K152722 · G.N.S Neolaser , Ltd. · GEX · Dec 8, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152722
Device Name
neoV980 & neoV1470 Diode Lasers
Applicant
G.N.S Neolaser , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The neoV980 & neoV1470 Diode Lasers, (and their delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialties including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation, and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.
Device Story
Solid-state infrared diode lasers (980nm and 1470nm) deliver continuous or pulsed energy for soft tissue surgical applications. System comprises optical block (laser diode, mirrors, lens, aiming beam), air cooling, and electronics. Operated via high-resolution color touch screen; allows selection of pulse modes, repetition rates, and aiming beam settings. Includes emergency shut-off, password protection, and external foot switch for activation. Used in clinical/surgical settings by physicians. Delivery via sterile/non-sterile fibers and handpieces. Output energy facilitates tissue vaporization, incision, excision, ablation, cutting, hemostasis, and coagulation. Benefits include precise tissue management and endovenous occlusion for venous reflux.
Clinical Evidence
No clinical data provided. Bench testing only, including compliance with IEC 60601-2-22 (safety of laser equipment), IEC 60601-1 (general safety/performance), IEC 60825-1 (laser product safety), and EN 60601-1-2 (electromagnetic compatibility).
Technological Characteristics
Solid-state infrared diode lasers; wavelengths 980nm and 1470nm; power 10-20 Watts. Air-cooled. Controlled via color touch screen interface. External low voltage power supply. Wired foot switch activation. Compliant with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and EN 60601-1-2.
Indications for Use
Indicated for patients requiring soft tissue surgery (vaporization, excision, ablation, cutting, hemostasis, coagulation) across various medical specialties including urology, dermatology, and orthopedics; specifically indicated for endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design. The eagle's head and neck are formed by the silhouettes of the human faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2015
G.N.S neoLaser Ltd. Mr. Gil Shapira Chief Executive Officer 7 Ha'Eshel Str. 38900 Caesarea ISRAEL, P.O. Box 3203
Re: K152722 Trade/Device Name: neoV980 & neoV1470 Diode Lasers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 15, 2015 Received: September 21, 2015
Dear Mr. Shapira:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Joshua C. Nipper -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152722
Device Name neoV980 & neoV1470 Diode Lasers
#### Indications for Use (Describe)
The neoV980 & neoV1470 Diode Lasers, (and their delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialties including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, Endovascular coagulation, and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------|----------------------------------------------|
| X Prescription Ise (Part 21 CER 801 Subnart D) | Over-The-Counter Use (21 C.FR 801 Subnart C. |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Image: neoLaser logo | neoV FDA 510(k) Submission K152722<br>Section 5 – 510(k) Summary | | | Revision #2 |
|----------------------------------------|------------------------------------------------------------------|--------------------------------|-------------------------------------|----------------------|
| Ha'Eshel St. 7, 38900, Caesarea ISRAEL | PO Box 3203, Phone:+972 4 677 9919, Fax:+972 4 8591505 | SOP Number:<br>DHF-001-R-025-5 | Effective<br>Date:<br>July 15, 2015 | Page 10<br>Section 5 |
# 510(k) SUMMARY
| Title: | neoV980 & neoV1470 Diode Lasers |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | G.N.S neoLaser Ltd.<br>Ha'Eshel St. 7, 38900, Caesarea ISRAEL, PO Box 3203 |
| Contact: | Gil Shapira, CEO<br>G.N.S neoLaser Ltd.<br>Ha'Eshel St. 7, 38900, Caesarea ISRAEL, PO Box 3203<br>Phone: +972 52 2246965<br>Fax: +972 4 859 1505<br>Email: shapirag@neo-laser.com |
| Date<br>Prepared: | July 15, 2015 |
| Device Trade<br>Name: | neoV980 & neoV1470 Diode Lasers |
| Common<br>Name: | Laser surgical instrument for use in general surgery and dermatology |
| Classification<br>Name: | Instrument, surgical, powered laser<br>GEX<br>21 CFR 878.4810 |
| Predicate<br>Devices: | Quanta Diode Laser Family (K100558), Ceralas 1470nm Diode Laser<br>(K112253) and Ceralas 980nm Diode Laser (K112324) |
| Device | Description: neoV980 & neoV1470, members of the neoV Diode Lasers family, cleared<br>under K133006, are medical grade, solid-state, infrared Diode lasers, designed<br>to deliver continuous or pulsed, infrared laser energy at wavelengths of 980nm<br>and 1470nm respectively, with power level ranging from 10 to 20 Watts<br>respectively. The lasers are controlled via a high-resolution color touch screen.<br>The touch screen display includes a user interface allowing selection of<br>continuous, repeat pulse, or single pulse modes of operation as well as<br>repetition rates, aiming beam settings, password key protection, and<br>standby/ready mode selection. The units have an emergency shut off button on |
{4}------------------------------------------------
| Image: neo Laser logo | neoV980 & neoV1470 Diode Lasers 510(k)<br>Submission<br>510(k) Summary | | | Revision #2 |
|----------------------------------------|------------------------------------------------------------------------|-------------------------------|----------------------------------|----------------------|
| Ha'Eshel St. 7, 38900, Caesarea ISRAEL | PO Box 3203, Phone:+972 4 677 9919, Fax:+972 4 8591505 | SOP Number:<br>DHF-001-R-025- | Effective Date:<br>July 15, 2015 | Page 11<br>Section 5 |
the front of the unit.
The Laser System: The laser system consists of an optical block which contains the laser diode, mirrors, lens, and aiming beam diode, an air cooling system, and electronics which include the color touch screen control panel. The unit utilizes an external low voltage power supply, as well as an external wired foot switch for laser activation.
The Delivery System: The delivery system consists of either sterile fibers (not provided with the system) or non-sterile fibers and hand pieces. Safety goggles, non sterile fibers, handpieces and a safety sign are provided with the unit.
### Intended
Use:
The neoV980 & neoV1470 Diode lasers are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialties including: Urology, Thoracic Surgery, Plastic Surgery and Dermatology, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Oral Surgery and Dental procedures, endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux.
## Substantial
- Equivalence: The neoV980 & neoV1470 Diode Lasers share the same intended use, technical and performance characteristics of the predicate devices Quanta System QUANTA Diode Laser Family (K100558), the Ceralas 1470nm Diode Laser (K112253) and the Ceralas 980nm Diode Laser (K112324).
- Conclusion: From a design and clinical perspective, the predicates and subject laser devices, have the same technological characteristics and share the same intended use. Accordingly, the safety and effectiveness of the neoV980 & neoV1470, is based upon a determination of the substantial equivalence to the predicate devices.
{5}------------------------------------------------
| Image: neo Laser | neoV980 & neoV1470 Diode Lasers 510(k)<br>Submission<br>510(k) Summary | | Revision #2 |
|--------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------|-------------|
| Ha'Eshel St. 7, 38900, Caesarea ISRAEL | | | Page 12 |
| PO Box 3203, Phone:+972 4 677 9919, Fax:+972 4 8591505 | SOP Number:<br>DHF-001-R-025- | Effective<br>Date:<br>July 15, 2015 | Section 5 |
### Non clinical Performance
Data:
The neoV Diode Laser family including neoV980 & neoV1470 has been tested for compliance to IEC 60601-2-22: 2007 (Third Edition), Medical electrical equipment Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment, IEC 60601-1: 2005 (Third Edition) Medical electrical equipment Part 1: General requirements for basic safety and essential performance and tested for compliance with all functional requirements, IEC 60825-1:2007 Safety of Laser Products - Part 1: Equipment classification, requirements and user's guide, EN 60601-1-2: 2007, Medical electrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests.
Clinical Performance Data: None
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.