K203788 · Lutronic Corporation · GEX · Mar 26, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203788
Device Name
DermaV Laser System
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1064 nm The DermaV Laser System is indicated for stable long term or permanent hair reduction and for treatment of PFB (Pseudofolliculitis Barbae) on all skin types, Fitzpatrick I – VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The 1064 nm laser is also indicated for coagulation and hemostasis of vasular lesions of vascular lesions such as but not limited to port wine stains, hemangiomna, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The 1064 mm laser system is also indicated for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, sebortheic keratoses, nevi, chloasma, verucae, skin tags, keratoses, (significant reduction in the intensity of black and/or blue- black tattoos) and plaques. The 1064 nm Laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. 532 nm For coagulation and hemostasis of vascular and cutaneous lesions in dermatology, including, but not limited to, the following general categories: vascular lesions [angiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions [nevi, lentigines, chloasma, café-au-lait, tattoos (red and green ink)]; verrucae; skin tags; keratoses; plaques; and cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
Device Story
DermaV Laser System is a pulsed solid-state laser (Nd:YAG resonator) emitting 1064 nm and 532 nm (via KTP) wavelengths. Device includes main body, optical fiber with handpiece, and footswitch; optional Intelligent Cooling Device (ICD) for skin cooling. Operated by clinicians in dermatology/surgical settings. Laser energy is delivered via lens-coupled optical fiber to target tissue; absorption causes rapid localized heating, destroying target materials (e.g., hair follicles, vascular/pigmented lesions). User interface is an LCD touch screen for parameter adjustment. Output allows for coagulation, hemostasis, and lesion reduction. Benefits include permanent hair reduction and treatment of various dermatological conditions.
Clinical Evidence
No clinical data submitted; bench testing only. Device conforms to IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60825-1 (laser safety), and ISO 10993-1, -5, -10 (biocompatibility).
Technological Characteristics
Nd:YAG resonator; 1064 nm and 532 nm (KTP) wavelengths; 635 nm aiming beam. Pulsed solid-state laser, Class IV. Optical fiber delivery with handpiece. Integrated cooling device (ICD). Touch LCD interface. Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and ISO 10993 standards.
Indications for Use
Indicated for stable long-term/permanent hair reduction and PFB treatment on all skin types (Fitzpatrick I-VI, including tanned skin); coagulation/hemostasis of vascular lesions (e.g., port wine stains, hemangiomas, telangiectasia, rosacea, leg veins); treatment of benign pigmented lesions (e.g., lentigos, café au lait macules, nevi, tattoos); and treatment of periocular/perioral wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Cutera Family of Coolglide Aesthetic Lasers (K153671)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 26, 2021
Lutronic Corporation Jhung Vojir Chief Operating Officer Lutronic Center. 219. Sowon-Ro Deogyang-Gu, Goyang-si, Geonggi-do 410220 Korea. South
Re: K203788
Trade/Device Name: DermaV Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 23, 2020 Received: December 28, 2020
Dear Jhung Vojir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203788
Device Name DermaV Laser System
#### Indications for Use (Describe)
1064 nm
The DermaV Laser System is indicated for stable long term or permanent hair reduction and for treatment of PFB (Pseudofolliculitis Barbae) on all skin types, Fitzpatrick I – VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The 1064 nm laser is also indicated for coagulation and hemostasis of vasular lesions of vascular lesions such as but not limited to port wine stains, hemangiomna, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The 1064 mm laser system is also indicated for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, sebortheic keratoses, nevi, chloasma, verucae, skin tags, keratoses, (significant reduction in the intensity of black and/or blue- black tattoos) and plaques. The 1064 nm Laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
#### 532 nm
For coagulation and hemostasis of vascular and cutaneous lesions in dermatology, including, but not limited to, the following general categories: vascular lesions [angiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions [nevi, lentigines, chloasma, café-au-lait, tattoos (red and green ink)]; verrucae; skin tags; keratoses; plaques; and cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Lutronic Corporation
DermaV Laser System
#### 510(k) Owner
Lutronic Corporation 219, Sowon-Ro Deogyang-Gu, Goyang-Si, KR 410220
#### Submission Correspondent
Jhung Won Vojir, PhD Chief Operating Officer, Lutronic Aesthetics, Inc. 19 Fortune Drive Billerica, MA 01821 Tel: 888-588-7644
Date Prepared: March 24, 2021
#### Trade Name of Device
DermaV Laser System
Common or Usual Name Surgical Laser
Classification Name Laser Surgical Instrument for Use in General and Plastic Surgery and Dermatology: 21 C.F.R. 878.4810 Class II Product Code: GEX
## Predicate Devices
Primary predicate: Lutronic Corporation CLARITY II cleared in K183566 Secondary predicate: Cutera Family of Coolglide Aesthetic Lasers cleared in K153671
#### Device Description
The DermaV laser system contains a Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates pulsed laser energy at the nominal wavelengths of 1064 mm and 532 nm using KTP. The outputs of each laser generator and the aiming beam (635 nm) are optically combined on the laser rail so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system for either the 532 nm or 1064 nm wavelength.
The control panel is equipped with an LCD touch screen so that users can easily adjust parameters for optimal settings. The software included provides all the function which is necessary to use the device.
{4}------------------------------------------------
The DermaV laser system is comprised of three main components: system main body, optical fiber with handpiece, and footswitch.
The DermaV laser system emits laser energy at 532 nm via a handpiece attached to an optical fiber. The DermaV laser system can be configured with an ICD (Intelligent Cooling Device) skin cooling device. The pulsed beam is directed to the treatment zone through a lenscoupled optical fiber attached to a handpiece. When the beam contacts human tissue, the energy of the beam is absorbed by the tissue, resulting in a very rapid and highly localized temperature increase in the target. The short but swift temperature increase causes selective heating and destroys of the target materials into smaller particles.
#### Indications for Use
#### 1064 nm
The DermaV Laser System is indicated for stable long term or permanent hair reduction and for treatment of PFB (Pseudofolliculitis Barbae) on all skin types. Fitzpatrick I - VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6. 9, and 12 months after the completion of a treatment regime.
The 1064 nm laser is also indicated for coagulation and hemostasis of soft tissue, for hemostasis of vascular lesions such as but not limited to port wine stains, hemangiomna, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The 1064 nm laser system is also indicated for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, (significant reduction in the intensity of black and/or blue- black tattoos) and plaques. The 1064 nm Laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
## 532 nm
For coagulation and hemostasis of vascular and cutaneous lesions in dermatology, including, but not limited to, the following general categories: vascular lesions [angiomas, hemangiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions [nevi, lentigines, chloasma, café-au-lait, tattoos (red and green ink)]; verrucae; skin tags; keratoses; plaques; and cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
## Nonclinical Performance Testing
The DermaV Laser System was tested for electrical safety and found to be in conformance with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and for electromagnetic compatibility and found to be in conformance with IEC 60601-1-2 Medical electrical equipment Part 1-2 General requirements for basic safety and essential performance- Collateral standard: Electromagnetic Disturbances-Requirements and tests.
{5}------------------------------------------------
# K203788
The DermaV Laser System has been tested and found in conformance with IEC 60825-1 Safety of laser products-Part 1: Equipment classification and requirements.
The DermaV Laser System was tested for biocompatibility and is in conformance with ISO 10993-1 "Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process ", ISO 10993-5 "Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity ", and ISO 10993-10 "Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization".
#### Clinical Performance Testing
Clinical performance testing was not needed for this submission; therefore, no clinical tests were submitted.
#### Comparison with the Predicate Devices
The 1064 nm and 532 nm functionality of the DermaV Laser System utilizes technology similar to the predicate devices. The DermaV Laser System and the Lutronic CLARITY II Laser System have the same intended use and similar technological characteristics for the 1064 nm functions and the Cutera Coolglide Laser System has the same intended use and similar technological characteristics for the 532 nm functions. Performance testing provided demonstrates that the device can function safely and effectively for its intended use. Differences in the proposed device's technical features, such as software, biocompatibility, and laser output parameters, do not raise new types of questions regarding the device's safety and efficacy for its indications for use.
| Characteristic | DermaV Laser System | CLARITY II Laser System |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Lutronic Corporation | Lutronic Corporation |
| 510(k) Number | K203788 | K183566 |
| Product Code | GEX | GEX |
| Laser Wavelength | 1064 nm, 532 nm | 1064 nm, 755 nm |
| Indications for Use at 1064 nm | 1064 nm: The DermaV Laser<br>System is indicated for stable<br>long term or permanent hair<br>reduction and for treatment of<br>PFB (Pseudofolliculitis Barbae)<br>on all skin types, Fitzpatrick I –<br>VI, including tanned skin.<br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9,<br>and 12 months after the<br>completion of a treatment regime.<br><br>The 1064 nm laser is also<br>indicated for coagulation and<br>hemostasis of soft tissue. for | 1064 nm: The DermaV Laser<br>System is indicated for stable<br>long term or permanent hair<br>reduction and for treatment of<br>PFB (Pseudofolliculitis Barbae)<br>on all skin types, Fitzpatrick I –<br>VI, including tanned skin.<br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6,<br>9, and 12 months after the<br>completion of a treatment<br>regime.<br><br>The 1064 nm laser is also<br>indicated for coagulation and |
| | | |
| | hemostasis of vascular lesions<br>such as but not limited to port<br>wine stains, hemangiomna, warts,<br>telangiectasia, rosacea, venus<br>lake, leg veins and spider veins.<br>The 1064 nm laser system is also<br>indicated for the treatment of<br>benign pigmented lesions such as,<br>but not limited to, lentigos (age<br>spots), solar lentigos (sun spots),<br>café au lait macules, seborrheic<br>keratoses, nevi, chloasma,<br>verrucae, skin tags, keratoses,<br>(significant reduction in the<br>intensity of black and/or blue-<br>black tattoos) and plaques. The<br>1064 nm Laser is also indicated<br>for the treatment of wrinkles such<br>as, but not limited to, periocular<br>and perioral<br>wrinkles. | hemostasis of soft tissue, for<br>hemostasis of vascular lesions<br>such as but not limited to port<br>wine stains, hemangiomna,<br>warts, telangiectasia, rosacea,<br>venus lake, leg veins and spider<br>veins. The 1064 nm laser system<br>is also indicated for the<br>treatment of benign pigmented<br>lesions such as, but not limited<br>to, lentigos (age spots), solar<br>lentigos (sun spots), café au lait<br>macules, seborrheic keratoses,<br>nevi, chloasma, verrucae, skin<br>tags, keratoses, (significant<br>reduction in the intensity of<br>black and/or blue- black tattoos)<br>and plaques. The 1064 nm Laser<br>is also indicated for the<br>treatment of wrinkles such as,<br>but not limited to, periocular and<br>perioral<br>wrinkles. |
| Laser Type | Pulsed Solid State Laser | Pulsed Solid State Laser |
| Beam Delivery | Optical Fiber with Handpiece | Optical Fiber with Handpiece |
| Aiming Beam | 635 nm | 515-535 nm |
| Pulse Energy | 1064 nm: Max 100 J | 1064 nm: Max 100 J |
| Pulse Width | 0.1 - 60 ms | 0.1 - 300 ms |
| Max Fluence (J/cm2) | 300 | 600 |
| Pulse Rate | 0.5-10 Hz | 0.5-10 Hz |
| CDRH Laser Class | IV | IV |
| User Interface | Touch LCD | Touch LCD |
| Spot Size | 2, 3, 5, 7, 8, 10, 12, 14, 16 nm | 2, 3, 5, 8, 10, 12, 15, 16, 18, 20,<br>22, 24 nm |
| Emission Control | Finger Switch or Footswitch | Finger Switch or Footswitch |
| Integrated Cooling | Yes | Yes |
#### Table 1: DermaV Laser System Substantial Equivalence (1064 nm Functions)
{6}------------------------------------------------
# Table 2: DermaV Laser System Substantial Equivalence (532 nm Functions)
| Characteristic | DermaV Laser System | Family of Coolglide Aesthetic Lasers |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Manufacturer | Lutronic Corporation | Cutera, Inc. |
| 510(k) Number | K203788 | K153671 |
| Product Code | GEX | GEX |
| Laser Wavelength | 1064 nm, 532 nm | 1064 nm, 532 nm |
| Indications for Use at 532 nm | 532 nm<br>For coagulation and hemostasis of<br>vascular and cutaneous lesions in<br>dermatology, including, but not | 532 nm<br>For coagulation and hemostasis<br>of vascular and cutaneous<br>lesions in dermatology, |
{7}------------------------------------------------
| | limited to, the following general<br>categories: vascular lesions<br>[angiomas, hemangiomas (port<br>wine), telangiectasia (facial or<br>extremities telangiectasias,<br>venous anomalies, leg veins)];<br>benign pigmented lesions [nevi,<br>lentigines, chloasma, café-au-lait,<br>tattoos (red and green ink)];<br>verrucae; skin tags; keratoses;<br>plaques; and cutaneous lesion<br>treatment (hemostasis, color<br>lightening, blanching, flattening,<br>reduction of lesion size). | including, but not limited to, the<br>following general categories:<br>vascular lesions [angiomas,<br>hemangiomas (port wine),<br>telangiectasia (facial or<br>extremities telangiectasias,<br>venous anomalies, leg veins)];<br>benign pigmented lesions [nevi,<br>lentigines, chloasma, café-au-<br>lait, tattoos (red and green ink)];<br>verrucae; skin tags; keratoses;<br>plaques; and cutaneous lesion<br>treatment (hemostasis, color<br>lightening, blanching, flattening,<br>reduction of lesion size). |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Laser Type | Pulsed Solid State Laser | Flash Pumped Solid State Rod |
| Beam Delivery | Optical Fiber with Handpiece | Optical Fiber with Handpiece |
| Aiming Beam | 635 nm | 630-680 nm |
| Pulse Energy | 532 nm: Max 13.5 J | 532 nm: Max 12 J |
| Pulse Width | 0.3 - 40 ms | 1.5 - 40 ms |
| Pulse Rate | 0.5-10 Hz | ≤ 10 Hz |
| Max Fluence (J/cm²) | 50 | 42 |
| CDRH Laser Class | IV | IV |
| User Interface | Touch LCD | Touch LCD |
| Spot Size | 2, 3, 5, 7, 8, 10, 12, 14, 16 nm | 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 nm |
| Emission Control | Finger Switch or Footswitch | Finger Switch or Footswitch |
| Integrated Cooling | Yes | Yes |
#### Conclusions
The DermaV Laser System's indications for use and technological characteristics do not raise new types of questions regarding safety and efficacy when compared to the predicates. Based on its technical characteristics, design, functional features, performance test data, and its indications for use as listed above, the DermaV Laser System is considered to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.