K233347 · Cryosurgery, Inc. · GEH · Dec 4, 2023 · General, Plastic Surgery
Device Facts
Record ID
K233347
Device Name
Verruca-Freeze® H
Applicant
Cryosurgery, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Dec 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Verruca-Freeze® H cryosurgical system is used for treatment of benign skin lesions.
Device Story
Verruca-Freeze® H is a cryosurgical system for treating benign skin lesions. The device consists of a pressurized aerosol canister containing a mixture of dimethyl ether, propane, and isobutane, and non-sterile, single-use polyurethane foam swabs. The operator (physician) sprays the cryogen onto a foam swab, which is then applied directly to the target lesion for 15-40 seconds. The device operates via evaporative cooling, reaching temperatures of -50°C to induce extracellular and intracellular ice formation, resulting in tissue destruction. The procedure is simple, with a typical thaw time of 20-40 seconds. The device is intended for prescription use in a clinical setting.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing comparing output temperature and effectiveness to the predicate device, including visual confirmation of ice ball formation on a neoprene pad.
Technological Characteristics
Pressurized aerosol canister containing dimethyl ether (95%), propane (2%), and isobutane (3%). Applicators are polyurethane foam swabs. Operates via evaporative cooling to reach -50°C. Non-sterile, single-use components. 2-year shelf life.
Indications for Use
Indicated for the treatment of actinic keratosis, genital warts, lentigo, molluscum contagiosum, seborrheic keratosis, skin tags, verruca plantaris, verruca vulgaris, and verruca plana.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
K252903 — Verruca-Freeze H Plus · Cryosurgery, Inc. · Oct 10, 2025
K243454 — Verruca-Freeze H · Cryosurgery, Inc. · Dec 3, 2024
K161337 — CryoDose H or Similar · Nuance Medical, LLC · Jul 12, 2016
K240106 — Histofreezer V · Cryoconcepts LP · Jun 10, 2024
K982506 — VERRUCA-FREEZE CRYOSURGERY DELIVERY SYSTEM · Cryosurgery, Inc. · Jan 27, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 4, 2023
CryoSurgery, Inc. % Rachel Paul Lead Consultant QA&RA Emergo Europe Consulting Prinsessegracht 20 The Hague, 2514AP Netherlands
Re: K233347
Trade/Device Name: Verruca-Freeze® H Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: September 29, 2023 Received: September 29, 2023
Dear Rachel Paul:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark Trumbore -S | Digitally signed by Mark Trumbore -S<br>Date: 2023.12.04 08:24:35 -05'00' |
|------------------|---------------------------------------------------------------------------|
|------------------|---------------------------------------------------------------------------|
| Mark Trumbore, Ph.D. |
|--------------------------------------------------------|
| Assistant Director |
| DHT4A: Division of General Surgery Devices |
| OHT4: Office of Surgical and Infection Control Devices |
| Office of Product Evaluation and Quality |
| Center for Devices and Radiological Health |
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K233347
Device Name
Verruca-Freeze® H
Indications for Use (Describe)
The Verruca-Freeze® H cryosurgical system is indicated for use in the following skin lesions:
- · Actinic Keratosis
- · Genital Warts
- Lentigo
- Molluscum Contaqiosum
- Seborrheic Keratosis
- Skin Tags
- · Verruca Plantaris
- · Verruca Vulgaris • Verruca Plana
- Verruca Plana
Type of Use (Select one or both, as applicable)
ر Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:
# 1. Submitter Information
| Company: | Ronald A. McDow, MD |
|------------------------|-------------------------------------------------------------------|
| | Medical Director |
| | CryoSurgery, Inc. |
| | 5829 Old Harding Pike |
| | Nashville, TN 37205 USA |
| | Telephone: (615) 354-0414 |
| | Fax: (615) 354-0466 |
| | rmcdow@comcast.net |
| Contact: | Rachel Paul |
| | Lead Consultant QA&RA |
| | Emergo Europe Consulting |
| | Prinsessegracht 20 |
| | The Hague, 2514AP The Netherlands |
| | Telephone: + 31 (0) 70 345 85 70/Direct: + 31<br>(0) 70 850 82 49 |
| | Fax: |
| | LST.AUS.ProjectManagement@ul.com |
| Date Summary Prepared: | November 27, 2023 |
| 2. Name of the Device | |
| Trade Name: | Verruca-Freeze® H |
| Common Name: | Unit, Cryosurgical, Accessories |
| Classification Name: | General & Plastic Surgery |
| Review Panel: | General & Plastic Surgery (SU) |
| Regulation: | 878.4350 |
| Class: | Class II |
| Product Code: | GEH |
{4}------------------------------------------------
# 3. Equivalence Claimed to Predicate Device
The Verruca-Freeze® H is equivalent to the CryoDose H Or Similar (K161337), manufactured by Nuance Medical, LLC.
# 4. Device Description
The Verruca-Freeze® H cryosurgical system is used for treatment of benign skin lesions.
The device consists of:
- Canister filled with a liquid mixture of the compressed gases dimethyl ether, propane, and isobutane
- . Polyurethane foam swabs
- . Detailed instructions including illustrated descriptions
The main component of the device is the pressurized aerosol canister containing cryogen spray. The canister is used with foam swabs, which are non-sterile, single-use applicators that are disposed after each use. The foam swabs allow the cryogen to be localized to a target lesion.
The one- to two-minute procedure begins by spraying cryogen onto a foam swab, which is then placed directly onto the lesion and held for 15-40 seconds. Upon removal of the swab from the skin, the frozen tissue appears white and begins thawing. The typical thaw time is 20-40 seconds.
This treatment methodology is simple and has been an accepted practice used by physicians for decades. It utilizes a standard cryogen composition profile to freeze lesions and destroy the underlying tissue through evaporative cooling. On the cellular level, both extracellular and intracellular ice formation are occurring.
| Substantial Equivalence Table | | |
|-------------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Characteristic | Proposed Device: Verruca-<br>Freeze® H | Primary Predicate Device:<br>CryoDose H Or Similar<br>(K161337) |
| Intended Use | The Verruca-Freeze H<br>cryosurgical system is used<br>for treatment of benign skin<br>lesions. | The CryoDose H is used for<br>treatment of benign skin<br>lesions. |
## 5. Substantial Equivalence Comparison
{5}------------------------------------------------
| Indication for Use | The Verruca-Freeze H<br>cryosurgical system is<br>indicated for use in the<br>treatment of the following<br>skin lesions: Actinic<br>Keratosis, Genital Warts,<br>Lentigo, Molluscum<br>Contagiosum, Seborrheic<br>Keratosis, Skin Tags, Verruca<br>Plantaris, Verruca Vulgaris,<br>and Verruca Plana | The CryoDose H indications<br>for use as follows: CryoDose<br>H is indicated for use in the<br>treatment of the following:<br>actinic keratosis, genital<br>warts, lentigo, molluscum<br>contagiosum, seborrheic<br>keratosis, skin tags, verruca<br>plantaris, verruca vulgaris,<br>verruca plana |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GEH | GEH |
| Regulation Number | 878.4350 | 878.4350 |
| Classification Name | Unit, Cryosurgical,<br>Accessories | Unit, Cryosurgical,<br>Accessories |
| Mechanism of Action | Application of extreme cold<br>causing tissue destruction | Application of extreme cold<br>causing tissue destruction |
| Cryogen Gas Material | Dimethyl Ether (95%),<br>Propane (2%) and Isobutane<br>(3%) | Dimethyl Ether (95%),<br>Propane (2%) and Isobutane<br>(3%) |
| Freeze Time | 15-40 seconds | 15-40 seconds |
| Applicators | Foam swabs | Foam swabs |
| Sterile | No | No |
| OTC or Rx | Rx | Rx |
| Design | Aerosol-filled canister and<br>parts within the canister<br>(valve, actuator, cover cap) | Aerosol-filled canister and<br>parts within the canister<br>(valve, actuator, cover cap) |
| Freezing temperature | -50°C | -50°C |
| Shelf life | 2-year | 2-year |
## 6. Additional Information
#### 6.1 Substantial Equivalence Discussion
The technological characteristics of Verruca-Freeze® H, including materials, design, and energy source, are identical to the predicate device. The table as above compares Verruca-Freeze® H to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence.
Other elements of Verruca-Freeze® H not mentioned in the table that have the same technical characteristics as the predicate device include:
{6}------------------------------------------------
- Canister ●
- . Valve
- Actuator ●
- Cover Cap
Verruca-Freeze® H has the same technological characteristics as the predicate device. Both devices use the same design, energy source, materials, and other features. When the cryogen is dispensed, it saturates the foam swab is then applied directly to the desired area of treatment. Subsequently, the skin lesion is frozen and destroyed. Verruca-Freeze® H does not raise any new questions of safety or effectiveness as compared to the predicate device.
It is concluded that Verruca-Freeze® H is substantially equivalent to the predicate device.
#### 6.2 Non-Clinical Performance Data
Performance testing was conducted to determine the output temperature and effectiveness of Verruca-Freeze® H as compared to the predicate device. The output temperature was measured in the cryogen-saturated foam-tipped applicators followed by a visual confirmation of an "ice ball" on a neoprene pad. The tests determined that Verruca-Freeze® H cryogen and applicators perform equivalently to the cryogen and applicators of the predicate device. Further, the tests demonstrated that the device is capable of reaching the minimum required temperature to destroy the underlying tissue of a skin lesion.
#### 6.3 Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use are equivalent to those of the predicate device. These types of devices have been on the market for decades and their clinical safety and efficacy has been established. The nonclinical testing detailed in this submission supports the substantial equivalence of the device.
#### 7. Substantial Equivalence Conclusion
Based on the device characteristics and testing, the Verruca-Freeze® H is substantially equivalent to the predicate device and raises no new questions related to safety or effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.