K112341 · Spiracur, Inc. · OKO · Aug 29, 2011 · General, Plastic Surgery
Device Facts
Record ID
K112341
Device Name
SNAP WOUND CARE SYSTEM
Applicant
Spiracur, Inc.
Product Code
OKO · General, Plastic Surgery
Decision Date
Aug 29, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4683
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SNaP Wound Care System is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material and tissue debris. The SNaP Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions, flaps and grafts.
Device Story
SNaP Wound Care System is non-powered, portable, single-use suction device for negative pressure wound therapy (NPWT). Device utilizes dedicated constant-force springs to mechanically generate negative pressure gradient; removes excess exudates, infectious material, and tissue debris from wound site. Used in conjunction with SNaP Dressing Kit, which incorporates hydrophobic polyurethane foam. System intended for wound management; promotes healing by maintaining negative pressure. Operated by clinicians or patients in various care settings. Output is physical removal of wound fluids; healthcare providers monitor wound status and exudate collection to guide treatment. Benefits include active wound treatment via portable, non-electric mechanism.
Clinical Evidence
Bench testing only. Evaluation included assessment of negative pressure delivery performance compared to the predicate device and biocompatibility testing.
Technological Characteristics
Non-powered, portable, single-use suction apparatus. Materials include hydrophobic polyurethane foam in the dressing kit. Principle of operation: constant-force springs for mechanical negative pressure generation. No electronic components or software.
Indications for Use
Indicated for patients requiring suction for wound healing; applicable to chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic/pressure), surgically closed incisions, flaps, and grafts.
Regulatory Classification
Identification
A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
SPIRACUR INC.
SNaP® WOUND CARE SY: SPECIAL 510(k) PREMARKET NOTIFICATION
# SECTION 6 510(k) SUMMARY
#### 510(k) Notification K (234) AUG 2 9 2011
## GENERAL INFORMATION
# Applicant:
Spiracur Inc. 1180 Bordeaux Drive Sunnyvale, CA 94089 U.S.A. Phone: 408-701-5300 408-701-5301 Fax:
## Contact Person:
Sarah L. Canio Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085 U.S.A. Phone: 408-400-0856 ext. 109 Fax: 408-400-0865 Email: sarah@experiengroup.com
Date Prepared: August 12, 2011
# Classification:
21 CFR§878.4683, Class II
#### Product Code:
OKO
Trade Name: SNaP® Wound Care System
#### Generic/Common Name:
Non-powered suction apparatus device intended for negative pressure wound therapy
Predicate Device: SNaP® Wound Care System (K111393)
{2}------------------------------------------------
SPIRACUR INC.
K 112341 Page 2/3
NaP® WOUND CARE SYSTEM SPECIAL 510(k) PREMARKET NOTIFICATION
#### SECTION 6 510(k) SUMMARY
#### Intended Use:
The SNaP Wound Care System is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material and tissue debris. The SNaP Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions, flaps and grafts.
# Product Description:
The SNaP Wound Care System is a non-powered, portable, single-use suction device intended for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. The SNaP Wound Care System is designed to provide active wound treatment through the removal of excess exudates, infectious material and tissue debris. The SNaP Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions, flaps and grafts. The SNaP Wound Care System utilizes dedicated constant-force springs to mechanically generate the negative pressure gradient.
The SNaP Wound Care System is used in conjunction with the SNaP Dressing Kit.
# Substantial Equivalence:
This Special 510(k) premarket notification is for the SNaP Wound Care System, which is a modified version of the cleared SNaP Wound Care System (K111393). The minor design modification implemented to develop the modified SNaP System includes a change to the SNaP Dressing Kit to incorporate the use of a hydrophobic polyurethane foam. There have been no changes to the other components of the SNaP System as a result of this design change. The design modifications outlined in this Special 510(k) premarket notification do not (1) affect the intended use or (2) alter the fundamental scientific technology of the device. The modified SNaP System shares the same intended use, the same technological characteristics and the same principles of operation as the predicate device. The modified SNaP System and the cleared SNaP System (K111393) are both non-powered, portable, single-use suction devices intended for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids and infectious materials. Both systems utilize dedicated constant-force springs to mechanically generate the negative pressure gradient. Any differences between the devices do not raise any new issues of safety or effectiveness.
{3}------------------------------------------------
Spiracur Inc.
## SECTION 6 510(k) Summary
SPECIAL 510(k) PREMARKET NOTIFICATION
# Testing in Support of Substantial Equivalence Determination:
The SNaP Wound Care System and its components were evaluated to ensure conformance to design specifications. The testing performed includes:
- Bench testing conducted on the modified SNaP Wound Care System to assess the . ability to deliver negative pressure wound therapy comparable to the predicate device (K111393)
- Biocompatibility testing .
# Summary:
The SNaP Wound Care System is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Spiracur, Inc. % Experien Group, LLC Ms. Sarah Canio 755 N. Mathilda Avenue Sunnyvale, California 94085
AUG 29 2011
Re: K112341
Trade/Device Name: SNaP® Wound Care System Regulation Number: 21 CFR 878.4683 Regulation Name: Non-Powered suction apparatus device intended for negative pressure wound therapy Regulatory Class: II Product Code: OKO Dated: August 12, 2011 Received: August 15, 2011
Dear Ms. Canio:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaused in addesice is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enorement date of the Medical Device Amendments, or to Conninered prilly to May 20, 1978, the econdance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, dicrerere, mains of the Act include requirements for annual registration, listing of general controls proficions of wactice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addition. Ticase note: "CDree books and device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to additional controls. Existing major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Ms. Sarah Canio
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Paul 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcectorYou/Industry/default.htm.
Sincerely vours.
fo
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
SPIRACUR INC.
## SECTION 5 INDICATIONS FOR USE STATEMENT
510(k) Number (if known): Klo
Device Name: SNaP® Wound Care System
# Indications For Use:
The SNaP® Wound Care System is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material and tissue debris. The SNaP Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), surgically closed incisions, flaps and grafts.
Prescription Use X (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Knoeffer MXM
Division Si (Division of Surgical, Orthoped and Restorative Devices
510(k) Number K112341
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.