K180993 · Omniguide, Inc. · GEX · Jun 28, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180993
Device Name
Beacon Advanced CO2 Laser System
Applicant
Omniguide, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Beacon Advanced CO2 Advanced Energy Laser System is intended to be used in conjunction with the OmniGuide WaveGuide Fibers or the Articulated Arm to be used for the incision, ablation, vaporization and coagulation of body soft tissues including intra-oral soft tissues.
Device Story
Beacon Advanced Energy CO2 Laser System is a surgical laser console utilizing an RF-excited CO2 laser tube. It delivers 10,600 nm laser energy via an articulated arm with handpieces/micromanipulators or through fiber adapters for OmniGuide WaveGuide fibers. The system features a touch-screen GUI for controlling lasing modes (continuous, single pulse, repeat pulse), waveforms (CW or Super Pulse), and output power. It includes a helium gas management system for fiber use. Operated by physicians in clinical settings, the device provides precise soft tissue interaction (incision/ablation/coagulation). The GUI is password-protected. The system uses passive optical components (mirrors, lenses) to direct the beam. By enabling controlled tissue removal and coagulation, the device assists surgeons in various surgical specialties, offering a versatile tool for soft tissue management.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluation included laser power output, beam quality, safety controls, emission indicators, switching mechanisms, and fiber/articulated arm integrity. Software verification and validation were performed per IEC 62304. Electrical safety and EMC testing were conducted per IEC 60601-1, IEC 60601-2-22, and IEC 60601-1-2.
Technological Characteristics
RF-excited CO2 laser; 10,600 nm wavelength. Materials: machined/cast aluminum, stainless steel, standard optics. Delivery: articulated arm or fiber adapter. Aiming beam included. GUI-controlled via touch screen. Software-based control. Standards: IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 62304, IEC 62366.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues in dermatology, general surgery, gynecology, head & neck surgery, neurosurgery, oral surgery, orthopedic surgery (soft tissue), otolaryngology (ENT), plastic & reconstructive surgery, podiatry, and urology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
The UltraPulse system (UltraPulse and UltraPulse Duo models, members of the modified Lumenis Family of UltraPulse SurgiTouch CO2 Surgical Lasers) (K151331)
Reference Devices
OmniGuide BeamPath®FELS 25A, CO2 Laser System (K093251)
K132661 — LIGHTSCALPEL · Lightscalpel, LLC · Jan 2, 2014
K121580 — SURELASE CO2 LASER SYSTEM · Clinicon Corp. · Sep 25, 2012
K101555 — CO2 LASER SYSTEM · Advanced Technology Laser Co., Ltd. · Sep 17, 2010
K063001 — SLIM EVOLUTION FAMILY OF CO2 LASERS AND DELIVERY DEVICE ACCESSORIES · Lasering S.R.L. · Mar 7, 2007
K022060 — LUMENIS ULTRAPULSE ENCORE CARBON DIOXIDE SURGICAL LASERS AND DELIVERY DEVICE ACCESSORIES · Lumenis, Inc. · Oct 30, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 28, 2018
OmniGuide, Inc Sam Wade Vice President, Regulatory Affairs and Quality Assurance 4 Maguire Road Lexington, Massachusetts 02421
Re: K180993
Trade/Device Name: Beacon Advanced CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 13, 2018 Received: April 16, 2018
Dear Sam Wade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -53 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180993
Device Name Beacon CO2 Advanced Energy Laser System
Indications for Use (Describe)
The Beacon Advanced Energy CO2 Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues in the following specialties: Dermatology, General Surgery, Gynecology, Head & Neck Surgery, Neurosurgery, Orthopedic Surgery, in soft tissue, Otolaryngology, Plastic & Reconstructive Surgery, Podiatry, Urology.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 7: 510(k) Summary
#### General Provisions
| 510(k) Owner's Name: | OmniGuide, Inc. |
|----------------------|----------------------------------------------------------------------------|
| Registration Number: | 3005350457 |
| Address: | 4 Maguire Road<br>Lexington, MA 02421 |
| Contact Person: | Samuel D. Wade<br>Vice President, Quality Assurance and Regulatory Affairs |
| Phone Number: | (617) 551 4810, Cell: (978) 844-7515 |
| Fax Number: | (617) 551-8445 |
| Classification Name: | Powered Laser Surgical Instrument |
| Regulation: | 21 CFR §878.4810 |
| Regulatory Class: | II |
| Product Code: | GEX |
| Proprietary Name: | Beacon Advanced Energy Laser System |
| Common Name: | Powered Laser Surgical Instrument |
### Name of Predicate Device(s)
- K151331 The UltraPulse system (UltraPulse and UltraPulse Duo models, members of the ● modified Lumenis Family of UltraPulse SurgiTouch CO2 Surgical Lasers)
## Name of Reference Device(s)
- K093251 OmniGuide BeamPath®FELS 25A, CO2 Laser System ●
- K100384 Lumenis FiberLase CO2 Laser WaveGuide ●
- K945648 TTI Medical AccuBeam COs Laser Handpiece ●
- K864378 TTI CO2 Laser Micromanipulator ●
## Indications for Use
The Beacon Advanced Energy CO2 Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues in the following specialties: Dermatology, General Surgery, Gynecology, Head & Neck Surgery, Neurosurgery, Oral Surgery, Orthopedic Surgery in soft tissue, Otolaryngology (ENT), Plastic & Reconstructive Surgery, Podiatry, Urology
## Intended Use
The Beacon Advanced CO2 Advanced Energy Laser System is intended to be used in conjunction with the OmniGuide WaveGuide Fibers or the Articulated Arm to be used for the incision, ablation, vaporization and coagulation of body soft tissues including intra-oral soft tissues.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OmniGuide. The logo consists of two concentric blue circles on the left side of the word "OmniGuide". The word "OmniGuide" is written in a sans-serif font, with the "O" in "OmniGuide" being partially enclosed by the two concentric circles. There is a trademark symbol to the right of the word "OmniGuide".
## Device Description
The Beacon Advanced Energy Laser System is a carbon dioxide laser system utilizing RF excited CO2 laser tube. The device is comprised of a system console, with or without an articulated arm terminated with a handpiece or a micromanipulator, a fiber adapter, system control electronics, a touch screen providing a graphical user interface, a covered footswitch, helium gas management system (for using the OmniGuide Waveguide fibers with the system (K140378, K093451, K093251 and K070157. Located in Appendix 6), and operating software.
The laser system delivers laser energy at 10,600 nm in three different output modes: continuous, single pulse and repeat pulse and two waveforms: Continuous Wave (CW) or Super Pulse (SP). The laser energy produced by the laser tube within the laser console is delivered through an articulated arm system terminating in a handpiece or multiple-use attachments or through a fiber adaptor coupling the laser energy into a OmniGuide. The articulating arm allows the laser energy to be delivered through a focusing handpiece or a micromanipulator achieving laser beam spot sizes in the range of 0.2-0.6 mm in the operating field. The Waveguide fibers have been cleared through K140378, K093451, K093251 and K070157.
The Beacon is operated and controlled via Graphical User Interface implemented by proprietary operating software running on single board computer in the laser console.
The operator can control various aspects of the Beacon operation through the Graphical User Interface on the touch screen. The access to the Graphical User Interface and laser operation is password protected to control access. The operator can adjust the lasing modes, pulse rate and duration in case of pulsed modes, set the waveform (CW or SP) and output power. They can also select whether to use fiber or the articulating arm with the micromanipulator or handpieces or other accessories.
Material used are mainly machined or cast aluminum, stainless steel, standard optics for the transmission or the reflection of the CO2 laser wave energy. It is worth noting that all components (mirrors, lenses, fiber and articulating arm) that the COs light travels through are passive and do not alter the wavelength or any other of the fundamental properties of the CO2 Laser.
## Technological Characteristics
The Beacon has the same intended use and similar indications for use, technological characteristics and operating principles as K151331 The UltraPulse system (UltraPulse and UltraPulse Duo models, members of the modified Lumenis Family of UltraPulse SurgiTouch CO2 Surgical Lasers). The design and components are very similar to the predicate devices have the Graphical User Interface on a touch screen in the front of the device has an articulating arm and an adapter port for the attachment of fibers. Both devices have the laser energy generated within the laser console and delivered to either the articulated arm or the fiber adapter. Each device has an aiming beam to assist in the proper delivery of the treatment beams but aiming beam does not deliver any therapeutic energy. Both devices have similar key design aspects including spot sizes, laser wavelength, ranges of laser operating parameters such as pulse widths and output power. The differences are minor and do not raise any new issues of safety or efficacy of the Beacon device as shown in the performance testing results.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for OmniGuide. The logo consists of the word "OmniGuide" in a simple, sans-serif font. To the left of the word is a graphic of two concentric circles, with the outer circle being slightly thicker than the inner circle. The color of the logo is a light blue.
## Performance Data
### Risk Analysis
Risk analysis was performed according to IEC 14971 and the FDA Guidance Document: Application of risk management to medical devices and reviewed by an independent third party (Intertek) and found to be in compliance.
## Electrical Safety and Electromagnetic Compatibility and other standard testing
Evaluation of the device was conducted by an independent testing laboratory (Intertek) and was found in compliance:
- IEC 60601-1, Medical electrical equipment, Part 1 General requirements for basic safety and essential performance.
- IEC 62366 Medical devices- Application of usability engineering to medical devices
- IEC 60601-2-22 Third Edition 2007-05, Medical electrical equipment-Part2: Particular ● requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1 Ed. 2.0 (2007), Safety of Laser Products- Part 1: Equipment classification and requirements
- IEC 62304:2006 Medical Device software- Software life cycle processes
- IEC 14971:2007/(R) 2010 Medical Devices- Application of Risk Management to Medical Devices
- IEC 60601-1 Medical Electrical Equipment, Part 1-6 General requirement for Safety: Collateral ● Standard Usability
- IEC 60601-1-2 ed 4. EMC
## Non-Clinical Performance Data
The Beacon Advanced Energy CO2 Laser System performance characteristics have been evaluated through testing and analysis of laser power output and beam quality. The ability of new device to withstand variant operation, storage and transportation was tested. The system testing (energy measurements, safety controls, emission indicator, switching mechanism, fiber and articulating arm and aiming beam was also completed. This type of testing complies with the respective section of the FDA Guidance on the Content and Organization of a Premarket Notification for a Medical Laser (1995) and is similar to the predicate device tests. The performance of the Beacon Advanced Energy CO2 Laser System and related parameters of the predicate device are substantially equivalent.
Software verification and validation testing was conducted, and documentation provided as recommended by FDA's "Guidance for the Content of Premarket Submissions Contained in Medical Devices" and IEC 62304:2006 Medical Device Software. Device software verification and validation results were found acceptable for software release
#### Clinical Performance Data
Formal clinical trials were not deemed necessary as the device is using the same technology and intended use as the predicate device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OmniGuide. The logo consists of the word "OmniGuide" in a simple, sans-serif font. To the left of the word is a circular graphic made up of three concentric circles. The logo is clean and professional in appearance.
### Biocompatibility
No part of the laser contacts the patient therefor no biocompatibility testing was necessary
### Summary of Substantial Equivalence
The Beacon Advanced Energy CO2 Laser System and the predicate device have the same intended use as surgical lasers with similar indications for use. The Beacon Advanced Energy CO2 Laser System presents similar technological characteristics as its predicate device, including the laser type, wavelength, device design, waveforms and pulsed mode characteristics, laser beam spot sizes. Although there might be minor differences in the details of the device design, they are not sufficient to raise new questions of safety or efficacy.
### Conclusion
The Beacon Advanced Energy CO2 Laser System is as safe and effective as the Lumenis UltraPulse and UltraPulse Duo Laser Family (K151331). The new device has similar intended use and indications for use, technological characteristics and the same principle of operation as the predicate device. The minor differences in the indications for use do not alter the intended surgical use of the device and do not affect the safety or efficacy when used as labeled. The minor design differences raise no new issues of safety or efficacy. Thus, the Beacon Advanced Energy CO2 Laser and the Lumenis UltraPulse and UltraPulse Duo are substantially equivalent.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for OmniGuide. The logo consists of the word "OmniGuide" in a simple, sans-serif font. The word is accompanied by a circular design to the left, which is made up of concentric rings in a light blue color. There is a trademark symbol to the right of the word "OmniGuide".
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.