Holmium Laser System (Potent HP90); Holmium Laser System (Potent HP100); Holmium Laser System (Potent HP9120); Holmium Laser System (Potent HP140); Holmium Laser System (Potent HP150)
K252771 · Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. · GEX · Nov 24, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252771
Device Name
Holmium Laser System (Potent HP90); Holmium Laser System (Potent HP100); Holmium Laser System (Potent HP9120); Holmium Laser System (Potent HP140); Holmium Laser System (Potent HP150)
Applicant
Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Holmium laser system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynecology, ENT and General Surgery.
Device Story
Potent HP Series is a family of pulsed solid-state Holmium:YAG laser systems (2100 nm wavelength) for soft tissue surgery and stone fragmentation. System uses flashlamp-pumped laser medium; energy transmitted via SMA905-compatible optical fibers. Operated by physicians in clinical settings (OR, clinic). Includes touchscreen interface, dual-pedal footswitch, fiber auto-recognition, and integrated liquid cooling. Safety features include emergency stop, door interlock, and real-time output monitoring. Device provides surgical precision for incision, excision, resection, ablation, vaporization, coagulation, and hemostasis. Benefits include versatile tissue interaction and stone fragmentation capabilities across multiple medical specialties.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser product safety). Software verification and validation conducted per IEC 62304 and FDA guidance.
Technological Characteristics
Pulsed Ho:YAG laser; 2100 nm wavelength; 90-150 W max average power; up to 6.0 J pulse energy; up to 1700 μs pulse duration; up to 120 Hz frequency. Closed-loop liquid cooling. SMA905 optical fiber delivery. Green diode aiming beam (<3 mW). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, IEC 62304.
Indications for Use
Indicated for patients requiring soft tissue surgical procedures including incision, excision, resection, ablation, vaporization, coagulation, and hemostasis in urology, urinary lithotripsy, gastroenterology, arthroscopy, discectomy, gynecology, ENT, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Litho 150 / Cyber Ho 150 Holmium Laser System (K201455)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 24, 2025
Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd.
Zhengzhou Li
Official Correspondent
Room 208, Building C, No. 3, Juquan Road,
Huangpu District
Guangzhou, Guangdong 510000
China
Re: K252771
Trade/Device Name: Holmium Laser System (Potent HP90); Holmium Laser System (Potent HP100); Holmium Laser System (Potent HP9120); Holmium Laser System (Potent HP140); Holmium Laser System (Potent HP150)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: September 2, 2025
Received: September 2, 2025
Dear Zhengzhou Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252771 - Zhengzhou Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252771 - Zhengzhou Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU-S
Digitally signed by YAN FU
Date: 2025.11.24 15:30:30
-05'00'
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252771 | |
| Device Name Holmium Laser System (Potent HP90); Holmium Laser System (Potent HP100); Holmium Laser System (Potent HP9120); Holmium Laser System (Potent HP140); Holmium Laser System (Potent HP150) | |
| Indications for Use (Describe) | |
| The Holmium laser system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynecology, ENT and General Surgery. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
#K252771
# 1 Submitter / Contact Information
| Field | Information |
| --- | --- |
| Date Preparation | November 24, 2025 |
| Submitter Name | Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. |
| Address | Room 208, Building C, No. 3 Juquan Road, Huangpu District, Guangzhou, Guangdong, 510000, China. |
| Phone | +86 -20-37396970 |
| Fax | +86 -20-37376979 |
| Contact Person | Zhengzhou Li |
| Email | potent_medical_public@potent-medical.com |
# 2 Device Information
| Field | Information |
| --- | --- |
| Device Trade Name | Holmium Laser System |
| Models | Potent HP90, Potent HP100, Potent HP120, Potent HP140, Potent HP150 |
| Common Name | Powered Laser Surgical Instrument |
| Regulation Name | Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology |
| Regulation Number | 21 CFR 878.4810 |
| Regulatory Class | Class II |
| Product Code | GEX |
| Panel | General and Plastic Surgery |
1 / 6
{5}
2 / 6
# 3 Predicate and Reference Devices
| Type | Device | Manufacturer | 510(k) Number |
| --- | --- | --- | --- |
| Primary Predicate | Litho 150 / Cyber Ho 150 Holmium Laser System | Quanta System S.p.A. | K201455 |
| Reference Device | MultiPulse HoPlus Holmium Laser System | Asclepion Laser Technologies GmbH | K161257 |
# 4 Device Description
The Potent HP Series is a family of pulsed solid-state Holmium:YAG laser systems that deliver energy at a wavelength of 2100 nm through optical fibers for soft tissue surgery and stone fragmentation. The system utilizes a flashlamp-pumped Ho:YAG laser medium, and the laser energy is transmitted to the target site via SMA905-compatible optical fibers. All models in the series share the same fundamental design and operational principles, differing in maximum average power output and frequency.
The system includes a touchscreen interface, dual-pedal footswitch, fiber auto-recognition, integrated liquid cooling, and multiple built-in safety features such as an emergency stop, door interlock, and real-time monitoring of laser output.
{6}
3 / 6
# 5 Intended Use
The Holmium laser system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including:
- Urology
- Urinary Lithotripsy
- Gastroenterology
- Arthroscopy
- Discectomy
- Gynecology
- ENT
- General surgery
# 6 Comparison of Technological Characteristics
| Feature | Subject Device (Potent HP Series) | Main Predicate (Litho 150 / Cyber Ho 150) | Reference Device (MultiPulse HoPlus) |
| --- | --- | --- | --- |
| 510(k) No. | / | K201455 | K161257 |
| Manufacturer | Potent Medical | Quanta System S.p.A. | Asclepion Laser Technologies GmbH |
| Regulation | 21 CFR 878.4810 | Same | Same |
{7}
| Classification | Class II | Class II | Class II |
| --- | --- | --- | --- |
| Product Code | GEX | GEX | GEX |
| Panel | General & Plastic Surgery | Same | Same |
| Indications for Use | The Holmium laser system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery. | Same | Same + adds pulmonary, dermatology, plastic surgery |
| Laser Source | Pulsed Ho:YAG (CHT:YAG) | Same | Same |
| Wavelength | 2100 nm | 2100 nm | 2100 nm |
| Emission Mode | Pulsed | Pulsed | Pulsed |
| Pulse Duration | Up to 1700 μs | Up to 1100 μs | Up to 1700 μs |
| Pulse Energy | Up to 6.0 J | Up to 5.0 J | Up to 6.0 J |
| Frequency | Up to 120 Hz | Up to 100 Hz | Up to 100 Hz |
| Max Avg. Power | 90–150 W (model dependent) | 152 W | 140 W |
| Delivery System | SMA905 optical fibers | Optical fibers | Optical fibers |
| Aiming Beam | Green diode < 3 mW | Green diode < 5 mW | Green diode < 5 mW |
| Cooling | Closed-loop liquid | Closed-loop liquid | Closed-loop liquid |
| Safety Systems | Key switch, door interlock, emergency stop, fiber detection, output monitoring | Equivalent | Equivalent |
# Conclusion:
The Potent HP Series falls within the same technological range as both predicate devices. Differences in frequency, pulse duration range, and aiming beam intensity do
{8}
not raise new questions of safety or effectiveness and are supported by performance testing and risk analysis.
## 7 Non-Clinical Performance Data
The Potent HP Series Holmium Laser System has undergone comprehensive performance testing to ensure compliance with applicable international standards and FDA guidance. The following evaluations were conducted:
## 7.1 Electrical Safety and Electromagnetic Compatibility
The device was tested and found to conform to the following standards:
- IEC 60601-1:2005 + A1:2012 + A2:2020
Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 + A1:2020
Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and tests.
- IEC 60601-2-22:2019
Medical electrical equipment – Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment.
- IEC 60825-1:2014
Safety of laser products – Part 1: Equipment classification and requirements.
5 / 6
{9}
6 / 6
## 7.2 Software Verification and Validation
Software development and validation were performed in accordance with:
- IEC 62304:2006 + A1:2015
Medical device software – Software life cycle processes.
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005).
## 8 Substantial Equivalence Conclusion
The Potent HP Series Holmium Laser System has the same intended use and similar technological characteristics as the predicate device. Minor differences between the subject and predicate device do not raise new risks regarding safety or effectiveness. The subject device is substantially equivalent to the identified predicate device.