br>in the aspiration, harvesting,<br>and reinjecting of autologous<br>fat. | The Vortech™ Adipose Transfer System<br>(VATS) is used in medical procedures<br>involving the harvesting and transferring of<br>autologous fat tissue. The VATS System is<br>used for concentrating fat harvested with a<br>legally marketed lipoplasty system. The<br>VATS System is intended for use in the<br>following surgical specialties when the<br>concentration of adipose tissue is desired.<br>Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecological Surgery Thoracic Surgery Laparoscopic Surgery |
| FEATURES | DEVICES | | |
| | AdiPrep Adipose Transfer<br>System<br>(This submission) | Lipose Fat Transfer System,<br>K081848, (Predicate device) | Biomet Vortech™ Adipose Transfer<br>System (VATS), K100114, (Predicate<br>device) |
| Processing Pack:<br>-Materials | Components-syringes with<br>removable plunger, centrifuge<br>tubes with filter, aspiration &<br>fat injection cannula, fat<br>injection syringes, skin<br>puncture needles, and oil<br>extraction syringe & needle.<br>Cannula & syringes composed<br>of medical grade plastics. | Components-syringes with<br>removable plunger, centrifuge<br>tubes with filter, aspiration &<br>fat injection cannula, fat<br>injection syringes, skin<br>puncture needles, and oil<br>extraction syringe & needle. | The materials used in the Fat Concentrator,<br>syringes, and accessories consist of medical<br>grade polymers, elastomers, and stainless steel<br>suitable for use in medical devices. All<br>components are packaged, labeled and<br>sterilized as indicated by the manufacturer<br>labeling. |
| -Fill volumes<br>(mL) | Volume = 5 mL to 25 mL | Volume = 5 mL to 20 mL | Volume = 180 mL |
| -Sterilization<br>Method | Ethylene-Oxide Gas (EtO) | Gamma Radiation | Gamma Radiation |
| Autologous<br>Adipose<br>Product*<br>Nucleated Cell<br>Count (x10³/ml of<br>Product) | Mean 1.823<br>Mean 83.5 | Mean 1.848<br>Mean 82.4 | Actual results are not available, however, in<br>K100114, Biomet reported that testing "by<br>direct comparison included verification that<br>the output of the Vortech™ Adipose Transfer<br>System (VATS) is substantially equivalent to<br>the Viafill™ System (Lipose Corp.), the<br>predicate device. Test results for both percent |
| -Nucleated Cell<br>Viability (%) | | | volume reduction and percent cell viability<br>show that the VATS System is substantially<br>equivalent to the Viafill System." |
| Processing<br>Capabilities | | | |
| Volume | Volume = 5 mL to 25 mL | Volume = 5 mL to 20 mL | Volume = 180 mL |
| Low-g-force | Centrifuge to spin for<br>approximately 4 minutes at<br>1250 g-force. | Centrifuge to spin for 2 minutes<br>at 50 g-force. | 4 ½ minute centrifugation process , g-force<br>unknown. |
| Laboratory<br>Centrifuge | SmartPReP2 Centrifuge is a<br>general purpose laboratory<br>centrifuge for clinical use. | Viafill™ System centrifuge is a<br>general purpose laboratory<br>centrifuge. | The VATS reusable portable tabletop base<br>unit is similar to a general purpose laboratory<br>centrifuge. |
# Table 1: SUBSTANTIAL EQUIVALENCE COMPARISON CHART
-
b
c
.
1
.
.
.
.
.
.
.
{4}------------------------------------------------
K121005/Page 4/4
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 2 1 2012
Harvest Technologies Corporation % Mr. Gabriel J. Muraca, Jr. Regulatory Affairs Manager 40 Grissom Road, Suite 100 Plymouth, Massachusetts 02360
Re: K121005
Trade/Device Name: AdiPrep™ Adipose Transfer System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: August 29, 2012 Received: August 30, 2012
Dear Mr. Muraca:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
### Page 2 - Mr. Gabriel J. Muraca, Jr.
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/dafault.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melker Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use
## 510(k) Number (if known): K121005
## Device Name: AdiPrep™ Adipose Transfer System
Indications for Use:
The AdiPrepTM Adipose Transfer System is used in medical procedures involving the harvesting and transferring of autologous adipose tissue. The AdiPrep system is used for concentrating adipose tissue harvested with a legally marketed lipoplasty system. The AdiPrep™ Adipose Transfer System is intended for use in the following surgical specialties when the concentration of harvested adipose tissue is desired.
- Neurosurgery .
- Gastrointestinal Surgery t
- Urological Surgery .
- · Plastic and Reconstructive Surgery
- General surgery ●
- . Orthopedic Surgery
- Gynecological Surgery .
- Thoracic Surgery .
- Laparoscopic Surgery
- Arthroscopic Surgery
Prescription Use (Part 21 CFR 801 Subpart D)
#### AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krene for MM
(Division Sign-Off)
Page of
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121005
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.