CureLight Medical Diode Laser Systems (CureLight F2-A15; CureLight F2-B15; CureLight F2-A30; CureLight F2-B30; CureLight F3-AB30; CureLight F3-AB60.)
K253965 · Wuhan PHOMED Technology Company , Ltd. · ILY · Mar 19, 2026 · Physical Medicine
Device Facts
Record ID
K253965
Device Name
CureLight Medical Diode Laser Systems (CureLight F2-A15; CureLight F2-B15; CureLight F2-A30; CureLight F2-B30; CureLight F3-AB30; CureLight F3-AB60.)
Applicant
Wuhan PHOMED Technology Company , Ltd.
Product Code
ILY · Physical Medicine
Decision Date
Mar 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The CureLight Medical Diode Laser Systems, include model: CureLight F2-A15 and CureLight F2-A30, are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation. The CureLight Medical Diode Laser Systems, model: CureLight F2-B15, CureLight F2-B30, CureLight F3-AB30 and CureLight F3-AB60 are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation. The CureLight Medical Diode Laser Systems, model: CureLight F2-B15, CureLight F2-B30, CureLight F3-AB30 and CureLight F3-AB60 are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Device Story
Device uses GaAlAs diode laser (810/980nm) to emit infrared energy for topical heating; fiber-coupled output. System comprises main unit, footswitch, and handpiece accessories. Operated by clinician via color LCD touch screen. MCU controls drive current, temperature, and laser output. Used in clinical settings for pain relief, muscle relaxation, increased circulation, and onychomycosis treatment. Output affects tissue temperature; healthcare provider monitors treatment via interface. Benefits include temporary pain relief and increased clear nail growth.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), and performance testing (IEC 60601-2-22).
Technological Characteristics
GaAlAs diode laser; 810nm/980nm wavelengths; Class 4 laser. Fiber-coupled transmission (400μm). Air-cooled. IPX1 main unit, IPX8 footswitch. MCU-controlled. Power supply 110-230V. Class I Type B safety classification.
Indications for Use
Indicated for patients with minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis; also indicated for temporary increase of clear nail in patients with onychomycosis (e.g., Trichophyton rubrum, T. mentagrophytes, Candida albicans).
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
K231997 — SIRONA 100 Therapy Laser System (OEM's Model EVANLAS) · Cure Point Therapeutics, Inc. · Oct 15, 2024
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 19, 2026
Wuhan PHOMED Technology Company, Ltd.
Huang Bill
General manager
Rm. 02, 3f, Bldg. 2, Gezhouba Sun City, # 40 Gaoxin
4th Rd., E. Lake New Technology Development Zone
Wuhan, Hubei 430200
China
Re: K253965
Trade/Device Name: CureLight Medical Diode Laser Systems (CureLight F2-A15; CureLight F2-B15; CureLight F2-A30; CureLight F2-B30; CureLight F3-AB30; CureLight F3-AB60.)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: PDZ, ILY
Dated: December 11, 2025
Received: December 11, 2025
Dear Huang Bill:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253965 - Huang Bill
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253965 - Huang Bill
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU -S
Digitally signed by YAN FU -S
Date: 2026.03.19 12:19:02
04'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253965 | ? |
| Please provide the device trade name(s). | | ? |
| CureLight Medical Diode Laser Systems (CureLight F2-A15;
CureLight F2-B15;
CureLight F2-A30;
CureLight F2-B30;
CureLight F3-AB30;
CureLight F3-AB60.) | | |
| Please provide your Indications for Use below. | | ? |
| (1) The CureLight Medical Diode Laser Systems, include model: CureLight F2-A15 and CureLight F2-A30,
are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating
tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness
associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local
blood circulation.
(2) The CureLight Medical Diode Laser Systems, model: CureLight F2-B15, CureLight F2-B30, CureLight
F3-AB30 and CureLight F3-AB60 are intended to emit energy in the infrared spectrum to provide topical
heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain,
muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and
to temporarily increase local blood circulation.
The CureLight Medical Diode Laser Systems, model: CureLight F2-B15, CureLight F2-B30, CureLight F3-
AB30 and CureLight F3-AB60 are also indicated for use for the temporary increase of clear nail in patients
with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts
Candida albicans, etc.). | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# K253965 510(k) Summary
## I SUBMITTER
Wuhan PHOMED Technology Co., Ltd.
Address: Room 02, 3F, Building 2, Gezhouba Sun City, No. 40 Gaoxin 4th Road, East Lake New Technology Development Zone, Wuhan, China.
Contact Person: Bill Huang
Telephone: +86-18607117569
Fax: +86-27-86750788
Email: bill.huang@phomedlaser.com
Date Prepared: December 10, 2025
## II DEVICE
Name of Device: CureLight Medical Diode Laser Systems
Model: CureLight F2-A15, CureLight F2-B15, CureLight F2-A30, CureLight F2-B30, CureLight F3-AB30, CureLight F3-AB60
Regulation number: 21 CFR 890.5500
21 CFR 878.4810
Regulation name: Infrared lamp
Laser surgical instrument for use in general and plastic surgery and in dermatology
Classification Name: Lamp, Infrared, Therapeutic Heating
Lasers For Temporary Increase Of Clear Nail In Patients With Onychomycosis
Regulatory class: Class II
Product code: ILY, PDZ
## III PREDICATE DEVICE
Primary Predicate Device
Trade name: Medical diode laser systems
Classification name: Lamp, Infrared, Therapeutic Heating
{5}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | Lasers For Temporary Increase Of Clear Nail In Patients With Onychomycosis |
| --- | --- |
| Regulation number: | 21 CFR 890.5500 |
| | 21 CFR 878.4810 |
| Regulation name: | Infrared lamp |
| | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Regulatory class: | Class II |
| Product code: | ILY, PDZ |
| Submitter: | Gigaalaser Company Ltd. |
| 510(k) number: | K242755 |
| Co-predicate Device | |
| Trade name: | Diowave Laser System |
| Classification name: | Lamp, Infrared, Therapeutic Heating |
| Regulation number: | 21 CFR 890.5500 |
| Regulation name: | Infrared lamp |
| Regulatory class: | Class II |
| Product code: | ILY |
| Submitter: | TECHNOLOGICAL MEDICAL ADVANCMENTS, INC. (TMA) |
| 510(k) number: | K121363 |
{6}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# IV. DEVICE DESCRIPTION
The "CureLight Medical Diode Laser Systems", include model: CureLight F2-A15, CureLight F2-A30, CureLight F2-B15, CureLight F2-B30, CureLight F3-AB30 and CureLight F3-AB60. Which consist of three main components:
(1) Main device;
(2) Foot switch;
(3) Accessories.
Diode laser is a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.
The diode laser for this unit is GaAlAs diode bar, and the wavelength is 810/980nm. It features impact structure, high efficiency and long lifetime. Generally, the beam shall be emitted as the big beam divergence of the laser from the diode. With the PHOMED'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber.

MCU (Micro Controller Unit) is the control center of system. MCU controls the drive board and temperature system by changing operating current. The operating power supply is the power source of the whole system. It supplies power for MCU, diode laser module by DC-DC conversion module and temp system by temp-control circuit. MCU controls the drive current of diode laser module by adjusting the DC-DC module and current-control circuit. The MCU
{7}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
has three external input sources: footswitch, remote interlock connector and information input from the touch units, it has one external output unit: display terminal. The diode laser module provides laser power output by fiber with optical fiber coupling system.
# V. INDICATIONS FOR USE
(1) The CureLight Medical Diode Laser Systems, include model: CureLight F2-A15 and CureLight F2-A30, are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation.
(2) The CureLight Medical Diode Laser Systems, model: CureLight F2-B15, CureLight F2-B30, CureLight F3-AB30 and CureLight F3-AB60 are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation.
The CureLight Medical Diode Laser Systems, model: CureLight F2-B15, CureLight F2-B30, CureLight F3-AB30 and CureLight F3-AB60 are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
4 / 47
{8}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Table 1 Substantial equivalence discussion (CureLight Medical diode laser systems, Model: CureLight F2-A15)
| Item | Proposed device | Primary predicate device | Discussion |
| --- | --- | --- | --- |
| 510k Number | / | K242755 | / |
| Proprietary Name | CureLight Medical diode laser systems | Medical diode laser systems | / |
| Model | CureLight F2-A15 | GBOX-15AB | / |
| Product Code | ILY | ILY | Identical |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | Identical |
| Classification | Class II | Class II | Identical |
| Indication for Use | The CureLight Medical Diode Laser Systems, model: CureLight F2-A15 is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation. | The Medical Diode Laser Systems, model: GBOX-15AB is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to | Identical |
{9}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | | temporarily increase local blood circulation. | |
| --- | --- | --- | --- |
| Laser type | diode laser | diode laser | Identical |
| Wavelength | 810nm±20nm | 810 ± 10nm; 980 ± 10nm; | Different Comment 1 |
| Output Power | 0-15W (±20%) | 1W-15W(±10%) | SE |
| Energy Density(ED) / (Fluence) | ≤1.55 J/cm² (Per second) | 0.05~1.55 J/cm2 (Per second) | SE |
| Power Density | ≤1.55 W/cm² | 0.05~1.55 W/cm2 | SE |
| Spot Diameter | 4~5 cm | 3.5~5 cm | Different Comment 2 |
| Working Distance | 0 cm~2 cm for therapy handpiece 1
2 cm~3 cm for therapy handpiece 2 | 3~5 cm | Different Comment 3 |
| Treatment Angle | Perpendicular to the treatment area | Perpendicular to the treatment area | Identical |
| Applied Part Moving Speed | 3~5 cm/s | 3~5 cm/s | Identical |
| Operation Mode | CW, repeat pulse | CW, single pulse, repeat pulse | Different Comment 4 |
| Pulse Width | 50μs-1s | 25μs-10s | Different Comment 5 |
| Pulse Repetition | 1Hz-10KHz | 1Hz-20KHz | Different |
6 / 47
{10}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Rate | | | Comment 6 |
| --- | --- | --- | --- |
| Treatment Methods | Contact, Non-contact. | Non-contact. | Different Comment 7 |
| Applied Part | Handpiece | Handpiece | Identical |
| Transmission System | 400μm fibers with metal protective armors. | fibers of 600μm with SMA905 connector. | Different Comment 8 |
| Aiming Beam | Diode laser of 635nm(±20nm), power <5mW, adjustable brightness. | Diode laser of 650nm, power < 5mW, adjustable brightness. | SE |
| Operation Interface | Color LCD touch screen | Color LCD touch screen | Identical |
| Power Supply | ~110-230V, 50-60Hz, 200VA | 100-240VAC, 50-60Hz, 350VA | SE |
| Laser Class | 4 | 4 | Identical |
| Safety Classification | Class I Type B | Class I Type B | Identical |
| Cooling | Air | Air | Identical |
| Waterproof Level | IPX1 | IPX1 | Identical |
| Footswitch Waterproof Level | IPX8 | IPX8 | Identical |
Discussion:
{11}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
Clinical:
Indication for Use claimed by proposed device is the same with the predicate device K242755 (GBOX-15AB).
Technology:
The subject device has the same output power of diode laser to achieve its intended use.
The energy density (ED) / (fluence), power density, treatment angle, applied part, moving speed, treatment methods and etc. are either identical to the predicate. The main differences are wavelength, spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods and transmission system.
Comment 1 Wavelength:
The wavelength of the proposed device: 810nm±20nm, which is different to the predicate device, K242755 (GBOX-15AB): 810 ± 10nm; 980 ± 10nm, but the wavelength of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
Comment 2 Spot Diameter:
The spot diameter of the proposed device: 4~5 cm, which is different to the predicate device, K242755 (GBOX-15AB): 3.5~5 cm, but the spot diameter of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
Comment 3 Working Distance:
The working distance of the proposed device: 0 cm~2 cm for therapy handpiece 1; 2 cm~3 cm for therapy handpiece 2, which is different to the predicate device, K242755: 3~5 cm.
The working distance of proposed device is shorter because the therapy handpiece head has a length of 3cm. Therefore, the actual distance from the laser light source at the end of the laser output port to the treatment end is 3 cm~5 cm for therapy handpiece 1
8 / 47
{12}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
which is identical with the predicate device, K242755 (GBOX-15AB): 3~5 cm; 5 cm~6 cm for therapy handpiece 2, a little longer: 1 cm than the working distance of predicate device, K242755 (GBOX-15AB): 3~5 cm, but the energy density (ED) / (fluence) and power density are all within the scope of what predicate device claims and it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
## Comment 4 Operation mode:
The operation mode of the proposed device: CW, repeat pulse is different to the predicate device, K242755 (GBOX-15AB): CW, single pulse, repeat pulse. but the operation mode of the proposed device is within the scope of what predicate device, K242755 (GBOX-15AB) claims. This difference does not affect the safety and effectiveness of the device.
## Comment 5 Pulse width:
The pulse width of the proposed device: 50μs-1s, which is different to the predicate device, K242755 (GBOX-15AB): 25μs-10s, but the pulse width of the proposed device is within the scope of what predicate device, K242755 (GBOX-15AB) claims. This difference does not affect the safety and effectiveness of the device.
## Comment 6 Pulse repetition rate:
The pulse repetition rate of the subject device: 1Hz-10KHz, which is different to the predicate device, K242755 (GBOX-15AB): 1Hz-20KHz, but the pulse repetition rate of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
## Comment 7 Treatment Methods:
The treatment methods of the subject device (Contact, Non-contact), which is different to the predicate device (Non-contact), but it meets the clinical needs. Contact mode is convenient for doctors to operate, which is more labor-saving. Moreover, the therapy handpiece head of therapy handpiece 1 has been evaluated and tested for biocompatibility as required, and the test results meet the requirements. This difference does not affect the safety and effectiveness of the device.
9 / 47
{13}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# Comment 8 Transmission system
The transmission system of the subject device (400μm fibers with metal protective armors), which is different to the predicate device (fibers of 600μm with SMA905 connector).
The transmission system only transmits laser energy. Although the optical fiber core diameter of optical fiber transmission system is different, but it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
# Conclusion:
The indication for use of the subject device claimed by proposed device is the same with the predicate device.
The proposed device uses same diode lasers technology as that used by the predicates. The core output parameters of the proposed device are identical to the predicates. The differences only exist in such contents: spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods and transmission system, that all can be controlled in range of application. These differences do not raise new types of questions regarding the safety and effectiveness when the device is used for the proposed indications for use.
So, the proposed devices are Substantially Equivalent (SE) to existing legally marketed predicate devices.
10 / 47
{14}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
Table 2 Substantial equivalence discussion (CureLight Medical diode laser systems, Model: CureLight F2-A30)
| Item | Proposed device | Primary predicate device | Co-predicate device | Discussion |
| --- | --- | --- | --- | --- |
| 510k Number | / | K242755 | K121363 | / |
| Proprietary Name | CureLight Medical diode laser systems | Medical diode laser systems | Diowave Laser System | / |
| Model | CureLight F2-A30 | VELAS II -30A | Diowave 30W | / |
| Product Code | ILY | ILY | ILY | Identical |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indication for Use | The CureLight Medical Diode Laser Systems, model: CureLight F2-A30 is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase | The Medical Diode Laser Systems, model: VELAS II -30A is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase | The Diowave Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating, tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to | Identical |
{15}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | local blood circulation. | local blood circulation. | and to temporarily increase local blood circulation. | |
| --- | --- | --- | --- | --- |
| Laser type | diode laser | diode laser | diode laser | Identical |
| Wavelength | 810nm±20nm | 810 ± 10nm | 810nm±10nm, 980nm±10nm | SE |
| Output Power | 0-30W (±20%) | 1W-30W(±20%) | N/A | SE |
| Energy Density(ED) / (Fluence) | ≤3.1 J/cm² (Per second) | 0.05~3.1 J/cm² (Per second) | N/A | SE |
| Power Density | ≤3.1 W/cm² | 0.05~3.1 W/cm² | N/A | SE |
| Spot Diameter | 4~5 cm | 3.5~5 cm | N/A | Different Comment 1 |
| Working Distance | 0 cm~2 cm for therapy handpiece 1
2 cm~3 cm for therapy handpiece 2 | 3~5 cm | N/A | Different Comment 2 |
| Treatment Angle | Perpendicular to the treatment area | Perpendicular to the treatment area | N/A | Identical |
| Applied Part Moving Speed | 3~5 cm/s | 3~5 cm/s | N/A | Identical |
| Operation Mode | CW, repeat pulse | CW, single pulse, repeat pulse | N/A | Different Comment 3 |
| Pulse Width | 50μs-1s | 10ms-1s | N/A | Different Comment 4 |
12 / 47
{16}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Pulse Repetition Rate | 1Hz-10KHz | 1Hz-50Hz | N/A | Different Comment 5 |
| --- | --- | --- | --- | --- |
| Treatment Methods | Contact, Non-contact. | Non-contact. | Non-contact | Different Comment 6 |
| Applied Part | Handpiece | Handpiece | Handpiece | Identical |
| Transmission System | 400μm fibers with metal protective armors. | fibers of 600μm with SMA905 connector. | fibers of 400μm, 600μm and 1000μm with SMA905 connector | Different Comment 7 |
| Aiming Beam | Diode laser of 635nm(±20nm), power < 5mW, adjustable brightness. | Diode laser of 650nm, power≤2mW, adjustable brightness. | Diode laser of 650nm power max. < 5mW, adjustable brightness. | SE |
| Operation Interface | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Identical |
| Power Supply | ~110-230V, 50-60Hz, 200VA | 100-240VAC, 50-60Hz, 350VA | 100-240VAC, 50/60Hz, 350VA | SE |
| Laser Class | 4 | 4 | 4 | Identical |
| Safety Classification | Class I Type B | Class I Type B | Class I Type B | Identical |
| Cooling | Air | Air | Air | Identical |
| Waterproof Level | IPX1 | IPX1 | IPX1 | Identical |
13 / 47
{17}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Footswitch
Waterproof
Level | IPX8 | IPX8 | IPX8 | Identical |
| --- | --- | --- | --- | --- |
## Discussion:
### Clinical:
Indication for use claimed by proposed device is the same with the predicate device K242755 (VELAS II -30A) and K121363.
### Technology:
The subject device has the same wavelengths and output power of diode laser to achieve its intended use.
The energy density (ED) / (fluence), power density, treatment angle, applied part, moving speed, treatment methods and etc. are either identical to the predicate. The main differences are spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods and transmission system.
## Comment 1 Spot Diameter:
The spot diameter of the proposed device: 4~5 cm, which is different to the predicate device, K242755 (VELAS II -30A): 3.5~5 cm, but the spot diameter of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
## Comment 2 Working Distance:
The working distance of the proposed device: 0 cm~2 cm for therapy handpiece 1; 2 cm~3 cm for therapy handpiece 2, which is different to the predicate device, K242755 (VELAS II -30A): 3~5 cm.
The working distance of proposed device is shorter because the therapy handpiece head has a length of 3cm. Therefore, the actual distance from the laser light source at the end of the laser output port to the treatment end is 3 cm~5 cm for therapy handpiece 1
14 / 47
{18}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
which is identical with the predicate device, K242755 (VELAS II -30A): 3~5 cm; 5 cm~6 cm for therapy handpiece 2, a little longer: 1 cm than the working distance of predicate device, K242755 (VELAS II -30A): 3~5 cm, but the energy density (ED) / (fluence) and power density are all within the scope of what predicate device, K242755 (VELAS II -30A) claims and it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
## Comment 3 Operation mode:
The operation mode of the proposed device: CW, repeat pulse is different to the predicate device, K242755 (VELAS II -30A): CW, single pulse, repeat pulse. but the operation mode of the proposed device is within the scope of what predicate device, K242755 (VELAS II -30A) claims. This difference does not affect the safety and effectiveness of the device.
## Comment 4 Pulse width:
The pulse width of the proposed device: 50 μs-1 s, which is different to the predicate device, K242755 (VELAS II -30A): 10 ms-1 s.
The difference in pulse width range (specifically the extension to 50 μs) is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30A). The expanded parameter range operates within the same thermal safety margins and is justified by the overlapping therapeutic effect of elevating tissue temperature.
## Comment 5 Pulse repetition rate:
The pulse repetition rate of the subject device: 1 Hz-10 KHz, which is different to the predicate device, K242755 (VELAS II -30A): 1 Hz-50 Hz.
The difference in pulse repetition rate: 1 Hz-10 KHz and 1 Hz-50 Hz is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30A).
15 / 47
{19}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# Comment 6 Treatment Methods:
The treatment methods of the subject device (Contact, Non-contact), which is different to the predicate device (Non-contact), but it meets the clinical needs. Contact mode is convenient for doctors to operate, which is more labor-saving. Moreover, the therapy handpiece head of therapy handpiece 1 has been evaluated and tested for biocompatibility as required, and the test results meet the requirements. This difference does not affect the safety and effectiveness of the device.
# Comment 7 Transmission system
The transmission system of the subject device (400μm fibers with metal protective armors), which is different to the predicate device (fibers of 600μm with SMA905 connector).
The transmission system only transmits laser energy. Although the optical fiber core diameter of optical fiber transmission system is different, but it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
# Conclusion:
The indication for use of the subject device claimed by proposed device is the same with the predicate device.
The proposed device uses same diode lasers technology as that used by the predicates. The core output parameters of the proposed device are identical to the predicates. The differences only exist in such contents: spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods and transmission system, that all can be controlled in range of application. These differences do not raise new types of questions regarding the safety and effectiveness when the device is used for the proposed indications for use.
So, the proposed devices are Substantially Equivalent (SE) to existing legally marketed predicate devices.
16 / 47
{20}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
Table 3 Substantial equivalence discussion (CureLight Medical diode laser systems, Model: CureLight F2-B15)
| Item | Proposed device | Primary predicate device | Co-predicate device | Discussion |
| --- | --- | --- | --- | --- |
| 510k Number | / | K242755 | K121363 | / |
| Proprietary Name | CureLight Medical diode laser systems | Medical diode laser systems | Diowave Laser System | / |
| Model | CureLight F2-B15 | VELAS II -30B | Diowave 30W | / |
| Product Code | ILY, PDZ | ILY, PDZ | ILY | Identical |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500, 21 CFR 878.4810 | 21 CFR 890.5500 | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indication for Use | The CureLight Medical Diode Laser Systems, model: CureLight F2-B15 are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase | The Medical Diode Laser Systems, model: VELAS II -30B is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily increase | The Diowave Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating, tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, | SE |
{21}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | local blood circulation. The CureLight Medical Diode Laser Systems, model: CureLight F2-B15 are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | local blood circulation. The Medical Diode Laser Systems, model: VELAS II -30B is also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | and to temporarily increase local blood circulation. | |
| --- | --- | --- | --- | --- |
| Laser type | diode laser | diode laser | diode laser | Identical |
| Wavelength | 980nm±20nm | 980 ± 10nm | 810nm±10nm, 980nm±10nm | SE |
| Output Power | Physical therapy mode: 0-15W (±20%) General mode: 0-10W (±20%) | Physical therapy mode: 1W-30W (±20%) General mode: 1W-10W (±20%), 980nm only | N/A | Different Comment 1 |
| Energy Density(ED) / (Fluence) | Physical therapy mode: ≤1.55 J/cm2 (Per second) General mode: ≤39.8 J/cm2 (Per second) | Physical therapy mode: 0.05~3.1 J/cm2 (Per second) General mode: 1.7~39.8 J/cm2 (Per second) | N/A | Different Comment 2 |
| Power Density | Physical therapy mode: ≤1.55 W/cm2 General mode: ≤39.8 W/cm2 | Physical therapy mode: 0.05~3.1 W/cm2 General mode: 1.7~39.8 W/cm2 | N/A | Different Comment 3 |
{22}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Spot Diameter | Physical therapy mode: 4~5 cm
General mode: 5~6 mm | Physical therapy mode: 3.5~5 cm
General mode: 4~5 mm | N/A | Different Comment 4 |
| --- | --- | --- | --- | --- |
| Working Distance | Physical therapy mode: 0 cm~2 cm for therapy handpiece 1
2 cm~3 cm for therapy handpiece 2
General mode: 3~4 mm | Physical therapy mode: 3~5 cm
General mode: 3~4 mm | N/A | Different Comment 5 |
| Treatment Angle | Perpendicular to the treatment area | Perpendicular to the treatment area | N/A | Identical |
| Applied Part Moving Speed | Physical therapy mode: 3~5 cm/s
General mode: 1~2 cm/s | Physical therapy mode: 3~5 cm/s
General mode: 1~2 cm/s | N/A | Identical |
| Operation Mode | Physical therapy mode: CW, repeat pulse
General mode: repeat pulse | Physical therapy mode: CW, single pulse, repeat pulse
repeat pulse: repeat pulse | N/A | Different Comment 6 |
| Pulse Width | Physical therapy mode: 50μs-1s
General mode: Ton: 10ms
Toff: 10ms, 20ms | Physical therapy mode: 10ms-1s
General mode: Ton: 10ms
Toff: 10ms, 20ms | N/A | Different Comment 7 |
| Pulse Repetition Rate | Physical therapy mode: 1Hz-10KHz
General mode: 33.3Hz/50Hz | General mode: 1Hz-50Hz
General mode: 33.3Hz/50Hz | N/A | Different Comment 8 |
| Treatment Methods | Physical therapy mode: Contact, Non-contact. | Non-contact. | Non-contact | Different Comment 9 |
19 / 47
{23}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | General mode: Non-contact | | | |
| --- | --- | --- | --- | --- |
| Treatment Mode | Physical therapy mode, general mode | Physical therapy mode, general mode | / | Identical |
| Applied Part | Handpiece | Physical therapy mode: Handpiece General mode: Fiber | Handpiece | Different Comment 10 |
| Transmission System | 400μm fibers with metal protective armors. | fibers of 600μm with SMA905 connector. | fibers of 400μm, 600μm and 1000μm with SMA905 connector | Different Comment 11 |
| Aiming Beam | Diode laser of 635nm(±20nm), power < 5mW, adjustable brightness. | Diode laser of 650nm, power≤2mW, adjustable brightness. | Diode laser of 650nm power max. < 5mW, adjustable brightness. | SE |
| Operation Interface | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Identical |
| Power Supply | ~110-230V, 50-60Hz, 200VA | 100-240VAC, 50-60Hz, 350VA | 100-240VAC, 50/60Hz, 350VA | SE |
| Laser Class | 4 | 4 | 4 | Identical |
| Safety Classification | Class I Type B | Class I Type B | Class I Type B | Identical |
| Cooling | Air | Air | Air | Identical |
| Waterproof Level | IPX1 | IPX1 | IPX1 | Identical |
20 / 47
{24}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Footswitch
Waterproof
Level | IPX8 | IPX8 | IPX8 | Identical |
| --- | --- | --- | --- | --- |
## Discussion:
### Clinical:
Indication for use claimed by proposed device is the same with the predicate device K242755 (VELAS II -30B).
### Technology:
The subject device has the same wavelength of diode laser to achieve its intended use.
The treatment angle, moving speed, treatment methods and etc. are either identical to the predicate. The main differences are output power, energy density (ED) / (fluence), power density, spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods, applied part and transmission system.
## Comment 1 Output Power:
### Physical therapy mode:
The output power of the proposed device: 0-15W (±20%), which is different to the predicate device, K242755 (VELAS II -30B): 1W-30W (±20%), but the output power of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
### General mode:
21 / 47
{25}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The output power of the proposed device: 0-10W (±20%) is substantial equivalence with the predicate device, K242755 (VELAS II - 30B): 1W-10W (±20%) claims.
## Comment 2 Energy Density (ED) / (Fluence):
Physical therapy mode:
The energy density (ED) / (fluence) of the proposed device: ≤1.55 J/cm² (Per second), which is different to the predicate device, K242755 (VELAS II -30B): 0.05~3.1 J/cm² (Per second), but the Energy Density (ED) / (Fluence) of the proposed device is within the scope of predicate device, K242755 (VELAS II -30B) claims. This difference does not affect the safety and effectiveness of the device.
General mode:
The energy density (ED) / (fluence) of the proposed device: ≤39.8 J/cm² (Per second) is substantial equivalence with the predicate device, K242755 (VELAS II -30B): 1.7~39.8 J/cm² (Per second) claims.
## Comment 3 Power Density:
Physical therapy mode:
The power density of the proposed device: ≤1.55 W/cm², which is different to the predicate device, K242755 (VELAS II -30B): 0.05~3.1 W/cm², but the power density of the proposed device is within the scope of predicate device, K242755 (VELAS II -30B) claims. This difference does not affect the safety and effectiveness of the device.
General mode:
22 / 47
{26}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The power density of the proposed device: ≤39.8 W/cm² is substantial equivalence with the predicate device, K242755 (VELAS II - 30B): 1.7~39.8 W/cm² claims.
## Comment 4 Spot Diameter:
Physical therapy mode:
The spot diameter of the proposed device: 4~5 cm, which is different to the predicate device, K242755: 3.5~5 cm, but the spot diameter of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
General mode:
The spot diameter of the proposed device: 5~6 mm, which is different to the predicate device, K242755: 4~5 mm, but the energy density (ED) / (fluence) and power density of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
## Comment 5 Working Distance:
Physical therapy mode:
The working distance of the proposed device: 0 cm~2 cm for therapy handpiece 1; 2 cm~3 cm for therapy handpiece 2, which is different to the predicate device, K242755: 3~5 cm.
The working distance of proposed device is shorter because the therapy handpiece head has a length of 3cm. Therefore, the actual distance from the laser light source at the end of the laser output port to the treatment end is 3 cm~5 cm for therapy handpiece 1 which is identical with the predicate device: 3~5 cm; 5 cm~6 cm for therapy handpiece 2, a little longer: 1cm than the working distance of predicate device: 3~5 cm, but the energy density(ED) / (fluence) and power density are all within the scope of what predicate device claims and it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
23 / 47
{27}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
General mode:
The working distance of the proposed device is the same with the predicate device.
## Comment 6 Operation mode:
Physical therapy mode:
The operation mode of the proposed device (CW, repeat pulse) is different to the predicate device (CW, single pulse, repeat pulse). but the operation mode of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
General mode:
The operation mode of the proposed device is the same with the predicate device.
## Comment 7 Pulse width:
Physical therapy mode:
The pulse width of the proposed device: 50μs-1s, which is different to the predicate device, K242755 (VELAS II -30B): 10ms-1s.
The difference in pulse width range (specifically the extension to 50 μs) is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30B). The expanded parameter range operates within the same thermal safety margins and is justified by the overlapping therapeutic effect of elevating tissue temperature.
General mode:
The pulse width of the proposed device is the same with the predicate device.
24 / 47
{28}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# Comment 8 Pulse repetition rate:
Physical therapy mode:
The pulse repetition rate of the subject device: 1Hz-10KHz, which is different to the predicate device, K242755 (VELAS II -30B): 1Hz-50Hz.
The difference in pulse repetition rate: 1Hz-10KHz and 1Hz-50Hz is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30B).
General mode:
The pulse repetition rate of the proposed device is the same with the predicate device.
# Comment 9 Treatment Methods:
Physical therapy mode:
The treatment methods of the subject device (Contact, Non-contact), which is different to the predicate device (Non-contact), but it meets the clinical needs. Contact mode is convenient for doctors to operate, which is more labor-saving. Moreover, the therapy handpiece head of therapy handpiece 1 has been evaluated and tested for biocompatibility as required, and the test results meet the requirements. This difference does not affect the safety and effectiveness of the device.
General mode:
The treatment methods of the proposed device are the same with the predicate device.
# Comment 10 Applied Part:
{29}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
Physical therapy mode:
The applied part of the proposed device are the same with the predicate device.
General mode:
The applied part of the subject device (Handpiece), which is different to the predicate device (Fiber), but it meets the clinical needs. The applied part: Handpiece and fiber have the same function of transmitting laser energy to the treatment area, but handpiece is convenient for doctors to operate. This difference does not affect the safety and effectiveness of the device.
## Comment 11 Transmission system
The transmission system of the subject device (400μm fibers with metal protective armors), which is different to the predicate device (fibers of 600μm with SMA905 connector).
The transmission system only transmits laser energy. Although the optical fiber core diameter of optical fiber transmission system is different, but it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
## Conclusion:
The indication for use of the subject device claimed by proposed device is the same with the predicate device.
The proposed device uses same diode lasers technology as that used by the predicates. The core output parameters of the proposed device are identical to the predicates. The differences only exist in such contents: output power, energy density (ED) / (fluence), power density, spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods, applied part and transmission system, that all can be controlled in range of application. These differences do not raise new types of questions regarding the safety and effectiveness when the device is used for the proposed indications for use.
So, the proposed devices are Substantially Equivalent (SE) to existing legally marketed predicate devices.
26 / 47
{30}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
Table 4 Substantial equivalence discussion (CureLight Medical diode laser systems, Model: CureLight F2-B30, CureLight F3-AB30)
| Item | Proposed device | Primary predicate device | Co-predicate device | Discussion |
| --- | --- | --- | --- | --- |
| 510k Number | / | K242755 | K121363 | / |
| Proprietary Name | CureLight Medical diode laser systems | Medical diode laser systems | Diowave Laser System | / |
| Model | CureLight F2-B30, CureLight F3-AB30 | VELAS II -30A, VELAS II -30B | Diowave 30W | / |
| Product Code | ILY, PDZ | ILY, PDZ | ILY | Identical |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500, 21 CFR 878.4810 | 21 CFR 890.5500 | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indication for Use | The CureLight Medical Diode Laser Systems, model: CureLight F2-B30 and CureLight F3-AB30 are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, | The Medical Diode Laser Systems, model: VELAS II -30A and VELAS II -30B is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of | The Diowave Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating, tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting | SE |
{31}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | promoting relaxation of muscle tissue, and to temporarily increase local blood circulation. The CureLight Medical Diode Laser Systems, model: CureLight F2-B30 and CureLight F3-AB30 are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | muscle tissue, and to temporarily increase local blood circulation. The Medical Diode Laser Systems, model: VELAS II -30B is also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | relaxation of muscle tissue, and to temporarily increase local blood circulation. | |
| --- | --- | --- | --- | --- |
| Laser type | diode laser | diode laser | diode laser | Identical |
| Wavelength | CureLight F2-B30: 980nm±20nm CureLight F3-AB30: 810nm±20nm, 980nm±20nm | VELAS II -30A: 810 ± 10nm VELAS II -30B: 980 ± 10nm | 810nm±10nm, 980nm±10nm | SE |
| Output Power | Physical therapy mode: 0-30W (±20%) General mode: 0-10W (±20%), 980nm only | Physical therapy mode: 1W-30W (±20%) General mode: 1W-10W (±20%), 980nm only | N/A | SE |
| Energy Density(ED) / (Fluence) | Physical therapy mode: ≤3.1 J/cm2 (Per second) General mode: ≤39.8 J/cm2 (Per | Physical therapy mode: 0.05~3.1 J/cm2 (Per second) General mode: 1.7~39.8 J/cm2 (Per | N/A | SE |
{32}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | second) | second) | | |
| --- | --- | --- | --- | --- |
| Power Density | Physical therapy mode: ≤3.1 W/cm²
General mode: ≤39.8 W/cm² | Physical therapy mode: 0.05~3.1 W/cm²
General mode: 1.7~39.8 W/cm² | N/A | SE |
| Spot Diameter | Physical therapy mode: 4~5 cm
General mode: 5~6 mm | Physical therapy mode: 3.5~5 cm
General mode: 4~5 mm | N/A | Different Comment 1 |
| Working Distance | Physical therapy mode: 0 cm~2 cm for therapy handpiece 1
2 cm~3 cm for therapy handpiece 2
General mode: 3~4 mm | Physical therapy mode: 3~5 cm
General mode: 3~4 mm | N/A | Different Comment 2 |
| Treatment Angle | Perpendicular to the treatment area | Perpendicular to the treatment area | N/A | Identical |
| Applied Part Moving Speed | Physical therapy mode: 3~5 cm/s
General mode: 1~2 cm/s | Physical therapy mode: 3~5 cm/s
General mode: 1~2 cm/s | N/A | Identical |
| Operation Mode | Physical therapy mode: CW, repeat pulse
General mode: repeat pulse | Physical therapy mode: CW, single pulse, repeat pulse
repeat pulse: repeat pulse | N/A | Different Comment 3 |
| Pulse Width | Physical therapy mode: 50μs-1s
General mode: Ton: 10ms
Toff: 10ms, 20ms | Physical therapy mode: 10ms-1s
General mode: Ton: 10ms
Toff: 10ms, 20ms | N/A | Different Comment 4 |
| Pulse | Physical therapy mode: 1Hz-10KHz | General mode: 1Hz-50Hz | N/A | Different |
29 / 47
{33}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Repetition Rate | General mode: 33.3Hz/50Hz | General mode: 33.3Hz/50Hz | | Comment 5 |
| --- | --- | --- | --- | --- |
| Treatment Methods | Physical therapy mode: Contact, Non-contact.
General mode: Non-contact | Non-contact. | Non-contact | Different Comment 6 |
| Treatment Mode | Physical therapy mode, general mode | Physical therapy mode, general mode | / | Identical |
| Applied Part | Handpiece | Physical therapy mode: Handpiece
General mode: Fiber | Handpiece | Different Comment 7 |
| Transmission System | 400μm fibers with metal protective armors. | fibers of 600μm with SMA905 connector. | fibers of 400μm, 600μm and 1000μm with SMA905 connector | Different Comment 8 |
| Aiming Beam | Diode laser of 635nm(±20nm), power < 5mW, adjustable brightness. | Diode laser of 650nm, power≤2mW, adjustable brightness. | Diode laser of 650nm power max. < 5mW, adjustable brightness. | SE |
| Operation Interface | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Identical |
| Power Supply | ~110-230V, 50-60Hz, 200VA | 100-240VAC, 50-60Hz, 350VA | 100-240VAC, 50/60Hz, 350VA | SE |
| Laser Class | 4 | 4 | 4 | Identical |
| Safety Classification | Class I Type B | Class I Type B | Class I Type B | Identical |
30 / 47
{34}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Cooling | Air | Air | Air | Identical |
| --- | --- | --- | --- | --- |
| Waterproof Level | IPX1 | IPX1 | IPX1 | Identical |
| Footswitch Waterproof Level | IPX8 | IPX8 | IPX8 | Identical |
## Discussion:
### Clinical:
Indication for use claimed by proposed device is the same with the predicate device K242755 (VELAS II -30A, VELAS II -30B).
### Technology:
The subject device has the same wavelengths and output power of diode laser to achieve its intended use.
The energy density (ED) / (fluence), power density, treatment angle, moving speed, treatment methods and etc. are either identical to the predicate. The main differences are spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods, applied part and transmission system.
## Comment 1 Spot Diameter:
### Physical therapy mode:
The spot diameter of the proposed device: 4~5 cm, which is different to the predicate device, K242755: 3.5~5 cm, but the spot diameter of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
### General mode:
31 / 47
{35}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The spot diameter of the proposed device: 5~6 mm, which is different to the predicate device, K242755: 4~5 mm, but the energy density (ED) / (fluence) and power density of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
## Comment 2 Working Distance:
Physical therapy mode:
The working distance of the proposed device: 0 cm~2 cm for therapy handpiece 1; 2 cm~3 cm for therapy handpiece 2, which is different to the predicate device, K242755: 3~5 cm.
The working distance of proposed device is shorter because the therapy handpiece head has a length of 3 cm. Therefore, the actual distance from the laser light source at the end of the laser output port to the treatment end is 3 cm~5 cm for therapy handpiece 1 which is identical with the predicate device: 3~5 cm; 5 cm~6 cm for therapy handpiece 2, a little longer: 1 cm than the working distance of predicate device: 3~5 cm, but the energy density (ED) / (fluence) and power density are all within the scope of what predicate device claims and it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
General mode:
The working distance of the proposed device is the same with the predicate device.
## Comment 3 Operation mode:
Physical therapy mode:
The operation mode of the proposed device (CW, repeat pulse) is different to the predicate device (CW, single pulse, repeat pulse). but the operation mode of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
General mode:
32 / 47
{36}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The operation mode of the proposed device is the same with the predicate device.
## Comment 4 Pulse width:
Physical therapy mode:
The pulse width of the proposed device: 50μs-1s, which is different to the predicate device, K242755 (VELAS II -30A, VELAS II -30B): 10ms-1s.
The difference in pulse width range (specifically the extension to 50 μs) is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30A, VELAS II -30B). The expanded parameter range operates within the same thermal safety margins and is justified by the overlapping therapeutic effect of elevating tissue temperature.
General mode:
The pulse width of the proposed device is the same with the predicate device.
## Comment 5 Pulse repetition rate:
Physical therapy mode:
The pulse repetition rate of the subject device: 1Hz-10KHz, which is different to the predicate device, K242755 (VELAS II -30A, VELAS II -30B): 1Hz-50Hz.
The difference in pulse repetition rate: 1Hz-10KHz and 1Hz-50Hz is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30A, VELAS II -30B).
General mode:
33 / 47
{37}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The pulse repetition rate of the proposed device is the same with the predicate device.
## Comment 6 Treatment Methods:
Physical therapy mode:
The treatment methods of the subject device (Contact, Non-contact), which is different to the predicate device (Non-contact), but it meets the clinical needs. Contact mode is convenient for doctors to operate, which is more labor-saving. Moreover, the therapy handpiece head of therapy handpiece 1 has been evaluated and tested for biocompatibility as required, and the test results meet the requirements. This difference does not affect the safety and effectiveness of the device.
General mode:
The treatment methods of the proposed device are the same with the predicate device.
## Comment 7 Applied Part:
Physical therapy mode:
The applied part of the proposed device are the same with the predicate device.
General mode:
The applied part of the subject device (Handpiece), which is different to the predicate device (Fiber), but it meets the clinical needs. The applied part: Handpiece and fiber have the same function of transmitting laser energy to the treatment area, but handpiece is convenient for doctors to operate. This difference does not affect the safety and effectiveness of the device.
## Comment 8 Transmission system
The transmission system of the subject device (400μm fibers with metal protective armors), which is different to the predicate device (fibers of 600μm with SMA905 connector).
34 / 47
{38}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The transmission system only transmits laser energy. Although the optical fiber core diameter of optical fiber transmission system is different, but it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
**Conclusion:**
The indication for use of the subject device claimed by proposed device is the same with the predicate device.
The proposed device uses same diode lasers technology as that used by the predicates. The core output parameters of the proposed device are identical to the predicates. The differences only exist in such contents: spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods, applied part and transmission system, that all can be controlled in range of application. These differences do not raise new types of questions regarding the safety and effectiveness when the device is used for the proposed indications for use.
So, the proposed devices are Substantially Equivalent (SE) to existing legally marketed predicate devices.
35 / 47
{39}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
Table 5 Substantial equivalence discussion (CureLight Medical diode laser systems, Model: CureLight F3-AB60)
| Item | Proposed device | Primary predicate device | Co-predicate device | Discussion |
| --- | --- | --- | --- | --- |
| 510k Number | / | K242755 | K121363 | / |
| Proprietary Name | CureLight Medical diode laser systems | Medical diode laser systems | Diowave Laser System | / |
| Model | CureLight CureLight F3-AB60 | VELAS II -30A, VELAS II -30B | Diowave 60W | / |
| Product Code | ILY, PDZ | ILY, PDZ | ILY | Identical |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500, 21 CFR 878.4810 | 21 CFR 890.5500 | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indication for Use | The CureLight Medical Diode Laser Systems, model: CureLight F3-AB60 are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily | The Medical Diode Laser Systems, model: VELAS II -30A and VELAS II -30B is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, and to temporarily | The Diowave Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating, tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with minor arthritis, promoting relaxation of muscle tissue, | SE |
{40}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | increase local blood circulation. The CureLight Medical Diode Laser Systems, model: CureLight F2-B30 and CureLight F3-AB30 are also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | increase local blood circulation. The Medical Diode Laser Systems, model: VELAS II -30B is also indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | and to temporarily increase local blood circulation. | |
| --- | --- | --- | --- | --- |
| Laser type | diode laser | diode laser | diode laser | Identical |
| Wavelength | 810nm±20nm, 980nm±20nm | VELAS II -30A: 810 ± 10nm VELAS II -30B: 980 ± 10nm | 810nm±10nm, 980nm±10nm | SE |
| Output Power | Physical therapy mode: 0-60W (±20%) General mode: 0-10W (±20%), 980nm only | Physical therapy mode: 1W-30W (±20%) General mode: 1W-10W (±20%), 980nm only | N/A | Different Comment 1 |
| Energy Density(ED) / (Fluence) | Physical therapy mode: ≤6.2 J/cm2 (Per second) General mode: ≤39.8 J/cm2 (Per second) | Physical therapy mode: 0.05~3.1 J/cm2 (Per second) General mode: 1.7~39.8 J/cm2 (Per second) | N/A | Different Comment 2 |
| Power Density | Physical therapy mode: ≤6.2 W/cm2 General mode: ≤39.8 W/cm2 | Physical therapy mode: 0.05~3.1 W/cm2 | N/A | Different Comment 3 |
{41}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| | | General mode: 1.7~39.8 W/cm² | | |
| --- | --- | --- | --- | --- |
| Spot Diameter | Physical therapy mode: 4~5 cm
General mode: 5~6 mm | Physical therapy mode: 3.5~5 cm
General mode: 4~5 mm | N/A | Different
Comment 4 |
| Working Distance | Physical therapy mode:
0 cm~2 cm for therapy handpiece 1
2 cm~3 cm for therapy handpiece 2
General mode: 3~4 mm | Physical therapy mode: 3~5 cm
General mode: 3~4 mm | N/A | Different
Comment 5 |
| Treatment Angle | Perpendicular to the treatment area | Perpendicular to the treatment area | N/A | Identical |
| Applied Part Moving Speed | Physical therapy mode: 3~5 cm/s
General mode: 1~2 cm/s | Physical therapy mode: 3~5 cm/s
General mode: 1~2 cm/s | N/A | Identical |
| Operation Mode | Physical therapy mode: CW, repeat pulse
General mode: repeat pulse | Physical therapy mode: CW, single pulse, repeat pulse
repeat pulse: repeat pulse | N/A | Different
Comment 6 |
| Pulse Width | Physical therapy mode:
50μs-1s
General mode:
Ton: 10ms
Toff: 10ms, 20ms | Physical therapy mode: 10ms-1s
General mode:
Ton: 10ms
Toff: 10ms, 20ms | N/A | Different
Comment 7 |
| Pulse Repetition Rate | Physical therapy mode: 1Hz-10KHz
General mode: 33.3Hz/50Hz | General mode: 1Hz-50Hz
General mode: 33.3Hz/50Hz | N/A | Different
Comment 8 |
| Treatment | Physical therapy mode: Contact, | Non-contact. | Non-contact | Different |
38 / 47
{42}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Methods | Non-contact.
General mode: Non-contact | | | Comment 9 |
| --- | --- | --- | --- | --- |
| Treatment Mode | Physical therapy mode, general mode | Physical therapy mode, general mode | / | Identical |
| Applied Part | Handpiece | Physical therapy mode: Handpiece
General mode: Fiber | Handpiece | Different Comment 10 |
| Transmission System | 400μm fibers with metal protective armors. | fibers of 600μm with SMA905 connector. | fibers of 400μm, 600μm and 1000μm with SMA905 connector | Different Comment 11 |
| Aiming Beam | Diode laser of 635nm(±20nm), power < 5mW, adjustable brightness. | Diode laser of 650nm, power≤2mW, adjustable brightness. | Diode laser of 650nm power max. < 5mW, adjustable brightness. | SE |
| Operation Interface | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Identical |
| Power Supply | ~110-230V, 50-60Hz, 200VA | 100-240VAC, 50-60Hz, 350VA | 100-240VAC, 50/60Hz, 350VA | SE |
| Laser Class | 4 | 4 | 4 | Identical |
| Safety Classification | Class I Type B | Class I Type B | Class I Type B | Identical |
| Cooling | Air | Air | Air | Identical |
| Waterproof | IPX1 | IPX1 | IPX1 | Identical |
39 / 47
{43}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
| Level | | | | |
| --- | --- | --- | --- | --- |
| Footswitch Waterproof Level | IPX8 | IPX8 | IPX8 | Identical |
## Discussion:
### Clinical:
Indication for use claimed by proposed device is the same with the predicate device K242755 (VELAS II -30A, VELAS II -30B).
### Technology:
The subject device has the same wavelength of diode laser to achieve its intended use.
The treatment angle, moving speed, treatment methods and etc. are either identical to the predicate. The main differences are output power, energy density (ED) / (fluence), power density, spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods, applied part and transmission system.
## Comment 1 Output Power:
### Physical therapy mode:
The output power of the proposed device: 0-60W (±20%), which is different to the predicate device, K242755 (VELAS II -30A and VELAS II -30B): 1W-30W (±20%), but the output power of the proposed device is within the scope of predicate device, K121363 (Diowave 60W): Max. 60W claims. This difference does not affect the safety and effectiveness of the device.
### General mode:
40 / 47
{44}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The output power of the proposed device: 0-10W (±20%) is substantial equivalence with the predicate device, K242755 (VELAS II - 30B): 1W-10W (±20%) claims.
## Comment 2 Energy Density (ED) / (Fluence):
Physical therapy mode:
The energy density (ED) / (fluence) of the proposed device: ≤6.2 J/cm² (Per second), which is different to the predicate device, K242755 (VELAS II -30A and VELAS II -30B): 0.05~3.1 J/cm² (Per second), but the Energy Density (ED) / (Fluence) of the proposed device will within the scope of predicate device, K121363 (Diowave 60W). This difference does not affect the safety and effectiveness of the device.
Energy Density (ED) / (Fluence):
$$
ED = P * Time / S * Duty \, J/cm^2
$$
All other conditions being equal, the higher the output power, the higher the energy density. At a maximum output power of 60W, the energy density will be twice as high as at a maximum output power of 30W. While the Max. ED≤3.1 J/cm² (Per second) at 30W, the Max. ED will ≤6.2 J/cm² (Per second) at 60W.
General mode:
The energy density (ED) / (fluence) of the proposed device: ≤39.8 J/cm² (Per second) is substantial equivalence with the predicate device, K242755 (VELAS II -30B): 1.7~39.8 J/cm² (Per second) claims.
## Comment 3 Power Density:
Physical therapy mode:
{45}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The power density of the proposed device: ≤6.2 W/cm², which is different to the predicate device, K242755 (VELAS II -30A and VELAS II -30B): 0.05~3.1 W/cm², but the power density of the proposed device will within the scope of predicate device, K121363 (Diowave 60W). This difference does not affect the safety and effectiveness of the device.
Power Density
$$
PD = P/S \times \text{Duty } W/cm^2
$$
All other conditions being equal, the higher the output power, the higher the power density. At a maximum output power of 60W, the power density will be twice as high as at a maximum output power of 30W. While the Max. PD≤3.1 W/cm² at 30W, the Max. PD will ≤6.2 W/cm² at 60W.
General mode:
The power density of the proposed device: ≤39.8 W/cm² is substantial equivalence with the predicate device, K242755 (VELAS II - 30B): 1.7~39.8 W/cm² claims.
## Comment 4 Spot Diameter:
Physical therapy mode:
The spot diameter of the proposed device: 4~5 cm, which is different to the predicate device, K242755: 3.5~5 cm, but the spot diameter of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
General mode:
{46}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The spot diameter of the proposed device: 5~6 mm, which is different to the predicate device, K242755: 4~5 mm, but the energy density (ED) / (fluence) and power density of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
## Comment 5 Working Distance:
Physical therapy mode:
The working distance of the proposed device: 0 cm~2 cm for therapy handpiece 1; 2 cm~3 cm for therapy handpiece 2, which is different to the predicate device, K242755: 3~5 cm.
The working distance of proposed device is shorter because the therapy handpiece head has a length of 3 cm. Therefore, the actual distance from the laser light source at the end of the laser output port to the treatment end is 3 cm~5 cm for therapy handpiece 1 which is identical with the predicate device: 3~5 cm; 5 cm~6 cm for therapy handpiece 2, a little longer: 1 cm than the working distance of predicate device: 3~5 cm, but the energy density (ED) / (fluence) and power density are all within the scope of what predicate device claims and it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
General mode:
The working distance of the proposed device is the same with the predicate device.
## Comment 6 Operation mode:
Physical therapy mode:
The operation mode of the proposed device (CW, repeat pulse) is different to the predicate device (CW, single pulse, repeat pulse). but the operation mode of the proposed device is within the scope of what predicate device claims. This difference does not affect the safety and effectiveness of the device.
43 / 47
{47}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
General mode:
The operation mode of the proposed device is the same with the predicate device.
## Comment 7 Pulse width:
Physical therapy mode:
The pulse width of the proposed device: 50μs-1s, which is different to the predicate device, K242755 (VELAS II -30A, VELAS II -30B): 10ms-1s.
The difference in pulse width range (specifically the extension to 50 μs) is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30A, VELAS II -30B). The expanded parameter range operates within the same thermal safety margins and is justified by the overlapping therapeutic effect of elevating tissue temperature.
General mode:
The pulse width of the proposed device is the same with the predicate device.
## Comment 8 Pulse repetition rate:
Physical therapy mode:
The pulse repetition rate of the subject device: 1Hz-10KHz, which is different to the predicate device, K242755 (VELAS II -30A, VELAS II -30B): 1Hz-50Hz.
The difference in pulse repetition rate: 1Hz-10KHz and 1Hz-50Hz is a technological modification that does not raise new questions of safety and effectiveness. The fundamental technology (infrared emission for topical heating) and the intended use remain unchanged from the predicate device K242755 (VELAS II -30A, VELAS II -30B).
44 / 47
{48}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
General mode:
The pulse repetition rate of the proposed device is the same with the predicate device.
## Comment 9 Treatment Methods:
Physical therapy mode:
The treatment methods of the subject device (Contact, Non-contact), which is different to the predicate device (Non-contact), but it meets the clinical needs. Contact mode is convenient for doctors to operate, which is more labor-saving. Moreover, the therapy handpiece head of therapy handpiece 1 has been evaluated and tested for biocompatibility as required, and the test results meet the requirements. This difference does not affect the safety and effectiveness of the device.
General mode:
The treatment methods of the proposed device are the same with the predicate device.
## Comment 10 Applied Part:
Physical therapy mode:
The applied part of the proposed device are the same with the predicate device.
General mode:
The applied part of the subject device (Handpiece), which is different to the predicate device (Fiber), but it meets the clinical needs. The applied part: Handpiece and fiber have the same function of transmitting laser energy to the treatment area, but handpiece is convenient for doctors to operate. This difference does not affect the safety and effectiveness of the device.
## Comment 11 Transmission system
45 / 47
{49}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
The transmission system of the subject device (400μm fibers with metal protective armors), which is different to the predicate device (fibers of 600μm with SMA905 connector).
The transmission system only transmits laser energy. Although the optical fiber core diameter of optical fiber transmission system is different, but it meets the clinical needs. This difference does not affect the safety and effectiveness of the device.
**Conclusion:**
The indication for use of the subject device claimed by proposed device is the same with the predicate device.
The proposed device uses same diode lasers technology as that used by the predicates. The core output parameters of the proposed device are identical to the predicates. The differences only exist in such contents: output power, energy density (ED) / (fluence), power density, spot diameter, working distance, operation mode, pulse width, pulse repetition rate, treatment methods, applied part and transmission system, that all can be controlled in range of application. These differences do not raise new types of questions regarding the safety and effectiveness when the device is used for the proposed indications for use.
So, the proposed devices are Substantially Equivalent (SE) to existing legally marketed predicate devices.
46 / 47
{50}
PHOMED
Wuhan PHOMED Technology Co., Ltd.
# VIII PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Medical Diode Laser System.
- Verify the conformity of the proposed devices with the requirements of IEC60601-1: 2005, AMD1 2012, AMD2 2020 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance).
- Verify the conformity of the proposed devices with the requirements of IEC60601-1-2: 2014 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic compatibility).
- Verify the conformity of the proposed devices to IEC 60825-1: 2014 (Safety of laser products - Part 1: Equipment classification and requirements).
- Verify the performance of the proposed devices according to IEC 60601-2-22: 2019 (Medical electrical equipment Part 2: Particular Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment).
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as an Enhanced Documentation level.
## Mechanical and acoustic testing
It is not applicable.
## Animal Study
It is not applicable.
## Clinical Studies
It is not applicable.
# VIII CONCLUSIONS
The CureLight Medical diode laser systems are substantially equivalent to its predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.