Oligio X

K240313 · Wontech Co., Ltd. · GEI · Feb 22, 2024 · General, Plastic Surgery

Device Facts

Record IDK240313
Device NameOligio X
ApplicantWontech Co., Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 22, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The 'Oligio X' is intended for use in dermatologic and general surgical procedures for non-invasive electrocoagulation and hemostasis of soft tissue.

Device Story

Oligio X is an electrosurgical device for non-invasive electrocoagulation and hemostasis of soft tissue in dermatologic and general surgical procedures. Operated by clinicians in a professional setting, the device delivers electrical energy to tissue. It functions similarly to the predicate Oligio (K221989), sharing the same electrification, principle, and method of operation. The device includes a main body, handpiece, and handpiece tip. Clinical benefit is achieved through controlled coagulation of soft tissue. Differences from the predicate are limited to the aesthetic color of the main body, handpiece, and tip, and updates to the graphical user interface (GUI).

Clinical Evidence

No clinical data.

Technological Characteristics

Electrosurgical cutting and coagulation device (Class II, GEI). Operates via electrical energy for soft tissue coagulation. Includes main body, handpiece, and handpiece tip. Features software-controlled GUI. Non-sterile, reusable device.

Indications for Use

Indicated for dermatologic and general surgical procedures requiring non-invasive electrocoagulation and hemostasis of soft tissue. Prescription use only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 22, 2024 Won Tech Co., Ltd. Yoon Hyunsik Team Leader 64 Techno8-ro, Yuseong-gu, Daejeon, 34028 Korea. South Re: K240313 Trade/Device Name: Oligio X Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: January 31, 2024 Received: February 2, 2024 Dear Yoon Hyunsik: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Mark Digitally signed by Mar Trumbore -S - Date: 2024.02.22 09:20:28 > -05'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) Device Name Oligio X Indications for Use (Describe) The 'Oligio X' is intended for use in dermatologic and general surgical procedures for non-invasive electrocoagulation and hemostasis of soft tissue. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circular graphic with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is in orange, matching the color of the graphic, while the "TECH" part is in gray. 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (31) 1670 1450 FAX: +82 (70) 7836 0110 # Executive Summary #### Indications for use 'Oligio X' is intended for use in dermatologic and general surgical procedures for non-invasive electrocoagulation and hemostasis of soft tissue. #### Design and Use of the Device | Question | Yes | No | |------------------------------------------------------------------------------------|-----|----| | Is the device intended for prescription use (21 CFR 801 Subpart D)? | √ | | | Is the device intended for over-the-counter use (21 CFR 807 Subpart C)? | | √ | | Does the device contain components derived from a tissue or other biologic source? | | √ | | Is the device provided sterile? | | √ | | Is the device intended for single use? | | √ | | Is the device a reprocessed single use device? | | √ | | If yes, does this device type require reprocessed validation data? | | √ | | Does the device contain a drug? | | √ | | Does the device contain a biologic? | | √ | | Does the device use software? | √ | | | Does the submission include clinical information? | | √ | | Is the device implanted? | | √ | ### Predicate Device Comparison The identified predicate device within this submission is shown as follow: The Oligio - Manufactured by WON TECH Co., Ltd. (K221989) The proposed device and the predicate device has same electrication, principle, method, indication for use, and same components. Differences limited to two features; 1. The color of main body, handpiece, and handpiece tip. 2. GUI of the device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...