Wartie Advanced Wart Remover or other proprietary name
K151309 · Youmedical Corp BV · GEH · Aug 14, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151309
Device Name
Wartie Advanced Wart Remover or other proprietary name
Applicant
Youmedical Corp BV
Product Code
GEH · General, Plastic Surgery
Decision Date
Aug 14, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Wartie® Advanced Wart Remover is intended for over-the-counter treatment of common warts and plantar warts to be used in adults and children 12 years of age and older.
Device Story
Device is an OTC cryosurgical system for home use. Input: pressurized canister containing dimethyl ether (DME) cryogen. Operation: user applies conductive gel to wart; activates canister to cool metal applicator tip; applies cold tip to wart to freeze skin tissue. Output: localized freezing of wart tissue, causing blister formation and subsequent removal of wart and virus. Includes disposable comfort pads for plantar wart protection. Healthcare provider role: none; patient self-use. Benefit: effective removal of common and plantar warts via controlled cryotherapy.
Clinical Evidence
No clinical data. Substantial equivalence supported by biocompatibility testing per ISO 10993 and bench testing comparing freezing performance to predicate devices.
Technological Characteristics
Cryosurgical unit using dimethyl ether (DME) as cryogen. Features a metal applicator tip for thermal energy removal, conductive gel for thermal coupling/moistening, and a mechanical locking ring for canister safety. Dimensions/form factor: handheld aerosol canister with applicator. Sterilization: not specified. Software: none.
Indications for Use
Indicated for the over-the-counter treatment of common warts and plantar warts in patients ages 12 years and older.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other, resembling a stylized bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 14, 2015
YouMedical Corp BV % Ms. Kathleen Johnson Medical Device Approvals Incorporated P.O. Box 1124 Bryn Mawr, Pennsylvania 19010
Re: K151309
Trade/Device Name: Wartie® Advanced Wart Remover Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: May 11, 2015 Received: May 18, 2015
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### Device Name
Wartie® Advaced Wart Remover
Indications for Use (Describe)
Wartie® Advanced Wart Remover is indicated for the over-the-counter treatment of common warts for patients ages 12 years and older.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
Image /page/2/Figure/13 description: The image shows a form section labeled "FOR FDA USE ONLY". Below this, the text "Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)" is present, indicating a space for a signature. The signature area is demarcated by a rectangular box with two diagonal lines forming an 'X' across it, suggesting that the area is reserved for official use and signature.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for YouMedical. The logo consists of a stylized graphic to the left of the text "YouMedical". The graphic is made up of intersecting blue and green shapes.
Sub. N .: TBD Version: 1.0 11-May-15 Date:
## Premarket Notification 510(k) summary (As Required by 21 CFR 807.92)
| Submitter<br>807.92(a)(1) | YouMedical Corp BV,<br>Rijnsburgstraat 9-11,<br>1059AT Amsterdam, The Netherlands |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Kathleen Johnson<br>Medical Device Approvals, Inc.<br>PO Box 1124<br>Bryn Mawr, PA 19010<br>kathleen@mdapprovals.com |
| Date Prepared | 06/May/15 |
| Trade Name<br>807.92(a)(2) | Wartie® Advanced Wart Remover |
| Common Name | OTC Wart Removal System |
| Classification Name<br>807.92(a)(2) | Cryosurgical unit, accessories |
| Classification Panel | General & Plastic Surgery |
| Product Code | GEH |
| Predicate Device<br>807.92(a)(3) | Wartie® Wart Remover K140314<br>Histofreezer® Wart Removal System K023487 |
| Predicate Comparison | Wartie® Wart Remover K140314: Intended use, technology, materials, label.<br><br>Histofreezer® Wart Removal System K023487: Intended use, technology, materials. |
| Device Description<br>807.92(a)(4) | The Wartie® Advanced Wart Remover is an over the counter cryosurgery product (for the treatment of warts), to be used at home.<br>The device consists of:<br>- A pressurized canister filled with 50mL of the compressed liquid gas dimethyl ether with a custom application unit used to administer the cold delivered by the cryogen to the wart.<br>- One 3 mL tube of conductive gel (also called gel).<br>- Six disposable comfort pads.<br>- Instructions for use in which you can read about the product how it works, how to use the product |
| | to treat warts, warnings and limitations. |
| Operating Principle | Wartie® Advanced Wart Remover technology involves freezing<br>a wart (common and/or plantar) using a very cold liquid gas<br>(cryogen) that is released into a patented-pending applicator.<br>The applicator becomes very cold and is able to freeze the skin.<br>A tube of conductive gel that maximizes the freezing potential<br>by directing and isolating the cold temperature onto the<br>application area, thereby moisturizing the skin.<br>The freezing of the skin causes damage to the cells housing and<br>protecting the wart. The wart then both develops a blister<br>underneath and then falls off, or the damaged skin cells are<br>discarded by the body, thereby also effectively removing the<br>wart and the virus.<br>Six disposable comfort pads used to protect the tender plantar<br>wart area after treatment has been completed in case of<br>particularly sensitive skin. |
| Intended Use<br>807.92 (a)(5) | Wartie® Advanced Wart Remover is intended for over-the-counter<br>treatment of common warts and plantar warts to be used in adults<br>and children 12 years of age and older. |
| Indications for Use<br>Technological<br>Characteristics as<br>compared to the<br>predicate devices<br>807.92 (a)(6) | The Wartie® Advanced Wart Remover employs a metal<br>applicator used as cold retraction vehicle. This allow for pen-pointed<br>accuracy in freezing the skin. Wartie® Advanced Wart<br>Remover makes use of a conductive gel to optimize relative<br>freezing performance, a secured locking ring in order to assure<br>mechanical safety of the pressurized canister, and thermal safety<br>precautions are associated with activation of the aerosol. Finally,<br>the product uses thermal energy removed from skin at the<br>anatomical site of a common wart and/or plantar wart.<br>DME applied to skin surface to freeze topical wart |
| Product Testing<br>807.92 (b)(1) | Biocompatibility per ISO 10993<br>Bench testing to show comparable freezing performance between<br>subject device and predicates |
| Clinical Testing<br>807.92 (b)(2) | N/A |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for YouMedical. The logo consists of a blue and green abstract symbol on the left, followed by the word "YouMedical" in a sans-serif font. The word "You" is in green, while "Medical" is in blue.
Date:
11-May-15
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for YouMedical. The logo consists of a blue and green abstract symbol on the left, followed by the text "YouMedical" in a combination of green and blue colors. The word "You" is in green, while "Medical" is in blue.
r
#### Premarket Notification 510(k) Summary Wartie® Advanced Wart Remover TBD
Sub. N .: Version:
Date: 11-May-15
| Substantial Equivalence Information | | | |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| | Wartie®<br>Advanced<br>Remover | Wartie®<br>Wart Remover | Histofreezer®<br>Wart<br>Removal System |
| 510(k) number | Not Assigned | K140314 | K023487 |
| Intended Use | OTC treatment of<br>common warts and<br>plantar warts. | OTC treatment of<br>common warts<br>and plantar warts. | OTC treatment of<br>common warts and<br>plantar warts. |
| Target population | Adults and Children<br>12+. | Adults and<br>Children 4+. | Adults and Children 4+. |
| Energy used and/or<br>delivered | Thermal energy<br>removed from skin<br>via a metal<br>interface. | Thermal energy<br>removed from<br>skin via a metal<br>interface. | Thermal energy<br>removed from skin via a<br>sponge containing liquid<br>gas. |
| Design | Device requiring<br>application of the<br>gel, activation and<br>application. Metal<br>tip provides pen-<br>pointed accuracy. | Device requiring<br>activation and<br>application. Metal<br>tip provides pen-<br>pointed accuracy. | Device requiring<br>assembly, activation,<br>and application. |
| Cryogen and<br>applicator materials | DME from aerosol<br>can applied to the<br>skin through a<br>metal core via a<br>Nickel tip providing<br>pen-point accuracy | DME from aerosol<br>can applied to the<br>skin through a<br>metal core via a<br>Nickel tip providing<br>pen-point accuracy | DME/Propane from<br>aerosol can applied to<br>the skin through a<br>sponge core mounted on<br>a plastic applicator. |
| Summary of<br>Technological<br>Differences | The only notable technological differences involve the type of<br>device applicator used to deliver the cryogen. The Histofreezer®<br>Wart Removal System device uses a sponge and the Wartie®<br>systems use a metal tip. The subject device uses a conductive gel<br>to moisten the skin whereas the predicates do not. Comfort pads<br>are included in Wartie® Advanced Wart Remover to protect tender<br>plantar warts after treatment. These technological differences do<br>not create any new risks of safety or effectiveness for the user. | | |
| Summary of<br>similarities<br>Conclusion<br>807.92 (b)(3) | All three devices have the same intended use and utilize the same<br>type of technology.<br>The subject device is substantially equivalent to the predicate<br>devices based on intended use, principle of operation, biological<br>evaluation and non-clinical bench testing. Any identified<br>differences do not constitute new risks of safety or effectiveness to<br>the user. | | |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.