CYNOSURE YAG FAMILY LASER (PHOTOGENICA YAG AND ACCLAIM)

K031690 · Cynosure, Inc. · GEX · Jul 8, 2003 · General, Plastic Surgery

Device Facts

Record IDK031690
Device NameCYNOSURE YAG FAMILY LASER (PHOTOGENICA YAG AND ACCLAIM)
ApplicantCynosure, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateJul 8, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

1064 nm: The Cynosure YAG Family laser is intended for the coagulation and hemostasis of bonion vascular lesions such as, but not limited to, port winc stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, stiae and psoriasis. The laser is also intended for the treatment of bonign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café' au lait macules, sebortheic keratoses, nevi, chloasma, verrucea, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or black tattoos) and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB). 1320 nm The Cynosure YAG Family laser is indicated for use in general surgery and dermatology for the incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue. It is also indicated for the treatment of perioral wrinkles. It is also indicated for the treatment of fine lines and wrinkles.

Device Story

Cynosure YAG Family Laser (PhotoGenica YAG and Acclaim) is a Nd:YAG laser system; utilizes Nd:YAG crystal rod as lasing medium. Operates at 1,064 nm and 1,320 nm wavelengths. Activated via foot or finger switch. Used in clinical settings by healthcare professionals for dermatological and general surgical procedures. Laser energy targets melanin in hair follicles, vascular lesions, or soft tissue depending on wavelength. Output facilitates coagulation, hemostasis, ablation, or vaporization of tissue. Device provides clinical benefit through permanent hair reduction, lesion treatment, and wrinkle reduction.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Nd:YAG laser system; wavelengths 1,064 nm and 1,320 nm. Dimensions: 112x48x71 cm; weight: 81 kg. Power: 220 VAC, 30A, 50-60 Hz. Activation via foot or finger switch.

Indications for Use

Indicated for patients requiring dermatological or general surgical procedures including hair reduction, treatment of vascular/pigmented lesions, benign cutaneous lesions, scars, psoriasis, tattoos, and wrinkles. 1064 nm wavelength targets melanin and vascular structures; 1320 nm wavelength used for soft tissue incision, excision, ablation, and coagulation.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K031690 1 of 1 ## JUL 0 8 2003 ## 510(K) Summary | Submitter: | Cynosure, Inc.<br>10 Elizabeth Drive<br>Chelmsford, MA 01824 | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | George Cho<br>Senior Vice President of Medical Technology | | Date Summary Prepared: | May 30, 2003 | | Device Trade Name: | Cynosure YAG Family Laser (PhotoGenical YAG and Acclaim) | | Common Name: | Medical Laser System | | Classification Name: | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR 878.48 | | Equivalent Devices: | Altus CoolGlide Family, ICN CoolTouch / CoolTouch II, Sciton<br>Profile 1320 | | Device Description: | The Cynosure YAG Family Laser is a Nd: Y AG laser, having a<br>Nd: Y AG crystal rod as the lasing medium. It is a laser with a<br>wavelength of 1,064 nm and 1,320 nm. | | | Laser activation is by foot switch or finger switch. Overall weight of<br>the laser is 81 Kg, and the size is 112x48x71 cm (HxWxD). | | | Electrical requirement is 220 VAC, 30A, 50-60 Hz, single phase. | | Intended Use: | The Cynosure YAG Family Laser is indicated for permanent hair<br>reduction and the treatment of vascular and pigmented lesions and<br>wrinkles. | | Comparison: | The Cynosure YAG Family Laser has an equivalent indication for<br>uses, the same principle of operation, the same wavelengths and<br>essentially the same pulse energy range as the predicate devices. | | Nonclinical Performance Data: | none | | Clinical Performance Data: | none | | Conclusion: | The Cynosure YAG Family Laser is another safe and effective device<br>for permanent hair reduction and treatment of pigmented lesions and<br>wrinkles. | | Additional Information: | none | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that resembles a ribbon or wave. The image is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 0 8 2003 Mr. George Cho Senior Vice President of Medical Technology Cynosure, Inc. 10 Elizabeth Drive Chelmsford, Massachusetts 01824 Re: K031690 Trade/Device Name: Cynosure YAG Family Laser (PhotoGenica YAG, Acclaim) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: May 30, 2003 Received: June 3, 2003 Dear Mr. Cho: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Page 2 - Mr. George Cho This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page 1 of 1 510(k) Number (if known): _ KO31690 Device Name: _ Cynosure YAG Family Laser (PhotoGcnica YAG, Acclaim) Indications For Use: 1064 nm: The Cynosure YAG Family laser is intended for the coagulation and hemostasis of bonion vascular lesions such as, but not limited to, port winc stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, stiae and psoriasis. The laser is also intended for the treatment of bonign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café' au lait macules, sebortheic keratoses, nevi, chloasma, verrucea, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or black tattoos) and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB). 1320 nm The Cynosure YAG Family laser is indicated for use in general surgery and dermatology for the incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue. It is also indicated for the treatment of perioral wrinkles. It is also indicated for the treatment of fine lines and wrinkles. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Miriam C. Provost and Neurological Devices 510(k) Number Prescription Use OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...