K052297 · Porex Surgical, Inc. · GEI · Mar 28, 2006 · General, Plastic Surgery
Device Facts
Record ID
K052297
Device Name
POREX ELECTROSURGERY NEEDLE
Applicant
Porex Surgical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 28, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Porex Electrosurgery Needle is an electrosurgery clectrode for use in monopolar electrosurgical handpieces. It is a single-use device intended for cutting, dissecting, and cauterizing of soft tissue.
Device Story
Porex Electrosurgery Needle is a single-use electrosurgical electrode designed for use with standard monopolar cautery handpieces. Device consists of a tungsten tip mounted on a gold-plated stainless steel shaft, featuring dual-layer insulation. During surgical procedures, the device delivers high-frequency electrical current to soft tissue to perform cutting, dissection, and coagulation. Operated by surgeons in clinical or OR settings, the device functions as a passive accessory to an electrosurgical generator. The provider manipulates the needle tip directly on the target tissue to achieve the desired surgical effect. By providing precise tissue interaction, the device facilitates surgical access and hemostasis, potentially reducing blood loss and improving procedural efficiency.
Clinical Evidence
Bench testing only.
Technological Characteristics
Tungsten tip; gold-plated stainless steel shaft/sleeve; dual-layer insulation. Monopolar electrosurgical electrode. Single-use. Passive accessory; no internal energy source or software.
Indications for Use
Indicated for cutting, dissecting, and cauterizing soft tissue using monopolar electrosurgical handpieces.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K163442 — Single Use Micro Electrode (Brass with gold plated shaft); Single Use Micro Electrode (Stainless Steel shaft) · Modern Medical Equipment Manufacturing Limited · Sep 28, 2017
{0}------------------------------------------------
## MAR 2 8 2006
#### 03/23/06
Page 1 of 2
### SUMMARY 510(k) NUMBER: K052297
# POREX SURGICAL ELECTROSURGICAL NEEDLE
### SUBMITTER
Porex Surgical, Inc. 15 Dart Road Newnan, GA 30265-1017 Tel: (678) 479-1610, Fax: (678)423-1435
#### CONTACT CONTACT CONTACT
Eric V. Hohenstein Vice President, Special Developmental Projects Tel: (678) 479-1610 / (323) 876-3767 Fax: (678) 423-1435 / (323) 876-5592 /
## PROPRIETARY NAME: Porex Electrosurgery Needle
COMMON NAME: Microdissection Needle, Electrosurgical Needle
| DEVICE CLASSIFICATION | | | |
|------------------------------------|-----------|-------|----------|
| DEVICE NAME | CLASS NO. | CLASS | REG NO. |
| Device, Electrosurgical, cutting & | JOS | 2 | 878.4400 |
ત્ coagulation & accessories
# IDENTIFICATION OF EQUIVALENT DEVICE___________________________________________________________________________________________________________________________________________
Porex Surgical, Inc. Electrosurgery Needles are substantially equivalent to the following predicate devices in their intended use, material, design, and surgical procedure:
Colorado MicroDissection Needle (K000348)
Megadyne Electrode, Electrosurgery (K903302)
Valleylab Coated Electrodes (K962044)
{1}------------------------------------------------
05297
## DESCRIPTION
The Porex Surgical, Inc. Electrosurgery Needle is an electrosurgical electrode for use with monopolar electrosurgical accessories (cautery handpieces). It is constructed of a tungsten tip held by a gold plated stainless shaft/sleeve. The electrode shaft has two layers of insulation.
## INTENDED USE
The Porex Surgical, Inc. Electrosurgery Needle is an electrosurgical electrode used for cutting, dissecting and cauterizing soft tissue by use of high-frequency electrical current.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The eagle is positioned within a circle that contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA." The text is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 8 2006
Porex Surgical Inc. c/o Mr. Eric V. Hohenstein V.P., Special Developmental Projects 15 Dart Road Newnan, Georgia 30265-1017
Re: K052297
Trade/Device Name: Porex Electrosurgery Needle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI, JOS Dated: March 6, 2006 Received: March 8, 2006
Dear Mr. Hohenstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I·DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 -- Mr. Eric V. Hohenstein
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Hubert Lehman
r / Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
052297
510(k) Number (if known):
Device Name: Porex Electrosurgery Needle
Indications For Use:
The Porex Electrosurgery Needle is an electrosurgery clectrode for use in monopolar electrosurgical handpieces. It is a single-use device intended for cutting, dissecting, and cauterizing of soft tissue.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IS NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
Julin Lemer
Division Sign-Off
Division of General, Restorative, and Neurological Devices
Page 1 of 1
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.