K170054 · Jaco, LLC Dba Macan Manufacturing and Bna Burz North America · GEI · Oct 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170054
Device Name
iSurg
Applicant
Jaco, LLC Dba Macan Manufacturing and Bna Burz North America
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 2, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cutting and coagulation of skin.
Device Story
iSurg (MC6B) is an isolated output monopolar electrosurgical generator; table-top standalone device. Generates RF current via AM modulated power oscillator using class C beam power vacuum tube Hartley oscillator. Power control via variable air capacitor attenuator. Operates at 3.0 MHz; produces 50 watts into 500 ohms. Used in clinical settings by physicians for skin cutting and coagulation. Device provides type BF floating patient applied parts; magnetic coupling provides MOPP. Low voltage control board isolates foot pedal from mains and earth. Healthcare providers use device with various electrodes (e.g., incision, excision, ball) to perform surgical procedures; output affects tissue via thermal cautery effect. Benefits include precise tissue incision and hemostasis during skin surgery.
Clinical Evidence
No clinical testing performed. Bench testing included histological evaluation of electrocautery effects on ex-vivo porcine skin using incision, excision, and ball electrodes at minimal, nominal, and maximal power levels. Results demonstrated distinct regions of cautery effect in the dermis at nominal and maximal power settings. Device demonstrated conformity to AAMI ANSI 60601-1 and IEC 60601-2-2 standards.
Technological Characteristics
Monopolar electrosurgical generator; 3.0 MHz operating frequency; 50W nominal power into 500 ohms. Analog power control via variable air capacitor. Class C beam power vacuum tube Hartley oscillator. Type BF floating patient applied parts. Magnetic coupling for isolation. Dimensions: 3.5in x 10.5in x 6.5in; Weight: 6lb. No software. Sterilization not specified for generator; accessories referenced from K052622.
Indications for Use
Indicated for cutting and coagulation of skin.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 2, 2017
JACO , LLC dba Macan Manufacturing and BNA Burz North America % Bill Mclain President and Principal Consultant Keystone Regulatory Services, LLC 342 East Main Street, Suite 207 Leola, Pennsylvania 17540
Re: K170054
Trade/Device Name: iSurg (MC6B) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 5, 2017 Received: January 6, 2017
Dear Bill McLain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/1/Picture/7 description: The image shows the signature of Jennifer R. Stevenson -S3. The signature is followed by the word "Sincerely,". The text is written in a clear, legible font.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Indications for Use | |
| 510(k) Number (if known)<br>K170054 | |
| Device Name<br>ESU (Electrosurgical Unit) Generator Model MC6B/iSurg | |
Indications for Use (Describe)
Cutting and coagulation of skin.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
## Section 5
# 510(k) Summary - K170054
#### 5.1 Submission Correspondent and Owner
## Sponsors (Co-Owners)
JACO, LLC dba Macan Manufacturing 21 Shay Lane Milton, DE 19968 Contact: John Marsillo Phone: 302-645-8068 Email: johnmarsillo@optonline.net
BNA Burz North America, Inc. 906 Wateredge Pl Hewlett, NY 11557 Contact: Jon Garito Phone: 1-800-971-2261 Email: jgarito@bnaburz.com
## Submission Correspondent
Keystone Regulatory Services, LLC 342 E. Main Street, Suite 207 Leola, PA 17540 Contact: William McLain Phone: 717-656-9656 E-Mail: bill.mclain@keystoneregulatory.com
#### 5.2 Date Summary Prepared
September 27, 2017
{4}------------------------------------------------
#### Device Trade Name 5.3
iSurg(MC6B)
#### Device Common Name 5.4
Electrosurgical Generator and Electrodes.
#### 5.5 Device Classification Name
Electrosurgical, Cutting and Coagulation and Accessories. Product code GEI. Classified at 21 CFR Part 878.4400.
### 5.6 Legally Marketed Device To Which The Device Is Substantially Equivalent
The iSurg(MC6B) is substantially equivalent to the Surgi-Max Ultra cleared under K170107and the JACO, LLC dba Macan Manufacturing RadioSurge™ Model MC6A cleared under K050735.
#### 5.7 Description Of The Device
The iSurg(MC6B) is an isolated output monopolar electrosurgical generator configured as a table top standalone device, Height 3.5in (89mm), Width 10.5in (267mm), Depth 6.5in (165mm), Weight 6lb (2,73kg) It produces 50 watts into 500 ohms at 3.0 MHz.
The iSurg(MC6B) generates RF current via an AM modulated power oscillator comprised of a class C beam power vacuum tube Hartley oscillator. The power oscillator runs at full power when actuated. Power control is by means of an attenuator in the form of variable air capacitor.
{5}------------------------------------------------
MOPP (Means of Patient Protection) is in the form of magnetic coupling to an isolated winding on a separate core in the RF coil providing type BF floating patient applied parts in accordance with IEC 60601-2-2. The RF coil primary serves as an intermediate stage with PE connection. MOOP is provided by the low voltage control board (61.010) which galvanically isolates the foot pedal from mains and earth and couples to the oscillator through opto-coupler and relay. SELV common on this board floats. The 81.001A Switch Board uses high isolation relays powered by the low voltage board to isolate the relay coils from the oscillator board. Oscillator actuation is by means of earthing the cathode of the beam power tube via a relay; oscillator OFF state is when the cathode of the beam power tube floats. Tube bias is by means of grid leak via the 10K grid load. The 10K grid load resistor is wire wound type with significant inductance: the current to ground through it is close to DC accordingly. It is current through this resistor that is sensed for the RF ON indicator and active tone (a function of the low voltage control board). The low voltage board also includes a warm up delay to prevent cathode stripping of the vacuum tube.
#### 5.8 Intended Use
The indication for use statement is as follows:
Cutting and coagulation of skin.
#### 5.9 Technological Characteristics
The iSurg(MC6B) has identical characteristics in relation to the predicate Surgi-Max Ultra and reference JACO, LLC dba Macan Manufacturing RadioSurge™ Model MCGA when considered together. The characteristics are identical in relation to:
- · Indication for Use,
- · Specifications,
- · Operating Modes,
- · Nominal Power
- · Cut, Blend, and Coag Output Energies,
- · Output Voltages,
- · Source Impedance,
- · Operating Frequency,
- · Power Control, and
- · Operational Duty Cycle.
#### Summary of Substantial Equivalence 5.10
The following table summarizes the basis for substantial equivalence.
{6}------------------------------------------------
| Features | iSurg(MC6B)(Proposed Device) | Surgi-Max Ultra (K170107) (Predicate) | JACO, Macan<br>RadioSurge TM Model MC6A<br>(K050735) (Predicate) |
|-------------------------|-------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------|
| Indications | Cutting and coagulating skin | Cutting, coagulation, hemostasis, skin surgery (Paraphrased) | Dental applications (Paraphrased) |
| Prescription or OTC Use | Prescription | Prescription | Prescription |
| FDA Code | GEI Electrosurgical | GEI Electrosurgical | EKZ Dental |
| Specifications | Monopolar, Type BF | Monopolar and Bi-Polar, Type BF | Monopolar, Type BF |
| Operating Mode | Cut, Cut/Coag, Coag | Cut, Blend, Hemo, Bipolar, Bipolar Turbo | Cut, Cut/Coag, Coag |
| Nominal Power | 50W Nominal | 170W Nominal | 50W Nominal |
| Cut Output Energy | 50W Nominal into 500 ohms | Not stated in the 510(k) Summary | 50W Nominal into 100 ohms |
| Blend Output Energy | 50W Nominal into 500 ohms | Not stated in the 510(k) Summary | 50W Nominal into 100 ohms |
| Coag Output Energy | 25W Nominal into 500 ohms | Not Stated in the 510(k) Summary | 25W Nominal into 100 ohms |
| Output Voltage | 520V pk | 1200V pk | 520V pk |
| Source Impedance | 500 ohms | Not stated in the 510(k) Summary | 500 ohms |
| Operating Frequency | 3.0 MHz | 4.0 MHz in Monopolar Mode | 3.0 MHz |
| Power Control | Analong, continously variable | Not stated in 510(k) Summary | Analong, continously variable |
| Operational Duty Cycle | 10 seconds ON, 20 seconds OFF | 10 seconds ON, 30 seconds OFF | 10 seconds ON, 20 seconds OFF |
{7}------------------------------------------------
| | Table 5.1: (continued) |
|--|------------------------|
| | |
| Features | Proposed Device | Predicate Device | Reference Device |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Software | None | Contains Software | None |
| Accessories | As cleared in K052622. Hand-<br>piece, Dispersive Pad, Disper-<br>sive Cord, Foot Switch, Elec-<br>trodes(RF1, RF11, RL33, RL34,<br>RC51, RC52) | Unknown | As cleared in K052622. Hand-<br>piece, Dispersive Pad, Disper-<br>sive Cord, Foot Switch, Elec-<br>trodes(RF1, RF11, RL33, RL34,<br>RC51, RC52) |
{8}------------------------------------------------
#### Non-Clinical Testing 5.11
JACO, LLC dba Macan Manufacturing and BNA Burz North America, Inc. have declared conformity to the following relevant consensus standards:
- · AAMI ANSI 60601-1 C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- · IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- · IEC 60601-2-2 Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories [Including: Technical Corrigendum 1 (2014)] 1/27/2015
## Histology Study
A histological evaluation of the cutting modes on porcine skin was conducted. The purpose of this study was to assess the length and depth of electrocautery effect (with these two dimensions defining the area of effect) on ex-vivo pig skin using histologic evaluation. Three different electrodes were used representing worst-case scenarios.
Five different electrocautery test article effects (comprising combinations of the specified electrodes with different waveforms designated as Groups 1-5) were each evaluated at minimal, nominal, and maximal power in triplicate. No cautery effect was appreciable within any group at minimal power. However, in all nominal and maximal power samples, distinct regions of cautery effect were evident within the dermis.
The incision electrode employing cut and blend waveforms had V shaped incisions that extended into the mid-dermis accompanied by small halos of cautery effect. Nominal and maximal power levels had similarly-sized areas of cautery effect for both waveforms.
The excision electrode employing cut and blend waveforms provided broad, shallow excisions with a curvilinear appearance and slightly raised margins extending into the superficial dermis accompanied by a thin halo of cautery effect. A direct relationship was observed with increased area of cautery effect occurring with increasing power levels.
The ball electrode employing a hard coagulation waveform affected a small focal area (approximately 2x1 mm), had a slightly convex central area on histology, and was accompanied by a small, focal halo of cautery effect within the superficial dermis. The relationship of cautery area to waveform power is somewhat uncertain; although average area of effect was slightly greater for nominal vs. maximal power, the small dimensions of test article effect could have lead to slight underestimation of maximal power effect due to sectioning variation.
At minimal power, no cautery effect was evident for any treatment group. For Group 1, with the incision electrode utilizing cut mode, the largest width of cautery effect was observed with maximal power as 0.421 mm (width); nominal power had slightly greater depth at 0.210 mm but was similar to maximal power (0.125 mm depth). For Group 2, with the incisoin electrode utilizing blend mode the largest area of cautery effect was observed with nominal power and was 0.473mm (width) and 0.190 mm (depth); but was similar to maximal values (0.369 mm for depth) and (0.148mm for width). For Group 3, with the excision electrode utilizing cut mode the largest area of cautery effect was observed with maximal power and was 6.239 mm (width) and 1.612 mm (depth). For Group 4, with the excision electrode utilizing cut mode the largest area of cautery effect was observed with
{9}------------------------------------------------
maximal power was 6.039 mm (width) and 0.917 mm (depth). For Group 5, with the ball electrode utilizing the hard coagulation mode the largest area of cautery effect was observed with with nominal power and was 2.256 mm (width) and 1.209 mm (depth), although small size of test article effect could have slightly underestimated cautery effect within the maximal cohort for this electrode and waveform combination.
#### 5.12 Biocompatibilty
Biocompatibility testing was not submitted in association with this 510(k). Rather, patient contacting materials associated with electrodes were referenced in relation to the reference device JACO, LLC dba Macan Manufacturing RadioSurge™ Model MC6A cleared under K050735.
#### 5.13 Clinical Testing
No clinical testing was performed in association with this submission.
#### 5.14 Conclusions
The results of the comparison of design, materials, intended use and technological characteristics demonstrate that the device is as safe and effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.