K170107 · Elliquence, LLC · GEI · May 3, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170107
Device Name
Surgi Max Ultra
Applicant
Elliquence, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
May 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Orthopedic, arthroscopic, spinal, and neurological For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist. Cutting Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty. Blended Cutting and Coagulation Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal Cell Carcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps. Hemostasis and Nonablative Coagulation Control of bleeding, Epilation, Telangiectasia Bipolar Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, sub-mucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.
Device Story
Surgi-Max Ultra is a high-frequency RF electrosurgical generator; provides monopolar and bipolar energy for soft tissue cutting, ablation, and coagulation. Device is a modification of Surgi-Max/Surgi-Max Plus (K100390); features updated internal electrical components for improved efficiency; maximum power output increased from 120W to 170W. Used in OR/clinical settings by physicians; operated via footswitch or fingerswitch. Input is electrical power; output is high-frequency RF energy delivered through compatible electrosurgical accessories. Healthcare providers use device to achieve precise hemostasis or tissue resection; benefits include effective hemorrhage control and surgical versatility across orthopedic, spinal, and ENT procedures.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety (AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2), usability assessment (IEC 62366), software verification/validation (IEC 62304), energy output verification, and ex-vivo thermal effect comparison on bovine tissue.
Technological Characteristics
High-frequency RF electrosurgical generator; 4 MHz (monopolar) and 1.7 MHz (bipolar) output frequency. Power output: 170W. Applied part type: CF. Connectivity: Standalone. Power input: 100-240V AC, 50/60Hz. Duty cycle: 10s ON/30s OFF. Complies with AAMI ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-2-2.
Indications for Use
Indicated for resection, ablation, and coagulation of soft tissues and hemostasis in orthopedic, arthroscopic, spinal, and neurological procedures, including specific arthroscopic procedures of the knee, shoulder, ankle, elbow, hip, and wrist, as well as various ENT and dermatological applications.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The text is left-aligned and in a simple font.
May 3, 2017
Elliquence, LLC Mr. Paul Buhrke IV QA/RA Manager 2455 Grand Avenue Baldwin, New York 11510
Re: K170107
Trade/Device Name: Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 03, 2017 Received: April 04, 2017
Dear Mr. Buhrke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K170107
Device Name
Surgi-Max Ultra (IEC6-SU170, IEC6-SV170, IEC6-SP170)
Indications for Use (Describe)
Orthopedic, arthroscopic, spinal, and neurological
For resection, ablation, and coagulation of soft tissues and hemostasis in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist.
#### Cutting
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty.
Blended Cutting and Coagulation
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal Cell Carcinoma. Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps.
Hemostasis and Nonablative Coagulation Control of bleeding, Epilation, Telangiectasia
Bipolar
Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, sub-mucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.
Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo of a person. The person is drawn in white lines on a dark blue background. The person is facing away from the viewer and has their right arm raised. The logo is circular in shape and has a glossy appearance.
# 510 (k) Summary
(As required by 21 CFR 807.92(a))
Date Prepared
May 1, 2017
# Submitter's Information (807.92(a)(1))
Company Name and Address:
Elliquence, LLC 2455 Grand Avenue Baldwin, NY 11510 Phone: (516) 277-9000 Fax: (516) 277-9001 www.elliquence.com
Establishment Registration for Elliquence, LLC is 3007024186.
Contact Information:
Daniel Connell - Requlatory Affairs Associate Phone: (516) 277-9036 Fax: (516) 277-9011 Email: dconnell@elliquence.com
Paul D. Buhrke IV- QA/RA Manager Phone: (516) 277-9010 Fax: (516) 277-9011 Email: pbuhrke@elliquence.com
# Device Information (807.92(a)(2))
Trade Name Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170)
Common/Usual Name Electrosurgical unit and accessories, electrosurgical cutting and coagulation device and accessories
> elliquence, LLC. 2455 Grand Avenue · Baldwin, New York 11510 (516) 277.9000 · Fax: (516) 277.9001 · www.elliquence.com
{4}------------------------------------------------
Classification Name and Regulation Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400
K170107 Page 2 of 5
Class FDA Classification: Class II FDA Product Code: GEI
#### Predicate Devices (807.92(a)(3))
- Surgi-Max® / Surgi-Max® Plus (K100390) ●
- . Bovie® IDS-310 Electrosurgical Generator (K134054)
#### Device Description (807.92(a)(4))
The elliquence Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170) is a modification to the existing Surgi-Max® / Surgi-Max® Plus (K100390) line of radiofrequency (RF) electrosurgical generators. It is a compact source of high-frequency RF energy employed for cutting and coagulation of soft tissue in a variety of surgical procedures. It is designed to enable the performance of monopolar and bipolar functions in conjunction with separatelymarketed electrosurgical accessories and ancillary equipment which have been developed by elliquence for mutually-exclusive use with the Surgi-Max® line of generators.
The Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170) is the result of an update to internal electrical components of the Surgi-Max® / Surgi-Max® Plus. As a result of this equipment update, device design as it relates to internal electrical system framework was simplified in many respects and overall efficiency was greatly improved upon. This increase in efficiency enabled an increase in the maximum power output of the device from 120W to 170W. Aside from this increase in maximum power output, all other operational aspects of the device were conserved. The only other modifications made to the device were aesthetic in nature.
The modifications enacted do not affect the device's intended use or alter its fundamental scientific technology.
{5}------------------------------------------------
#### Intended Use (807.92(a)(5))
#### Orthopedic, arthroscopic, spinal, and neurological
For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist.
#### Cutting
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty.
#### Blended Cutting and Coagulation
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal Cell Carcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps.
### Hemostasis and Nonablative Coagulation
Control of bleeding, Epilation, Telangiectasia
### Bipolar
Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, sub-mucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.
### Substantial Equivalence Comparison (807.92(a)(6))
The Surgi-Max® Ultra is substantially equivalent in both intended use and technological characteristics to the Surgi-Max® / Surgi-Max® Plus (K100390) as well as the Bovie® IDS-310 Electrosurgical Generator (K134054). An extensive comparison chart concerning key characteristics is provided on the following pages (Table 5.1).
{6}------------------------------------------------
# Table 5.1 Substantial Equivalence Comparison Chart
| | Surgi-Max® Ultra<br>(K170107) - SUBJECT<br>DEVICE | Surgi-Max® / Surgi-Max<br>Plus (K100390) –<br>PREDICATE | Bovie® IDS-310<br>Electrosurgical Generator<br>(K134054) - PREDICATE |
|-----------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | elliquence, LLC | elliquence, LLC | Bovie Medical Corp. |
| Product code | GEI | GEI | GEI |
| Regulation<br>number | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Regulation<br>name | Electrosurgical Cutting and<br>Coagulation Device and<br>accessories | Electrosurgical Cutting and<br>Coagulation Device and<br>accessories | Electrosurgical Cutting and<br>Coagulation Device and<br>accessories |
| Indications for<br>use | See previous page | See previous page | The IDS-310 High<br>Frequency Electrosurgical<br>Generator is used to deliver<br>RF energy via an assortment<br>of surgical devices to cut and<br>coagulate different kinds of<br>tissue. |
| Prescription<br>device<br>(21 CFR<br>801.109) | Yes | Yes | Yes |
| Principle of<br>Operation | Conversion of electrical<br>energy to high-frequency RF<br>energy<br>(monopolar and bipolar) | Conversion of electrical<br>energy to high-frequency RF<br>energy<br>(monopolar and bipolar) | Conversion of electrical<br>energy to high-frequency RF<br>energy<br>(monopolar and bipolar) |
| Modes of<br>operation | Cut<br>Blend<br>Hemo<br>Bipolar<br>Bipolar Turbo | Cut<br>Blend<br>Hemo<br>Bipolar<br>Bipolar Turbo | 9 Monopolar and 4 bipolar<br>modes |
| Output<br>frequency | 4 MHz (monopolar)<br>1.7 MHz (bipolar) | 4 MHz (monopolar)<br>1.7 MHz (bipolar) | 350 - 490 kHz (monopolar)<br>490 kHz (bipolar) |
| Maximum power<br>output | 170W | 120W | 300W |
| Voltage<br>(peak-to-peak) | 1,200V | 1,600V | 8,000V |
| Duty Cycle | 10 sec On / 30 sec OFF | 10 sec On / 30 sec OFF | 10 sec On / 30 sec OFF |
| Supply input<br>power | 100 V AC – 240 V AC, 50/60<br>Hz | 100 V AC – 240 V AC, 50/60<br>Hz | 100 V AC – 240 V AC, 50/60<br>Hz |
elliquence, LLC. 2455 Grand Avenue • Baldwin, New York 11510 (516) 277.9000 • Fax: (516) 277.9001 • www.elliquence.com
{7}------------------------------------------------
| | Surgi-Max® Ultra<br>(K170107) – SUBJECT<br>DEVICE | Surgi-Max® / Surgi-Max<br>Plus (K100390) –<br>PREDICATE | Bovie® IDS-310<br>Electrosurgical Generator<br>(K134054) - PREDICATE |
|---------------------------------------------------|-------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------|
| Power<br>activation<br>control | Footswitch and/or<br>fingerswitch | Footswitch and/or<br>fingerswitch | Footswitch and/or<br>fingerswitch |
| Electrical safety<br>and EMC testing<br>standards | AAMI ANSI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | AAMI ANSI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 |
| Applied part<br>type | CF | BF | CF |
# Non-Clinical Testing (807.92(b)(1))
The modifications enacted for the Surgi-Max® Ultra were made in accordance with the design control requirements of 21 CFR 820.30. Appropriate non-clinical verification and validation activities were planned and conducted in order to address identified risks and ensure the safety and effectiveness of the device with the modifications enacted. This testing included:
- . Electrical safety and EMC testing in accordance with AAMI ANSI ES60601-1, IEC 60601-1-2. and IEC 60601-2-2
- Usability assessment in accordance with IEC 62366
- Software verification and validation in accordance with IEC 62304 and the FDA Final . Guidance document titled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- Energy output verification of each of the five modes and comparison to predicate ●
- . Thermal effect comparison for representative modes of the subject device and predicate on ex-vivo bovine tissue
### Clinical Testing (807.92(b)(2))
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
### Conclusion (807.92(b)(3))
Based upon a comparison of the intended uses, technological characteristics, and performance testing, we have concluded that the subject device is substantially equivalent to the selected predicate devices.
> elliquence, LLC. 2455 Grand Avenue · Baldwin, New York 11510 (516) 277.9000 · Fax: (516) 277.9001 · www.elliquence.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.