K990452 · Ybeam Technologies, Inc. · GEX · Apr 29, 1999 · General, Plastic Surgery
Device Facts
Record ID
K990452
Device Name
ORLIGHT 2000 SURGICAL LASER SYSTEM
Applicant
Ybeam Technologies, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 29, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The OrLight 2000 CO2 Surgical Laser System is intended for use in the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, including aesthetic surgery and resurfacing. The specific indications are as follows: - Ablation, vaporization, excision, incision, and coagulation of soft tissue in laser skin resurfacing, laser derm-abrasion, and laser burn debridement. - Laser skin resurfacing (ablation and/or vaporization) for the treatment of wrinkles, rhytids, and furrows. - Laser skin resurfacing (ablation, vaporization, excision, incision, and coagulation) for the reduction, removal and/or treatment of actinic keratosis, solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas, tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts). - Laser incision and/or excision of soft tissue for the performance of laser blepharoplasty.
Device Story
OrLight 2000 is a carbon dioxide (CO2) surgical laser system utilizing a radio frequency (RF) amplifier. Device delivers laser energy to target tissue via an articulated arm or hollow wave guide. Used by physicians in clinical settings for dermatology and plastic surgery procedures, including skin resurfacing, derm-abrasion, and burn debridement. Laser energy enables precise ablation, vaporization, excision, incision, and coagulation of soft tissue. Clinical benefit includes reduction or removal of skin lesions, scars, and aesthetic imperfections. Physician operates system to achieve desired tissue effect based on visual assessment of surgical site.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance was validated through design testing and comparison of soft tissue ablation characteristics against predicate CO2 laser systems.
Technological Characteristics
Carbon dioxide (CO2) laser system; radio frequency (RF) amplifier; delivery via articulated arm or hollow wave guide. Class II device. No specific material standards or software algorithm details provided.
Indications for Use
Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, or coagulation in dermatology and plastic surgery. Includes treatment of wrinkles, scars, keloids, hemangiomas, tattoos, various skin lesions (e.g., actinic keratosis, squamous cell carcinoma), and performance of blepharoplasty.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Coherent Ultrapulse Carbon Dioxide Lasers
Sharplan Carbon Dioxide SilkLaser
Related Devices
K980398 — KAPLAN PENDULASER 115 SURGICAL LASER WITH OPTOSCAN II · Optomedic Medical Technologies , Ltd. · Sep 22, 1998
{0}------------------------------------------------
Laser System
Premarket Notification February 2, 1999
K990452
## SECTION 7
# 5104 SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
- Submitter's name, address, telephone number, contact person, and date 1. summary prepared:
- a. Y-Beam Technologies, Inc. 216 Technology Drive, Suite L Irvine, CA 92618
- Joseph Neev, Ph.D. b. Contact Person: President
- January 20, 1999 Date Summary Prepared: C.
#### 2. Name of device, including trade name and classification name:
| a. Trade/Proprietary Name: | OrLight 2000 Surgical Laser System |
|----------------------------|----------------------------------------|
| b. Classification Name: | Surgical laser system (21 CFR 878.4810 |
### Identification of the predicate device or legally marketed device or devices to 3. which substantial equivalence is being claimed:
The OrLight 2000 Surgical Laser is substantially equivalent to the following predicate devices : Coherent Ultrapulse Carbon Dioxide Lasers and Sharplan Carbon Dioxide SilkLaser.
- A description of the device that is the subject of the 510(k), including 4. explanation of how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):
{1}------------------------------------------------
The OrLight 2000 Surgical Laser system is a carbon dioxide laser that utilizes a radio frequency amplifier. Laser treatment is delivered to the target tissue through an articulated arm or a hollow wave guide.
#### 5. Statement of intended use:
The OrLight 2000 Surgical Laser System is intended for use in the ablation, vaporization, excision, incision and coagulation of soft tissue in dermatology and plastic surgery, including aesthetic surgery and resurfacing.
## Statement of how the technological characteristics of the device compare to 6. those of the predicate or legally marketed device:
The OrLight 2000 Surgical Laser System has similar design features, functional features, and laser parameters as the predicate, legally marketed devices (Sharplan and Coherent carbon dioxide lasers).
#### 7. Brief summary of nonclinical tests and results:
The OrLight 2000 Surgical Laser System has been designed and tested to applicable safety standards. In addition, the OrLight 2000 Surgical Laser System was found to perform equivalently to the predicate devices with respect to soft tissue ablation.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 9 1000
Y-Beam Technologies, Inc. c/o Dr. Judy F. Gordon, D.V.M. Judy Gordon Consulting 18732 Saginaw Drive Irvine, California 92612
> K990452 Trade Name: OrLight 2000 Surgical Laser System Regulatory Class: II Product Code: GEX Dated: February 2, 1999 Received: February 12, 1999
Dear Dr. Gordon:
Re:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 – Dr. Judy F. Gordon
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
[signature]
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page *_ of _*
510(k) Number (if known): K990452
Device Name: Or Light 2000 Surgical Laser System
Indications For Use:
The OrLight 2000 CO2 Surgical Laser System is intended for use in the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, including aesthetic surgery and resurfacing. The specific indications are as follows:
- Ablation, vaporization, excision, incision, and coagulation of soft tissue in ● laser skin resurfacing, laser derm-abrasion, and laser burn debridement.
- Laser skin resurfacing (ablation and/or vaporization) for the treatment of . wrinkles, rhytids, and furrows.
- Laser skin resurfacing (ablation, vaporization, excision, incision, and ● coagulation) for the reduction, removal and/or treatment of actinic keratosis, solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas, tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts).
- Laser incision and/or excision of soft tissue for the performance of laser . blepharoplasty.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | K990452 |
| Prescription Use<br>(Per 21 CFR 801/109) | X | OR | Over-The-Counter Use |
|------------------------------------------|---|----|----------------------|
|------------------------------------------|---|----|----------------------|
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.