K221989 · Wontech Co., Ltd. · GEI · Oct 13, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221989
Device Name
The Oligio
Applicant
Wontech Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oligio is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis of soft tissue.
Device Story
The Oligio is a radio frequency (RF) therapy system used in dermatologic and general surgical procedures. It delivers 6.78 MHz RF energy to soft tissue for selective coagulation and hemostasis while conductively cooling the epidermis. The system comprises a main unit with a touch LCD monitor, an RF handpiece, return pad cable, non-sterile treatment tips, cooling gas, and coupling fluids. The device employs RF tuning to deliver energy across varying tissue impedances. Operated by a clinician, the system is controlled manually or via footswitch. The output provides therapeutic coagulation; the cooling mechanism protects the skin surface. The device is intended for clinical settings.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
RF therapy system operating at 6.78 MHz ± 1% with 145 W max output. Components include touch LCD, RF handpiece, return pad, and disposable tips. RF tip material is polyethylene terephthalate (ISO 10993-1 compliant for limited contact). Features conductive epidermal cooling, manual/footswitch operation, and adjustable handpiece vibration (30-50 dB). Complies with IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, and IEC 61000-3-2.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis of soft tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.
10/13/2022
WON TECH Co., Ltd. Hyun Yoon General Manager 64 Techno 8-ro Yuseong-gu, Daejeon 34028 Korea, South
### Re: K221989
Trade/Device Name: The Oligio Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: October 13, 2022 Received: October 13, 2022
## Dear Hyun Yoon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221989
Device Name The Oligio
### Indications for Use (Describe)
'The Oligio' is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis of soft tissue.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is orange, while the "TECH" part is gray.
# 510(k) Summary
[As required by 21 CFR 807.92]
K221989
#### 1. Date Prepared [21 CFR 807.92(a)(a)]
September 13, 2022
#### 2. Submitter's Information & Contact Person [21 CFR 807.92(a)(1)]
- Name of Manufacturer: WON TECH Co., Ltd.
- Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028, Republic of Korea - Contact Name: Hyun Sik Yoon - Telephone No.: +82-10-6750-5346
- Fax No.: +82-70-7836-0110
- Email Address: yoonhs21@wtlaser.com
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Common name: Radio Frequency Therapy System Trade name: The Oligio
| Classification Description | 21 CFR Section | Product Code |
|-------------------------------------------------------------------|----------------|--------------|
| Electrosurgical Cutting and Coagulation Device and<br>Accessories | 878.4400 | GEI |
As stated in 21 CFR, parts 878.4400, this generic types of devices has been classified as Class II.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is orange, while the "TECH" part is gray. The logo appears to be for a technology company.
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow:
Predicate device
- 510(k) Number: K170758
- Applicant: Solta Medical Inc. ●
- Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories ●
- Trade Name: ● Thermage FLX System
#### ട്. Description of the Device [21 CFR 807.92(a)(4)]
'The Oligio' is a radio frequency therapy system. The radio frequency output of the device is 6.78 MHz, and the maximum power is 145 W. "The Oligio" consists of a main body including touch LCD monitor, a RF handpiece, return pad cable, non-sterile treatment tips, cooling gas, coupling fluids and a power cable.
'The Oligio' RF System delivers radio frequency energy for selective coagulation of tissue while conductively cooling the epidermis. "The Oligio' delivers energy form the disposable tip to the patient, and employs radio frequency turning to provide radio frequency energy across a range of impedances for delivery to the patient through the tip.
#### Indications for Use [21 CFR 807.92(a)(5)] 6.
'The Oligio' is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis of soft tissue.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the word "WONTECH" in orange and gray. The "WON" part of the word is in orange, while the "TECH" part is in gray.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
# 7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]
There are no significant differences between The Oligio and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.
| | Proposed Device | Predicate Device | SE Decision |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number | K221989 | K170758 | - |
| Manufacturer | WON TECH Co., Ltd. | Solta Medical Inc. | - |
| Model | The Oligio | Thermage FLX System | - |
| Intended Use | Dermatologic and general surgical<br>procedures for electrocoagulation<br>and hemostasis of soft tissue. | Dermatologic and general surgical<br>procedures for electrocoagulation<br>and hemostasis. | Same |
| Principle/Method of<br>Operation | 'The Oligio' RF System delivers<br>radio frequency energy for selective<br>coagulation of tissue while<br>conductively cooling the epidermis.<br>'The Oligio' delivers energy form<br>the disposable tip to the patient, and<br>employs radio frequency turning to<br>provide radio frequency energy<br>across a range of impedances for<br>delivery to the patient through the<br>tip. | The Thermage FLX System<br>delivers RF energy for selective<br>coagulation of tissue while<br>conductively cooling the epidermis.<br>The Thermage FLX System<br>delivers energy from the disposable<br>tip to the patient. The System and its<br>Handpiece monitor skin contact<br>during treatment. The System<br>employs RF tuning to provide RF<br>energy across a range of<br>impedances for delivery to the<br>patient through single and multiple<br>pass stamping motions of the tip. | Same |
| Output frequency | 6.78MHz ± 1% | 6.78MHz | Same |
| Max output power | 145 W | 400 W | Different |
| Different - Max output power<br>The Max output power for proposed device is different from the predicate devices. However, the Max output power<br>of the proposed The Oligio is covered by the range of the Max output power of the predicate device. Therefore, this<br>difference does not affect substantially equivalence on safety and effectiveness. | | | |
| Mode of Operation | Manual or footswitch | Manual or footswitch | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH" in orange and gray. The "WON" part of the text is in orange, while the "TECH" part is in gray.
### Non-Clinical Test Summary [21 CFR 807.92(b)(1)]
### 1) Electrical Safety, Electromagnetic Compatibility Testing
Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
| Standard<br>(Edition) | Standard Title |
|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005,<br>AMD1:2012 | Amendment 1 - Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance |
| IEC 60601-2-2:2017 for use in<br>conjunction with IEC 60601-<br>1:2005, IEC 60601-<br>1:2005/AMD1:2012 | Medical electrical equipment - Part 2-2: Particular requirements for the<br>basic safety and essential performance of high frequency surgical<br>equipment and high frequency surgical accessories |
| IEC 60601-1-2:2014 | Medical electrical equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
| IEC 61000-3-2:2018 | Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for<br>harmonic current emissions (equipment input current ≤16 A per phase) |
### 2) Software Validation
The Oligio contains Basic Documentation Level software. Software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: Content of premarket submissions for Device Software Functions, on November 04, 2021.
### 3) Biocompatibility
| Part | Material | Patient<br>Contact | Duration of Contact<br>by ISO 10993-1 | Bio-<br>compatibility |
|--------|-------------------------------|--------------------|---------------------------------------|-----------------------|
| RF tip | Polyethylene<br>terephthalate | Intact Skin | Limited<br>(< 24 hours) | Yes |
### 4) Performance Testing
The performance of The Oligio has been defined as follows.
- RF output frequency: 6.78MHz ± 1% -
- -Max output power: 145 W
- Handpiece vibration: Low (30dB), Medium (40dB), High (50dB) -
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside, followed by the text "WON" in orange and "TECH" in gray. The "W" in the circle is made up of two curved lines that intersect in the middle.
### Clinical Test Summary [21 CFR 807.92(b)(2)]
No clinical studies were considered necessary and performed.
### Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that The Oligio is substantially equivalent to predicate devices as described herein.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.