K162114 · Zolar Technology & Mfg Co., Inc. · GEX · Dec 12, 2016 · General, Plastic Surgery
Device Facts
Record ID
K162114
Device Name
Photon, Photon Plus
Applicant
Zolar Technology & Mfg Co., Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Photon Series of Dental Diode Lasers are intended for use by dentists for excision, incision, vaporization, ablation and coagulation of oral soft tissue procedures, including Tooth Whitening and the temporary relief of pain. The specific indications are as follows: 1. DENTAL SOFT TISSUE INDICATIONS The indications are identical to the previously cleared predicate systems. Incision, excision, vaporization, ablation and coagulation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingival and the following specific indications: - Exposure of Unerupted teeth - Fibroma removal - Frenectomy - Gingival Troughing for crown impressions - Gingivectomy - Gingivoplasty - Hemostasis and coagulation - Gingival incision and excision hemostasis - Implant recovery - Incision and drainage of abscess - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy as an adjunct to root canal therapy - Reduction of gingival hypertrophy - Soft tissue crown lengthening - Gingival bleeding index - Excisional and incisional biopsies - Treatment of aphthous ulcer canker sores and herpetic. - Vestibuloplasty 2. LASER PERIODONTAL INDICATIONS: - Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including: probe depth, attachment loss and tooth mobility) - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium. 3. TOOTH WHITENING INDICATIONS - Laser Assisted whitening/bleaching of teeth. - Light activation for bleaching materials for teeth whitening. 4. PAIN INDICATIONS: - Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle.
Device Story
Solid-state diode laser system; emits infrared energy (810nm/980nm) via flexible fiber optic cable to handpiece; includes visible aiming beam. Comprised of base console, detachable delivery system, tips, and wireless footswitch. Used by dentists in clinical settings for oral soft tissue surgery, periodontal debridement, tooth whitening, and therapeutic heating. Laser energy targets soft tissue for surgical procedures; thermal energy elevates tissue temperature for pain relief. Healthcare providers operate device to perform procedures; output provides surgical precision and therapeutic heat. Benefits include minimally invasive soft tissue management and temporary pain relief.
Clinical Evidence
Performance assessment conducted with human subjects for therapeutic heating indications. Data demonstrates device increases skin temperature to 40–45°C and maintains effect for at least 10 minutes safely and effectively.
Technological Characteristics
Solid-state diode laser; 810nm (Photon) or 980nm (Photon Plus) wavelength; 3W (Photon) or 10W (Photon Plus) max power; continuous/pulse operation modes. Includes wireless 2.4 GHz footswitch. Complies with IEC 60825-1 and IEC 60601-2-22. Class IV laser.
Indications for Use
Indicated for dental professionals performing oral soft tissue procedures (incision, excision, vaporization, ablation, coagulation), periodontal pocket debridement, tooth whitening, and topical heating for temporary relief of minor muscle/joint pain, stiffness, arthritis, muscle spasms, sprains, and strains.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 12, 2016
Zolar Technology & Mfg Co. Inc. Paul Atkins Executive Officer 6315 Shawson Drive Mississauga, L5T 1J2 CA
Re: K162114
Trade/Device Name: Photon, Photon Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 17, 2016 Received: November 17, 2016
Dear Paul Atkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) 162114
Device Name Photon, Photon Plus
### Indications for Use (Describe)
### 1. DENTAL SOFT TISSUE INDICATIONS
The indications are identical to the previously cleared predicate systems.
Incision, excision, vaporization, ablation of oral soft tissues including marginal and inter-dental givgival and epithelial lining of free gingival and the following specific indications:
- · Exposure of Unerupted teeth
- · Fibroma removal
- · Frenectomy
- · Gingival Troughing for crown impressions
- · Gingivectomy
- · Gingivoplasty
- · Hemostasis and coagulation
- · Gingival incision and excision hemostasis
- · Implant recovery
- · Incision and drainage of abscess
- · Leukoplakia
- · Operculectomy
- · Oral papillectomies
- · Pulpotomy as an adjunct to root canal therapy
- · Reduction of gingival hypertrophy
- · Soft tissue crown lengthening
- · Gingival bleeding index
- · Excisional and incisional biopsies
- · Treatment of aphthous ulcer canker sores and herpetic.
- · Vestibuloplasty
- 2. LASER PERIODONTAL INDICATIONS:
· Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including: probe depth, attachment loss and tooth mobility )
· Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.
- 3. TOOTH WHITENING INDICATIONS
- · Laser Assisted whitening/bleaching of teeth.
- · Light activation for bleaching materials for teeth whitening.
### 4. PAIN INDICATIONS:
· Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness minor arthritis pain, or muscle spasm, minor sprains and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/14)
Page 1 of 2
PSC Publishing Services: (703) 403-6740 EF
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Exhibit # 6
# 510 (K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA l990 and 21 CFR §807.92.
The assigned 5l0(k), number is: K162114.
- 1. Submitter's Identification: ZOLAR TECHNOLGY & MFG CO. INC. 6315 SHAWSON DRIVE, UNIT # 7-8, MISSISSAUGA, ON, L5T L5T 1J2, CANADA
Contact: Paul Atkins Executive Officer Tel: 905-593-3605 Fax: 1-888-426-7728 e-mail: patkins@zolartek.com
Alternate Contact Dr. Kapil Surana President Tel: 905-593-3605 Fax: 1-888-426-7728 e-mail: kapilsurana@hotmail.com
Date Summary Prepared: November 17, 2016
- 2. Name of the Device: Photon/Photon Plus
- 3. Common or Usual Name: Dental Diode Laser
- 4. Predicate Device Information: Biolase Epic 10 510(k) 121286 Cao Precise SHP 510(k) 123443 Photon/Photon Plus 510(k) 122020
{5}------------------------------------------------
#### ட். Device Description:
The Zolar Technology & MFG Co. Inc. Soft Tissue Dental Laser PHOTON/PHOTON PLUS uses a solid state diode to emit infrared laser energy which is transmitted via a flexible fiber optic cable to the hand piece that emits the energy to the target site. A visible light (aiming beam) is emitted at the same time to visually assist in identifying the specific treatment location. The Photon/Photon Plus Laser Systems are comprised of a base consol, a detechable delivery system, tips and a wireless footswitch. The Photon/Photon Plus laser is intended to be used by dentist for a variety of oral soft tissue procedures
The original Operator Manual has been revised to reference the additional Pain Indication for use.
#### 6. Indications for Use:
The Photon Series of Dental Diode Lasers are intended for use by dentists for excision, incision, vaporization, ablation and coagulation of oral soft tissue procedures, including Tooth Whitening and the temporary relief of pain. The specific indications are as follows:
# 1. DENTAL SOFT TISSUE INDICATIONS
The indications are identical to the previously cleared predicate systems.
Incision, excision, vaporization, ablation and coagulation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingival and the following specific indications:
- Exposure of Unerupted teeth
- Fibroma removal ●
- Frenectomy
- Gingival Troughing for crown impressions
- Gingivectomy
- Gingivoplasty
- Hemostasis and coagulation
- Gingival incision and excision hemostasis
- Implant recovery
- Incision and drainage of abscess
- Leukoplakia
- Operculectomy
- Oral papillectomies
- . Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
- Soft tissue crown lengthening
- Gingival bleeding index
- Excisional and incisional biopsies
- Treatment of aphthous ulcer canker sores and herpetic.
- . Vestibuloplasty
## 2. LASER PERIODONTAL INDICATIONS:
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including: probe depth, attachment loss and tooth mobility)
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.
{6}------------------------------------------------
# 3. TOOTH WHITENING INDICATIONS
- · Laser Assisted whitening/bleaching of teeth.
- Light activation for bleaching materials for teeth whitening.
#### 4. PAIN INDICATIONS:
- Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle.
#### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The subject devices have been tested for thermal, electrical and mechanical safety, as demonstrated in the previously approved clearance 510(k) 122020 and conform to applicable medical device safety standards, IEC 60825-1 and IEC 60601-2-22.
#### 8. Performance Assessment:
The clinical test for the therapeutic heating device indications listed in item 6 of the IFU have been conducted with human subjects and the performance data table demonstrates that the device can increase the skin temperature to 40 – 45 "c and maintain the effect for at least 10 minutes safely and effectively.
#### 9. Substantial Equivalence:
The purpose of this submission is to consolidate the current Photon Plus Laser Systems (K122020) with a new Pain Indication.
The Subject Device Photon Dental Diode Laser is of a comparable type and substantially equivalent to the currently marketed and approved Photon Plus 510(k) 122020, Biolase Epic 510(k) 121286 and Cao Precise SHP 510(k) 123443 in regard to the primary features of soft tissue surgical and minor pain indications for use. The subject device retains similar overall size and weight together with the similar construction materials and performance in addition to equivalent key features. The subject device has the same intended uses as the predicate devices.
The device modifications to include Pain Indications (IFU) do not potentially alter the fundamental scientific technology of the device. Substantial Equivalence for the Photon/Photon Plus Laser Systems has also been determined through comparison to the cleared predicate devices i.e. Biolase Epic 510(k) 121286 and Cao Precise SHP 510(k) 123443. This submission demonstrates that the Photon Plus Laser systems are safe, as effective and performs as well as the predicate devices indicated above.
{7}------------------------------------------------
#### 10. Conclusion:
The Device modifications do not potentially alter the fundamental scientific technology of the device. Substantial Equivalence for the Photon/Photon Plus Laser Systems has also been determined through the technical comparison to the cleared predicate devices. This submission demonstrates that the Photon Plus Laser systems are as safe, as effective and performs as well as the predicate devices.
We suggest that there are no significant differences between the subject device, (Photon/Photon Plus) and the predicate devices (Biolase Epic 10/Cao Precise SHP) in terms of indications for use, safety and effectiveness based on electrical, mechanical and environmental test results. With this information and in our opinion the subject device is considered to be substantially equivalent to the predicate devices.
{8}------------------------------------------------
# Subject Device Technological Comparison
| DEVICE NAME | Photon/Photon<br>Plus | Photon/Photon<br>Plus | Precise SHP | Epic 10 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER | Zolar Technology Inc | Zolar Technology Inc | Cao Group Inc | Biolase Inc |
| 510(K) | 162114 | 122020 | 123443 | 121286 |
| DIMENSION | 187 -138 -148mm | 187 -138 -148mm | 139 - 120 - 61mm | 145 - 112 - 165mm |
| WEIGHT | <2.0 kg | <2.0 kg | <2.0 kg | 1.1 kg |
| LASER TYPE | Diode | Diode | Diode | Diode |
| LASER<br>CLASSIFICATION | 4 | 4 | 4 | 4 |
| MAXIMUM<br>POWER | 3 watts Photon<br>10 watts Photon Plus | 3 watts Photon<br>10 watts Photon Plus | 3 watts | 10 watts |
| OPERATION<br>MODE | Continuous / Pulse | Continuous / Pulse | Continuous / Pulse | Continuous / Pulse |
| WAVELENGTH | 810 nm Photon<br>980 nm Photon Plus | 810 nm Photon<br>980 nm Photon Plus | 810 nm | 940 nm |
| AUDIBLE<br>NOTIFICATION | Yes | Yes | Yes | Yes |
| VISUAL<br>NOTIFICATION | Yes | Yes | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.