K170777 · Bovie Medical Corporation · GEI · May 1, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170777
Device Name
Bovie J-Plasma Precise FLEX Handpiece
Applicant
Bovie Medical Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
May 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bovie J-Plasma Precise Flex Handpiece is used for the delivery of radiofrequency energy and/or a helium gas plasma for electrosurgical cutting, coagulation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures. The Bovie J-Plasma Precise Flex Handpiece is compatible with Bovie Ultimate Electrosurgical Generators BVX-200H, and BVX-200P.
Device Story
Electrosurgical accessory; connects to Bovie Ultimate Electrosurgical Generators (BVX-200H, BVX-200P). Delivers helium gas plasma (J-Plasma) and monopolar coagulation energy to soft tissue. Features flexible shaft for advancement through cannula; manipulated by graspers/forceps during minimally invasive surgery. Activated via generator footswitch. Enables surgeon to perform cutting, coagulation, and ablation. Benefits include precise tissue effect in minimally invasive settings.
Clinical Evidence
Bench testing only. Included electrical verification, plasma stream characterization, mechanical functionality, and electromagnetic compatibility (EMC) testing. Simulated use testing demonstrated device maneuverability with grasping accessories.
Technological Characteristics
Flexible shaft electrosurgical handpiece; ceramic nozzle tip; blade/needle tip configurations. Energy: helium gas plasma and monopolar coagulation. Compatible with Bovie Ultimate Generators. Footswitch activation. 500mm working length; 8mm cannula diameter. Complies with IEC electrical safety standards.
Indications for Use
Indicated for cutting, coagulation, and ablation of soft tissue during minimally invasive surgical procedures using a grasper.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
OmniGuide Laser System with FlexGuide Ultra (K140378)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is often associated with medicine and healthcare. The caduceus is composed of three abstract human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2017
Bovie Medical Corporation Ms. Rubiela Maldonado Regulatory Affairs Manager 5115 Ulmerton Road Clearwater, Florida 33760
Re: K170777
Trade/Device Name: Bovie J-Plasma Precise Flex Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 14, 2017 Received: March 15, 2017
Dear Ms. Maldonado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170777
Device Name Bovie J-Plasma Precise Flex Handpiece
#### Indications for Use (Describe)
The Bovie J-Plasma Precise Flex Handpiece is used for the delivery of radiofrequency energy and/or a helium gas plasma for electrosurgical cutting, coagulation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures.
The Bovie J-Plasma Precise Flex Handpiece is compatible with Bovie Ultimate Electrosurgical Generators BVX-200H, and BVX-200P.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CER 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CER 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Bovie" in a bold, blue font. Above the word is a curved, light blue line with a blue dot at the end. The "®" symbol is located at the bottom right of the word.
# 510(k) SUMMARY
### (Submitted As Required per 21 CFR 807.92)
#### GENERAL INFORMATION:
| Submitter Name: | Bovie Medical Corporation |
|------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3007593903 |
| Submitter Address: | 5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America |
| Submitter Telephone Number: | (727) 803-8508 |
| Submitter FAX Number: | (727) 322-4465 |
| Contact Person: | Rubiela Maldonado<br>Regulatory Affairs Manager |
| Date Prepared: | March 14, 2017 |
| DEVICE IDENTIFICATION: | |
| Proprietary Name: | Bovie® J-Plasma® Precise Flex Handpiece |
| Common Name: | Electrosurgical Generator Accessory |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Model Numbers: | Multiple |
| Classification: | 21CFR 878.4400; Class II; Product Code GEI |
| Legally Marketed<br>Predicate Device(s): | 510(k) Numbers: K152325, K152570, K170188<br>Predicate Device Name: Bovie J-Plasma Precise<br>Handpieces |
{5}------------------------------------------------
B
510(k) K170777
# 510(k) SUMMARY
510(k) Number: K140378
Predicate Device Name: OmniGuide Laser System with FlexGuide™ Ultra
Manufacturer: OmniGuide Incorporated
#### INTENDED USE/INDICATIONS
The Bovie J-Plasma Precise Flex handpiece is used for the delivery of radiofrequency energy and/or helium gas plasma for electrosurgical cutting, coagulation, and ablation of soft tissue. It is intended for use with grasping instruments during minimally invasive surgical procedures.
The Bovie J-Plasma Precise Flex handpiece is compatible with Bovie Ultimate Electrosurgical Generators BVX-200H, and BVX-200P.
#### DEVICE DESCRIPTION
The Bovie J-Plasma Precise Flex handpiece, is an electrosurgical accessory that is connected to a Bovie Ultimate Electrosurgical Generator and used to deliver helium gas plasma (J-Plasma) and standard electrosurgical monopolar coagulation during surgical procedures. The accessory has a flexible shaft that can be advanced through a cannula during minimally invasive surgical procedures. The handpiece is activated by the generator footswitch.
The primary difference between the Bovie J-Plasma Precise Flex handpiece and currently available Bovie J-Plasma handpieces is that the Bovie J-Plasma Precise Flex handpiece has a flexible shaft that can be controlled with graspers, forceps, or grasping instruments during minimally invasive surgical procedures.
The Bovie J-Plasma Precise Flex handpiece uses the same basic technology and delivers the same J-Plasma helium plasma energy as currently available J-Plasma handpieces (K151325, K152570, K170188) and are compatible only with Bovie Ultimate electrosurgical generators.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Bovie" in a bold, sans-serif font. The word is a dark blue color. Above the word is a light blue curved line with a dot at the end. The registered trademark symbol is located to the bottom right of the word.
# 510(k) SUMMARY
#### PERFORMANCE TESTING
Performance testing to assure that the Bovie J-Plasma Precise Flex Handpiece meets performance requirements was performed and is summarized in the following tables:
| Test | Objective | Protocols |
|----------------------------------|----------------------------------------------------------------------------------------------|-----------|
| Electrical Verification | Verify the electrical functionality and safety of the<br>handpiece | VR-1510 |
| Bench Equivalency | Measure effect of generator settings on the plasma<br>stream, plasma stream characterization | VR-1511 |
| Mechanical<br>Performance | Verify the mechanical functionality of the handpiece | VP 1529 |
| Electromagnetic<br>Compatibility | EMC testing for emissions and immunity | VR-1512 |
### SUBSTANTIAL EQUIVALENCE
The Bovie J-Plasma Precise Flex handpiece is intended for cutting, coagulation, and ablation of soft tissue. The Bovie J-Plasma Precise FlexHandpiece has the same intended use and energy source as the predicate device, J-Plasma Handpiece. There are no changes to the principle of operation or the method of application. There is no new technology or differences that would raise new or different questions of safety or efficacy and no changes to the principle of operation.
The Bovie J-Plasma Precise Flex handpiece it is intended for use with a grasping accessory during minimally invasive surgical procedures as the other predicate device, OmniGuide Laser System with FleGuide™ Ultra.
{7}------------------------------------------------
# 510(k) SUMMARY
| Feature/<br>Characteristic | Bovie J-Plasma Precise® Flex | Predicate Bovie® J-Plasma<br>Handpiece<br>(K170188, K151325,<br>K152570) | OmniGuide Laser System<br>with FlexGuide Ultra<br>K140378 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | The Bovie J-Plasma Precise®<br>Flexible Handpiece is used for<br>the delivery of radiofrequency<br>energy and/or a helium gas<br>plasma for electrosurgical<br>cutting, coagulation, and<br>ablation of soft tissue. It is<br>intended for use with a<br>grasping instruments during<br>minimally invasive surgical<br>procedures.<br>The Bovie J-Plasma Precise<br>Open Handpiece is compatible<br>with Bovie Ultimate Generators<br>BVX-200H and BVX-200P. | K170188- intended to be<br>used in conjunction with<br>the Bovie Ultimate®<br>electrosurgical generator<br>for the delivery of<br>radiofrequency energy<br>and/or a helium gas<br>plasma for electrosurgical<br>cutting,<br>coagulation and ablation of<br>soft tissue during open<br>surgical procedures.<br>The Bovie J-Plasma Precise<br>Open® Handpiece is<br>compatible with Bovie<br>Ultimate® Electrosurgical<br>Generators BVX-200H, and<br>BVX-200P. | The OmniGuide Laser<br>System with FlexGuide™<br>Ultra, which includes the<br>FELS 25A Laser, BeamPath<br>Fibers and the<br>FlexGuide™ Ultra<br>handpiece, is indicated for<br>use to enable the surgeon<br>to perform incision,<br>excision, ablation,<br>vaporization and<br>coagulation of body soft<br>tissues including intra-oral<br>tissues. This system is<br>intended for use with a<br>grasper in the following<br>medical specialties: general<br>laparoscopic surgical<br>procedures, gynecologic<br>laparoscopic surgical<br>procedures, urologic<br>surgical procedures,<br>otorhinolaryngology |
| General Description<br>of Procedures | Laparoscopic or minimally<br>invasive surgical procedures | Open and Laparascopic,<br>minimally invasive<br>procedure through cannula | surgical procedures.<br>Medical specialties: general<br>laparoscopic surgical<br>procedures, gynecologic<br>laparoscopic surgical<br>procedures, urologic<br>surgical procedures,<br>otorhinolaryngology<br>surgical procedures. |
| Energy Type | Helium gas plasma<br>Monopolar coagulation | Helium gas plasma<br>Monopolar coagulation | Laser |
| Device Activation | Footswitch activation only, no<br>hand activation | Hand Activation or<br>Footswitch Activation | Footswitch activation only,<br>no hand activation |
| Feature/<br>Characteristic | Bovie J-Plasma Precise® Flex | Predicate Bovie® J-Plasma<br>Handpiece<br>(K170188, K151325,<br>K152570) | OmniGuide Laser System<br>with FlexGuide Ultra<br>K140378 |
| Grasping Feature | Yes | No | Yes |
| Shaft | Flexible | Straight | Flexible |
| Shaft Working Length | 500mm | 18mm, 270mm, 150mm,<br>330mm, 450mm | 600mm |
| Compatible Cannula<br>diameter | 8mm | 5mm | 12mm |
| Tip Configuration | Blade, Needle | Same | Spatula |
| Blade Extension | Maximum 10mm in 2mm<br>increments | Same | N/A |
| Blade Width x<br>Thickness | 0.4mm x 0.08mm | Same | 4 mm |
| Tip protector | Ceramic nozzle | Same | Cycloolefin Polymer (COP) |
| Plasma Settings | Maximum 40 watts, 1-5 lpm<br>gas flow | Same | N/A |
| Compatibility | Compatible only with Bovie<br>Ultimate Generator (BVX-300H,<br>BVX-200P) | Compatible with Original<br>Ultimate (BVX-100H) and<br>Ultimate Generators (BVX-<br>200H, BVX-200P) | OmniGuide Laser System |
| Connector | Bovie Proprietary | Same | OmniGuide |
| Electrical Safety | Complies with IEC Electrical<br>Safety Standards | Same | Same |
| Transformer | None | None — K170188 | N/A |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Bovie" in a stylized font. The word is in a dark blue color. Above the word is a light blue curved line with a dot at the end.
## 510(k) SUMMARY
### CONCLUSION
The Bovie J-Plasma Precise Flex Handpiece is intended for cutting, coagulation, and ablation of soft tissue. The Bovie J-Plasma Precise Flex Handpiece has the same intended use and energy source as the predicate device, J-Plasma Handpiece. There are no changes to the principle of operation and there is no new technology or technology differences that would raise new or different questions of safety or efficacy.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Bovie" in a dark blue font. Above the word is a light blue curved line with a dot at the end. The word "Bovie" is the logo for a company.
## 510(k) SUMMARY
The primary differences between the subject device J-Plasma handpiece and the predicate J-Plasma handpieces is the flexible shaft that can be manipulated via grasping accessories. The grasping feature is the same as the predicate device OmniGuide Laser System with FlexGuide Ultra.
The Bovie J-Plasma Precise® Flex Handpiece was subjected to verification testing that confirmed device performance and demonstrated equivalent tissue effect as the predicate device J-Plasma handpiece. Simulated use testing demonstrated the ability of the J-Plasma Precise Flex device to be grasped and maneuvered.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.