K192185 · Ulthera, Inc. · OUP · Oct 9, 2019 · General, Plastic Surgery
Device Facts
Record ID
K192185
Device Name
The Cellfina System
Applicant
Ulthera, Inc.
Product Code
OUP · General, Plastic Surgery
Decision Date
Oct 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4790
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cellfina System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits through five years of observation.
Device Story
Cellfina System provides precise focal release of subcutaneous tissue to improve cellulite appearance; consists of sterile, single-use disposable kit (CK1) and electromechanical, 50-use motor module (CM1). Used by physicians in clinical settings. Device mechanically releases fibrous septae causing cellulite dimples. Output is physical tissue release; healthcare provider uses device to treat specific cellulite depressions. Benefits include long-term reduction in cellulite severity, as demonstrated by clinical observation up to five years.
Clinical Evidence
Prospective, multi-center, non-randomized open-label study (IDE G120116) with 55 female subjects (ages 18-55, BMI <35). Primary endpoint: ≥1 point reduction in Cellulite Severity Scale at 5 years; mean improvement 1.8 points (p<0.0001). Secondary endpoint: >60% of subjects showed ≥1-grade improvement; 86.5% achieved this (95% LCL 74.7%). 100% of subjects free from serious adverse events at 5 years. 78% of subjects satisfied/very satisfied.
Technological Characteristics
Electromechanical powered surgical instrument. System comprises a sterile, single-use disposable kit (CK1) and a non-sterile, 50-use motor module (CM1). Compliant with IEC 60601-1-2:2014 (4th Edition) for electromagnetic compatibility. Biocompatible patient-contacting components. Sterility and shelf-life validated for disposable components.
Indications for Use
Indicated for long-term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females.
Regulatory Classification
Identification
A powered surgical instrument for improvement in the appearance of cellulite is a prescription device that is used for the controlled release of subcutaneous tissue for improvement in the appearance of cellulite. The device consists of a cutting tool powered by a motor and a means for instrument guidance to control the areas of subcutaneous tissue cutting underneath the cellulite depressions or dimples.
Special Controls
In addition to the general controls of the FD&C Act, the Cabochon System is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.
(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 9, 2019
Ulthera, Inc Lisa Pray Regulatory Affairs Manager 1840 S. Stapley Dr. Suite 200 Mesa, Arizona 85204
Re: K192185
Trade/Device Name: The Cellfina System Regulation Number: 21 CFR 878.4790 Regulation Name: Powered Surgical Instrument for Improvement in the Appearance of Cellulite Regulatory Class: Class II Product Code: OUP Dated: August 9, 2019 Received: August 12, 2019
Dear Lisa Pray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192185
Device Name Cellfina System
Indications for Use (Describe)
The Cellfina System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits through five years of observation.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <span style="font-size: large;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: large;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
This 510(k) Summary for the Cellfina System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.
| Applicant: | Ulthera, Inc. |
|-----------------------------|-----------------------------------------------------------------------------------|
| Address: | 1840 South Stapley Drive<br>Suite 200<br>Mesa, AZ 85204 |
| Contact Person: | Lisa Pray, Regulatory Affairs Manager |
| Telephone: | (480) 828-8856 |
| Fax: | (480) 619-4071 |
| Submission Date: | August 9, 2019 |
| Device Trade Name: | The Cellfina System |
| Common Name: | Powered surgical instrument for improvement in the appearance<br>of cellulite. |
| Classification: | Regulatory Class II |
| Classification Name: | Powered surgical instrument for improvement in the appearance<br>of cellulite. |
| Regulation Number: | 21CFR 878.4790 |
| Product Code: | OUP |
| Legally Marketed Predicate: | The Cellfina System, 510(k): K161885 |
| Applicable Guidance: | The following guidance special controls are applicable to the<br>Cellfina System: |
#### Class II (special controls). The special controls for this device are:
(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
Image /page/3/Picture/9 description: The image shows the word "ulthera" in a stylized font. The "ul" is stacked on top of the "thera". The "ul" is in a light blue color and is made up of three vertical bars. To the right of the "ul" and above the "thera" are three orange circles in a horizontal row. To the right of the "thera" is a small trademark symbol.
{5}------------------------------------------------
# K192185 The Cellfina System
(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.
(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.
- Device Description: The Cellfina System is intended to provide precise focal release of subcutaneous tissue for improvement in the appearance of cellulite. The system consists of a sterile, single-use, disposable kit (CK1) and an electromechanical, non-sterile, 50-use motor module (CM1).
- Indications for Use: The Cellfina System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits through five years of observation.
#### Substantial Equivalence Comparison:
The subject device remains unchanged from the predicate device. There has been no design, material, shelf life, cleaning and disinfection, software, or packaging updates since the predicate 510(k) submission (K161885). The subject device instructions for use have been updated to reflect compliance to IEC 60601-1-2 and to include the expanded indications for use statement.
#### Non-clinical Performance Data:
There have been no device changes to the subject device since the predicate device 510(k) K161885 submission. As such, all non-clinical performance data presented within the predicate device submission remains applicable to the subject device.
The only non-clinical performance data included in this submission is updated electromechanical safety testing. The predicate device of this Traditional 510(k) submission was compliant to IEC 60601-1-2:2007 3rd Edition; however, the subject device is now compliant to IEC 60601-1-2:2014 4th Edition.
#### Clinical Performance Data (In-Vivo):
The safety and effectiveness of the Cellfina System was evaluated in a pivotal clinical study conducted under IDE G120116. No additional treatments were performed at any followup time points. The overall study success was met (as defined within the protocol), despite the reduction in follow up participation at 5 years with a clinically significant, long term improvement in the appearance of cellulite in the treated subjects as shown within Table 05-1.
Image /page/4/Picture/12 description: The image shows the word "ulthera" in a sans-serif font. The letters are a light gray color. Above the "era" portion of the word are three orange circles. To the right of the word is a small trademark symbol.
{6}------------------------------------------------
# K192185 The Cellfina System
| Study Design: | Prospective, multi-center, non-randomized open label, safety<br>and<br>effectiveness study |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample Size | 55 Subjects participated in the baseline assessment and treatment |
| Follow up (N) | 3 Day (55), 14 Day (54), 1 Month(54), 3Month (55), 6 Month (52), 1 year<br>(50), 2 year (52), 3 year (45), 5 year (37) |
| Inclusion<br>Criteria | Female between the ages of 18 and 55, BMI less than 35 |
| Exclusion<br>Criteria | No prior drug treatment for contouring or cellulite during prior 90 days; no<br>prior liposuction of thighs or buttocks, subject has had >10% weight loss<br>in prior 6 months; subject has known difficulty with local anesthesia,<br>subject is pregnant, other minor criteria |
| Primary<br>Endpoint result<br>at 5 years | The primary endpoint was met: achievement of ≥1 point average reduction<br>in the 0-5 point Cellulite Severity Scale as determined by independent<br>physician assessment of subject photographs taken before and (5 years)<br>after treatment.<br>The mean improvement was 1.8 points (p<0.0001) with a 97.5%<br>upper confidence limit improvement of 1.4 points. |
| Secondary<br>Efficacy result<br>at 5 years | The powered secondary endpoint was met: improvement of one grade or<br>more in severity (none, mild, moderate, severe) in >60% of treated subjects<br>as determined by independent physician assessment of subject photographs<br>taken before and 5 years after treatment.<br>86.5% of treated subjects had ≥1-grade improvement with a 95%<br>lower confidence limit of 74.7%. |
| Primary Safety<br>Result | The safety endpoint was met: 100% of subjects were free from Serious<br>Adverse Events (SAE) directly attributable to the Cellfina System or<br>procedure at 5 years |
| Additional<br>Measures | All additional secondary measures were also achieved or acceptable at the<br>5 year time point:<br>The mean overall rate of correct selection was 92.8% with a lower<br>95% confidence limit of 86.3%.100% of the subjects had noticeable improvement by the GAIS and<br>35% were characterized as having Much Improved or Very Much<br>Improved78% of the subjects were either Satisfied or Very Satisfied.No subjects reported pain in the treatment area using an 11-point<br>numerical rating scale at the 5-year follow-up visit. |
### Table 05-1: Clinical Study Design Characteristics
### Conclusion:
This Traditional 510(k) Premarket Notification is for an expanded indication for use statement. The included clinical data demonstrates that the Cellfina System is both safe and effective for the long term improvement in the appearance of cellulite in the buttocks
Image /page/5/Picture/5 description: The image shows the word "ulthera" in a stylized font. The word is in lowercase letters and is colored in a light blue-gray color. Above the "era" portion of the word are three yellow dots arranged horizontally. To the right of the dots is the trademark symbol.
{7}------------------------------------------------
# K192185 The Cellfina System
and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits through five years of observation.
No design modifications to the Cellfina System motor module (CM1) or disposable kit (CK1) have occurred since the predicate device submission, 510(k) K161885. Updated labeling has been provided as part of this 510(k) submission for updated compliance to IEC 60601-1-2:2014.
These updated indications for use and labeling do not pose any new questions of safety, product output or efficacy. The subject device Cellfina System is as safe, as effective, and performs as well as the legally marketed predicate device (K161885), and as such, is substantially equivalent.
Image /page/6/Picture/4 description: The image shows the logo for Ulthera. The logo consists of the word "ulthera" in a sans-serif font, with the "u" and "l" in a light blue color and the rest of the word in a darker blue. To the right of the word, there are three orange circles arranged horizontally, followed by the "TM" symbol in a smaller font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.