K120040 · Sterilmed, Inc. · NUJ · Oct 5, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120040
Device Name
REPROCESSED SURGICAL ELECTRIC INSTRUMENT
Applicant
Sterilmed, Inc.
Product Code
NUJ · General, Plastic Surgery
Decision Date
Oct 5, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reprocessed Surgical Electric Instrument (hereinafter sealer/divider) is indicated for use in open surgical procedures to seal vessels up to and including 7 mm, lymphatics, and tissue bundles. When used with the ForceTriad™ energy platform the sealer/divider can also be used to seal pulmonary vasculature. The sealer/divider should not be used for tubal sterilization or tubal coagulation as it has not been shown effective for sterilization procedures.
Device Story
Reprocessed surgical electric instrument (sealer/divider) used in open surgery to seal vessels (up to 7 mm), lymphatics, and tissue bundles; compatible with ForceTriad energy platform for pulmonary vasculature sealing. Device features 18 cm shaft, 13.5 mm diameter, 180-degree rotation, and 36 mm curved jaw with 34 mm cutting length. Operated by surgeon via handle latching and button activation or foot switch. Device functions by delivering energy to tissue to achieve hemostasis. Reprocessing restores used devices to original functional specifications.
Clinical Evidence
Bench testing only. Validation included cleaning and sterilization efficacy, biocompatibility (ISO 10993-1), ethylene oxide residuals (ISO 10993-7), packaging integrity (ASTM D4169, F88, F1929, F2096), shelf life (ASTM 1980-07), and functional performance testing via simulated use on beef tissue, visual inspection, and fatigue testing.
Technological Characteristics
Electrosurgical cutting and coagulation device. 18 cm shaft, 13.5 mm diameter, 36 mm curved jaw. Materials and design identical to predicate. Sterilization via ethylene oxide. Standards: ISO 11135, USP <71>, ISO 10993-1/7, ASTM D4169, F88, F1929, F2096, ASTM 1980-07.
Indications for Use
Indicated for use in open surgical procedures for patients requiring vessel sealing (up to 7 mm), lymphatics, and tissue bundle sealing. Can be used for pulmonary vasculature sealing when used with ForceTriad energy platform. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Covidien LigaSure Impact™ Hand Activated Sealer/Divider (K070162)
Related Devices
K111600 — REPROCESSED HAND ACTIVATE SEALER/DIVIDER · Ascent Healthcare Solutions · Nov 15, 2011
K230251 — Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A) · Stryker Sustainability Solutions · Mar 28, 2023
K123096 — REPROCESSED VESSEL SEALER · Sterilmed, Inc. · Jan 25, 2013
K152134 — Reprocessed Vessel Sealer · Sterilmed, Inc. · Jan 15, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K120040" on the first line and "PAGE 1 OF 1" on the second line. The text is written in a handwritten style with thick, dark strokes. The text appears to be a document identifier and page number.
ОСТ
5 2012
Medical Device Reprocessing
Small Equipment & Instrument Repair
### 510(K) PREMARKET NOTIFICATION SUBMISSION 04 JANUARY 2012 For Reprocessed Surgical Electric Instruments
## II. SUMMARY AND CERTIFICATION
### A. 510(k) Summary
| Submitter: | SterilMed, Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jason Skramsted<br>11400 73rd Avenue North<br>Maple Grove, MN 55369<br>Phone: 763-488-3483<br>Fax: 763-488-4491 |
| Date Prepared: | 04 January 2012 |
| Trade Name: | Reprocessed Surgical Electric Instrument |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories, Reprocessed |
Classification Number: Class II, 21 CFR 878.4400
Product Code: NUJ
| Predicate<br>Devices: | The reprocessed sealer/divider is substantially equivalent to the Covidien LigaSure Impact™ Hand Activated<br>Sealer/Divider (K070162). |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description: | The sealer/divider is for use in open surgical procedures to seal vessels up to and including 7 mm, lymphatics,<br>and tissue bundles. The sealer/divider can be used to seal pulmonary vasculature when used with the<br>ForceTriad™ energy platform. |
| | The sealer/divider has an 18 cm shaft with a 13.5 mm diameter and can be rotated 180 degrees. The curved<br>jaw is 36 mm long with a 34 mm cutting length. The sealer/divider is activated by pulling the handle until<br>latched and pressing the button on the device or depressing the foot switch. |
| Intended Use: | The Reprocessed Surgical Electric Instrument (hereinafter sealer/divider) is indicated for use in open surgical<br>procedures to seal vessels up to and including 7 mm, lymphatics, and tissue bundles. When used with the<br>ForceTriad™ energy platform the sealer/divider can also be used to seal pulmonary vasculature. The<br>sealer/divider should not be used for tubal sterilization or tubal coagulation as it has not been shown effective<br>for sterilization procedures. |
| Technological<br>Characteristics: | The reprocessed surgical electric instruments are identical to the predicate devices in design, materials of<br>construction, and intended use. There are no changes to the clinical applications, patient population,<br>performance specifications, or method of operation. |
| Functional and<br>Safety Testing: | Representative samples of reprocessed sealer/dividers were tested to demonstrate appropriate functional<br>characteristics. Process validation testing was performed to validate the cleaning and sterilization procedures<br>as well as device packaging. In addition, the manufacturing process includes visual and validated functional<br>testing of all products produced. |
| Summary of<br>Non-clinical<br>Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization validation (ISO 11135, USP<br><71>), biocompatibility testing (ISO 10993-1), ethylene oxide residual testing (ISO 10993-7), packaging<br>validation (ASTM D4169, ASTM F88, ASTM F1929, ASTM F2096), and shelf life validation (ASTM 1980-<br>07). In addition, validation of functional performance (bench testing) was performed through simulated use<br>on beef tissue, visual inspection, fatigue testing, and function testing. Performance testing shows the<br>reprocessed sealer/dividers to perform as intended. |
| Conclusion: | The reprocessed sealer/divider is substantially equivalent to the Covidien LigaSure Impact™ Hand Activated<br>Sealer/Divider (K070162). This conclusion is based upon the devices' similarities in functional design<br>(principles of operation), materials, indications for use and methods of construction. |
18
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" written around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT
5
2012
SterilMed, Incorporated % Mr. Jason Skramsted Regulatory Affairs Specialist 11400 73rd Avenue North Maple Grove, Minnesota 55369
Re: K120040
Trade Name: Reprocessed Surgical Electric Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NUJ Dated: September 28, 2012 Received: October 2, 2012
Dear Mr. Skramsted:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
### Page 2 - Mr. Jason Skramsted
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Eric Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
**STERIL**MED, INC.
*Medical Device Reprocessing*
*Small Equipment & Instrument Repair*
. (
# Indications for Use
510(k) Number (if known):
Device Name: Reprocessed Surgical Electric Instrument
Indications for Use:
The Reprocessed Surgical Electric Instrument (hereinafter sealer/divider) is indicated for use in open surgical procedures to seal vessels up to and including 7 mm, lymphatics, and tissue bundles. When used with the ForceTriad 101 energy platform the sealer/divider can also be used to seal pulmonary vasculature. The sealer/divider should not be used for tubal sterilization or tubal coagulation as it has not been shown effective for sterilization procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Off)*
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120047
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.