CERAMIC BLADE ELECTRODE
Device Facts
| Record ID | K960255 |
|---|---|
| Device Name | CERAMIC BLADE ELECTRODE |
| Applicant | Utah Medical Products, Inc. |
| Product Code | GEI · General, Plastic Surgery |
| Decision Date | Mar 27, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4400 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Utah Medical Products, Inc. (UMP) Ceramic Blade Electrodes (CBE) are intended for use in virtually every surgical discipline where flat, paddle-type electrosurgical blades are used for making straight cuts through tissue. The CBE-2XX blades are intended for use in procedures that require precise cuts where only light or no hemostasis is desired. The CBE-1XX blades are intended for use in procedures where a greater level of hemostasis is desired. When an electrosurgical generator blend mode is used for cutting with the CBE-1XX blades, cauterization of the wound will occur while the cut is being made.
Device Story
Ceramic blade electrode; paddle-type electrosurgical accessory. Input: electrical energy from standard electrosurgical generator via electrosurgical pen. Operation: blade conducts energy to tissue to perform cutting and cauterization. Output: physical tissue incision and wound cauterization. Used in surgical settings by clinicians. Benefits: precise cutting; controlled hemostasis; minimizes bleeding.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Ceramic blade electrode; paddle-type form factor. Compatible with standard electrosurgical pens and generators. Sterilization via gamma radiation. Materials selected for biocompatibility, electrical/thermal properties, and strength.
Indications for Use
Indicated for surgical procedures requiring straight tissue cuts and varying levels of hemostasis, ranging from light/no hemostasis (CBE-2XX) to greater hemostasis (CBE-1XX) via electrosurgical generator blend modes.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
- UMP DLP-L11 Loop Electrode (K905263)
- Mega Dyne E-Z Clean Blade (K913473)
- Aspen Labs Blade Electrode #60-0180-001 (K780977)