Diode Laser System model Dawn-S
K243037 · Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd · GEX · Dec 26, 2024 · General, Plastic Surgery
Device Facts
| Record ID | K243037 |
| Device Name | Diode Laser System model Dawn-S |
| Applicant | Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | Dec 26, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Diode Laser System model Dawn-S is indicated for incision, ablation, vaporization, and coagulation of body soft tissues in surgical application. The 980 nm diode laser is generally indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue, including: general surgery, ENT (ear, nose, and throat) and oral surgery (otolaryngology), dental procedures, dermatology, plastic surgery, podiatry, urology, gastroenterology, and gynecology. The laser is further indicated for laser assisted lipolysis. The 1470nm diode laser is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular, coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The laser is further indicated for laser assisted lipolysis.
Device Story
Diode Laser System model Dawn-S delivers dual-wavelength (980nm and 1470nm) laser energy to soft tissue for surgical applications. System comprises optical block (laser diodes, mirrors, lens, aiming beam), air cooling system, and power supply. Operated via high-resolution color touch screen for selection of CW/Pulse modes, duty cycle, frequency, and power. External footswitch triggers laser activation. Used in clinical settings (OR, clinic) by medical specialists. Laser energy delivered via fiber optic cable to target tissue; facilitates vaporization, incision, excision, ablation, and coagulation. Clinical benefit includes precise tissue management and hemostasis during surgical procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and adherence to recognized international safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22).
Technological Characteristics
Dual-wavelength (980nm/1470nm) diode laser system. Air-cooled. Fiber-optic delivery. 8-inch color touch screen interface. Power supply 100-240V, 50-60Hz. Aiming beam 650nm (<5mW). Operation modes: CW and Pulsed (10-990ms). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and IEC 60601-2-22.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues in surgical applications including general surgery, ENT, oral/dental, dermatology, plastic surgery, podiatry, urology, gastroenterology, gynecology, thoracic surgery, ophthalmology, orthopedics, arthroscopy, spinal surgery, pulmonary surgery, neurosurgery, and endovascular procedures (e.g., saphenous vein occlusion). Also indicated for laser-assisted lipolysis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- Diode Laser Therapy Device (K212734)
- Medical Diode Laser (K240747)
- Medical Diode Laser (K240644)
Related Devices
- K060304 — PHOTEX DIODE LASER SERIES, MODELS 980, 810 AND 940 · Biotex, Inc. · Mar 21, 2006
- K133006 — NEOV DIODE LASER FAMILY · G.N.S Neolaser , Ltd. · Apr 17, 2014
- K252658 — Medical Diode Laser Systems (TR-B) · Baoding Te'Anzhou Electronic Technology Co., Ltd. · Nov 28, 2025
- K202041 — Opera Evo · Quanta System Spa · Oct 15, 2020
- K152722 — neoV980 & neoV1470 Diode Lasers · G.N.S Neolaser , Ltd. · Dec 8, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
December 26, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd Lily Zhou Management Representative 2F, Building 11, Lane 1175, Tongpu Rd, Shanghai, 200333 China
Re: K243037
Trade/Device Name: Diode Laser System model Dawn-S Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 22, 2024 Received: September 27, 2024
Dear Lily Zhou:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Digitally signed by Shlomit Shlomit Halachmi -S Date: 2024.12.26 17:23:42 Halachmi -S -05'00'
Shlomit Halachmi For Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243037
Device Name Diode Laser System model Dawn-S
#### Indications for Use (Describe)
The Diode Laser System model Dawn-S is indicated for incision, ablation, vaporization, and coagulation of body soft tissues in surgical application.
The 980mm diode laser is generally indicated for use in surgical applications requiring the vaporization, incision, ablation, cutting and hemostasis, or coagulation of soft tissue, including: general surgery, ENT (ear, nose, and throat) and oral surgery (otolaryngoloty), dental procedures, dermatology, plastic surgery, podiatry, urology, gastroenterology, and gynecology. The laser is further indicated for laser assisted lipolysis.
The 1470mm diode laser is indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular, coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The laser is further indicated for laser assisted lipolysis.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.## This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K243037
Date Prepared: December 26, 2024
This summary of 510 (k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 801.92
#### 1. Submitter's Information
2.
| Name of Sponsor: | Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. |
|------------------|------------------------------------------------------------------------------|
| Address: | 2F, Building 11, Lane 1175, Tongpu Rd.,<br>Shanghai 200333, China |
| Contact Name: | Lily Zhou |
| Telephone No.: | 0086-021-62642623 |
| Fax No.: | 0086-021-52827988 |
| Email Address: | <a href="mailto:regulatory@wonderful-sh.com">regulatory@wonderful-sh.com</a> |
**Correspondent's Information**
| Company Name: | Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. |
|---------------|---------------------------------------------------|
|---------------|---------------------------------------------------|
| Company Name: | Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. |
|---------------------|---------------------------------------------------|
| Correspondent Name: | Lily Zhou |
| Telephone No.: | 0086-021-62642623 |
| Email Address: | regulatory@wonderful-sh.com |
#### 3. Trade Name, Common Name, Classification
| Trade Name: | Diode Laser System |
|----------------------------|----------------------------------------------------------------------------------------|
| Common Name: | Powered Laser Surgical Instrument |
| Model: | Dawn-S |
| Regulation Name: | Laser surgical instrument for use in general and plastic surgery and<br>in dermatology |
| Product Code: | GEX |
| Classification Panel: | General and Plastic Surgery |
| Device Class: | II |
| Classification Regulation: | 21 CFR 878.4810 |
#### 4. Identification of Predicate Device(s)
The identified predicates within this submission are as follows:
{5}------------------------------------------------
| Predicate<br>Devices: | 510 (k) number: K212734 |
|-----------------------|----------------------------------------------------------------------------------|
| | Trade name: Diode Laser Therapy Device<br>Product code: GEX |
| | 510(K) number:K240747<br>Device name: Medical Diode Laser<br>Product code: GEX |
| | 510 (k) number: K240644<br>Device name: Medical Diode Laser<br>Product code: GEX |
#### 5. Description of the Device
The Diode Laser System model Dawn-S is intended for delivery of dual wavelength laser beam to soft tissue. The Diode Laser System model Dawn-S generates a 1470nm wavelength laser and 980nm wavelength laser.
The laser is controlled via a high-resolution color touch screen display includes a user interface allowing selection of CW and Pulse mode as well as duty cycle and frequency of operation, laser power level and Stand-by/Ready selection.
The laser system consists of an optical block with contains the laser diode, mirrors, lens, and aiming beam diode, an air cooling system, and the laser power system, as well as an external footswitch for laser activation.
#### 6. Intended Use/Indication for Use
The Diode Laser System model Dawn-S is indicated for incision, ablation, vaporization, and coagulation of body soft tissues in surgical application.
The 980 nm diode laser is generally indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue, including: general surgery, ENT (ear, nose, and throat) and oral surgery (otolaryngology), dental procedures, dermatology, plastic surgery, podiatry, urology, gastroenterology, and gynecology, The laser is further indicated for laser assisted lipolysis.
The 1470nm diode laser is indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ ENT and Radiology, Oral Surgery and Dental procedures, Endovascular, coagulation and endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The laser is further indicated for laser assisted lipolysis.
#### 7. Technological Characteristics
The Diode Laser System model Dawn-S uses the same technology that is utilized in the predicate device. A comparison of technological characteristics is provided in the following table:
{6}------------------------------------------------
| Item | Diode Laser System<br>model Dawn-S | Predicate Device | | | Comparison |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Information | | | | | |
| Product Code | GEX | GEX | GEX | GEX | Identical |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Class | II | II | II | II | Identical |
| Product Name | Diode Laser System model Dawn-S | Diode laser therapy device | Medical Diode Laser | Medical Diode Laser | |
| 510 (K) No. | K243037 | K212734 | K240747 | K240644 | |
| Models | Dawn-S | ST-AR | M2-GK | S1Pro | |
| Indications for<br>Use | The Diode Laser System model<br>Dawn-S is indicated for incision,<br>excision, ablation, vaporization, and<br>coagulation of body soft tissues in<br>surgical application.<br><br>The 980 nm diode laser is generally<br>indicated for use in surgical<br>applications requiring the<br>vaporization, incision, excision,<br>ablation, cutting and hemostasis, or<br>coagulation of soft tissue, including:<br>general surgery, ENT (ear, nose,<br>and throat) and oral surgery<br>(otolaryngology), dental procedures,<br>dermatology, plastic surgery,<br>podiatry, urology, gastroenterology,<br>gynecology. The laser is further<br>indicated for laser assisted lipolysis.<br><br>The 1470 nm diode laser is indicated<br>for use in surgical applications<br>requiring the vaporization, incision,<br>excision, ablation, cutting and<br>hemostasis, or coagulation of soft<br>tissue in conjunction with<br>endoscopic equipment for medical<br>specialist including: Urology,<br>Thoracic Surgery, Plastic Surgery<br>and Dermatology, General Surgery,<br>Ophthalmology, Orthopedics,<br>Podiatry, Arthroscopy, Spinal<br>Surgery, Gynecology, Pulmonary<br>Surgery, Neurosurgery,<br>Gastroenterology, Head/neck/ENT<br>and Radiology, Oral Surgery and<br>Dental procedures | The Diode laser therapy<br>device is intended for delivery<br>of laser light to soft tissue in<br>the contact and non contact<br>mode during surgical<br>procedures. The device 's<br>980 nm laser is generally<br>indicated for use in incision,<br>excision, vaporization,<br>ablation, hemostasis or<br>coagulation of soft tissue in<br>ear, nose and throat and oral<br>surgery(otolaryngology),<br>dental procedures,<br>gastroenterology,<br>general surgery,<br>dermatology, plastic surgery,<br>podiatry, urology,<br>gynecology. The device is<br>further indicated for laser<br>assisted lipolysis.<br><br>The device's 1470nm laser<br>is intended for delivery of<br>laser light to soft tissue in<br>non-contact mode during<br>general surgery procedures,<br>indicated for the treatment of<br>reflux of the saphenous veins<br>associated with varicose<br>veins and varicosities. | The Medical Diode Laser<br>(Model: M2-GK) is indicated<br>for use in surgical<br>applications requiring the<br>vaporization, incision,<br>excision, ablation, cutting and<br>hemostasis, or coagulation of<br>soft tissue in conjunction with<br>endoscopic equipment for<br>medical specialist including:<br>Urology, Thoracic Surgery,<br>Plastic Surgery and<br>Dermatology, General<br>Surgery, Ophthalmology,<br>Orthopedics, Podiatry,<br>Arthroscopy, Spinal Surgery,<br>Gynecology, Pulmonary<br>Surgery, Neurosurgery,<br>Gastroenterology,<br>Head/neck/ENT and<br>Radiology, Oral Surgery and<br>Dental procedures,<br>Endovascular coagulation<br>and endovenous occlusion of<br>the greater saphenous vein in<br>patients with superficial vein<br>reflux. The Medical Diode<br>Laser (Model: M2-GK) is<br>further indicated for laser<br>assisted lipolysis. | The Medical Diode<br>Laser(Model: S1Pro) is<br>indicated for use in surgical<br>applications requiring the<br>vaporization, incision,<br>excision, ablation, cutting<br>and hemostasis, or<br>coagulation of soft tissue in<br>conjunction with endoscopic<br>equipment for medical<br>specialist including: Urology,<br>Thoracic Surgery, Plastic<br>Surgery and Dermatology,<br>General Surgery,<br>Ophthalmology,<br>Orthopedics, Podiatry,<br>Arthroscopy, Spinal Surgery,<br>Gynecology, Pulmonary<br>Surgery, Neurosurgery,<br>Gastroenterology,<br>Head/neck/ENT and<br>Radiology, Oral Surgery and<br>Dental procedures,<br>Endovascular coagulation<br>and endovenous occlusion<br>of the greater saphenous<br>vein in patients with<br>superficial vein reflux. The<br>Medical Diode Laser(Model:<br>S1 Pro) is further indicated<br>for laser assisted lipolysis. | Indications for<br>use of the<br>980nm diode<br>laser are the<br>same as the<br>indications of<br>the 980nm<br>laser in<br>K212734.<br><br>Indications for<br>use of the<br>1470nm diode<br>laser are<br>identical to the<br>indications for<br>use of the<br>1470nm laser<br>in K240747<br>and K240644. |
| | coagulation and endovenous<br>occlusion of the greater saphenous<br>vein in patients with superficial vein<br>reflux. The laser is further indicated<br>for laser assisted lipolysis. | | | | |
| Performance Information | | | | | |
| Wavelength | 980nm, 1470nm | 980nm, 1470nm | 980nm, 1470nm | 1470nm | Identical |
| Output Power | 15W (980nm)<br>15W (1470nm) | 16W/980nm±20%,<br>4.5W/1470nm±20% | 30W-980nm<br>15W-1470nm | 12W-1470nm | 980nm power<br>is similar to<br>K212734, only<br>1W lower.<br>1470nm power<br>is identical to<br>K240747 and<br>slightly higher<br>than K240644 |
| Aiming Beam | 650nm, <5mW, adjustable | 650 nm, red 0.5 mW, user<br>controlled intensity | Red 650nm (<2mW) | Diode laser of 650 nm, power<br><2mW,<br>adjustable<br>brightness. | Same<br>wavelength,<br>red light, the<br>aiming beam<br>of subjected<br>and predicated<br>devices are all<br>classified as<br>the lowest<br>safety level,<br>Class I. |
| Operation Mode | Continuous or Pulsed | Continuous or Pulsed | CW, Pulsed, Single Pulse…