K140378 · Omniguide, Inc. · GEX · Sep 25, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140378
Device Name
FLEXGUIDE ULTRA BEAMPATH ROBOTIC FIBER CONDUIT
Applicant
Omniguide, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OmniGuide Laser System with FlexGuide™ Ultra, which includes the FELS 25A Laser, BeamPath Fibers and the FlexGuide™ Ultra handpiece, is indicated for use to enable the surgeon to perform incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. This system is intended for use with a grasper in the following medical specialties: general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, urologic surgical procedures, otorhinolaryngology surgical procedures.
Device Story
System comprises FELS 25A CO2 laser, BeamPath fibers, and FlexGuide Ultra handpiece. Laser energy transmits through fiber; handpiece guides/protects fiber. Surgeon grasps distal tip of handpiece to manipulate tissue; operates in contact or non-contact mode. Used in laparoscopic/surgical settings by surgeons to stabilize hand motion and provide precision control during soft tissue procedures. Benefits include controlled tissue incision/ablation. Device is non-sterile, reusable, stainless steel.
Clinical Evidence
Bench testing only. Engineering analysis and performance testing validated the FlexGuide Ultra handpiece, confirming it can be grasped and manipulated without affecting performance compared to the predicate handpiece.
Technological Characteristics
Materials: 304 stainless steel. Energy source: CO2 laser (FELS 25A). Form factor: Flexible handpiece (50cm length) with proximal sealing cap, body/cannula, and distal tip. Connectivity: None. Sterilization: Reusable/sterilizable. No software/algorithm.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues (including intra-oral) in general, gynecologic, and urologic laparoscopic procedures, and otorhinolaryngology procedures. Used with a grasper.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
OmniGuide BeamPath Fiber Optic Handpiece System & Sterilization Tray (K081939)
OmniGuide BeamPath FELS 25A CO2 Laser System (K093251)
OmniGuide BeamPath CO2 Mark III WaveGuide Fiber (K070157)
OmniGuide BeamPath CO2 Mark III WaveGuide Fiber w/Low Loss Tip (K093451)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be flowing or merging together.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2014
OmniGuide Incorporated Ms. Nicole Rasmussen Senior Regulatory Specialist One Kendall Square, Suite B1301 Cambridge, Massachusetts 02139
Re: K140378
Trade Name: OmniGuide Laser System with FlexGuide™ Ultra Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 29, 2014 Received: September 2, 2014
Dear Mr. Rasmussen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K140378
Device Name OmniGuide Laser System with FlexGuide™ Ultra
### Indications for Use (Describe)
The OmniGuide Laser System with FlexGuide™ Ultra, which includes the FELS 25A Laser, BeamPath Fibers and the FlexGuide™ Ultra handpiece, is indicated for use to enable the surgeon to perform incision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. This system is intended for use with a grasper in the following medical special laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, urologic surgical procedures, otorhinolaryngology surgical procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the OmniGuide Surgical logo. The logo consists of the words "OmniGuide" in black text, with the word "SURGICAL" in smaller blue text underneath. To the left of the text is a blue circular graphic. There are two light blue and gray curved lines in the background.
## 510(k) Summary
This 510(k) summary is being submitted in accordance with 21 CFR 807.92
#### 1. Submitter's Information
| Name: | OmniGuide, Inc. |
|----------|-----------------------------------------------------------------------------------------------------------------------------|
| Address: | OmniGuide, Inc.<br>One Kendall Square<br>Suite B1301<br>Cambridge, MA 02139<br>Phone: (617) 551-8426<br>Fax: (617) 551-8445 |
| Contact: | Nicole Boeyink |
Contact: Nicole Rasmussen Sr. Regulatory Affairs Specialist
Date Prepared: September 24, 2014
### 2. Device Information
| | Trade/Proprietary Name: OmniGuide Laser System with FlexGuide™ Ultra |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Fiber Optic Handpiece<br>CO2 Laser Powered Surgical Instrument |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology (21 CFR § 878.4810) |
| Product Code: | GEX |
#### 3. Predicate Devices
## Handpiece: See Tables 1 & 2
OmniGuide, Inc. OmniGuide BeamPath Fiber Optic Handpiece System & Sterilization Tray K081939
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OmniGuide Surgical. The logo consists of two concentric blue circles on the left, with the text "OmniGuide" in black to the right of the circles. Below "OmniGuide" is the word "SURGICAL" in blue, written in a smaller font. There is a registered trademark symbol to the right of the word "Guide".
## Laser: No Changes
OmniGuide, Inc. OmniGuide BeamPath FELS 25A CO2 Laser System K093251
### Fibers: No Changes, see Table 1.
OmniGuide, Inc. OmniGuide BeamPath CO2 Mark III WaveGuide Fiber K070157
OmniGuide, Inc. OmniGuide BeamPath CO2 Mark III WaveGuide Fiber w/Low Loss Tip K093451
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo for OmniGuide Surgical. The logo consists of a stylized "O" formed by two concentric blue circles, with the text "OmniGuide" in black to the right of the circle. Below "OmniGuide" is the word "SURGICAL" in smaller, uppercase letters. The background includes two curved, gradient bands in shades of blue and gray, adding a modern and clean aesthetic to the overall design.
#### Intended Use 4.
The OmniGuide Laser System with FlexGuide™ Ultra, which includes the FELS 25A Laser, BeamPath Fibers and the FlexGuide™ Ultra handpiece, is indicated for use to enable the surgeon to perform incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. This system is intended for use with a grasper in the following medical specialties: general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, urologic surgical procedures, otorhinolaryngology surgical procedures.
### Device Description న.
The OmniGuide Laser System with FlexGuide™ Ultra, which includes the FELS 25A Laser, BeamPath Fibers and the FlexGuide™ Ultra handpiece are used together to perform the intended use above. The FELS 25A Laser and BeamPath Fibers remain unchanged. Additionally, the BeamPath fibers that deliver the laser energy, are cleared for laparoscopic use. The BeamPath Fiber Optic Handpiece (K081939) has been modified to the FlexGuide™ Ultra handpiece, which includes a sterilization tray and are sold as non-sterile reusable devices. The FlexGuide Ultra is a flexible handpiece that consists of three main sections: the proximal rear sealing cap, handpiece body/cannula and distal tip. The material of the FlexGuide Ultra handpiece is stainless steel. The OmniGuide BeamPath fiber (K070157, K093451) is inserted through the flexible handpiece proximal rear sealing cap and is fixated so that it is visualized through two site holes at the handpiece distal tip. The OmniGuide FELS 25A Laser System (K093251) still generates the CO2 laser energy, which is transmitted through the fiber. During the procedure the surgeon grasps and holds the distal tip of the FlexGuide Ultra with a grasper. The FlexGuide Ultra distal tip may be used in either non-contact mode or contact mode. The surgeon may also use the distal tip to enable tissue manipulation.
### Performance Testing 6.
Performance testing has validated the FlexGuide Ultra handpiece. Engineering analysis and bench testing concluded that the FlexGuide Ultra can be grasped, manipulated and confirmed the modified handpiece does not affect its purpose and performance when compared to the OmniGuide predicate handpiece. Additionally, the review of the technical characteristics, indications for use, risk analysis,
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image features the logo for OmniGuide Surgical. The logo consists of a circular design with the word "OmniGuide" in bold, black letters, with the word "SURGICAL" in smaller, blue letters underneath. The logo is positioned on the left side of the image, with two curved, light blue and gray lines extending horizontally to the right.
verification and validation information provided in this 510(k) Premarket Notification demonstrates the OmniGuide FlexGuide Ultra handpiece, when used in conjunction with OmniGuide BeamPath fibers, are as safe and effective as its predicate device.
#### 7. Substantial Equivalence
The FlexGuide Ultra Handpiece is substantially equivalent to its predicate device when used according to its intended use. This statement is based on the information provided in this 510(k) Premarket Notification which demonstrates the FlexGuide Ultra handpiece shares the similar intended uses, operating principles, similar technical characteristics and incorporates the same basic handpiece design when compared to the original cleared handpiece. The handpiece also shares a similar intended use as the predicate fibers, which are used in conjunction with the FlexGuide Ultra to transmit CO2 laser energy to the treatment site.
| | OmniGuide Laser<br>System with<br>FlexGuide™ Ultra<br>(K140378 pending) | OmniGuide<br>BeamPath Fiber<br>Optic Handpiece<br>System<br>(K081939) | OmniGuide Beam<br>Path CO 2 Mark III<br>WaveGuide Fiber<br>(K070157) | OmniGuide<br>BeamPath CO2<br>Mark III<br>WaveGuide Fiber<br>w/Low Loss Tip<br>(K093451) |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The OmniGuide<br>Laser System with<br>FlexGuide™ Ultra,<br>which includes the<br>FELS 25A Laser,<br>BeamPath Fibers<br>and the FlexGuide™<br>Ultra handpiece, is<br>indicated for use to<br>enable the surgeon<br>to perform incision,<br>excision, ablation,<br>vaporization and<br>coagulation of body<br>soft tissues<br>including intra-oral<br>tissues. This system<br>is intended for use<br>with a grasper in the<br>following medical<br>specialties: general<br>laparoscopic | The OmniGuide<br>BeamPath® Fiber<br>Optic Handpiece<br>System and<br>Sterilization Tray is<br>indicated for use in<br>conjunction with<br>OmniGuide's<br>waveguide fibers to<br>enable the surgeon to<br>perform incision,<br>excision, ablation,<br>vaporization and<br>coagulation of body<br>soft tissues including<br>intra-oral tissues. It is<br>indicated in the<br>medical specialties of<br>general and plastic<br>surgery,<br>oral/maxillofacial<br>surgery, dentistry, | The OmniGuide<br>Mark Ill WaveGuide<br>Fiber is indicated for<br>the incision, excision,<br>ablation, vaporization<br>and coagulation of<br>body soft tissues<br>including intra-oral<br>tissues. It<br>is indicated in the<br>medical specialties of<br>general and plastic<br>surgery,<br>oral/maxillofacial<br>surgery, dentistry,<br>dermatology,<br>gynecology,<br>otorhinolaryngology,<br>gastroenterology,<br>neurosurgery,<br>urology, and<br>pulmonology. and | The OmniGuide<br>Beam Path CO 2<br>Mark II WaveGuide<br>Fiber with Low<br>Profile/Low Loss<br>Tip is indicated for<br>the incision, excision,<br>ablation, vaporization<br>and coagulation of<br>body<br>soft tissues including<br>intra-oral tissues. It is<br>indidated in the<br>medical specialties of<br>general<br>and plastic surgery,<br>oral/maxillofacial<br>surgery, dentistry,<br>dermatology,<br>gynecology,<br>otorhinolaryngology,<br>gastroenterology. |
| | Table 1. Predicate Device Indications for Use Comparison | | |
|--|----------------------------------------------------------|--|--|
| | | | |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for OmniGuide Surgical. The logo consists of three concentric blue circles on the left side of the image. To the right of the circles is the word "OmniGuide" in black font, with the word "SURGICAL" in blue font underneath. There is a registered trademark symbol to the right of the word "OmniGuide".
| surgical procedures,<br>gynecologic<br>laparoscopic<br>surgical procedures,<br>urologic surgical<br>procedures,<br>otorhinolaryngology<br>surgical procedures. | dermatology,<br>gynecology,<br>otorhinolaryngology,<br>gastroenterology,<br>neurosurgery,<br>urology, and<br>pulmonology. | can be used in open<br>surgical procedures<br>as well as endoscopic<br>minimally invasive<br>procedures in<br>conjunction with<br>rigid or flexible<br>endoscopes, such as<br>in laryngoscopy,<br>gastroscopy,<br>colonoscopy,<br>laparoscopy,<br>thoracoscopy,<br>hysteroscopy and<br>bronchoscopy.<br>The indications for<br>use for which the<br>delivery system is<br>used for are<br>dependent upon the<br>cleared indications<br>for use of the laser<br>system and those<br>laser system<br>accessories to which<br>it is attached. | neurosurgery,<br>urology, and<br>pulmonology, and<br>can be used in open<br>surgical procedures<br>as well as endoscopic<br>minimally invasive<br>procedures<br>in conjunction with<br>rigid or flexible<br>endoscopes, such as<br>in laryngoscopy,<br>gastroscopy,<br>colonoscopy,<br>laparoscopy,<br>thoracoscopy,<br>hysteroscopy and<br>bronchoscopy.<br>The indications for<br>use for which the<br>delivery system is<br>used for are<br>dependent upon the<br>cleared indications<br>for use of the laser<br>system and those<br>laser system<br>accessories to which<br>it is attached. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## Table 2. Predicate Handpiece Specification Comparison
| | OmniGuide Laser System with<br>FlexGuide™ Ultra<br>(K140378 pending) | OmniGuide BeamPath Fiber Optic<br>Handpiece System<br>(K081939) |
|--------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------|
| Handpiece<br>Lumen<br>Material | 304 Stainless Steel | 304 Stainless Steel Hypodermic Tubing |
| Lumen<br>Length | 50cm | Up to 60cm |
| Handpiece<br>Design | Distal tip, handpiece body/cannula, rear<br>sealing cap | Distal tip, handpiece body/cannula, rear<br>sealing cap |
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for OmniGuide Surgical. The logo consists of a blue circle with the words "OmniGuide" in black letters inside the circle. Below the circle, the word "SURGICAL" is written in blue letters. The logo is simple and modern, and it is likely used to represent a company that specializes in surgical products or services.
| Distal Tip | Used in either non-contact mode or contact<br>mode, and may be used to enable tissue<br>manipulation. | Used in either non-contact mode or contact<br>mode, and may be used to enable tissue<br>manipulation. |
|------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Mechanism of<br>Action | Handpiece is used to guide and protect<br>BeamPath® fibers | Handpiece is used to guide and protect<br>BeamPath® fibers |
| Operating<br>Principle | To enable precision control and to stabilize<br>surgeon hand motion | To enable precision control and to stabilize<br>surgeon hand motion |
#### 8. Conclusions
OmniGuide conducted device performance non-clinical verification and validation testing. Results from the testing shows similar performance profiles when comparing the modified flexible handpiece to the original handpiece design. As confirmed in this 510(k) submission, testing was performed to confirm the modifications to the flexible handpiece when compared to its predicate device did not raise any new concerns of safety or effectiveness, alter the fundamental scientific technology of the device, nor affect its mode of use. OmniGuide believes the modified handpiece is substantially equivalent to its predicate handpiece based on similar specifications, performance data, intended use, and the handpiece mechanism of action. Thus, the FlexGuide Ultra handpiece is as safe, as effective and performs similar to the predicate handpiece.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.