UNIMED COATED LAPAROSCOPIC/ARTHROSCOPY/SUCTION-IRRIGATION ELECTRODE

K970066 · Unimed Surgical Products, Inc. · GEI · Feb 18, 1997 · General, Plastic Surgery

Device Facts

Record IDK970066
Device NameUNIMED COATED LAPAROSCOPIC/ARTHROSCOPY/SUCTION-IRRIGATION ELECTRODE
ApplicantUnimed Surgical Products, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 18, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

These electrosurgical devices are designed for cutting of soft tissue and coagulation.

Device Story

Electrosurgical electrodes (laparoscopic, arthroscopy, suction-irrigation) coated with Emralon 333 or Teflon 420-104; dry lubricants reduce friction and eschar build-up; minimizes need for high power settings; reduces scraping requirements; improves OR efficiency; sterile, single-use; used by surgeons for tissue cutting and coagulation.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Electrosurgical electrodes; stainless steel base; coated with Emralon 333 or Teflon 420-104; available in various diameters, lengths, and tip configurations; sterile, single-use.

Indications for Use

Indicated for cutting of soft tissue and coagulation during laparoscopic and arthroscopic surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K 970066 # SAFETY AND EFFECTIVENESS SUMMARY COMMON/USUAL NAME: Electrosurgical Electrodes FEB 18 1997 PROPRIETARY NAME: *unimed* Coated Laparoscopic Electrode *unimed* Coated Arthroscopy Electrode *unimed* Coated Suction-Irrigation Electrode CLASSIFICATION: CLASS II MATERIALS: All materials used to manufacture the Coated Electrodes are non-toxic and have been used in other previously marketed devices. DESCRIPTION: The *unimed* Coated Laparoscopic Electrode, *unimed* Coated Arthroscopy Electrode and the *unimed* Coated Suction-Irrigation Electrode are coated with Emralon 333 or Teflon 420-104. Both are dry lubricants which provide a low coefficient of friction and reduce eschar build-up during surgery and the need for higher power settings. Without eschar build-up that takes place with a standard stainless steel tip less scraping and inconvenience takes place resulting in greater OR efficiency. Each coating is non-toxic, corrosion resistant, and therefore safe for the patient and the OR staff. These electrosurgical devices are designed for cutting of soft tissue and coagulation. The electrodes are available in a variety of diameters, lengths and tip configurations. The electrodes are available sterile, single use only. SUBSTANTIAL EQUIVALENCE: This 510(k) notification represents the adding of a coating that is identical to the coatings on *unimed’s* Coated Electrode (K962935) to *unimed* Electrodes identified as *unimed* Laparoscopic Electrodes (K945030) and *unimed* Suction-Irrigation Electrodes (K963338). -28-
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...