SNAP WOUND CARE DEVICE

DEN080011 · Spiracur, Inc. · OKO · Aug 7, 2009 · General, Plastic Surgery

Device Facts

Record IDDEN080011
Device NameSNAP WOUND CARE DEVICE
ApplicantSpiracur, Inc.
Product CodeOKO · General, Plastic Surgery
Decision DateAug 7, 2009
DecisionDENG
Submission TypePost-NSE
Regulation21 CFR 878.4683
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Indicated for patients who would benefit from a suction device particularly as the device may promote wound healing. The SNaP Wound Care Device is further indicated for removal of small amounts of exudates from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions and flaps.

Device Story

SNaP Wound Care Device is a non-powered suction apparatus for negative pressure wound therapy (NPWT). Device applies negative pressure to wound sites to remove fluids, including exudate, irrigation fluids, and infectious materials; promotes wound healing. Intended for use on chronic, acute, traumatic, subacute, and dehisced wounds, burns, ulcers, and surgical incisions. Device is non-powered, distinguishing it from traditional electric NPWT pumps. Operates by creating suction to manage wound environment. Used in clinical or home settings; managed by healthcare providers or patients. Benefits include fluid removal and potential acceleration of healing process.

Technological Characteristics

Non-powered suction apparatus for NPWT. Mechanical suction mechanism. Subject to Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).

Indications for Use

Indicated for patients requiring negative pressure wound therapy for wound management, including removal of exudate, irrigation fluids, and infectious materials from chronic, acute, traumatic, subacute, and dehisced wounds, partial-thickness burns, ulcers (diabetic or pressure), and surgically closed incisions and flaps.

Regulatory Classification

Identification

A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.

Related Devices

Submission Summary (Full Text)

{0} DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center – WO66-G609 Silver Spring, MD 20993-0002 AUG - 7 2009 Spiracur, Inc. % Ms. Janice Hogan Regulatory Counsel Hogan & Hartson LLP 1835 Market Street, Suite 2820 Philadelphia, Pennsylvania 19103 Re: K081406 SNaP Wound Care Device Evaluation of Automatic Class III Designation Regulation Number: 21 CFR 878.4683 Classification: II Product Code: OKO Dear Ms. Hogan: The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA) has completed its review of your Evaluation of Automatic Class III Designation Petition (de novo) for classification of the SNaP Wound Care Device, a prescription device under 21 CFR Part 801.109 that is indicated for patients who would benefit from a suction device particularly as the device may promote wound healing. The SNaP Wound Care Device is further indicated for removal of small amounts of exudates from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions and flaps. FDA concludes that this device, and substantially equivalent devices of this generic type, should be classified into class II. This order, therefore, classifies the SNaP Wound Care Device, and substantially equivalent devices of this generic type into class II under the generic name, Non-powered suction apparatus device intended for negative pressure wound therapy (NPWT). FDA identifies this generic type of device as: "A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps and grafts." In accordance with section 513(f)(1) of the Federal Food, Drug, and Cosmetic Act (21 U.S.C. 360c(f)(1)) (the act), devices that were not in commercial distribution prior to May 28, 1976 (the date of enactment of the Medical Device Amendments of 1976 (the {1} 2019 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 ^{}[] {2} Page 2 – Ms. Janice Hogan amendments)), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval, unless and until the device is classified or reclassified into class I or II or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the act (21 U.S.C. 360c(i)), to a predicate device that does not require premarket approval. The agency determines whether new devices are substantially equivalent to previously marketed devices by means of premarket notification procedures in section 510(k) of the act (21 U.S.C. 360(k)) and Part 807 of the FDA regulations (21 CFR 807). Section 513(f)(2) of the act provides that any person who submits a premarket notification under section 510(k) for a device may, within 30 days after receiving an order classifying the device in class III under section 513(f)(1), request FDA to classify the device under the criteria set forth in section 513(a)(1). FDA shall, within 60 days of receiving such a request classify the device. This classification shall be the initial classification of the device type. Within 30 days after the issuance of an order classifying the device type, FDA must publish a notice in the Federal Register classifying the device type. On November 3, 2008, FDA filed your petition requesting classification of the SNaP Wound Care Device into class II. The petition was submitted under section 513(f)(2) of the act. In accordance with section 513(f)(1) of the act, FDA issued an order on October 28, 2008 automatically classifying the SNaP Wound Care Device in class III, because it was not within a type of device which was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, nor which was subsequently reclassified into class I or class II. In order to classify the SNaP Wound Care Device into class I or II, it is necessary that the proposed class have sufficient regulatory controls to provide reasonable assurance of the safety and effectiveness of the device type for its intended use. After review of the information submitted in the petition, FDA has determined that the SNaP Wound Care Device indicated for patients who would benefit from a suction device particularly as the device may promote wound healing and further indicated for removal of small amounts of exudates from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions and flaps can be classified in class II with the establishment of special controls. FDA believes that class II special controls provide reasonable assurance of the safety and effectiveness of the device type. In addition to the general controls of the act, the SNaP Wound Care Device is subject to the following special controls: the guidance document entitled, “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT),” to address the specific risks to health associated with the non-powered suction apparatus device intended for NPWT. The risks identified in the “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device {3} Page 2 / 26. James Hogan **مستقبلات (1).** **الخديوية (2).** **الخديوية (3).** **الخديوية (4).** **الخديوية (5).** **الخديوية (6).** **الخديوية (7).** **الخديوية (8).** **الخديوية (9).** **الخديوية (10).** **الخديوية (11).** **الخديوية (12).** **الخديوية (13).** **الخديوية (14).** **الخديوية (15).** **الخديوية (16).** **الخديوية (17).** **الخديوية (18).** **الخديوية (19).** **الخديوية (20).** **الخديوية (21).** **الخديوية (22).** **الخديوية (23).** **الخديوية (24).** **الخديوية (25).** **الخديوية (26).** **الخديوية (27).** **الخديوية (28).** **الخديوية (29).** **الخديوية (30).** **الخديوية (31).** **الخديوية (32).** **الخديوية (33).** **الخديوية (34).** **الخديوية (35).** **الخديوية (36).** **الخديوية (37).** **الخديوية (38).** **الخديوية (39).** **الخديوية (40).** **الخديوية (41).** **الخديوية (42).** **الخديوية (43).** **الخديوية (44).** **الخديوية (45).** **الخديوية (46).** **الخديوية (47).** **الخديوية (48).** **الخديوية (49).** **الخديوية (50).** **الخديوية (51).** **الخديوية (52).** **الخديوية (53).** **الخديوية (54).** **الخديوية (55).** **الخديوية (56).** **الخديوية (57).** **الخديوية (58).** **الخديوية (59).** **الخديوية (60).** **الخديوية (61).** **الخديوية (62).** **الخديوية (63).** **الخديوية (64).** **الخديوية (65).** **الخديوية (66).** **الخديوية (67).** **الخديوية (68).** **الخديوية (69).** **الخديوية (70).** **الخديوية (71).** **الخديوية (72).** **الخديوية (73).** **الخديوية (74).** **الخديوية (75).** **الخديوية (76).** **الخديوية (77).** **الخديوية (78).** **الخديوية (79).** **الخديوية (80).** **الخديوية (81).** **الخديوية (82).** **الخديوية (83).** **الخديوية (84).** **الخديوية (85).** **الخديوية (86).** **الخديوية (87).** **الخديوية (88).** **الخديوية (89).** **الخديوية (90).** **الخديوية (91).** **الخديوية (92).** **الخديوية (93).** **الخديوية (94).** **الخديوية (95).** **الخديوية (96).** **الخديوية (97).** **الخديوية (98).** **الخديوية (99).** {4} Page 3 – Ms. Janice Hogan Intended for Negative Pressure Wound Therapy (NPWT)” are: adverse tissue reaction, material degradation, improper function of suction apparatus, non-compatibility with other therapeutics and diagnostics, uncontrolled bleeding, transmission of infectious agents, and unsafe use of device. Section 510(m) of the act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type. FDA has determined premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of this device type and, therefore, the device type is not exempt from the premarket notification requirements of the act. Thus, persons who intend to market this device type must submit to FDA a premarket notification submission containing information on the non-powered suction apparatus device intended for NPWT they intend to market prior to marketing the device and receive clearance from FDA prior to marketing. A notice announcing this classification order will be published in the Federal Register. A copy of this order and supporting documentation are on file in the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Room 1061, Rockville, MD 20852 and are available for inspection between 9 a.m. and 4 p.m., Monday through Friday. As a result of this order, you may immediately market your device as described in the de novo, subject to the general control provisions of the act and the special controls identified in this order. If you have any questions concerning this classification order, please contact Jiyoung M. Dang, Ph.D. at (240) 276-3555. Sincerely yours, Christy Foreman Christy Foreman Deputy Director for Science and Regulatory Policy Office of Device Evaluation Center for Devices and Radiological Health {5} Page 1 - Ms. Imice Hogan Page 2 - Ms. Imice Hogan 2018/07/28 10:45 Page 3 - Ms. Imice Hogan 2018/07/28 10:45 Page 4 - Ms. Imice Hogan 2018/07/28 10:45 Page 5 - Ms. Imice Hogan 2018/07/28 10:45 2018/07/28 10:45 2018/07/28 10:45 2018/07/28 10:45 2018/07/28 10:45 2018/07/28 10:45 2018/07/28 2018/07/28 2018/07/28 2018/07/28
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...