K083899 · Hoya Conbio, Inc. · GEX · Jun 1, 2009 · General, Plastic Surgery
Device Facts
Record ID
K083899
Device Name
REVLITE Q-SWITCHED ND:YAG LASER SYSTEM
Applicant
Hoya Conbio, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis.
Device Story
RevLite Q-Switched Nd:YAG laser system delivers laser energy for dermatological procedures. System console houses laser source; energy delivered via articulated arm with MultiSpot handpiece (532 nm/1064 nm) or optional Multilite Dye Laser handpiece (650 nm/585 nm). Physician-operated via footswitch. Laser pulses (5-20 ns) target specific chromophores (pigment, vascular) to induce thermal effects for tissue ablation, coagulation, or pigment destruction. Used in clinical settings for skin resurfacing, tattoo removal, and lesion treatment. Benefits include precise tissue interaction and minimal thermal damage due to short pulse duration.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Q-switched Nd:YAG laser system; wavelengths: 1064 nm, 532 nm (optional 650 nm, 585 nm via dye handpiece); pulse duration: 5-20 ns; delivery: articulated arm with MultiSpot or Multilite handpieces; energy source: facility electrical power; console-based; footswitch activation.
Indications for Use
Indicated for patients requiring dermatological or surgical procedures including tattoo removal (dark/light inks), treatment of epidermal/dermal pigmented lesions (e.g., Nevus of Ota, cafe-au-lait, lentigines, freckles), vascular lesions (e.g., port wine stains, telangiectasias, angiomas), skin resurfacing for acne scars and wrinkles, hair removal, and reduction of red pigmentation in hypertrophic/keloid scars.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Alma, Harmony XL Multi-Platform (Q-Switched Nd:YAG hand piece) (K072564)
K063834 — REVLITE C6 Q-SWITCHED ND:YAG LASER SYSTEM · Hoya Conbio, Inc. · Jan 25, 2007
K102954 — CUTERA QSWITCH LASER SYSTEM · Cutera, Inc. · Jun 22, 2011
K161670 — Lucid Q-PTP / HWA 55 · Bison Medical Co, Ltd. · Nov 2, 2016
K172908 — Q-Switched Nd:YAG Laser System · Ami, Inc. · Dec 7, 2017
K103118 — REVLITE Q-SWITCHED ND: YAG LASER SYSTEM WITH NADPIECE ADAPTOR · Hoya Photonics, Inc. · Nov 19, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Premarket Notification - Original Submission
RevLite™ Q-Switched Nd:YAG Laser System, New Indication for Use
December 19, 2008
**HOYA ConBio™**
47733 Fremont Boulevard, Fremont, C
510-445-4500 www.conbio.com 510
| Submitter: | HOYA ConBio, Inc.<br>47733 Fremont Blvd.<br>Fremont, California 94538<br>Phone: 510-445-4500<br>Fax: 510-445-4550 | JUN - 1 2009 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact: | Mr. Jim Green<br>Vice President of Engineering | |
| Date Summary Prepared: | April 3, 2009 | |
| Device Trade Name: | RevLite™ Q-Switched Nd:YAG Laser Systems | |
| Common Name: | Dermatology Laser System | |
| Classification Name: | Instrument, surgical, powered, laser<br>79-GEX | |
| Classification Code: | 878.4810 Laser surgical instrument for use in general and plastic<br>surgery and in dermatology (1) A carbon dioxide laser for use in<br>general surgery and in dermatology is a laser device intended to<br>cut, destroy, or remove tissue by light energy emitted by carbon<br>dioxide.<br>(2) An argon laser for use in dermatology is a laser device<br>intended to destroy or coagulate tissue by light energy emitted by<br>argon. | |
| Equivalent Device(s): | Alma, Harmony XL Multi-Platform (Q-Switched Nd:YAG hand<br>piece) (K072564)<br>Candela VBeam, VBeam Laser (K033461, K021180) and<br>Sclerolaser (K013748)<br>Cynosure, Photogenica/VLS- Star Laser (K053608)<br>Iridex, Diolite2- 532 (K964074) | |
| Device Description: | The entire laser unit and controls are contained in a single<br>console. Electrical power is supplied to the console by the<br>facility's power source. Laser energy produced within the device<br>is delivered to the tissue by means of an articulated arm and a<br>specially designed MultiSpot Handpiece (532 nm and 1064 nm)<br>or optional Multilite Dye Laser Handpiece (650 nm and 585 nm).<br>The user activates laser emission by means of a footswitch. | |
Proprietary
Section 5: Page 2
{1}------------------------------------------------
510(k) Premarket Notification -- Original Submission RevLite™ Q-Switched Nd:YAG Laser System, New Indication for Use December 19, 2008
> The RevLite Systems are designed to provide laser energy for use in a variety of dermatological procedures (see indications for use). The 532 nm and 1064 nm wavelengths and optional 650 nm and 585 nm wavelengths are absorbed by pigment and other chromophores within the skin to create the desired clinical effect. The laser incorporates very narrow laser pulses (5-20 ns) designed to apply higher peak power over a very short period to minimize the time to absorb heat into the tissue. .
General: Incision, Excision, Ablation and Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis
Specific: For use for the following indications: Treatment of Epidermal Pigmented Lesions, Treatment of Dermal Pigmented Lesions. Nevus of Ota. Laser skin resurfacing procedures for the treatment of acne scars and wrinkles, Tattoo Removal: Dark Ink: (Black & Blue) Light Ink: (Red, Sky Blue, Green), Treatment of Vascular Lesions, Removal or lightening of unwanted hair with or without adjuvant preparation. Benign cutaneous lesions, such as, but not limited to: striae and scars. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength).
The RevLite™ Laser Systems are comparable to their predicate devices in terms of their indications for use, technical specifications, operating performance features, and general design features.
Nonclinical Performance Data:
Intended Use:
Comparison:
None
None.
Clinical Performance Data:
Additional Information:
None requested at this time.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an emblem of an eagle with its wings spread, clutching what appears to be a caduceus, a symbol often associated with medicine and healthcare.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
HOYA ConBio, Incorporated % Ms. Lisa Burns Liza Burns and Associates 19722 Westview Drive Twain Harte, California 95383
JUN - 1 2009
Re: K083899
Trade/Device Name: RevLite™Q-Switched Nd:YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery
And In Dermatology
Regulatory Class: II Product Code: GEX Dated: May 15, 2009 Received: May 18, 2009
Dear Ms. Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting
{3}------------------------------------------------
Page 2-Ms. Liza Burns
(reporting of medical device related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
**Device Name:**
# RevLite™ Q-Switched Nd: YAG Laser System
Intended Use: Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis.
183896
## Specific Indications :.
### 1064 nm wavelength
- Tattoo Removal (dark ink: blue and black)
- Nevus of Ota
- Removal or lightening of hair with or without adjuvant preparation.
- Skin Resurfacing for Acne Scars and Wrinkles
- Benign cutaneous lesions, such as, but not limited to: striae and scars (excludes the 650nm wavelength)
- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
532 nm wavelength (nominal delivered energy of 585 nm and 650 nm with the Optional Multilite Dye Laser Handpiece)
- . Tattoo removal (light ink: red, sky blue, green)
- Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, . spider angioma, cherry angioma, spider nevi
- Epidermal Pigmented lesions including but not limited to: cafe-au-lait birthmarks, . solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus
- Skin Resurfacing for Acne Scars and Wrinkles
- Benign cutaneous lesions, such as, but not limited to: striae and scars (excludes the 650nm wavelength)
- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
AND/OR
KA O gilling to m King (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K083899
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Proprietary
Section 4: Page 2
{5}------------------------------------------------
510(k) Premarket Notification – Original Submission RevLite™ Q-Switched Nd: YAG Laser System, New Indication for Use March 23, 2009
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
( 879年32 816 国産品番号 2007年度の2007年
Neils Dale Forman
(Division Sign-Off) (Division Sign of Surgical. Orthopedic, and Restorative Devices
510(k) Number K083899
Proprietary
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.