K201455 · Quanta System Spa · GEX · Jun 30, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201455
Device Name
Litho 150, Cyber Ho 150
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 30, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Multicavity Holmium laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery.
Device Story
Multicavity Holmium laser system (2100nm) utilizes pulsed laser energy delivered via optical fibers for soft tissue surgery and lithotripsy. System comprises Holmium laser source, power electronics, control electronics, cooling system, and footswitch. Modifications from predicate include increased max power (152W) and frequency (100Hz) via hardware change to 4 laser cavities. Operated by physicians in clinical settings (OR, endoscopy suites). User selects parameters via GUI; software controls emission. Output is laser energy for tissue ablation/coagulation or stone fragmentation. Benefits include precise tissue management and stone removal. Device is used with cleared compatible optical fibers.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing (IEC 60601-2-22, IEC 60825-1). Software verification and validation were performed per FDA guidance.
Technological Characteristics
Pulsed Holmium:YAG laser; 2100nm wavelength; 4 laser cavities; max energy 5J; max power 152W; max frequency 100Hz; pulse duration up to 1100µs. Delivery via optical fibers. Aiming beam: green diode laser < 5mW. Cooling system included. Software-controlled user interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for surgical procedures (incision, resection, ablation, vaporization, coagulation, haemostasis) in urology (e.g., BPH, lithotripsy, tumors), gastroenterology (e.g., polyps, ulcers, neoplasms), arthroscopy (soft/cartilaginous tissue, spinal decompression), gynaecology, ENT (endonasal/sinus), and general surgery. Applicable to open, laparoscopic, and endoscopic approaches.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
June 30, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Quanta System Spa Francesco Dell'antonio Vice President Regulatory Affairs and QA Via Acquedotto 109 Samarate (Va), 21017 Italy
Re: K201455
Trade/Device Name: Litho 150, Cyber Ho 150 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 25, 2020 Received: June 1, 2020
Dear Francesco Dell'antonio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K201455
Device Name Multicavity Holmium laser
#### Indications for Use (Describe)
The Multicavity Holmium laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery.
## Urology
Open and endoscopic surgery (incision, excision, resection, ablation,
vaporization, coagulation and haemostasis) including:
- · Urethral Strictures
- Bladder Neck Incisions (BNI)
- · Ablation and resection of Bladder Tumors. Uretheral Tumors and Ureteral Tumors.
- · Ablation of Benign Prostatic Hypertrophy (BPH),
- · Transurethral incision of the prostate (TUIP)
- · Holmium Laser Resection of the Prostrate (HoLRP)
- · Holmium Laser Enucleation of the Prostate (HoLEP)
- · Holmium laser Ablation of the Prostate (HoLAP)
- Condylomas
- · Lesions of external genitalia
Lithotripsy and Percutaneous Urinary Lithotripsy
· Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate,
- monohydrate and calcium oxalate
- · dehydrate stones.
- · Endoscopic fragmentation of kidney calculi
- · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
## Gastroenterology
Open and endoscopic Gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- · Appendectomy
- Polyps
- · Biopsy
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- · Gastric ulcers
## FORM FDA 3881 (1/14)
{4}------------------------------------------------
- Duodenal ulcers
- · Non Bleeding Ulcers
- Pancreatitas
- Haemorrhoids
- · Cholecystectomy
- Benign and Malignant Neoplasm
- Angiodysplasia
- Colorectal cancer
- · Telangiectasias
- · Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- Gastritis
- · Esophagitis
- Esophageal ulcers
- Varices
- · Colitis
- Mallory-Weiss tear
- Gastric Erosions
## Arthroscopy
Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- · Ligament and tendon Release
- · Contouring and sculpting of articular surfaces
- · Capsulectomy in the Knee
- · Chondreplasty in the Knee
- Debridement of inflamed synovial tissue
- Chondromalacia Ablation
- Chondromalacia and tears
- Plica Removal
- · Meniscectomy
- Loose Body Debridement
- Lateral retinecular release
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- · Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy
- Percutaneous Thoracic Disc Decompression/Discectomy
## Gynaecology
Open and laparoscopic gynaecological surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) of soft tissue
## ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization and haemostasis of soft tissue and cartilage) including:
- Endonasal/sinus Surgery
- Partial turbinectomy
- Polypectomy
{5}------------------------------------------------
- · Dacryocystorhinostomy
- Frontal Sinusotomy
- · Ethmoidectomy
- Maxillary antrostomy
- · Functional endoscopic sinus surgery
#### General Surgery
Open, laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- · Skin incision
- Excision of external and internal lesions
- · Complete of partial resection of internal organs, tumors and lesions
- · Biopsy
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
#### 5. 510(K) SUMMARY
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy, 21017 |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA |
| | Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | May 25th 2020 |
| Brand Names: | Litho 150, Cyber Ho 150 |
| Common name: | Multicavity Holmium laser |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Basis for Submission: | Device modifications |
| Predicate Device | Litho 100/60 (K192600) – Quanta System SPA |
| Reference Device | MultiPulse HoPlus (K161257), Asclepion Laser Technologies<br>Gmbh |
The subject device is derived from the legally marketed (predicate) deviced Litho 100 (K192600).
## Performance Standards:
There are no mandatory performance standards for this device.
## Description of the device:
{7}------------------------------------------------
The devices belonging to Multicavity Holmium laser family are laser devices based on a Holmium laser source.
The main parts (subsystems) of the device are the Holmium laser source, the power electronics, the optical delivery system, the control electronics and the cooling system. A specific software controls the device functions and allows the user selections. Laser emission is triggered by a footswitch.
## Description of the modifications:
The device has the same technological characteristics as the predicate device and only differ for maximum power and maximum frequency.
This Special 510(k) is submitted due to Device Modifications of the already cleared device Litho 100 (K192600) due to hardware changes to support a broadening of the range of some laser emission parameters such as power and frequency.
The modified device has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
Based on the nature of the changes implemented, the device underwent and successfully passed electrical safety, EMC and performance testing.
| | Subject device | K192600 |
|---------------------|-----------------------------|--------------------|
| Laser Source | Unchanged | - |
| Laser cavity | Unchanged | - |
| Technical | Wavelength 2100nm | Wavelength 2100nm |
| specifications | Max energy 5J | Max energy 5J |
| | Max power 152W | Max power 100W |
| | Max frequency 100Hz | Max frequency 80Hz |
| Software | Same with minor improvement | - |
| Hardware | 4 laser cavities | 3 laser cavities |
| External appearance | Unchanged | - |
| Graphic user | Unchanged | - |
| interface | | |
| Virtual basket mode | present | present |
## Summary of the technological characteristics
| | Subject device | predicate device | additional reference<br>device |
|--------------------|--------------------------------|-----------------------------------|-----------------------------------|
| 510(k) | | K192600 | K161257 |
| model name | Litho 150, Cyber Ho 150 | Litho 100, Cyber Ho 100 | MultiPulse HoPlus |
| manufacturer | Quanta System SpA | Quanta System SpA | Asclepion Laser |
| | | | Technologies Gmbh |
| Laser Source | Pulsed Holmium laser (CHT:YAG) | Pulsed Holmium laser<br>(CHT:YAG) | Pulsed Holmium laser<br>(CHT:YAG) |
| Wavelength<br>(nm) | 2.1 µm | 2.1 µm | 2.1 um |
{8}------------------------------------------------
| Emission | pulsed | pulsed | pulsed |
|-----------------------------|--------------------------|--------------------------|--------------------------|
| Pulse duration | up to 1100 µs | up to 1100 µs | up to 1700 µs |
| Energy per<br>pulse | up to 5.0 Joule | up to 5.0 Joule | up to 6.0 J |
| Frequency | up to 100 Hz | up to 80 Hz | from 5 to 100 Hz |
| Max average<br>power | 152W | 105 W | 140 W |
| Pulse<br>modulation<br>mode | Virtual basket mode | Virtual basket mode | N/A |
| | | | |
| Delivery<br>system | Optical fibers | Optical fibers | Optical fibers |
| Aiming beam | Green diode laser < 5 mW | Green diode laser < 5 mW | Green diode laser < 5 mW |
#### Indications for Use
The Multicavity Holmium laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery.
#### Urology
Open and endoscopic surgery (incision, excision, resection, ablation,
vaporization, coagulation and haemostasis) including:
- Urethral Strictures
- Bladder Neck Incisions (BNI)
- Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors,
- Ablation of Benign Prostatic Hypertrophy (BPH),
- Transurethral incision of the prostate (TUIP)
- Holmium Laser Resection of the Prostrate (HoLRP)
- Holmium Laser Enucleation of the Prostate (HoLEP)
- Holmium laser Ablation of the Prostate (HoLAP)
- Condylomas
- Lesions of external genitalia ●
## Lithotripsy and Percutaneous Urinary Lithotripsy
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate
- dehydrate stones.
- Endoscopic fragmentation of kidney calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
{9}------------------------------------------------
#### Gastroenterology
Open and endoscopic Gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- Polyps
- Biopsy
- Gall Bladder calculi ●
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non Bleeding Ulcers
- Pancreatitas
- Haemorrhoids
- Cholecystectomy
- Benign and Malignant Neoplasm
- Angiodysplasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Renu disease ●
- Vascular Malformation ●
- Gastritis
- . Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
## Arthroscopy
Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- Ligament and tendon Release
- Contouring and sculpting of articular surfaces
- Capsulectomy in the Knee
- Chondreplasty in the Knee
- Debridement of inflamed synovial tissue
- Chondromalacia Ablation
- Chondromalacia and tears
{10}------------------------------------------------
- Plica Removal
- Meniscectomy
- Loose Body Debridement
- Lateral retinecular release
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, 0 including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy
## Gynaecology
Open and laparoscopic gynaecological surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) of soft tissue
# ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and cartilage) including:
- Endonasal/sinus Surgery
- Partial turbinectomy
- Polypectomy
- Dacryocystorhinostomy
- Frontal Sinusotomy
- Ethmoidectomy
- Maxillary antrostomy
- o Functional endoscopic sinus surgery
# General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- Skin incision
- Excision of external and internal lesions ●
- Complete of partial resection of internal organs, tumors and lesions
- Biopsy
## Accessories
This device is intended to be used together with delivery optical fiber that separately received a FDA clearance for an intended use compatible with the one of this device.
{11}------------------------------------------------
#### Performance testing
The subject device was subject to testing according to the following recognized consensus standards related to electromagnetic compatibility, electrical safety and performances.
- -IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- -IEC 60601-1-2 Edition 4: 2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests.
- -IEC 60601-2-22: 2012-10 ed 3.1, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment
- -IEC 60825-1 Ed. 3.0 (2014) Safety of laser products - Part 1: Equipment classification and requirements
The following testing were performed on the modified device:
- -Software Verification and Validation Testing: Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### Comparison with predicate device:
The subject and unmodified devices have the same intended use and the same fundamental scientific technology, based on Holmium laser sources.
#### Summary
The subject device is substantially equivalent to its identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.