K141425 · Cynosure, Inc. · GEX · Sep 11, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141425
Device Name
ELITE + LASER SYSTEM
Applicant
Cynosure, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 11, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
755 nm: The Cynosure Elite+ Laser is indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles. 1064 nm: The Cynosure Elite+ Laser is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Cynosure Elite+ Laser is a dual-wavelength (755 nm Alexandrite, 1064 nm Nd:YAG) medical laser system. Device inputs include user-selected parameters via control interface; laser activation via finger or foot switch. System transforms electrical energy into laser output for dermatological procedures. Modifications include four larger handpieces (18, 20, 22, 24 mm) and internal adjustments (rod diameter, lamp fill pressure) to Alexandrite laser to maintain fluence at larger spot sizes. Used in clinical settings by trained professionals. Software automatically detects attached handpiece to control energy delivery, preventing fluence increases for existing spot sizes. Output is laser energy applied to patient skin for coagulation, hemostasis, or hair reduction. Benefits include expanded treatment area capabilities while maintaining performance specifications of predicate device.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by non-clinical bench testing and software verification confirming the device meets product system requirements.
Technological Characteristics
Dual-wavelength laser system: 755 nm (Alexandrite crystal rod) and 1064 nm (Nd:YAG). Power: 208/240 VAC, 30A, 50/60 Hz. Dimensions: 41 x 15 x 25 in; Weight: 180 lbs. Spot sizes: 3, 5, 7, 10, 12, 15, 18, 20, 22, 24 mm. Pulse duration: 0.1-300 ms. Repetition rate: 1-5 Hz (755 nm), 1-10 Hz (1064 nm). Software-controlled energy delivery based on handpiece identification.
Indications for Use
Indicated for permanent hair reduction (all Fitzpatrick skin types I-VI, including tanned skin), treatment of vascular lesions (e.g., port wine stains, hemangiomas, telangiectasia, leg veins), benign pigmented lesions (e.g., lentigos, nevi, seborrheic keratoses), wrinkles, scars, striae, psoriasis, and pseudofolliculitis barbae.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 11, 2014
Cynosure Incorporated Ms. Kelli McMillan Regulatory Affairs Specialist 5 Carlisle Road Westford, Massachusetts 01886
Re: K141425 Trade/Device Name: Cynosure Elite+ Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 11, 2014 Received: August 13, 2014
Dear Ms. McMillan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141425
Device Name Cynosure Elite+ Laser
### Indications for Use (Describe)
755 nm:
The Cynosure Elite+ Laser is indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowting when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.
#### 1064 nm:
The Cynosure Elite+ Laser is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Type of Use (*Select one or both, as applicable*)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k Summary Elite+ Laser Special 510(k) K141425
| 807.92(a)(1) Submitter Information | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Cynosure, Inc |
| Address | 5 Carlisle Road |
| | Westford, MA 01886 |
| Phone Number | (781) 993-2454 |
| Fax Number | (978) 256-6556 |
| Establishment Registration Number | 1222993 |
| Contact Person | Kelli McMillan |
| Preparation Date | September 10, 2014 |
| 807.92(a)(2) Name of Device | |
| Trade or Proprietary Name | Elite+ Laser |
| Common or Usual Name | Medical Laser System |
| Classification | II |
| Regulation Name | Laser Surgical Instrument for Use in General and Plastic Surgery<br>and in Dermatology |
| Classification Panel | General & Plastic Surgery |
| Regulation | 878.4810 |
| Product Code(s) | GEX |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed | |
| | Apogee Elite K034030 |
| 807.92(2)(4) Device Description | |
| | Cynosure Elite+ laser has an Alexandrite crystal rod as the lasing<br>medium, with a wavelength of 755 nm. The 1064 nm wavelength<br>is generated from the Nd:YAG laser head. |
| | Laser activation is by either by finger or foot switch. The overall<br>weight of the laser is 180 lbs and the size is 41 x 15 x 25 in<br>(HxWxD). |
| | Electrical equipment is 208/240 VAC, single phase, 30A, 50/60<br>Hz. |
| | The modifications to this device are four handpieces allowing<br>for a larger treatment area; 18 mm, 20 mm, 22 mm and 24 mm.<br>The 1064 Nd:YAG laser did not require any modifications to<br>support the new handpieces. Minor modifications, including<br>rod diameter change and lamp fill pressure change, were made<br>to the Alexandrite laser to increase the energy to support the<br>larger spot sizes. The software continues to control the energy<br>based on the handpiece attached. This prevents the increase in<br>available energy from the laser from increasing the fluence of<br>the existing spot sizes. |
| 807/92(a)(5) Intended Use of the Device | |
| | 755 nm:<br>The Cynosure Elite+ Laser is indicated for stable long-term, or<br>permanent hair reduction. Permanent hair reduction is defined<br>as long-term stable reduction in the number of hairs regrowing |
| K141425 | |
| when measured 6, 9, or 12 months after the completion of a<br>treatment regime. It is used for all skin types (Fitzpatrick I-VI)<br>including tanned skin. It is also indicated for the treatment of<br>vascular lesions, benign pigmented lesions, and wrinkles. | |
| 1064 nm: | |
| The Cynosure Elite+ Laser is intended for the coagulation and<br>hemostasis of benign vascular lesions such as, but not limited to,<br>port wine stains, hemangiomas, warts, telangectasia, rosacea,<br>venus lake, leg veins, spider veins and poikiloderma of civatte;<br>and treatment of benign cutaneous lesions such as warts, scars,<br>striae and psoriasis. The laser is also intended for the treatment<br>of benign pigmented lesions such as, but not limited to, lentigos<br>(age spots), solar lentigos (sun spots), cafe au lait macules,<br>seborrheic keratoses, nevi, chloasma, verrucae, skin tags,<br>keratosis and plaques. | |
| The laser is also indicated for the treatment of wrinkles such as,<br>but not limited to, periocular and perioral wrinkles. Additionally,<br>the laser is indicated for the treatment of pseudofolliculitis<br>barbae (PFB) and for stable long-term, or permanent hair<br>reduction. Permanent hair reduction is defined as long-term<br>stable reduction in the number of hairs regrowing when<br>measured at 6, 9 and 12 months after the completion of a<br>treatment regime. | |
{4}------------------------------------------------
# 510k Summary Elite+ Laser Special 510(k)
{5}------------------------------------------------
# 510k Summary Elite+ Laser Special 510(k) K141425
| 807.92(a)(6) Summary of the Technological Characteristics of the Device compared to the Predicate | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------|-------------------------------|------------|--|
| | Elite+<br>K141425 | | Apogee Elite<br>K034030 | | |
| Laser Type | Alexandrite and Nd:YAG Lasers | | Alexandrite and Nd:YAG Lasers | | |
| Wavelength | 755 nm | 1064 nm | 755 nm | 1064 nm | |
| Maximum Fluence | 60 J/cm | 300 J/cm | 60 J/cm | 300 J/cm | |
| Repetition Rate | 1 to 5 Hz | 1 to 10 Hz | 1 to 5 Hz | 1 to 10 Hz | |
| Pulse Duration | 0.1-300 ms | | 0.1-300 ms | | |
| Spot Sizes (mm) | 3, 5, 7, 10, 12, 15, 18, 20, 22, 24 | | 3, 5, 7, 10, 12.5, 15 | | |
| 807.92(b)(1) Non-clinical tests submitted | | | | | |
| Test | Result | | | | |
| Software | The testing confirmed that the software specifications were met. | | | | |
| 807.92(b)(2) Clinical tests submitted | | | | | |
| None | | | | | |
| 807.92(b)(3) Conclusions drawn from non-clinical data | | | | | |
| Testing confirmed that the performance of the Elite+ Laser meets the product system requirements, which is based on the predicate device. | | | | | |
| Therefore, the modification resulted in a device that performs within the same specifications of the predicate device, and is therefore | | | | | |
substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.