K242932 · Cryoconcepts LP · GEH · Apr 15, 2025 · General, Plastic Surgery
Device Facts
Record ID
K242932
Device Name
Skin Clinic Nitro Clear Wart Remover
Applicant
Cryoconcepts LP
Product Code
GEH · General, Plastic Surgery
Decision Date
Apr 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Skin Clinic Nitro Clear Wart Remover is intended for the OTC treatment of common warts and plantar warts in adults and children 4 years of age and older.
Device Story
Device utilizes extreme cold to remove common and plantar warts via cryosurgery. System consists of a container of nitrous oxide cryogen in a plastic holder with an applicator tip. User dispenses cryogen into the tip, which acts as a reservoir and cools to as low as -89°C. Applicator is placed against the wart for 10-40 seconds; freezing destroys target tissue, allowing the wart to fall away and new epidermis to grow. Intended for OTC consumer self-use or caregiver use on children. Treatment may require 1-4 applications at two-week intervals. Benefits include non-invasive wart removal.
Clinical Evidence
Bench testing only. In vitro testing demonstrated the device kills cells embedded in an agar matrix comparable to predicate devices. Bench testing confirmed the device achieves equivalent or lower temperatures than predicates. Biocompatibility testing performed per ISO 10993. Human factors usability study conducted to support OTC indication.
Technological Characteristics
Cryosurgical unit; Nitrous Oxide cryogen; plastic holder/applicator tip; cooling to -89°C; manual operation; disposable unit; 4-year shelf life; biocompatibility per ISO 10993.
Indications for Use
Indicated for OTC treatment of common warts and plantar warts in adults and children 4 years of age and older.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
K060859 — COMPOUND W FREEZE OFF WART REMOVAL SYSTEM · OraSure Technologies, Inc. · Jul 18, 2006
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 15, 2025
CryoConcepts LP
Sam Niedbala
CEO
1100 Conroy Place
Easton, Pennsylvania 18040
Re: K242932
Trade/Device Name: Skin Clinic NitroClear & Private Label Versions
Regulation Number: 21 CFR 878.4350
Regulation Name: Cryosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: GEH
Dated: March 16, 2025
Received: March 19, 2025
Dear Sam Niedbala:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K242932 - Sam Niedbala
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K242932 - Sam Niedbala
Page 3
Sincerely,
James H. Jang -S
Digitally signed by
James H. Jang -S
Date: 2025.04.15
22:50:02 -04'00'
For
Long Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242932 | |
| Device Name Skin Clinic Nitro Clear Wart Remover | |
| Indications for Use (Describe) The Skin Clinic Nitro Clear Wart Remover is intended for the OTC treatment of common warts and plantar warts in adults and children 4 years of age and older. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K242932
# 510(k) SUMMARY
(As Required by 21 CFR 807.92)
## I. SUBMITTER
CryoConcepts LP
1100 Conroy Place
Easton, PA 18040
Phone: (215) 853-6276
Contact: Sam Niedbala, Ph.D.
April 14, 2025
## II. DEVICE
Name of Device: Skin Clinic Nitro Clear Wart Remover
Usual Name – Cryosurgical unit and accessories
Classification Name – General & Plastic Surgery
Regulatory Class: II
Code of Federal Regulation: 878.4350
Product Code: GEH
## III. PREDICATE DEVICES – Class II
Skin Clinic Freeze n’ Clear for Skin Tags & Warts K211099
Compound W Nitro-Freeze Wart Remover K172373
## IV. DEVICE DESCRIPTION
The Skin Clinic Nitro Clear Wart Remover device utilizes extreme cold to facilitate the removal of common and plantar warts by freezing. Each device contains a container of cryogen gas in a plastic holder along with instructions for use. The Skin Clinic Nitro Clear Wart Remover device is for OTC use and utilizes Nitrous Oxide cryogen delivered from the cartridge into a tip which acts as a reservoir for the cryogen gas. The applicator tip is cooled to as low as $-89^{\circ}\mathrm{C}$. Depending on the location and size, the applicator is placed against the common wart or
Page 1 of 5
{5}
K242932
plantar wart between 10-40 seconds which freezes the targeted area. The frozen wart falls away over time and new epidermis grows in its place. One to four treatments with intervals of two weeks may be required.
## V. INDICATIONS FOR USE
The Skin Clinic Nitro Clear Wart Remover is intended for the OTC treatment of common warts and plantar warts in adults and children 4 years of age and older.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A summary of the technological characteristics for the Skin Clinic Nitro Clear Wart Remover vs Freeze 'n Clear Skin Clinic for Warts/Skin Tags and the Compound W Nitro-Freeze are provided in the following table. Each of these products utilize similar technological characteristics as shown in the table below.
| Technological Characteristics | Freeze 'n Clear Skin Clinic for Warts and Skin Tags (K211099) | Compound W Nitro-Freeze Wart Remover (K172373) | Skin Clinic Nitro Clear Wart Remover (Subject Device) |
| --- | --- | --- | --- |
| -Intended Use | The Freeze 'n Clear Skin Clinic for Warts and Skin Tags product is intended for the OTC treatment of common warts, plantar warts, and skin tags.
Warts: For use with children 4 years of age and older | For the over-the-counter treatment of common warts and plantar warts. For use with Adults and children 4 years and older | The Skin Clinic Nitro Clear Wart Remover is intended for the OTC treatment of common warts and plantar warts in adults and children 4 years of age and older. |
| -Cryogen | Mixture of Dimethyl Ether, Propane and Butane | Nitrous Oxide | Nitrous Oxide |
Page 2 of 5
{6}
K242932
| Technological Characteristics | Freeze 'n Clear Skin Clinic for Warts and Skin Tags (K211099) | Compound W Nitro-Freeze Wart Remover (K172373) | Skin Clinic Nitro Clear Wart Remover (Subject Device) |
| --- | --- | --- | --- |
| -Materials | -canister containing cryogen
-Foam Applicators | -Plastic device holding canister of nitrous oxide cryogen
-Foam Tip Applicator | - Plastic device holding canister of nitrous oxide cryogen
-Plastic Tip Applicator |
| -Mode of Use | Cryogen dispensed into foam applicator which is then applied to wart | Table-top device with disposable foam applicators | Cryogen dispensed into reusable applicator tip and applied to the target wart |
| -Mechanism of action | Extreme cold destroys the target tissue | Extreme cold destroys the target tissue | Extreme cold destroys the target tissue |
| -Storage & Safety Conditions | -Keep away from fire or flame
-Do not smoke while using the product
-Do not puncture or incinerate canister
-Do not expose to heat or store at temperatures above 120°F.
-Store at room temperature away from heat | -The device is under high pressure and cannot be opened.
-Protect from sunlight and do not expose to temperature above 50°C(120°F) before first use or 35°C(95°F) after first use
-Store at room temperature away from heat
-Keep away from flames or materials that burn easily, or are sources of sparks or ignition, including mobile phones, radios, and other electrical appliances.
-Do not smoke or use near and open flame. | -The device is under high pressure and cannot be opened.
-Protect from sunlight and do not expose to temperature above 50°C(120°F) before first use or 35°C(95°F) after first use
-Store at room temperature away from heat
-Keep away from flames or materials that burn easily, or are sources of sparks or ignition, including mobile phones, radios, and other electrical appliances.
-Do not smoke or use near and open flame. |
Page 3 of 5
{7}
K242932
| Technological Characteristics | Freeze 'n Clear Skin Clinic for Warts and Skin Tags (K211099) | Compound W Nitro-Freeze Wart Remover (K172373) | Skin Clinic Nitro Clear Wart Remover (Subject Device) |
| --- | --- | --- | --- |
| | | -Do not inhale nitrous oxide | -Do not inhale nitrous oxide |
| -Treatment Procedure | Spray the cryogen into the applicator and then place it directly onto the wart for a specified number of seconds | Following Device Activation, dispense the cryogen into the applicator and then place it directly onto the wart for a specified number of seconds | Following Device Activation, dispense the cryogen into the applicator and then place it directly onto the wart for a specified number of seconds |
| Shelf Life | 3 Years | 3 Years | 4 Years |
| -Disposal | Entire unit is disposable after emptied of cryogen. | Entire unit is disposable after emptied of cryogen. | Entire unit is disposable after emptied of cryogen. |
| -Defined Operators | OTC for consumer use | OTC for consumer use | OTC for consumer use |
| -Service / Repair | None | None | None |
## VII. PERFORMANCE DATA
As part of the submission, the product was tested for biocompatibility according to FDA's biocompatibility 2020 guidance and meets the requirements of ISO 10993.
Bench-testing compared the temperatures attained by the Compound W Nitro-Freeze and Freeze n Clear Skin Clinic predicate devices to the Skin Clinic Nitro Clear Wart Remover product to demonstrate that they were comparable. The subject device bench testing showed that equivalent or lower temperatures could be attained using the device. An in vitro test was also performed demonstrating that the subject device and predicate devices were able to kill cells embedded in an agar matrix over the time course of a standard treatment time.
A human factors usability study was performed in support of the over-the-counter indication for the Skin Clinic Nitro Clear Wart Remover product to treat common and plantar warts.
## VIII. CONCLUSIONS
Page 4 of 5
{8}
K242932
The Skin Clinic Nitro Clear Wart Remover product is substantially equivalent to the Freeze n Clear Skin Clinic Wart/Skin Tag product (K211099) and the Compound W Nitro-Freeze wart treatment product (K172373) predicate devices. All devices are intended for OTC treatment of common warts and plantar warts. The combination of studies and performance data presented demonstrates the subject device is as safe and effective as compared to the predicate device(s) for the OTC treatment of common warts and plantar warts by consumers and for use with adults and children 4 years of age and older.
Page 5 of 5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.