K2 MOBILE

K200693 · Hulaser, Inc. · GEX · Nov 19, 2020 · General, Plastic Surgery

Device Facts

Record IDK200693
Device NameK2 MOBILE
ApplicantHulaser, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateNov 19, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

K2 mobile laser is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures, including Tooth Whitening and the temporary relief of pain. The Specific indications are as follows: Endodontic procedures: Pulpotomy, Root canal therapy Periodontal procedures: Sulcular debridement Implant procedures: Implant recovery Surgery procedures: Biopsies, Crown lengthening, Exposure of unerupted teeth, Fibroma removal, Frenectomy, Gingivoplasty, Incision and drainage of abscess, Operculectomy, Papillectomies, Reduction of gingival hypertrophy, Vestibuloplasty Other procedures: Gingival troughing, Hemostasis and coagulation, Leukoplakia, Treatment of aphthous-ulcer canker sores and herpetic, Laser Assisted whitening, Topical heating for the purpose of elevating tissue temperary relief of minor muscle and joint pain and stiffness minor arthritis pain, or muscle spasm, minor muscular back pain; the temporary increase in local blood circulation; the temporary relation of muscle

Device Story

K2 Mobile is a handpiece-style dental diode laser; uses invisible infrared radiation (980nm) delivered via optical fiber to treatment site. Operates in continuous or pulse modes; powered by internal rechargeable lithium-ion battery. Used by dentists in clinical settings for soft tissue surgery, periodontal procedures, and pain relief via topical heating. Device provides precise tissue interaction; clinician controls output via handpiece. Benefits include minimally invasive tissue management, hemostasis, and pain relief. Safety features include aiming beam (635nm).

Clinical Evidence

Bench testing only. No clinical data presented. Performance validated via electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1, 21 CFR 1040.10/11), battery safety (IEC 62133), and biocompatibility (ISO 10993 series).

Technological Characteristics

Diode laser; 980nm wavelength; Class IV laser. Handpiece form factor with internal 3.7V 1700mA Li-ion battery. Optical fiber delivery (400 μm core). Aiming beam 635nm (5mW). Continuous and pulse operation modes. Biocompatible patient-contact tips. Standards: IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 62133, ISO 10993.

Indications for Use

Indicated for dentists performing oral soft tissue procedures (excision, vaporization, ablation, coagulation), endodontic procedures (pulpotomy, root canal), periodontal procedures (sulcular debridement), implant recovery, and various surgeries (biopsies, frenectomy, etc.). Also indicated for tooth whitening, treatment of aphthous/herpetic ulcers, and topical heating for temporary relief of minor muscle/joint pain, stiffness, arthritis, muscle spasm, and increased local blood circulation.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products. November 19, 2020 HULASER, Inc. % Dongha Lee Regulatory Affairs Consultant KMC, Inc. Room no. 904, 27, Digital-ro 27ga-gil, Guro-gu Seoul, 08375 Korea, Republic Of Re: K200693 Trade/Device Name: K2 Mobile Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 10, 2020 Received: March 16, 2020 Dear Dongha Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200693 #### Device Name K2 MOBILE Indications for Use (Describe) K2 mobile laser is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures, including Tooth Whitening and the temporary relief of pain. The Specific indications are as follows: Endodontic procedures: Pulpotomy, Root canal therapy Periodontal procedures: Sulcular debridement Implant procedures: Implant recovery Surgery procedures: Biopsies, Crown lengthening, Exposure of unerupted teeth, Fibroma removal, Frenectomy, Gingivoplasty, Incision and drainage of abscess, Operculectomy, Papillectomies, Reduction of gingival hypertrophy, Vestibuloplasty Other procedures: Gingival troughing, Hemostasis and coagulation, Leukoplakia, Treatment of aphthous-ulcer canker sores and herpetic, Laser Assisted whitening, Topical heating for the purpose of elevating tissue temperary relief of minor muscle and joint pain and stiffness minor arthritis pain, or muscle spasm, minor muscular back pain; the temporary increase in local blood circulation; the temporary relation of muscle Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for HU Laser, which includes a red cross symbol integrated into the "H" of the company name. The text "LASER SCIENCE FOR HUMANS" is written in smaller font below the company name. To the left of the logo, the text "SINCE 2002" is written vertically. # 510(k) Summary (K200693) This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: September 28, 2020 # 1. Applicant / Submission Sponsor HULASER, Inc. Address: 1110, DAERUNG TECHNOTOWN 17, 25, GASAN DIGITAL 1-RO, GEUMCHEON-GU, SEOUL, 08594, KOREA Tel: +82-2-3012-5678 Fax: +82-2-544-6540 ## 2. Submission Correspondent DongHa Lee (Consultant, KMC, Inc.) Address: Room no. 904, 27, Digital-ro 27ga-gil, Guro-gu, Seoul, 08375, Korea Tel: +82-70-8965-5554 Fax: +82-2-856-5904 Email: dhlee@kmcerti.com # 3. Device Identification Trade/Proprietary Name: K2 MOBILE Common Name: Dental Diode Laser Classification Regulation: 21CFR 878.4810 Product Code: GEX, Device Class: 2 ## 4. Predicate Devices | | Predicate Device 1 | Predicate Device 2 | |---------------|----------------------|--------------------------| | Manufacturer | Zolar Technology Inc | Biolase Technology, Inc. | | Device Name | Photon Plus | iLaseTM | | 510(k) number | K162114 | K093852 | # 5. Description The dental diode laser (Model: K2 MOBILE) is a surgical device designed for a wide variety of dental soft tissue procedures. Diode laser is used as a source of invisible infrared radiation for this device and delivered to the treatment area through optical fiber is {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for "HU LASER", with the tagline "LASER SCIENCE FOR HUMANS". The logo features the letters "HU" in a bold, sans-serif font, with a red cross symbol incorporated into the "U". To the left of the "HU" is the text "SINCE 2002" in a smaller font size, arranged vertically. The words "LASER SCIENCE FOR HUMANS" are placed below the "HU LASER" text, also in a sans-serif font. incorporated into a tip. The dental diode laser (Model: K2 MOBILE) is handpiece type using an internal rechargeable battery. The rechargeable battery charger are provided. # 6. Indication for use K2 mobile laser is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures, including Tooth Whitening and the temporary relief of pain. The Specific indications are as follows: #### Endodontic procedures: Pulpotomy, Root canal therapy #### Periodontal procedures: Sulcular debridement #### Implant procedures: Implant recovery #### Surgery procedures: Biopsies, Crown lengthening, Exposure of unerupted teeth, Fibroma removal, Frenectomy, Gingivoplasty, Incision and drainage of of abscess, Gingivectomy, Operculectomy, Papillectomies, Reduction of gingival hypertrophy, Vestibuloplasty ## Other procedures: Gingival troughing, Hemostasis and coagulation, Leukoplakia, Treatment of aphthous-ulcer canker sores and herpetic, Laser Assisted whitening, Topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain and stiffness minor arthritis pain, or muscle spasm, minor sprains and minor muscular back pain; the temporary increase in local blood circulation; the temporary relation of muscle ## 7. Substantial Equivalence Dental Diode Laser (Model: K2 MOBILE) is substantially equivalent to the predicate devices, Photon Plus (K162114, Zolar Technology Inc) and iLaseTM (K093852, Biolase Technology, Inc.) The following comparison table is presented to demonstrate substantial equivalence. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for HUTLASER. The logo is gray and features a red plus sign in the middle of the "U". Below the logo is the text "LASER SCIENCE FOR HUMANS". To the left of the logo is the text "SINCE 2002". | | Subject Device | Predicate Device 1 (PD1) | Predicate Device 2 (PD2) | Comparison | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | Manufacturer | HULASER, Inc. | Zolar Technology Inc | Biolase Technology, Inc. | - | | Device Name | K2 MOBILE | Photon Plus | iLaseTM | - | | 510(k) number | K200693 | K162114 | K093852 | - | | Product Code | GEX | GEX | GEX | Same | | Regulatory Class | 2 | 2 | 2 | Same | | Indications for Use | K2 mobile laser is intended for use<br>by dentists for excision, incision,<br>vaporization, ablation and<br>coagulation of oral soft tissue<br>procedures, including Tooth<br>Whitening and the temporary relief<br>of pain. The Specific indications are<br>as follows:<br><br>Endodontic procedures:<br>Pulpotomy,<br>Root canal therapy<br><br>Periodontal procedures:<br>Sulcular debridement<br><br>Implant procedures:<br>Implant recovery<br><br>Surgery procedures:<br>Biopsies,<br>Crown lengthening,<br>Exposure of unerupted teeth,<br>Fibroma removal,<br>Frenectomy,<br>Gingivectomy,<br>Gingivoplasty,<br>Incision and drainage of abscess | Dental Soft Tissue Indications<br>Incision, excision, vaporization,<br>ablation and coagulation of oral soft<br>tissues including marginal and inter-<br>dental gingival and epithelial lining<br>of free gingival and the following<br>specific indications:<br>Exposure of Unerupted teeth<br>Fibroma removal<br>Frenectomy<br>Gingival Troughing for crown<br>impressions<br>Gingivectomy<br>Gingivoplasty<br>Hemostasis and coagulation<br>Gingival incision and excision<br>hemostasis<br>Implant recovery<br>Incision and drainage of abscess<br>Leukoplakia<br>Operculectomy<br>Oral papillectomies<br>Pulpotomy as an adjunct to root<br>canal therapy<br>Reduction of gingival hypertrophy<br>Soft tissue crown lengthening<br>Gingival bleeding index | The iLaseTM dental soft tissue laser<br>is a surgical device designed for a<br>wide variety of dental soft tissue<br>procedures and laser periodontal<br>procedures.<br><br>Dental Soft Tissue Indications:<br>Incision, excision, vaporization,<br>ablation and coagulation of oral soft<br>tissues, including marginal and inter-dental gingival<br>and epithelial lining of free gingiva<br>and the following specific<br>indications:<br>Excisional and incisional biopsies<br>Exposure of unerupted teeth<br>Fibroma removal<br>Frenectomy<br>Frenotomy<br>Gingival troughing for crown<br>impressions<br>Gingivectomy<br>Gingivoplasty<br>Gingival incision and excision<br>Hemostasis and coagulation<br>Implant recovery<br>Incision and drainage of abscess<br>Leukoplakia | Equivalent<br>(Included in the<br>indications of PD1) | | | Papillectomies,<br>Reduction of gingival hypertrophy,<br>Vestibuloplasty<br><br>Others procedures:<br>Gingival troughing<br>Hemostasis and coagulation<br>Leukoplakia<br>Treatment of aphthous-ulcer canker<br>sores and herpetic<br>Laser Assisted whitening<br>Topical heating for the purpose of<br>elevating tissue temperature for<br>temporary relief of minor muscle<br>and joint pain and stiffness minor<br>arthritis pain, or muscle spasm,<br>minor sprains and minor muscular<br>back pain; the temporary increase in<br>local blood circulation; the<br>temporary relation of muscle | Treatment of aphthous ulcer canker<br>sores and herpetic.<br>Vestibuloplasty.<br><br>Laser Periodontal Indications<br>Sulcular debridement (removal of<br>diseased or inflamed soft tissue in<br>the periodontal pocket to improve<br>clinical indices including: probe<br>depth, attachment loss and tooth<br>mobility)<br>Removal of highly inflamed<br>edematous tissue affected by<br>bacteria penetration of the pocket<br>lining and junctional epithelium.<br><br>Tooth Whitening Indications<br>Laser Assisted whitening/bleaching<br>of teeth.<br>Light activation for bleaching<br>materials for teeth whitening.<br><br>Pain Indications<br>Topical heating for the purpose of<br>elevating tissue temperature for a<br>temporary relief of minor muscle<br>and joint pain and stiffness minor<br>arthritis pain, or muscle spasm,<br>minor sprains and minor muscular<br>back pain; the temporary increase in<br>local blood circulation; the<br>temporary relaxation of muscle. | Oral papillectomies<br>Pulpotomy<br>Pulpotomy as an adjunct to root<br>canal therapy<br>Reduction of gingival hypertrophy<br>Soft tissue crown lengthening<br>Treatment of canker sores, herpetic<br>and aphthous ulcers of the oral<br>mucosa.<br>Vestibuloplasty<br>Tissue retraction for impression<br><br>Laser Periodontal Procedures,<br>including:<br>Laser soft tissue curettage<br>Laser removal of diseased, infected,<br>inflamed and necrosed soft tissue<br>within the periodontal pocket<br>Sulcular debridement (removal of<br>diseased, infected, inflamed and<br>necrosed soft tissue in the<br>periodontal pocket to improve<br>clinical indices including gingival<br>index, gingival bleeding index,<br>probe depth, attachment loss and<br>tooth mobility). | | | Principle of operation | Laser diode is used as a source of<br>invisible infrared radiation for this<br>device and delivered to the treatment<br>area through optical fiber | Laser diode is used as a source of<br>invisible infrared radiation for this<br>device and delivered to the treatment<br>area through optical fiber | Laser diode is used as a source of<br>invisible infrared radiation for this<br>device and delivered to the treatment<br>area through optical fiber | Same | | Laser Medium | Diode Laser | Diode Laser | Diode Laser | Same | | Laser Classification | Class IV (4) Laser | Class IV (4) Laser | Class IV (4) Laser | Same | | Laser Wavelength | 980nm | 980nm | 940nm | Same as the PD1 | | Laser Power range | 0~6W | 0~10W | 0~5W | Equivalent<br>(Included in the rage<br>of PD1)<br>Similar<br>(Refer to Note 1.) | | Operation mode | Continuous (CW) mode,<br>Pulse mode | Continuous (CW) mode,<br>Pulse mode | Continuous (CW) mode,<br>Pulse mode | Same | | Pulse width/interval | On/off time: 50ms (Pulse mode 1)<br>On/off time: 5ms (Pulse mode 2) | 0.1ms to 9.9s | Unknown | Equivalent<br>(Included in the rage<br>of PD1)<br>(Refer to Note 1.) | | Pulse repetition rate | 10Hz (Pulse mode 1)<br>100Hz (Pulse mode 2) | 1Hz to 5,000Hz | Unknown | Different<br>(Refer to Note 1.) | | Aiming Beam | 635nm, 5 mW | 650nm, 2mW | 635nm, 5 mW | Same as the PD2 | | Fiber core diameter | 400 μm | 400 μm | 400 μm | Same | | Power Input | Rechargeable Lithium Ion battery:<br>3.7V, 1700 mA<br>(Battery charger: AC 100-240V, 47-<br>63Hz) | Rechargeable lithium polymer<br>battery: 7.4V, 5A<br>(Battery charger: AC 100-240V,<br>50/60Hz) | Rechargeable Lithium Ion battery:<br>3.6V, 650 mA<br>(Battery charger: AC 90-230V, 50-<br>60Hz) | Different<br>(Refer to Note 2.) | | Prescription or OTC | Prescription | Prescription | Prescription | Same | | Intended User | Licensed practitioner | Licensed practitioner | Licensed practitioner | Same | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for HU LASER. The logo is gray and features a red cross in the letter U. Below the logo, the text "LASER SCIENCE FOR HUMANS" is written in a smaller font. The text "SINCE 2002" is written vertically on the left side of the logo. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for HU Laser. The logo is gray and features a red cross in the upper left corner. Below the logo, the text "LASER SCIENCE FOR HUMANS" is written in gray. To the left of the logo, the text "SINCE 2002" is written vertically. #### Note 1. : Laser output power range of the subject device is 0 to 6.0W. (The maximum output power is 3.5W on CW mode and the maximum peak power is 6.0W on pulsed model) The range is within the range (0 to 10W) of the predicate device 1 (K16214) and similar to the predicate device 2 (K093852, 0 to 5W). Pulse width/interval and pulse repetition rate of that onloff time is 50ms (10Hz) on pulse mode 1 and onloff time is 5ms (100Hz) on pulse mode 2. The predicate device 1 (K162114) has a specification of the pulse repetition rate (0.1ms to 9.9s 1Hz to 5,000Hz). The specifications of the predicate device 2 (K093852) is unknown. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for HU LASER. The logo is in gray and features a red cross in the upper left corner. Below the logo, the text "LASER SCIENCE FOR HUMANS" is written in a smaller font. The laser safety and performance test was conting to IEC 60825-1. The device safety and performance test was conducted acording to IEC 60601-1, IEC 60601-2-2, 21CFR1040.10 and 21CFR1040.11 with the above differences. The testing results show that these differences do not raise different questions of safety and effectiveness. (The safety and performance test reports are attached in this submission). # Note 2. : The subject device uses an internal rechargeable lithium-ion battery (DC 3.7V,1700mA). The predicate device 1 (K16214) uses an internal rechargeable lithium-polymer battery (DC 7.4V, 5A), and the predicate device 2 (K093852) uses an internal recharged be lithing-ion battery (DC 3.6V, 650mA). The internal battery and performance test was conding to IEC 62133. The device safety and performance test was conducted according to IEC 60601-1 and IEC 60601-2-22. The testing results show that these different questions of safety and effectiveness. (The safety and performance test reports are attached in this submission). {9}------------------------------------------------ # 8. Reprocessing We tested and validated the reprocessing process according to FDA reprocessing guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling.". The results show that the designed reprocessing processes are effective to remove soil and contaminants by cleaning and to inactivate microorganisms by disinfection or sterilization. # 9. Biocompatibility The biocompatibility tests of patient contact part (Tip, optical fiber is incorporated into tip, External communicating device - Blood path, indirect / limited contact duration: <24 hours) were performed in accordance with the following FDA recognized standards - ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process - ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro । cytotoxicity - -ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. - -ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity - ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood # 10. Electrical Safety and Electromagnetic compatibility The Electrical Safety and Electromagnetic compatibility tests were performed in accordance with the following FDA recognized standards. - -IEC 60601-1:2005+A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-2-22:2007+A1:2012, Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency. - -IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements - -21 CFR 1040.10 and 1040.11 - IEC 62133:2012, Secondary cells and batteries containing alkaline or other non-acid electrolytes. Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility - Requirements and tests # 11. Performance Testing - Nonclinical 1) Laser performance - Classification, Accessible emission level The test was performed in in accordance with the FDA recognized standard, IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements, Clause 9, and 21CFR1040.10. {10}------------------------------------------------ 2) Laser output power testing - Protection against unwanted and excessive output, Accuracy of control The test was performed in accordance with the FDA recognized standard, IEC 60601-2-22:2007+A1:2012, Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment, Clause 201.10 and 201.12, and 21CFR1040.11 # 12. Conclusion In comparing between the subject device and the predicate device, there are the same product code, regulatory classification, indications for use, principle of operation, laser classification, laser medium, laser operation mode, using a rechargeable battery, prescription use and intended user. Although there are some differences (laser output specifications and internal battery specifications), the safety and performance test reports are supported to the safety and effectiveness of the subject device. In this regard, we conclude that the subject device is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...