ADAPTABLE MONOPOLAR ADAPTABLE AND/OR RETRACTABLE ELECTRODES

K062047 · PAJUNK GmbH Medizintechnologie · GEI · Oct 5, 2006 · General, Plastic Surgery

Device Facts

Record IDK062047
Device NameADAPTABLE MONOPOLAR ADAPTABLE AND/OR RETRACTABLE ELECTRODES
ApplicantPAJUNK GmbH Medizintechnologie
Product CodeGEI · General, Plastic Surgery
Decision DateOct 5, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

Pajunks HF Electrodes are instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures.

Device Story

Pajunk HF Electrodes are modular, reusable, non-sterile surgical instruments used in laparoscopic procedures. The device consists of a stainless steel guidance rod, an insulated tube, and a stainless steel electrode tip (insulated with ceramic or polyamide). It connects to a handle via click-screw, Luer, or M8x1 connector. The device features a retractable/extendable tip mechanism; the electrode can be retracted into the tube to allow for suction/irrigation, then extended for monopolar coagulation or surgical intervention. Operated by a surgeon in an OR setting, the device allows for multi-functional use (suction, irrigation, and coagulation) within a single instrument, reducing the need for multiple tools. The device is supplied non-sterile and requires sterilization before use.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material, and technological comparison to predicate devices. Bench testing was performed to validate the retraction/extension mechanism and sterilization cycles.

Technological Characteristics

Modular monopolar electrosurgical instrument. Materials: stainless steel rod/tip, ceramic or polyamide insulation. Connectivity: M8x1, Luer, or click-screw connectors. Energy: Monopolar electrosurgical energy. Form factor: Retractable/extendable tip with integrated suction/irrigation channel. Sterilization: Reusable, requires sterilization (validated for up to five cycles).

Indications for Use

Indicated for use by surgeons during laparoscopic procedures to grasp, manipulate, dissect, retrieve, biopsy, cut, or coagulate internal tissue or organs.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ \$\omega b24)\$ Image /page/0/Picture/2 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below the word "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, white letters. MEDIZINTECHNOLOGIE page 1 of 4 # Special 510(k) Premarket Notification Submission: Summary of Safety and Effectiveness Date of Preparation: September 19th, 2006 OC7 - 5 2006 | Submitter Information/ production site: | USA Contact: | |----------------------------------------------|--------------------------------------| | Pajunk GmbH | Pajunk USA | | Karl-Hall-Strasse 01 | German American Trade Center | | 78187 Geisingen | 5126 South Royal Atlanta Drive | | Germany | 30084 Tucker, Georgia | | Fon: +49(0)7704-9291-0 | USA | | Fax: +49(0)7704-9291-610 | | | Establishment Registration Number: 9611612 | | | Contact: | Contact | | Christian Quass, Director Regulatory Affairs | Stefan Dayagi | | Fon: +49(0)7704-9291-586 | Fon: +01(0)770-493-9305 | | Fax: +49(0)7704-9291-605 | | | E-Mail: christian.quass@pajunk.com | E-Mail: stefan.dayagi@pajunk-usa.com | # Device Information: | Trade Names: | Pajunks HF Electrodes | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Monopolar Electrodes | | Classification Name: | Electrosurgical cutting and coagulation device and accessories | | Classification Reference: | 21 CFR 878.4400, April 1. 2005 | | Proposed Classification: | Regulatory Class: II | | Proposed Product Code: | GEI | | Predicate Devices: | 1. Pajunks monopolar Ceramic Tip electrodes K053282<br>2. Pajunks Modular valve handles for adaptable tubes K011997<br>3. Pajunks monopolar electrodes marketed under K033249 | {1}------------------------------------------------ K.obw2 **PAJUNK** MEDIZINTECHNOLOGIE Page 2 of 4 ### Device Description: Pajunks MIS-Instruments and HF-electrodes are used during minimal invasive surgery, according to the professional surgeon and the instructions for use. The Pajunk adaptable and retractable HF electrodes are multiple uses, non-sterile delivered and latex free medical devices. They are intended for transient delivery of energy required for coaqulation of tissue. The modular system for optional monopolar coagulation enables a surgeon to manipulate, The mound by commer of of on coagulate internal tissue or organs while performing laparoscopic procedures. The instruments are insulated for optional monopolar coagulation. The Pajunk MIS-System contains but is not limited to HF-electrodes for comfortable and safe rne i ajain with and without suction/ irrigation and microsurgical instruments with and without suction/irrigation. Pajunk provides electrodes with and without suction and irrigation element and electrodes with r ajurik provides ticonooo with retractable/ extendable tips. Tubes for suction/irrigation are available seperately also. Pajunks electrodes consist of a stainless steel guidance rod, insulated tube, ceramic or polyamide insulation at tip and stainless steel electrode tip for coagulation. Isolation at tip is made by ceramic or poly amid and already cleared for market. The electrodes are fixed to handle via click-screw, Luer or Pajunks M8x1 connector. Pajunks electrode and instrument series are modular and interchangeable as stated in the user instructions. Disposable inserts are also available. #### Substantial equivalenve: basis for submission The subject of Pajunks K011997 are Pajunks multipurpose handles (Modular valve handles) already designed for use with the HF Electrodes subject of this submission. In K033249 FDA cleared HF Electrodes and instruments for either coagulation, grasping and so on or suction/irriragtion (channel only). The subject of this PMN is the combination of both: HF Electrodes within a suction/irrigation channel and optional retraction-extension. There are no changes in material or technique in fact; the subject of this PMN is a rners are no onarges in the predicates. Pajunks HF electrodes subject of this PMN optionally are for suction/irrigation and coagulation at the same time with the same electrode enabled by retraction-extension feature. The electrode can be drawn back into the tube – re-tracted – for suction/irrigation procedures. Afterwards it can be comfortably extended to go on coagulating (or whatever the surgical intervention is intended to). The combination of this two cleared technologies does increase safety and comfortability of the rne odminitiation it it is weriod the surgical intervention takes place. Insttead of employing two different instrument or an instrument inside a working channel The retractable electrode combine both, suction/irrigation and surgical intervention. {2}------------------------------------------------ bzw4) Page 3 of 4 Pajunk's adaptable and retractable HF electrodes are substantially equivalent to Pajunk's r guilk o additional and rotratia under 510(k) number K033249. All materials, dimensions, fixation specifications and intended use are identical. A retraction/extension feature has been added. Pajunk's monopolar adaptable and retractable HF electrodes are equivalent in design, physical dimensions, metal and plastics materials, to Pajunks monopolar ceramic/ polyamid tip electrodes marketed under 510(k) number K053282, K033249 and are intended to be used with Pajunks Modular Valve handle system with different interchangeble fixation methods (click-screw, Luer, M8x1) cleared for market under K011997. #### Technological Equivalence | | Predicate device | Subject device | Predicate device | |-------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | | K033249, K053282 | K062047 | K011997 | | Shape of electrode tip<br>tip | grasping forceps,<br>biopsy, clamps and<br>scissors, hook, right-<br>angle, spatula, knife,<br>45°, needle | Retractable/ extendable:<br>hook, right-angle,<br>spatula, knife. 45°,<br>needle<br>Suction/ Irrigation:<br>grasping forceps,<br>biopsy, clamps and<br>scissors, hook, right-<br>angle, spatula, knife,<br>45°, needle | Not applicable | | Connector | M8 x 1 or Luer or click-<br>screw | M8 x 1 or Luer or click-<br>screw | M8 x 1 or Luer or click-<br>screw | | Features | Coagulation<br>Suction/ irrigation | Suction/ Irrigation<br>Coagulation<br>Retraction/ Extension | Suction/ Irrigation<br>Coagulation | #### Sterilization Pajunks HF Electrodes are supplied non-sterile. They have to be sterilized before initial use and before each further use. They can be sterilized up to five times (validation report in sect. 14 of this submission). #### Biocompatibility Biocompatibility information for Pajunk's adaptable and retractable HF is located in Section 15.0 of this submission. #### Indications for use Pajunks HF Electrodes are instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures. {3}------------------------------------------------ Kob2047 Image /page/3/Picture/2 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, black letters. Below the company name is the word "MEDIZINTECHNOLOGIE" in a smaller font size. The logo appears to be extracted from a document. ﺮ ## Conclusion: The comparison between the predicate devices and the proposed device demonstrates that The companson between the predicate devices and the proposed and substantially equivalent to Pajunks predicate device. ッ {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wings, representing health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT - 5 2006 Pajunk GmbH % Mr. Christian Quass Director Regulatory Affairs Karl-Hall-Strasse 01 78187 Geisingen Germany Re: K062047 Trade/Device Name: Pajunks HF Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 25, 2006 Received: September 27, 2006 Dear Mr. Quass: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FIDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2 - Mr. Christian Quass forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, . "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Barbave Buchum Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ : Image /page/6/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below that, in smaller letters, is the word "MEDIZINTECHNOLOGIE". | Indications for use | |---------------------| |---------------------| | 510(k) Number: | K062047 | |----------------|-----------------------| | Device Name: | Pajunks HF Electrodes | | Indications for Use: | |----------------------| |----------------------| **Pajunks HF Electrodes** are instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures. | Prescription Use<br>(Per 21 CFR 801.109) | <b>X</b> AND/OR Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |------------------------------------------|----------------------------------------------------------------| |------------------------------------------|----------------------------------------------------------------| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | K062047 | |---------------|---------| |---------------|---------| | Page 1 of 1 | |-------------| |-------------| : : 0 0 4 · : ・・
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...