K221336 · Revelle Aesthetics, Inc. · OUP · Aug 5, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221336
Device Name
Avéli
Applicant
Revelle Aesthetics, Inc.
Product Code
OUP · General, Plastic Surgery
Decision Date
Aug 5, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4790
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Avéli is indicated for long-term reduction in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating treatment benefits through one year of observation.
Device Story
Avéli is a sterile, single-use manual surgical instrument for cellulite reduction. Used in clinics by physicians, the device features a handle and a distal end with an integrated light source. The distal end is inserted subcutaneously through a small incision. The user advances the device to the procedure site, using the light for tracking. A blocker mechanism forms a hook to capture fibrous septa; the user confirms engagement by feeling resistance. The user then exposes a blade to release the septa while applying manual pressure to the skin. The device is retracted after the blocker is withdrawn. The procedure is repeated for each cellulite depression. The device provides controlled mechanical cutting of fibrous tissue. Clinical benefits include long-term reduction in cellulite appearance, with durability demonstrated through one year of observation.
Clinical Evidence
Prospective, nonrandomized, multi-center study (n=74) in US/Australia. Participants (females 21-55, BMI < 30) received a single treatment. Primary effectiveness endpoint: mean change in Cellulite Severity Score (CSS) > 1 at 3 months, assessed by blinded physician reviewers. Results showed CSS improvement sustained through 12 months. Safety endpoint: no device-related SAEs at 30 days. Common AEs: ecchymosis (86.8%), tenderness (51.5%), pain (38.2%), induration (36.8%), numbness (17.6%).
Technological Characteristics
Manual, single-use surgical instrument. Distal end made of stainless steel. Features integrated light source for illumination. Mechanism: controlled mechanical cutting of fibrous septa via hook and blade. Sterilized per ISO 11135-1. Complies with IEC 60601-1 (electrical safety) and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for long-term reduction in the appearance of cellulite in the buttocks and thigh areas of adult females (ages 21-55, BMI < 30).
Regulatory Classification
Identification
A powered surgical instrument for improvement in the appearance of cellulite is a prescription device that is used for the controlled release of subcutaneous tissue for improvement in the appearance of cellulite. The device consists of a cutting tool powered by a motor and a means for instrument guidance to control the areas of subcutaneous tissue cutting underneath the cellulite depressions or dimples.
Special Controls
In addition to the general controls of the FD&C Act, the Cabochon System is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.
(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.
K232153 — Avéli · Revelle Aesthetics, Inc. · Aug 18, 2023
K150505 — Cellfina System · Ulthera, Inc. · Jul 29, 2015
K134010 — CABOCHON SYSTEM · Cabochon Aesthetics, Inc. · Apr 14, 2014
K192185 — The Cellfina System · Ulthera, Inc. · Oct 9, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration".
August 5, 2022
Revelle Aesthetics, Inc. Melissa Viotti Sr. Director, Quality & Regulatory Affairs 2570 W. El Camino Real, Suite 310 Mountain View, California 94040
Re: K221336
Trade/Device Name: Avéli Regulation Number: 21 CFR 878.4790 Regulation Name: Powered Surgical Instrument For Improvement In The Appearance Of Cellulite Regulatory Class: Class II Product Code: OUP Dated: May 6, 2022 Received: May 9, 2022
Dear Melissa Viotti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221336
Device Name Avéli
Indications for Use (Describe)
Aveli is indicated for long-term reduction in the appearance of cellulite in the buttocks and thigh areas as supported by clinical data demonstrating treatment benefits through one year of observation.
**Type of Use (Select one or both, as applicable)**X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Revelle Aesthetics, Inc. Traditional 510(k) The Avéli Precision Cellulite Release Device
#### Submitter Information
| Submitter: | Revelle Aesthetics, Inc.<br>2570 W. El Camino Real, Suite 310<br>Mountain View, CA 94040 | | |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact: | Melissa Viotti<br>Senior Director, Quality and Regulatory Affairs<br>Telephone and Fax: 650-336-5985<br>Email: mviotti@revelleax.com | | |
| Date Summary Prepared: | August 5, 2022 | | |
| Subject Device Information | | | |
| Device Trade Name: | Avéli(TM) | | |
| Common Name: | Avéli(TM) | | |
| Regulation Number: | 21 CFR 878.4790 | | |
| Product Code: | OUP | | |
| Predicate Device Information | | | |
| Predicate Device: | Avéli (K212399) | | |
| Predicate Device Manufacturer: | NC8, Inc. (now Revelle Aesthetics) | | |
#### Device Description
Avéli is a sterile, single-use manual instrument that releases fibrous tissue (septa) beneath cellulite for long-term reduction in the appearance of cellulite in the buttocks and thigh areas of adult females. The device consists of a Handle and a Distal End. The Handle houses components used to actuate the moving parts at the distal end of the device. The Distal End is advanced into subcutaneous tissue through a small incision to a procedure location. An integrated light source provides illumination and allows the user to track and advance to the procedure location. The Distal End contains a Blocker forming a Hook. When the Handle is moved in a retrograde fashion, the Hook next to fibrous septa captures the septa resulting in tugging. The user feels the resistance, confirming that septa under a cellulite depression have been identified and then exposes the Blade at the Distal End. The user pushes the skin distally with the free hand while maintaining the device stable or applies additional retrograde motion with the device to release the fibrous septa. The user then retracts the Blocker into the device, allowing
{4}------------------------------------------------
removal without further tissue engagement. The user can verify all contributing septa have been released by passing through the area again with the Hook. The step is repeated for each visible cellulite depression.
## Indications for Use
Avéli is indicated for long-term reduction in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating treatment benefits through one year of observation.
### Comparison of Technological Characteristics with the Predicate Device
A comparison of Avéli to the predicate device is provided in the table below. This table demonstrates that the subject device is substantially equivalent to the predicate device for its indication for use, performance, and technological characteristics.
| | SUBJECT DEVICE | PREDICATE DEVICE |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Avéli | Avéli |
| 510(k) Number | TBD | K212399 |
| Manufacturer | Revelle Aesthetics, Inc. | Same, NC8, Inc. (name changed to<br>Revelle Aesthetics, Inc.) |
| Device Class | II | Same, II |
| Regulation Number | 878.4790 | Same, 878.4790 |
| Product Code | OUP | Same, OUP |
| Clearance Date | TBD | October 22, 2021 |
| Intended Use | Avéli is intended for long-term<br>reduction in the appearance of cellulite<br>in the buttocks and thigh areas of adult<br>females as supported by clinical data<br>demonstrating treatment benefits<br>through one year of observation. | Avéli is intended for temporary<br>reduction in the appearance of cellulite<br>in the buttocks and thigh areas of adult<br>females as supported by clinical data<br>demonstrating benefits through three<br>months of observation. |
| Indications for Use | Avéli is indicated for long-term<br>reduction in the appearance of cellulite<br>in the buttocks and thigh areas of adult<br>females as supported by clinical data<br>demonstrating treatment benefits<br>through one year of observation. | Avéli is indicated for temporary<br>reduction in the appearance of cellulite<br>in the buttocks and thigh areas of adult<br>females as supported by clinical data<br>demonstrating benefits through three<br>months of observation. |
| Where Used | Clinic/Doctor's office | Same, Clinic/Doctor's office |
| Anatomical Site | Buttocks and thighs | Same, Buttocks and thighs |
| Technological<br>Characteristics | Blade dissects soft tissue | Same, Blade dissects soft tissue |
{5}------------------------------------------------
| Device Name | SUBJECT DEVICE | PREDICATE DEVICE |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| | Avéli | Avéli |
| Material of Distal End<br>Dissecting Zone | Stainless steel | Same, Stainless steel |
| Illumination Feature | Yes | Same, Yes |
| Controlled Dissection | Yes | Same, Yes |
| Biological Effect | Controlled mechanical cutting of the<br>fibrous tissue which contributes to the<br>appearance of cellulite. | Same, Controlled mechanical cutting of<br>the fibrous tissue which contributes to<br>the appearance of cellulite. |
| Single Use | Yes | Same, Yes |
| Sterile | Yes | Same, Yes |
| Electromagnetic<br>Compatibility<br>Standards | Compatible | Same, Compatible |
| Medical Electrical<br>Equipment Safety<br>Standards | Compatible | Same, Compatible |
## Performance Data
The following performance testing was conducted to verify that the subject device meets all design specifications in support of the substantial equivalence determination.
-Verification Testing:
- 1) Visual Inspections
- 2) Dimensional Inspections
- 3) Functional testing including:
- Simulated use testing .
- Force measurements .
- . Tensile testing
- Mechanical testing .
- Electrical testing ●
- Optical output testing ●
-Electrical Safety Testing (IEC 60601-1:2005 +A1:2012, IEC 60601-1-2:2014, IEC 60601-1-6:2010 + AMD1:2013/IEC 62366-1:2020)
-Biocompatibility Testing (ISO 10993-1:2018)
-Sterilization Validation (ISO 11135-1:2014, ISO 14161:2009)
{6}------------------------------------------------
### Clinical Performance Data
The in vivo performance of Avéli was evaluated in a pivotal clinical study. A prospective, nonrandomized, multi-center study was conducted to evaluate the safety and effectiveness of Avéli for the treatment of cellulite in the buttock and thigh areas of 74 participants across nine (9) investigational sites in the US and Australia. All participants served as their own control and underwent a single treatment with Avéli. Participants underwent follow-up assessments at 1 day (virtual or clinio), 30 days, 3 months, 6 months, and one year post-treatment.
The participant inclusion/exclusion criteria limited inclusion to females between the ages of 21 and 55 with moderate to severe cellulite in the thighs and/or buttocks and a body mass index (BMI) less than 30. The enrolled study population included females between the ages of 26 and 54 with moderate to severe cellulite and BMI between 19 and 29.8. In addition, participants were excluded that smoked or had recently quit smoking (within the last 6 months). The participants were asked to rate their pain and satisfaction with their appearance. Photographs were taken under standardized conditions in accordance with a study photography manual at baseline and each follow-up visit. Outcomes were assessed by three, independent blinded reviewers using photographs before treatment and 3 months, and one year post-treatment to verify the effectiveness of the procedure. The one year results demonstrate safety and durability through one year.
No Unanticipated Adverse Device Effects (UADE), related Serious Adverse Events (SAE) nor severe adverse events occurred in the study. The primary safety endpoint for the study was achieved with no device related SAEs at 30 days. Any undesirable medical occurrence was considered an adverse event. There were three adverse device effects (ADEs) that occurred in two participants: 1) an extended incision to facilitate removal of device in Hook Position, 2) a skin laceration (~1mm), and 3) a small scar from the skin laceration. The types and rates of the observed events are typical of this mechanism of action and were generally mild, transient, and only four events involved interventions to resolve: (1) aeration of ~2x1cm discreet subcutaneous mass (2) seroma requiring aspiration (3) hematoma requiring aspiration, and (4) stitching an extended device entry incision. The most common AEs were ecchymosis (86.8%), tenderness (51.5%), pain (38.2%), induration (36.8%) and numbness (17.6%). Ecchymosis had a median duration of 27 days, tenderness a median duration of 21 days, induration a median of 172 days, and numbness a median of 58 days. Many participants (83.8%, 57/68) experienced tenderness or pain within the first 24 hours. Some participants (13/68, 19.1%) could return to normal activities the day of the procedure, and most (50/68, 73.5%) within a week. Of the most common adverse events reported, induration had the longest mean duration. Generally, the induration was described as small areas of firmness, not visible or painful, and not associated with the incision site. A small proportion of adverse events (6.0%) were ongoing at study exit (one year) and included induration, numbness, skin hyperpigmentation, skin dysaesthesia, and skin indentation (also described as a depression or other irregularity). The skin indentations (2/68, 2.9%) presented at 3 months and 6 months, respectively. No medical intervention was required.
In the clinical study, effectiveness was evaluated with an improvement assessment by independent physician evaluation of participant photographs. Participants were overall satisfied with their cellulite procedure results. The participant photo evaluation was conducted by an independent firm in accordance
{7}------------------------------------------------
with the study protocol whereby the evaluators were disclosed nothing about the sponsor, the investigational device, or the clinical investigators. A total of three independent physician evaluators were selected, individually trained, and monitored throughout the evaluation. In the evaluation, blinded before (baseline) and after photographs were provided side by side in randomized orientation (L-R) and the physician evaluators were asked to identify the baseline and after photographs and rate the overall improvement according to a Global Aesthetic Improvement Scale GAIS and the Cellulite Severity Scale (CSS). All reliability and repeatability measures were met at 3 months and validated the methodology.
The overall study success criterion was achievement of a safe and clinically significant improvement in the appearance of cellulite in the treated subjects. The primary effectiveness endpoint was to demonstrate that the mean change (improvement) in the Cellulite Severity Score (CSS) is more than 1 for the study population, as determined by the independent, blinded physician evaluators comparing baseline and 3month photos. The primary effectiveness endpoint was achieved. The CSS evaluations at 6 and 12 months demonstrated durability through one year.
In conclusion, the study primary endpoints were achieved, and results were sustained at one year. The in vivo performance data collected in the pivotal study demonstrates that Avéli is both safe and effective for treating cellulite in the buttock and thigh locations.
## Conclusion
This Traditional 510(k) Premarket Notification is to extend treatment benefits in the indications for use through one year of observation. The updated indications for use any new questions of safety or efficacy. Clinical study data demonstrates treatment benefits through one year of observation. Therefore, Avéli device is substantially equivalent to the predicate device in terms of safety and effectiveness for long-term reduction in the appearance of cellulite in the buttocks and thigh areas of adult females.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.