K151890 · Wuhan Gigaa Optronics Technology Co., Ltd. · GEX · Dec 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151890
Device Name
Medical Diode Laser Systems
Applicant
Wuhan Gigaa Optronics Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The "GBOX-15A/15B" are indicated for use in surgical applications requiring the hemostasis, ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), gastroenterology, general surgery, genitourinary surgery (urology), qynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, and thoracic surgery; and Laser Assisted Lipolysis (980 nm only). The "VELASII-30A/30B" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (UJrology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology Endovascular coagulation, Oral Surgery and Dental procedures, laser assisted lipolysis. The "VELASII -60A/60B/60C" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue. Such as: Dermatology, Aesthetics, Plastic Surgery, Vascular Surgery, Podiatry, Endovenous Occlusion of the greater saphenous vein of the thigh in patients with superficial vein reflux.
Device Story
Medical diode laser systems (GaAlAs diode bar) utilize semiconductor technology to generate laser energy at 810nm, 940nm, or 980nm. The device includes a main unit with a color LCD touch screen, power source, and fiber-coupling technology to deliver laser energy via optical fibers. Operated by physicians in clinical settings, the system is activated via a footswitch after parameter selection. The laser energy is used for soft tissue surgical applications (ablation, coagulation, etc.). The device provides visual feedback via the LCD and requires the use of protective safety goggles. By enabling precise tissue interaction, the device facilitates surgical procedures, potentially reducing patient trauma and recovery time compared to traditional methods.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-2-22, IEC 60601-1-2, IEC 60825-1, and ISO 10993 (biocompatibility) standards was demonstrated. Software verification was performed per FDA guidance.
Technological Characteristics
GaAlAs diode laser; wavelengths 810/940/980nm; output power 1-60W; CW, single pulse, and repeat pulse modes. Fiber-coupled delivery (400-600µm core). Class I Type B safety classification. IPX1 main unit, IPX8 footswitch. Microprocessor-controlled. Standards: IEC 60601-1, IEC 60601-2-22, ISO 10993.
Indications for Use
Indicated for patients requiring soft tissue surgery (hemostasis, ablation, vaporization, excision, incision, coagulation) across various medical specialties including dermatology, plastic surgery, urology, gynecology, ENT, and others. Specific indications include BPH, vascular lesions, hair removal, and endovenous occlusion of the greater saphenous vein for superficial vein reflux.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K060304 — PHOTEX DIODE LASER SERIES, MODELS 980, 810 AND 940 · Biotex, Inc. · Mar 21, 2006
K243037 — Diode Laser System model Dawn-S · Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd · Dec 26, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 23, 2015
Wuhan Gigaa Optronics Technology Company, Ltd Nie Xinxing Regulation Control Manager Hi-Tech Medical Device Industrial Park 818 Gauxin Avenue 5.6/f Unit A. B. Building B8 East Lake Development Zone Wuhan. CHINA 430206 Hubei
Re: K151890
Trade/Device Name: Medical Diode Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 27, 2015 Received: November 27, 2015
Dear Nie Xinxing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
#### K151890
Device Name
Medical Diode Laser Systems
Indications for Use (Describe)
The "GBOX-15A/15B" are indicated for use in surgical applications requiring the hemostasis, ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), gastroenterology, general surgery, genitourinary surgery (urology), qynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, and thoracic surgery; and Laser Assisted Lipolysis (980 nm only).
The "VELASII-30A/30B" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (UJrology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology Endovascular coagulation, Oral Surgery and Dental procedures, laser assisted lipolysis.
The "VELASII -60A/60B/60C" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue. Such as: Dermatology, Aesthetics, Plastic Surgery, Vascular Surgery, Podiatry, Endovenous Occlusion of the greater saphenous vein of the thigh in patients with superficial vein reflux.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Section 5 - 510(k) Summary
Date of Summary Preparation: August 06, 2015
#### 1. Submitter's Identifications
Submitter's Name: Wuhan Gigaa Optronics Technology Co., Ltd Address: 5, 6/F, Unit A, B, Building B8, Hi-Tech Medical Device Industrial Park, #818 Gaoxin Avenue, East Lake Development Zone, Wuhan 430206,China Contact Person: Nie Xin xing Contact Email Address: fdagigaalaser(@sohu.com Telephone: +86-27-67848871-8045 Fax: +86-27-67848873
#### 2. Correspondent's Identifications
Correspondent's Name: Wuhan Gigaa Optronics Technology Co., Ltd Address: 5, 6/F, Unit A, B, Building B8, Hi-Tech Medical Device Industrial Park, #818 Gaoxin Avenue, East Lake Development Zone, Wuhan 430206, China Contact Person: Nie Xinxing Contact Email Address: fdagigaalaser(@sohu.com Telephone: +86-27-67848871-8045 Fax: +86-27-67848873
#### 3. Name of the Device
Device Classification Name: POWERED LASER SURGICAL INSTRUMENT Product Name: POWERED LASER SURGICAL INSTRUMENT Trade Name: Medical diode laser systems Model: GBOX-15A/15B, VELAS II -30A/30B, VELAS II -60A/60B/60C Classification Panel: General & Plastic Surgery Product Code: GEX Device Classification: Class II
#### 4. The Predicate Devices
LaserPro Diode Surgical Laser System, LaserPro 810, LaserPro 980 K082721 JOULE 810/940/980 Multi-Platform System, JOULE 810, JOULE 940, JOULE 980 K122567 Quanta System QUANTA Diode Laser Family, QUANTA 808, QUANTA 980 K100558
#### 5. Device Description
Medical diode laser systems are a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.
The medical diode laser systems for this unit is GaAlAs diode bar, and the wavelength are 810mm/980mm of GBOX-15A/15B, 810/980mm of VELAS II -30A/30B, 810/980/940nm of
{5}------------------------------------------------
VELAS II -60A/60B/60C. It features impact structure, high efficiency and long lifetime. The beam divergence angle is too large lead to poor beam quality. In order to better meet the needs of Laser Diode practical applications, Laser Diode carried beam shaping is very necessary. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber.
To start the laser unit, turn the main switch ON into " | " position. Immediately the power indicator will be green with the system fans working. At the same time, the LCD screen lights up as the picture shows below.
After finishing setting the parameters, press the "Ready" button and then the system will remind you to wear the safety goggles (protect wavelength from 800nm to 1100nm). At this time when you press down the footswitch, the laser will emit.
#### 6. Intended Use of Device
Intended Use of Medical Diode Laser Systems "GBOX-15A" and "GBOX-15B" are in surgical applications requiring the hemostasis, ablation, vaporization, excision, and coagulation of tissue in medical specialties including: aesthetic (dermatology and plastic surgery), soft gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, and thoracic surgery; and Laser Assisted Lipolysis (980 nm only).
Intended Use of Medical Diode Laser Systems "VELAS II-30A" and "VELAS II-30B" are in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (UJrology), Thoracic Surgery, Plastic Surgery Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, and Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology Endovascular coagulation, Oral Surgery and Dental procedures, laser assisted lipolysis.
Intended Use of Medical Diode Laser Systems "VELAS II-60A", "VELAS II-60B" and "VELAS II-60C" are in surgical applications requiring the vaporization, incision, ablation, cutting and hemostasis, or coagulation of soft tissue. Such as: Dermatology, Aesthetics, Plastic Surgery, Vascular Surgery, Podiatry, Endovenous Occlusion of the greater saphenous vein of the thigh in patients with superficial vein reflux.
#### 7. Summary of Substantial Equivalence
Table 1: The difference between GBOX-15A/15B and Predicate LaserPro 810, LaserPro 980 (K082721)
| | Proposed Device | Predicate device | Comparison |
|--------------|-----------------|------------------|------------|
| 510k Number | K151890 | K082721 | ------ |
| Product Code | GEX | GEX | Same |
Table 1
{6}------------------------------------------------
| Proprietary Name | Medical Diode Laser Systems | LaserPro Diode Surgical Laser System | |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | GBOX-15A, GBOX-15B | LaserPro 810, LaserPro 980 | ------- |
| Manufacturer | Wuhan Gigaa Optronics<br>Technology Co., Ltd. | PhotoMedex, Inc. | ------- |
| Indications for use | The Medical Diode Laser Systems “GBOX-15A/15B”<br>are indicated for use in surgical applications requiring the<br>ablation, hemostasis, vaporization, excision, incision,<br>and coagulation of soft tissue in medical specialties including:<br>aesthetic (dermatology and plastic surgery),<br>gastroenterology, general surgery, genitourinary surgery<br>(urology), gynecology (GYN), neurosurgery, otolaryngology<br>(ENT), ophthalmology, arthroscopy, podiatry,<br>pulmonology, and thoracic surgery; and Laser Assisted<br>Lipolysis (980 nm only). | The PhotoMedex LaserPro 810, 940 and 980 Surgical<br>Diode Laser Systems, (and the fiber delivery systems<br>and accessories used to deliver laser energy), are<br>indicated for use in surgical applications requiring the<br>ablation, hemostasis, vaporization, excision, incision,<br>and coagulation of soft tissue in medical<br>specialties including: aesthetic(dermatology and<br>plastic surgery), gastroenterology, general<br>surgery, genitourinary surgery (urology), gynecology<br>(GYN), neurosurgery, otolaryngology (ENT),<br>ophthalmology, arthroscopy, podiatry, pulmonology, and<br>thoracic surgery; and Laser Assisted Lipolysis (980 nm<br>only). | Same |
| Laser Type | diode laser | diode laser | Same |
| Components | Laser system, Color touch screen, Fiber, Foot switch,<br>Power detector | Laser system, Color touch screen, Fiber and the handpiece,<br>Foot switch, Power detector | Similar.<br>Handle as the laser transmission system,<br>should be used with optical fiber. So GBOX<br>only uses fiber as the laser transmission<br>system does not affect the safety and<br>effectiveness. |
| Wavelength | GBOX-15A: 810nm±10nm<br>GBOX-15B: 980nm±10nm | LaserPro 810: 810nm±10%<br>LaserPro 980: 980nm±10% | Similar.<br>Center wavelength is |
| | | | consistent, and GBOX<br>wavelength accuracy is<br>better than that of the |
| | | | predicate device. |
| Output power | 1-15W | 1-15W<br>1-20W<br>1-25W | Similar.<br>Same as 1-15W. GBOX<br>design output power<br>has claimed to meet<br>intended use. |
| Operation mode | CW, single pulse, repeat pulse | CW, single pulse, repeat pulse | Same |
| Pulse width | $\le$ 2.5s | 0.05s -10s | Similar.<br>Pulse width range of<br>the device is contained<br>in that of the predicate<br>device, and it has been<br>designed to meet the<br>intended use<br>requirement claimed. |
| Application / Light<br>delivery system | Bare fiber, long as 3m<br>Fiber core diameter: 400µm,<br>600μm<br>NA $\ge$ 0.22<br>With SMA905 connector<br>The fiber is use on a single<br>patient for once<br>Before use must sterile<br>contact: fibers of 400μm, 600μm<br>with SMA905 connector;<br>Non-contact: fibers and tips | SMA905 connector<br>Silica fiber core diameters<br>greater than or equal to 365<br>microns and Fiber numerical<br>aperture greater than or equal to<br>0.29 NA | Similar. |
| Aiming Beam | Diode laser of 635nm, power<br>max. $<$ 5mW, adjustable<br>brightness | 635-650nm, adjustable intensity,<br>either continuous or pulsed | Similar. |
| Laser Class | 4 | 4 | Same |
| Operation interface | Color LCD touch screen | Color LCD touch screen | Same |
| Power Supply | 100-240VAC, 200VA, 50-60Hz | 100-240 VAC, 540VA max,<br>50/60Hz | Similar.<br>Only rated power is<br>different, not affect<br>safety and<br>effectiveness. |
| Safety classification | Class I Type B | Class I Type B | Same |
| Dimensions | 215(W)*245(L)*315(H)mm | 228(W)*406(L)*356(H)mm | Similar.<br>GBOX is more light |
| | | | and portable. |
| Weight | 4 kg | 11.2kg | Similar.<br>GBOX is more light<br>and portable. |
| Waterproof level | IPX1 | IPX1 | Same |
| Footswitch<br>Waterproof level | IPX8 | IPX8 | Same |
| Standard | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-22<br>IEC60825-1<br>ISO10993-1<br>ISO10993-4<br>ISO10993-5<br>ISO10993-10<br>ISO10993-11 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-22<br>IEC60825-1<br>ISO10993-1<br>ISO10993-4<br>ISO10993-5<br>ISO10993-10<br>ISO10993-11 | Same |
| Non- sterile | Main engine is Non- sterile, but<br>fiber is sterile for single use. | Non- sterile | Similar.<br>Fiber steriled for single<br>use of GBOX is more<br>effective to ensure the<br>safety of patients. |
| Microprocessor<br>Control | Yes | Yes | Same |
{7}------------------------------------------------
{8}------------------------------------------------
Table 2: The difference between VELAS II -30A/30B, VELAS II-60A/60B/60C and Predicate JOULE 810, JOULE 940, JOULE 980 (K122567), QUANTA 808, QUANTA 980 (K100558)
| Table | 2 |
|-------|---|
|-------|---|
| | Proposed Device | Predicate device | Predicate device | Comparation |
|---------------------|------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------|
| 510k Number | K151890 | K122567 | K100558 | ------ |
| Product Code | GEX | GEX | GEX | Same |
| Proprietary Name | Medical Diode Laser<br>Systems | JOULE 810/940/980<br>Multi-Platform System | Quanta System<br>QUANTA Diode Laser<br>Family | ------ |
| Model | VELAS II -30A, VELAS<br>II -30B, VELAS II -60A,<br>VELAS II -60B, VELAS<br>II -60C | JOULE 810, JOULE<br>940, JOULE 980 | QUANTA 808,<br>QUANTA 980 | ------ |
| Manufacturer | Wuhan Gigaa Optronics<br>Technology Co., Ltd | Sciton, Inc. | Quanta System SpA | ------ |
| Indications for use | The<br>"VELAS II -30A/30B"<br>are indicated for use in | The JOULE 810/940/980<br>Multi-Platform System<br>are indicated for use in | The Quanta System<br>QUANTA Diode Laser<br>Family are indicated | Same.<br>The indications for |
{9}------------------------------------------------
applications surgical surgical applications for use in surgical of use requiring requiring the applications requiring "VELAS II -30A/3 the vaporization, incision, vaporization, incision, the vaporization, 0B" are same to excision, ablation. excision, ablation, incision, excision, that of Quanta cutting and hemostasis. ablation, cutting and System QUANTA cutting and hemostasis, or coagulation of soft or coagulation of soft hemostasis, or Diode Laser tissue in conjunction Family. tissue. Such as: coagulation of soft with endoscopic Dermatology, tissue in conjunction equipment for medical Aesthetics,Plastic with endoscopic The indications for specialist including: Surgery, Vascular equipment for medical use of Urology (BPH). Surgery, Podiatry, specialist including: "VELAS II -60A/6 0B/60C" are same Genitourinary Endovenous Occlusion Urology (BPH), (UJrology), Thoracic of the greater saphenous Genitourinary to that of JOULE Surgery, Plastic Surgery vein of the thigh in (UJrology), Thoracic 810/940/980 and Dermatology, patients with superficial Surgery, Plastic Surgery Multi-Platform Aesthetics including vein reflux. and Dermatology, System. vascular lesions and hair Aesthetics including removal, General vascular lesions and hair Surgery, removal, Ophthalmology, General Surgery, Podiatry. Orthopedics. Ophthalmology, Arthroscopy, Spinal Orthopedics, Podiatry, Gynecology, Arthroscopy , Spinal Surgery, Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Pulmonary Surgery, Gastroenterology, Neurosurgery(PLDD), Head/neck/ENT Gastroenterology, and Radiology Endovascular Head/neck/ENT and coagulation, Oral Radiology Endovascular Surgery and coagulation, Oral Dental procedures, laser assisted Surgery and Dental procedures, laser assisted lipolysis. lipolysis. The "VELAS II -60A/60B/60 C" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue. Such as: Dermatology, Aesthetics,
{10}------------------------------------------------
| | | | | | | | | | designed to meet<br>the intended use<br>requirement<br>claimed. |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Plastic Surgery, Vascular<br>Surgery, Podiatry,<br>Endovenous Occlusion<br>of the greater saphenous<br>vein of the thigh in<br>patients with superficial<br>vein reflux. | | | | Operation mode | CW, single pulse, repeat<br>pulse | CW, single pulse, repeat<br>pulse | CW, single pulse, repeat<br>pulse | Same |
| Laser Type | diode laser | diode laser | diode laser | Same | Pulse width | 10ms-2.5s | ≤2500ms | 0.25-10000ms | Similar.<br>Pulse width range<br>of<br>"VELAS II -30A/3<br>0B,<br>VELAS II -60A/60<br>B/60C"<br>is<br>contained in that of<br>the two types<br>predicate device,<br>and it has been<br>designed to meet<br>the intended use<br>requirement<br>claimed. |
| Components | Laser system, Color<br>touch screen, Fiber, Foot<br>switch, Power calibrator | Laser system, Color<br>touch screen, Fiber, Foot<br>switch, Power calibrator | Laser system, Color<br>touch screen, Fiber, Foot<br>switch, Power calibrator | Similar.<br>Handle as the laser<br>transmission<br>system, should be<br>used with optical<br>fiber. So<br>VELAS II only<br>uses fiber as the<br>laser transmission<br>system does not<br>affect the safety<br>and effectiveness. | Application / Light<br>delivery system | Fiber core diameter:<br>400µm, 600μm<br>NA ≥0.22<br>With SMA905 connector<br>The fiber is use on a<br>single patient for once<br>Before use must sterile<br>contact: fibers of 400µm,<br>600µm with SMA905<br>connector; Non-contact:<br>fibers and tips | K122567<br>Fiber optic with<br>handpiece | K100558<br>Bare fier size:<br>200,300,320,400,600,80<br>0,1000μm | Similar.<br>VELAS II is<br>described more<br>detailed.<br>Fiber core diameter<br>range of VELAS II<br>is contained in that<br>of the two types<br>predicate device,<br>and it has been<br>designed to meet<br>the intended use<br>requirement<br>claimed. |
| Wavelength | VELAS II-30A, VELAS<br>II-60A: 810nm $\pm$ 10nm<br>VELAS II-30B, VELAS<br>II-60B: 980nm $\pm$ 10nm<br>VELAS II-60C:<br>940nm $\pm$ 10nm | JOULE 810: 810nm<br>JOULE 940: 940nm<br>JOULE 980: 980nm | QUANTA 808: 808nm<br>QUANTA 980: 980nm | Similar.<br>Center wavelength<br>is consistent.<br>Reasonable<br>accuracy range of<br>VELAS II does not<br>affect safety and<br>effectiveness. | Aiming Beam | Diode laser of<br>635/532nm power max.<br>< 5mW, adjustable<br>brightness. | --- | red (650nm), adjustable -<br><5mW | Similar.<br>VELAS II has two<br>types of indication<br>light which not<br>affect safety and<br>effectiveness. |
| Output power | VELAS II-30A/ VELAS<br>II-30B: 1-30W<br>VELAS II-60A, VELAS<br>II-60B, VELAS II-60C:<br>1-60W | $\leq$ 100W | 30W | Similar.<br>Output power of<br>"VELAS II -30A/3<br>0B" is Same to that<br>for "QUANTA<br>808, QUANTA<br>980".<br>Output power<br>range of<br>"VELAS II -60A/6<br>0B/60C" is<br>contained in that of<br>"JOULE 810,<br>JOULE 940,<br>JOULE 980", and<br>it has been | Laser Class | 4 | 4 | 4 | Same |
| Operation interface | Color LCD touch screen | Color LCD touch screen | Color LCD touch screen | Same | | | | | |
{11}------------------------------------------------
{12}------------------------------------------------
| | Wuhan Gigaa Optronics Technology Co., Ltd | |
|--|-------------------------------------------|--|
| | | |
| Power Supply | VELAS II -30A, VELAS<br>II -30B: 100-240VAC,<br>50/60Hz, 350VA<br>VELAS II-60A, VELAS<br>II-60B, VELAS II-60C:<br>100-240VAC, 50/60Hz,<br>650VA | 230VAC, 50/60Hz | 100-240VAC, 50-60Hz,<br>350VA | Similar.<br>Only rated power<br>is different, not<br>affect safety and<br>effectiveness. |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Safety classification | Class I Type B | Class I Type B | Class I Type B | Same |
| Dimensions | 400(W)*385(L)*200(H)<br>mm | 762 (W)*406<br>(L)*1092(H)mm | 330(W)*390(L)*250(H)<br>mm | ---- |
| Weight | 12.9kg | ---------- | 8kg | ---- |
| Waterproof level | IPX1 | IPX1 | IPX1 | Same |
| Footswitch<br>Waterproof level | IPX8 | IPX8 | IPX8 | Same |
| Standard | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-22<br>IEC60825-1<br>ISO10993-1<br>ISO10993-5<br>ISO10993-10 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-22<br>IEC60825-1<br>ISO10993-1<br>ISO10993-5<br>ISO10993-10 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-22<br>IEC60825-1<br>ISO10993-1<br>ISO10993-5<br>ISO10993-10 | Same |
| Non- sterile | Main engine is Non-<br>sterile, but fiber is sterile<br>for single use. | Non- sterile | Non- sterile | Similar.<br>Fiber steriled for<br>single use of<br>VELAS II is more<br>effective to ensure<br>the safety of<br>patients. |
| Microprocessor<br>Control | Yes | Yes | Yes | Same |
#### 8. Substantial Equivalence:
The proposed devices of GBOX-15A, GBOX-15B have the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The differences only exist in The differences only exist in such contents: component of handpiece that do not influence use even if without it, wave rang of wavelength, output power, pulse width, and aiming beam that all can be controlled in range of application. These differences are slight and do not influence the effectiveness and safety of the device. According to the non-clinical test results, the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.
{13}------------------------------------------------
The proposed devices of VELAS II-30A, VELAS II-30B, VELAS II-60A, VELAS II-60B, and VELAS II -60C have the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The differences only exist in such contents: component of handpiece that do not influence use even if without it, wave rang of wavelength, output power, pulse width, and aiming beam that all can be controlled in range of application. These differences are slight and do not influence the effectiveness and safety of the device. According to the non-clinical and clinical test results, the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.
#### 9. Non-Clinical Tests Performed:
The following testing was performed on the "GBOX" and "VELAS II" Medical Diode Laser Systems in accordance with the requirements of the design control regulations and established quality assurance procedures.
IEC60601-1:2005+A1:2012+National differences of US ANSI/AAMI Medical electrical equipment-Part 1: General requirements for basic safety and essential performance Incorporates Amendment 1:2012; Incorporates National differences of US ANSI/AAMI
ES60601-1:2005+C1:2009+A2:2010+A1:2012 Medical equipment-Part 1: General requirements for basic safety and essential performance Consolidated reprint C1:2009; Incorporates Amendment 2:2010; Incorporates Amendment 1:2012
IEC60601-2-22:2007+A1:2012 Medical electrical equipment-Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment Incorporates Amendment 1:2012
IEC60601-1-2:2007 Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC60825-1:2007 Safety of laser products-Part 1: Equipment classification and requirements
ISO10993-1:2009/C1:2010, Biological evaluation of medical devices-Part 1: evaluation and testing within a risk management process.
ISO10993-4:2002/A1:2006, Biological evaluation of medical devices - Part 4: selection of tests for interactions with blood.
ISO10993-5:2009, Biological evaluation of medical devices-Part 5: tests for in vitro cytotoxicity.
ISO10993-10:2010, Biological evaluation of medical devices-Part10: tests for irritation and skin sensitization.
ISO10993-11:2006, Biological evaluation of medical devices - part 11: tests for systemic toxicity. In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
#### 10. Conclusion:
The proposed devices of GBOX-15A, GBOX-15B are determined to be Substantially Equivalent (SE) to the predicate device, LaserPro Diode Surgical Laser System, LaserPro 810/980 in respect of safety and effectiveness.
{14}------------------------------------------------
The proposed devices of VELAS II-30A, VELAS II-60A, VELAS II-60A, VELAS II-60B, VELAS II-60C are determined to be Substantially Equivalent (SE) to the predicate device, JOULE 810/940/980 Multi-Platform System, JOULE 810, JOULE 940, JOULE 980 and Quanta System QUANTA Diode Laser Family, QUANTA 808, QUANTA 980 in respect of safety and effectiveness.
--- End of this section ---
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.