K183566 · Lutronic Corporation · GEX · Jun 14, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183566
Device Name
CLARITY II Laser System
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CLARITY II 755 nm Laser System is indicated for temporary hair reduction and stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime on all skin types (Fitzpatrick 1-VI) including tanned skin. The 755 nm laser is also indicated for treatment of benign pigmented lesions, vascular lesions and wrinkles. The CLARITY II 1064 nm Laser System is indicated for stable long term or permanent hair reduction and for treatment of PFB (Pseudofolliculitis Barbae) on all skin types, Fitzpatrick I -VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The 1064 nm laser is also indicated for coagulation and hemostasis of soft tissue, for hemostasis of vascular lesions such as but not limited to port wine stains, hemangiomna, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The 1064 nm laser system is also indicated for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, (significant reduction in the intensity of black and/or blue- black tattoos) and plaques. The 1064 nm Laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
Device Story
CLARITY II Laser System utilizes Alexandrite (755 nm) and Nd:YAG (1064 nm) resonators to generate pulsed laser energy; energy delivered via optical fiber and handpiece to target tissue; user-controlled via finger switch or footswitch; system includes LCD touch screen for parameter adjustment; features include skin temperature monitoring and Intelligent Tracking Tip for pulse overlap; used in clinical settings by trained professionals; laser energy absorption causes localized tissue heating for hair reduction, coagulation, and lesion treatment; benefits include stable long-term hair reduction and dermatological lesion management.
Clinical Evidence
No clinical performance data provided; substantial equivalence established via bench testing and comparison to previously cleared predicate devices.
Technological Characteristics
Dual-resonator laser (755 nm Alexandrite, 1064 nm Nd:YAG); optical fiber delivery; LCD touch screen interface; skin temperature monitoring; Intelligent Tracking Tip; cooling via Chilled Air or ICD; electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for temporary or permanent hair reduction, treatment of benign pigmented/vascular lesions, wrinkles, and PFB in patients of all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Lutronic Corporation James Childs Clinical Science Director Lutronic Center, 219, Sowon-ro Deogyang-gu. Goyang-si, 410-220 Republic of Korea
June 14, 2019
Re: K183566
Trade/Device Name: CLARITY II Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Regulatory Class: Class II Product Code: GEX Dated: May 8, 2019 Received: May 13, 2019
Dear James Childs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden, MS Acting Assistant Director, Light Based Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183566
Device Name
Clarity II Laser System
Indications for Use (Describe) 755 nm
The CLARITY II 755 nm Laser System is indicated for temporary hair reduction and stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime on all skin types (Fitzpatrick 1-VI) including tanned skin.
The 755 nm laser is also indicated for treatment of benign pigmented lesions, vascular lesions and wrinkles.
1064 nm
The CLARITY II 1064 nm Laser System is indicated for stable long term or permanent hair reatment of PFB (Pseudofolliculitis Barbae) on all skin types, Fitzpatrick I - VI, including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The 1064 nm laser is also indicated for coagulation and hemostasis of vasular lesions of vascular lesions such as but not limited to port wine stains, hemangiomna, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The 1064 nm laser system is also indicated for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, (significant reduction in the intensity of black and/or blue- black tattoos) and plaques.
The 1064 nm Laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td></tr><tr><td><div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | |
| <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 6. 510(k) Summary
The Company's 510(k) Summary for the CLARITY II Laser System is as follows:
## 510(k) Summary for the Lutronic Corporation CLARITY II Laser System
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## 1. General Information
Applicant:
Lutronic Corporation 219, Sowon-ro Haengsin-dong, Deogyang-gu. Goyang-si, Gyeonggi-do, 410-722
Republic of Korea Tel: (82) 31-908-3440 Fax: (82) 31-907-3440
Contact Person:
Mark W. Sheehan Regulatory Affairs Specialist Lutronic Global 19 Fortune Drive Billerica, MA 01821 Tel: (888) 588-7644 ext. 6253
#### Summary/Preparation Date: December 20, 2018
#### 2. Names
Trade Name:
CLARITY II Laser System
Common Name:
Classification Name:
Dermatology Laser
Laser instrument, Surgical powered
Product Code: GEX Panel: General and Plastic Surgery
## 3. Predicate Devices
The CLARITY II Laser System is substantially equivalent to the CLARITY LPC Laser System (K130199) and the Elite + Laser (K141425).
## 4. Device Description
The CLARITY II Laser System contains two separate resonators, an Alexandrite resonator and an Nd: YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates pulsed laser energy at the nominal wavelength 755 nm and 1064 nm,
{5}------------------------------------------------
respectively. The outputs of each laser generator and the aiming beam are optically combined on the laser rail so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system for either the 755 nm or 1064 nm wavelength. It is a medical device, designed for specific clinical applications, emitting laser energy via a handpiece and controlled by the user with a handpiece finger switch or optional footswitch. The CLARITY II Laser System consists of a system main body with an LCD touch screen, optical fiber with an AC or ICD handpiece.
The control panel is equipped with an LCD touch screen so that users may easily adjust parameters for optimal settings.
The CLARITY II Laser System is a medical device, designed for specific clinical applications, emitting laser energy at 755 nm and 1064 nm via a handpiece attached to an optical fiber. The pulsed beam is directed to the treatment zone through a lenscoupled optical fiber attached to a handpiece. When the beam contacts human tissue, the energy of the beam is absorbed by the tissue, resulting in very rapid highly localized temperature increase in the target.
The CLARITY II Laser System is composed of three main components, including the system main body, optical fiber (with optional AC handpiece or ICD handpiece), and the footswitch. The system main body is composed of the system control module, power supply, laser module, control panel (LCD touch screen), emergency stop button, and key switch. Laser radiation will be emitted from the handpiece through the optical fiber when the footswitch or finger switch is pressed. Safety goggles are provided with the system (optical density of 7 or higher at 1064 nm and 755 nm) for ancillary personnel. The casters and handle on the main body allows for transportation of the laser system within the treatment room. The laser console measures 563.6 mm (W) x 952.3 mm (L) x 998.5 mm (H).
The laser delivery system begins at the laser aperture, which is the fiber socket, at the upper front of the device. At this point, the invisible and infrared laser beam (755 nm or 1064 nm) and the visible 515-535 nm aiming beam are fed into and pass through the optical fiber and exit at the handpiece aperture.
Cartridges are available for the AC handpiece and ICD handpiece. The system is fitted with a window to protect the lens inside the cartridges from foreign material. The cartridge irradiation diameters available within the system are 2, 3, 5, 8, 10, 12, 15, 16, 18, 20, 22, 24mm. Users can use a cartridge fitted with an intelligent tracking tip (roller), O-type tip, C-type tip, or I-type tip.
The AC handpiece can be used with the CryoMini air cooling system that was cleared April 2008 (K080735).
{6}------------------------------------------------
#### 5. Indications for Use
#### 755 nm
The CLARITY II 755 nm Laser System is indicated for temporary hair reduction and stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime on all skin types (Fitzpatrick 1-VI) including tanned skin. The 755 nm laser is also indicated for treatment of benign pigmented lesions, vascular lesions and wrinkles.
#### 1064 nm
The CLARITY II 1064 nm Laser System is indicated for stable long term or permanent hair reduction and for treatment of PFB (Pseudofolliculitis Barbae) on all skin types, Fitzpatrick I -VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The 1064 nm laser is also indicated for coagulation and hemostasis of soft tissue, for hemostasis of vascular lesions such as but not limited to port wine stains, hemangiomna, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The 1064 nm laser system is also indicated for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, (significant reduction in the intensity of black and/or blue- black tattoos) and plaques. The 1064 nm Laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
#### 6. Substantial Equivalence
The CLARITY II Laser System is substantially equivalent to the CLARITY LPC Laser System (K130199) and the Elite + Laser (K141425). It has the same technological characteristics as the predicate devices. All three devices use two separate laser resonators, an Alexandrite resonator and an Nd: Y AG (Neodymiumdoped Yttrium Aluminum Garnet) resonator which generates pulsed laser energy at the nominal wavelengths of 755 nm and 1064 nm. They have the same intended uses and similar operating principles. They also have the same spot sizes, pulse widths and fluences. The laser systems differ in the features offered. Unlike its predicated devices, the Subiect device includes a skin temperature monitoring feature for informational purposes only. In addition, the Subject device includes an Intelligent Tracking Tip that insures selectable overlap of contiguous pulses during treatment. These features, however, do not affect the safety and effectiveness of the subject device and therefore it is substantially equivalent to the Predicate devices.
{7}------------------------------------------------
## 7. Performance Data
The Company's performance data for the CLARITY II Laser System is as follows:
| Device Name | | Clarity II | Clarity LPC | Elite + |
|----------------|------------------------------------------------------|-------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Manufacturer | | Lutronic Corporation | Lutronic<br>Corporation | Cynosure |
| | 510(k) Number | NA | K130199 | K141425 |
| | Max Fluence<br>(J/cm²) | 600 | 600 | 60 |
| | Max Energy (J) | 60 | 60 | 30 |
| 755 nm | Max Rep Rate (Hz) | 10 | 10 | 5 |
| | Pulse Duration (ms) | 0.1 - 300 | 0.35 - 300 | 0.1 - 300 |
| | Spot Sizes (mm) | 2,3,5,8,10,12,15,16,18,20<br>,22,24 | 2,3,5,8,10,12,15,18,2<br>0 | 3,5,7,10,12,15,18,20,2<br>2,24 |
| | Max Fluence (J/cm²) | 600 | 600 | 300 |
| | Max Energy (J) | 100 | 100 | 50 |
| | Max Rep Rate (Hz) | 10 | 10 | 10 |
| | Pulse Duration (ms) | 0.1 - 300 | 0.35 - 300 | 0.1 - 300 |
| 1064 nm | Spot Sizes (mm) | 2,3,5,8,10,12,15,16,18.20<br>,22,24 | 2,3,5,8,10,12,15,18,2<br>0 | 3,5,7,10,12,15,18,20,2<br>2,24 |
| | Aiming Beam | Green | Green | Red |
| 755/1064<br>nm | Electrical<br>Requirements<br>(Power<br>consumption) | AC220-230V 50/60 Hz<br>Fuse: 250V/30A, Circuit<br>Breaker: 30A<br>Power Consumption: 6.0<br>kVa | VAC 220-230, single<br>phase, 30A, 50/60 Hz | VAC 208/240, single<br>phase, 30A, 50/60 Hz |
| | Size (mm) | 563.6 (W) x 952.3 (L) x<br>998.5 (H) mm | 434 (W) x 776.8 (D)<br>x 1148.8 (H) | 381 (W) x 635 (D) x<br>1041.4 (H) |
| | Weight (kg) | 110 | 110 | 81.8 |
| | Optical Delivery<br>System | Optical fiber with<br>handpiece | Optical fiber with<br>handpiece | Optical fiber with<br>handpiece |
| | Cooling System | Chilled Air or ICD | Chilled Air or ICD | Chilled Air |
## Bench Testing
The CLARITY II Laser System complies with all applicable standards, including ISO 13485:2016, ISO 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility.
## Clinical Testing
No performance data has been provided since the CLARITY II Laser System is equivalent to the previously cleared predicate devices with no new issues regarding safety and effectiveness.
## 8. Conclusion
The intended use of the CLARITY II Laser System is virtually identical to the intended use of the predicate devices and the technological characteristics of the CLARITY II Laser System. Any differences between the CLARITY II Laser System and the predicate devices have no
{8}------------------------------------------------
significant influence on safety or effectiveness of the CLARITY II Laser System. Therefore, the CLARITY II Laser System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.