MODULAR VALVE HANDLES FOR ADAPTABLE TUBES, MODEL 1298-XX-XX, 1295-XX-XX

K011997 · Pajunk GmbH · GCJ · Jul 10, 2001 · Gastroenterology, Urology

Device Facts

Record IDK011997
Device NameMODULAR VALVE HANDLES FOR ADAPTABLE TUBES, MODEL 1298-XX-XX, 1295-XX-XX
ApplicantPajunk GmbH
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 10, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

Systems for irrigation and aspiration of rinsing liquid in the operating area.

Device Story

Modular multifunction suction/irrigation system; includes pistol handle, maintenance-free 2-way valves, and adaptable tubes. Features lockable instrument channel (5mm) for laparoscopic procedures; channel opens automatically upon instrument insertion and closes upon removal; manual override button allows channel opening for instrument exchange (e.g., hook-type or biopsy). System components are interchangeable via M8x1 mm connectors or LuerLock connectors. Used in laparoscopic surgery; operated by surgeons/clinical staff. Device facilitates fluid management in the operating field; allows simultaneous irrigation and aspiration; enables instrument passage through the same handle. Dismountable/reassemblable without tools for cleaning.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Modular suction/irrigation system; pistol handle design; maintenance-free 2-way valves; M8x1 mm and LuerLock connectors; 5mm instrument channel. Mechanical operation; no energy source required. Manual assembly/disassembly without tools.

Indications for Use

Indicated for irrigation and aspiration of rinsing liquid in the laparoscopic operating area.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Submission Summary (Full Text)

{1}------------------------------------------------ ## 510(k) Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92. | Applicant: | Pajunk GmbH<br>Am Holzplatz 5 - 7<br>D-78187 Geisingen<br>Germany<br>Tel. 07704 9291-0 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Martin Hauger<br>Technical Director | | Device Identification: | Common Name:<br>modular valve handles for adaptable tubes, adaptable<br>suction/irrigation tubes<br>Trade Name:<br>modular valve handles for adaptable tubes, adaptable<br>suction/irrigation tubes | # Indication and Device Description: ### General Oenchal modular multifunction suction/irrigation system with pistol handle, maintenance i he Fajant modular many different adaptable tubes and lockable instrument channel for iree & way varro, marry af rinsing liquid in the laparoscopic operating area. The instrument imgation and aspliation of this instruments are inserted and closes when they are channel adomatioally opene would the instrument channel can be opened manually by removed. Allemativery "are xample to remove a hook-type instrument or a biopsy. The prossing a pachbutter.inted/ assembled for cleaning without any tools needed. #### Modular valve handles for adaptable tubes Modular Turve fiction handles with trumpet valves or maintenance-free 2-way valves The T ajurit Suchon inigation hardison her in the dismounted without any tools. Each handle can be used with any tube having a M8x1 mm connector; i.e. all handles of the series 1298-XX-00 are completely interchangeable. #### Adaptable suction/irrigation tubes The Pajunk adaptable suction/irrigation tubes with connector M8x1 mm can be used with any handle having a M8x1 mm connector. Adaptable suction/irrigation tubes with large LuerLock connector female can be used with any handles of series 1295-XX-00. ### Modular multifunction suction/irrigation system with instrument channel Modular multifunction suction/irrigation system with pistol handle, maintenance-free 2-way walve, many different adaptable tubes and lockable instrument channel Ø 5mm for irrigation valve, many direront adaptable the laparoscopic operating area. The instrument channel automatically opens when instruments are inserted and closes when they are removed. automativally the closure of the instrument channel can be opened manually by pressing a Akerhutton for example to repove a hook-type instrument or a biopsy. The system can be dismounted/assembled for feaning without any tools need Signed: Mr. Horst Pajunk General Manager tools needed. Mr.Heinrich Rajunk General Manager {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 1 0 2001 Pajunk GmbH c/o Mr. Mark Job TÜV Product Service 1775 Old Highway 8 New Brighton, Minnesota 55112 Re: K011997 Trade/Device Name: Modular Valve Handles for Adaptable Tubes Regulation Number: 876.1500 Regulatory Class: II Product Code: GCJ Dated: June 13, 2001 Received: June 27, 2001 Dear Mr. Job: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Mr. Mark Job This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonification. THE I Drivin a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and IT you desire specific arriver Jamostic devices), please contact the Office of Compliance at additionally 607.10 for questions on the promotion and advertising of your device, (2017) 594-1639. entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mormation on your responsive at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerety yours, Mark N Melkinn Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Medizintechnologie Page_1_of_i 510(k) Number (if known): KO 11997 Device Name: _modular valve handles for adaptable tubes Indications For Use: _suction/irrigation Systems for irrigation and aspiration of rinsing liquid in the operating area. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation le Mark n. Melleker Restorative d Neurologic 510(k) Number K011497 (Optional Format 3-10-98)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...