MateLaser Medical Laser Therapy Systems (ML-HIL-30, ML-HIL-60)
K250195 · Matelaser, Inc. · GEX · May 30, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250195
Device Name
MateLaser Medical Laser Therapy Systems (ML-HIL-30, ML-HIL-60)
Applicant
Matelaser, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
810 nm and 980 nm wavelength: MateLaser Medical Laser Therapy Systems (Model: ML-HIL-30/60) emit energy in the invisible and near infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and the temporary relaxation of muscle. 980 nm wavelength: MateLaser Medical Laser Therapy Systems (Model: ML-HIL-60) are also indicated for use in surgical applications requiring hemostasis, ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including aesthetic (dermatology and plastic surgery), gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, thoracic surgery; and Laser Assisted Lipolysis. 980 nm wavelength: MateLaser Medical Laser Therapy Systems (Model: ML-HIL-60) are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and Trichophyton mentagrophytes, and/or yeasts Candida albicans, etc.)
Device Story
MateLaser Medical Laser Therapy Systems (ML-HIL-30, ML-HIL-60) utilize 810 nm and 980 nm laser energy to elevate tissue temperature. Device operates by emitting near-infrared energy to provide topical heating for pain relief, muscle relaxation, and increased circulation. ML-HIL-60 model additionally supports surgical soft tissue procedures (ablation, coagulation, incision) and treatment for onychomycosis. Used in clinical settings by healthcare professionals. Output consists of controlled laser energy delivery to target tissue. Benefits include non-invasive pain management, surgical precision in soft tissue procedures, and clearing of nail fungal infections.
Technological Characteristics
Laser therapy system utilizing 810 nm and 980 nm wavelengths. Energy source: laser diode. Form factor: medical laser console (ML-HIL-30/60).
Indications for Use
Indicated for patients requiring topical heating for relief of minor muscle/joint pain, stiffness, arthritis, muscle spasms, or increased local blood circulation. ML-HIL-60 indicated for soft tissue surgical applications (hemostasis, ablation, excision, etc.) across multiple medical specialties and for temporary increase of clear nail in patients with onychomycosis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Related Devices
K123014 — LITECURE THERAPY SYSTEM MODEL LTS-1500 · Litecure, LLC · Mar 29, 2013
K190034 — REMY Medical Therapy Laser System · Footdocprenur, LLC · May 15, 2019
K253965 — CureLight Medical Diode Laser Systems (CureLight F2-A15; CureLight F2-B15; CureLight F2-A30; CureLight F2-B30; CureLight F3-AB30; CureLight F3-AB60.) · Wuhan PHOMED Technology Company , Ltd. · Mar 19, 2026
K231997 — SIRONA 100 Therapy Laser System (OEM's Model EVANLAS) · Cure Point Therapeutics, Inc. · Oct 15, 2024
K242755 — Medical Diode Laser Systems (VELAS II-30B, VELAS II-30A, GBOX-15AB, GBOX-20B, CHEESE II-10B) · Gigaalaser Company , Ltd. · May 19, 2025
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
MateLaser, Inc.
Ruth Garcia
Quality Manager
3257 Nw 7th Ave Cir
Miami, Florida 33127
Re: K250195
Trade/Device Name: MateLaser Medical Laser Therapy Systems (ML-HIL-30, ML-HIL-60)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX, ILY, PDZ
Dated: May 29, 2025
Received: May 29, 2025
Dear Ruth Garcia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250195 - Ruth Garcia
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250195 - Ruth Garcia
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.05.30
15:45:30 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250195
Device Name
MateLaser Medical Laser Therapy Systems (ML-HIL-30, ML-HIL-60)
Indications for Use (Describe)
810 nm and 980 nm wavelength:
MateLaser Medical Laser Therapy Systems (Model: ML-HIL-30/60) emit energy in the invisible and near infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and the temporary relaxation of muscle.
980 nm wavelength:
MateLaser Medical Laser Therapy Systems (Model: ML-HIL-60) are also indicated for use in surgical applications requiring hemostasis, ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including aesthetic (dermatology and plastic surgery), gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, thoracic surgery; and Laser Assisted Lipolysis.
980 nm wavelength:
MateLaser Medical Laser Therapy Systems (Model: ML-HIL-60) are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and Trichophyton mentagrophytes, and/or yeasts Candida albicans, etc.)
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.