Sutter Arrowtip Monopolar Electrodes

K192128 · Sutter Medizintechnik GmbH · GEI · Oct 4, 2019 · General, Plastic Surgery

Device Facts

Record IDK192128
Device NameSutter Arrowtip Monopolar Electrodes
ApplicantSutter Medizintechnik GmbH
Product CodeGEI · General, Plastic Surgery
Decision DateOct 4, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ARROWtip™ Single-Use monopolar electrode is an electrosurgical instrument for precision soft tissue dissection. It is a single-use device intended for cutting, dissecting and cauterizing soft tissue. The ARROWtip™ Single-Use monopolar electrode is not intended for use in the central nervous system or in the central circulatory system.

Device Story

Sutter ARROWtip Monopolar Electrodes are single-use electrosurgical instruments designed for precision soft tissue dissection, cutting, and cauterization. The device consists of an electrode tip (available in various lengths and angles) that connects to a standard monopolar electrosurgical generator via a compatible handpiece. During surgery, the clinician uses the electrode to deliver electrical energy to soft tissue to achieve the desired surgical effect. The device is provided sterile and is intended for single-use only. By providing a specialized tip, the device assists surgeons in performing precise tissue manipulation, potentially improving surgical outcomes through controlled cutting and coagulation. It is intended for use in clinical settings by trained healthcare professionals.

Clinical Evidence

Bench testing only. No clinical or animal studies performed. Performance verified through electromagnetic compatibility (IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2), mechanical strength, and thermal effects on tissue testing. Histological morphometric measurements confirmed equivalent cutting/coagulation performance to the predicate. Biocompatibility evaluated per ISO 10993-1.

Technological Characteristics

Materials: Tungsten, stainless steel, PTFE. Energy: Monopolar electrosurgical (RF). Form factor: Single-use electrode with various lengths/angles. Connectivity: Connects to external electrosurgical generator. Sterilization: Ethylene Oxide (ISO 11135). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 11607-1, ISO 10993-1.

Indications for Use

Indicated for precision soft tissue dissection, cutting, and cauterizing in surgical procedures. Contraindicated for use in the central nervous system or central circulatory system.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 4, 2019 SUTTER MEDIZINTECHNIK GmbH Mr. Ulrike Zeissler Manager, Regulatory Affairs Tullastrasse 87 79108 Freiburg, Germany Re: K192128 Trade/Device Name: Sutter ARROWtip Monopolar Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 1, 2019 Received: August 7, 2019 Dear Mr. Zeissler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K192128 Device Name Sutter ARROWtip™ Monopolar Electrodes #### Indications for Use (Describe) The ARROWtip™ Single-Use monopolar electrosurgical instrument for precision soft tissue dissection. It is a single-use device intended for cutting, dissecting and cauterizing soft tissue. The ARROWtip™ Single-Use monopolar electrode is not intended for use in the central nervous system or in the central circulatory system. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Section 807.92 | Date: | October 2nd, 2019 | | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Name:<br>Address: | SUTTER MEDIZINTECHNIK GmbH<br>Tullastrasse 87<br>79108 Freiburg<br>Germany | | | Contact person:<br>Titel:<br>Telephone:<br>Fax: | Ulrike Zeissler<br>Manager Regulatory Affairs<br>+49 (0) 761 51551-14<br>+49 (0) 761 51551-30 | | Product: | Trade Name:<br>Common Name:<br>Classification name:<br>Product Code:<br>Regulation Number:<br>Classification:<br>Classification Panel: | Sutter ARROW tip™ Monopolar Electrodes<br>Monopolar Electrodes<br>Electrosurgical, cutting & coagulation & Accessories<br>GEI<br>CFR 21 § 878.4400<br>Class II<br>General and Plastic Surgery | | Predicate Device: | Predicate device to which Sutter ARROWtip™ Monopolar Electrodes are claimed to be substantially equivalent is manufactured by Stryker Leibinger, Stryker Leibinger Colorado MicroDissection Needles® (K033232) | | | Device Description: | Sutter ARROW tip™ Monopolar Electrodes are electrosurgical instruments. The electrodes are provided with different angled electrode front parts and in different lengths. They are to be connected through an appropriate monopolar handpiece with the monopolar output of an electrosurgical generator. The electrodes are provided sterile and are single-use instruments. | | | Intended Use: | The ARROWtip™ Single-Use monopolar electrode is an electrosurgical instrument for precision soft tissue dissection. It is a single-use device intended for cutting, dissecting and cauterizing soft tissue. The ARROWtip™ Single-Use monopolar electrode is not intended for use in the central nervous system or in the central circulatory system. | | | Technological<br>Characteristics: | The table below provides a comparison between the subject device<br>and the predicate device | | | | Sutter ARROWtip™<br>Monopolar Electrodes | Predicate device<br>K033232 | | | Straight / angled | Straight / angled | | | Ø 0.8 | Ø 0.5 mm | | | Tungsten, stainless<br>steel, PTFE | Tungsten, stainless<br>steel, PTFE, Polyolefin | | | yes | yes | | | yes | yes | | | yes | yes | | | yes | yes | | | 10-6 | 10-6 | | | Ethylene Oxide | Ethylene Oxide | | Non-Clinical<br>Performance<br>Data: | The following performance data has been obtained for the<br>substantial equivalence determination. | | | | <b>Bench Testing</b> | | | | Performance testing has been executed in line with the internal<br>R&D process and in compliance with the proposals and<br>recommendations of the FDA guidance: "Premarket Notification<br>[510(k)] Submissions for Electrosurgical Devices for General<br>Surgery" - Guidance for Industry and Food and Drug Administration<br>Staff, August 15, 2016. | | | | In particular, tests were carried out with respect to following subject<br>areas: | | | | Performance Data Bench Tests | | | | Test | Conclusion | | | Electromagnetic Compatibility and<br>Electrical Safety | Pass | | | Mechanical strength and<br>functionality performance testing | Pass | | | Thermal effects on tissue | Pass | | | <b>Electromagnetic Compatibility and Electrical Safety Testing</b> | | | | Electrical and electromagnetic tests were performed to demonstrate<br>that design specifications and performance requirements are met.<br>Compliance to the voluntary standards IEC 60601-1 (AAMI/ANSI<br>ES60601-1:2005), IEC 60601-2-2 and IEC 60601-1-2 has been<br>shown. | | | | <b>Mechanical strength and functionality performance testing</b> | | | | Mechanical strength and functionality performance testing was | | | | Mechanical stress tests showed that the design specification are<br>met.<br>Thermal effects on tissue<br>Thermal effects on tissue testing was performed to determine<br>thermal effects cased by the Sutter ARROW tip™ Monopolar<br>Electrodes at different modes and power levels in comparison to<br>the predicate device. Three different types of tissue were used and<br>tests were performed in triplicate. Visual comparison as well as<br>digital morphometric measurement using histology showed<br>equivalent coagulation / cutting performance of subject device and<br>predicate device. | | | | Sterilization<br>Sutter ARROW tip™ Monopolar Electrodes are sterilized by using a<br>validated ethylene oxide cycle. The sterilization cycle has been<br>validated to ensure a sterility level of (SAL) 10-6 in accordance with<br>ISO 11135. | | | | Shelf Life Testing<br>Shelf-life testing has been conducted in accordance ISO 11607-1.<br>The aging studies established that the Sutter ARROW tip™<br>Monopolar Electrodes and packaging remain functional and<br>maintain sterility for up to 3 years. | | | | Biocompatibility Testing<br>Biological evaluation and Biocompatibility testing has been<br>performed in compliance to ISO 10993-1<br>Fourth edition 2009-10-15, "Biological evaluation of medical devices<br>- Part 1: Evaluation and testing within a risk management process". | | | | Animal Studies<br>No animal studies have been performed as appropriate verification<br>and validation of the subject device has been achieved based on<br>comparison to the predicate device and from results of the bench<br>testing, biocompatibility evaluation, and electrical / safety testing. | | | | Clinical Studies<br>No clinical studies have been performed as appropriate verification<br>and validation of the subject device has been achieved based on<br>comparison to the predicate device and from results of the bench<br>testing, biocompatibility evaluation, and electrical / safety testing. | | | Conclusion: | Sutter ARROW tip™ Monopolar Electrodes are substantially<br>equivalent to the predicate device since the intended use, design,<br>material and basic features are the same. The minor differences<br>raise no new issues of safety and effectiveness, as the design<br>differences have no effect on the performance, function or intended<br>use. | | {4}------------------------------------------------ # APPLICANT: SUTTER MEDIZINTECHNIK GMBH DEVICES: SUTTER ARROWtip™ MONOPOLAR ELECTRODES {5}------------------------------------------------ ## APPLICANT: SUTTER MEDIZINTECHNIK GMBH DEVICES: SUTTER ARROWtip™ MONOPOLAR ELECTRODES
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%