FAS-CLEAN ELECTROSURGICAL COATED ACTIVE BLADE
K073616 · Electromedics Co. , Ltd. · GEI · Jul 23, 2008 · General, Plastic Surgery
Device Facts
| Record ID | K073616 |
| Device Name | FAS-CLEAN ELECTROSURGICAL COATED ACTIVE BLADE |
| Applicant | Electromedics Co. , Ltd. |
| Product Code | GEI · General, Plastic Surgery |
| Decision Date | Jul 23, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4400 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The FAS-CLEAN Electrosurgical Coated Active Blade is intended for use by surgeous as an alternative to uncoated stainless steel blades or "non-stick" coated electrodes where monopolar electrosurgical cutting and coagulation blades are normally used.
Device Story
FAS-CLEAN Electrosurgical Coated Active Blade is a sterile, single-use, monopolar electrosurgical electrode; consists of Emralon-coated stainless steel blade with insulated polyethylene sleeve. Used by surgeons in OR settings for cutting and coagulation during general, neurosurgical, laparoscopic, orthopedic, and gynecologic procedures. Device functions as an active electrode in a monopolar electrosurgical circuit; non-stick coating reduces tissue adhesion during use. Provides surgeons with an alternative to standard stainless steel or existing coated electrodes; intended to improve surgical efficiency by minimizing electrode cleaning requirements.
Clinical Evidence
No clinical data submitted. Bench testing only; device relies on substantial equivalence to legally marketed predicate devices.
Technological Characteristics
Materials: Emralon-coated stainless steel blade, insulated polyethylene sleeve. Energy: Monopolar electrosurgical energy. Form factor: Blade electrode. Sterilization: Radiation (ISO 11137). Biocompatibility: ISO 10993.
Indications for Use
Indicated for use in general, neurosurgical, laparoscopic, orthopedic, and gynecologic surgical procedures requiring monopolar electrosurgical cutting and coagulation. Intended as an alternative to standard uncoated or other non-stick coated stainless steel electrodes.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
- Unimed Coated Blade Electrode (K962935)
- Valley Lab's EDGE (K962044)
- ConMed's UltraClean (K052104)
- Megadyne's EX Clean (K943055)
Related Devices
- K962935 — UNIMED COATED NEEDLE ELECTRODE/COATED BLADE ELECTRODE/COATED BALL ELECTRODE · Unimed Surgical Products, Inc. · Sep 3, 1996
- K962044 — VALLEYLAB COATED ELECTRODES · Valleylab, Inc. · Aug 30, 1996
- K061878 — ELECTRODE COATED BALL, MODEL AB302C; NEEDLE, MODEL AN301C; BLADE, MODEL AE301HC · E Surgical, LLC · Oct 24, 2006
- K991855 — CONMED STEALTH NON-STICK ACCESSORY ELECTRODE, MODEL 138XXX · Conmedcorp · Jun 15, 1999
- K081791 — E-Z CLEAN ELECTROSURGICAL ELECTRODE · Megadyne Medical Products, Inc. · Oct 21, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 2
510(k) Summary
K073616
JUL 2 3 2008
General Information: Submitter:
ELECTROMEDICS Co. LTD. 55 Anderson Avenue Moonachie, NJ 07074
Contact Person:
John Scholz Tel : 201-935-0770 201-935-2812 Fax:
Summary Date of Preparation:
December 1, 2007
#### Names:
Proprietary Name: Common Name: Classification Name: FAS-CLEAN Electrosurgical Coated Active Blade Electrosurgical Electrode Electrosurgical cutting and coagulation device and accessories
## Predicate Devices:
Unimed Coated Blade Electrode (K962935) Valley Lab's EDGE (K962044) ConMed's UltraClean (K052104) Megadyne's EX Clean (K943055)
ELECTROMEDICS' FAS-CLEAN Electrosurgical Coated Active Blade is substantially equivalent to the Unimed Coated Blade Electrode, Valley Lab EDGE, ConMed's Ultra Clean and Megadyne's EZ Clean electrode. The FAS-CLEAN Electrosurgical Coated Active Blade shares the same intended uses, indications for use and the same or similar characteristics including: single use, non-stick properties, sterilization method, and various configurations.,
# Device Description
ELECTROMEDICS' FAS-CLEAN Electrosurgical Coated Active Blade is a sterile, single use, coated stainless steel blade. The main parts of the device are the Emralon coated stainless steel blade and the insulated polyethylene sleeve.
# Intended Use
The FAS-CLEAN Electrosurgical Coated Active Blade is intended for use by surgeous as an alternative to uncoated stainless steel blades or "non-stick" coated electrodes where monopolar electrosurgical cutting and coagulation blades are normally used.
{1}------------------------------------------------
KO 73616
## Technological Characteristics
Page 2 of (2)
The FAS-CLEAN Electrosurgical Coated Active Blade is substantially equivalent in materials and operation to the predicate devices on the market. This device has been designed to comply with the Sterilization of health care products - Requirements for validation and routine control- radiation Sterilization, ISO 11137 and Biocompatibility ISO 10993. There are no new technological characteristics and therefore no new questions of safety and effectiveness.
#### Performance Data:
No performance data is required for this Class II device nor requested by the Food and Drug Administration (Office of Device Evaluation). A database search has been conducted to evaluate any adverse effects of the device that is currently marketed.
No data submitted for section 807.92 6[(b)(1)(2)(3c)]. See attached documentation of adverse effects.
#### Conclusion:
The FAS-CLEAN Electrosurgical Coated Active Blade is substantially equivalent to Unimed's Coated Blade Electrode, Valley Lab's EDGE, ConMed's Ultra Clean and Megadyne's EZ Clean Electrodes. The FAS-CLEAN Electrosurgical Active Coated Blade shares the same intended use, indications for use and same or similar technological characteristics as the predicate electrosurgical electrodes.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, to the right. Encircling the caduceus are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 2 3 2008
Electromedics Co., Ltd. % Mr. John Scholz 55 Anderson Avenue Moonachie, New Jersey 07074
Re: K073616
Trade/Device Name: FAS-CLEAN Electrosurgical Coated Active Blade Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: April 10, 2008 Received: April 11, 2008
Dear Mr. Scholz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. John Scholz
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M Millican
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
## 510(k) K073616
# Device Name: FAS-CLEAN Electrosurgical Coated Active Blade
Indications for Use:
The FAS-CLEAN Electrosurgical Coated Active Blade, a single use, sterile, monopolar, Emralon coated electrode is intended for use:
- In surgical procedures (general, neurosurgical, laparoscopic, orthopedic, and ● gynecologic) where monopolar electrosurgical cutting and coagulation are normally used.
- As an alternative to uncoated stainless steel electrodes or "non-stick" coated electrodes, . which are used for these indications.
# Prescription Use X (Part 21 CFR 801 Subpart D)
And/Or
Over- The-Counter Use No (21 CFR 801 Subpart C)
R
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number