K252903 · Cryosurgery, Inc. · GEH · Oct 10, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252903
Device Name
Verruca-Freeze H Plus
Applicant
Cryosurgery, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Oct 10, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Verruca-Freeze-H Plus cryosurgical system is intended for use by trained medical professionals to treat the following benign skin lesions: - Actinic Keratosis - Genital Warts - Lentigo - Molluscum Contagiosum - Seborrheic Keratosis - Skin Tags - Verruca Plantaris - Verruca Vulgaris - Verruca Plana
Device Story
Verruca-Freeze H Plus is a prescription cryosurgical system; consists of pressurized aerosol canister containing cryogen and non-sterile, single-use foam swabs. Used by trained medical professionals in clinical settings to treat benign skin lesions. Procedure involves saturating foam swab with cryogen; applying directly to lesion for 15–40 seconds; evaporative cooling induces extracellular and intracellular ice formation; destroys underlying tissue. Thaw time is 20–40 seconds. Device provides localized cryotherapy; benefits patient through simple, established tissue destruction method. Modification limited to cryogen formulation; no changes to device design, intended use, or operating principle.
Clinical Evidence
Bench testing only. Evaluated modified cryogen formulation performance against acceptance criteria including cryogen temperature, freeze charge hold time, and ice-ball formation. All criteria met.
Technological Characteristics
Pressurized aerosol canister with valve, actuator, and non-sterile foam swabs (2mm arrow, 5mm round). Operates via evaporative cooling (cryogen spray). Class II device, 21 CFR 878.4350, Product Code GEH.
Indications for Use
Indicated for trained medical professionals to treat benign skin lesions including actinic keratosis, genital warts, lentigo, molluscum contagiosum, seborrheic keratosis, skin tags, verruca plantaris, verruca vulgaris, and verruca plana.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 10, 2025
CryoSurgery, Inc.
Kaitlyn Clements
Regulatory Affairs Coordinator
5829 Old Harding Pike
Nashville, Tennessee 37205
Re: K252903
Trade/Device Name: Verruca-Freeze H Plus
Regulation Number: 21 CFR 878.4350
Regulation Name: Cryosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: GEH
Dated: September 9, 2025
Received: September 11, 2025
Dear Kaitlyn Clements:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252903 - Kaitlyn Clements
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252903 - Kaitlyn Clements
Page 3
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.10.10
14:35:11 -04'00'
Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252903 | ? |
| Please provide the device trade name(s). | | ? |
| Verruca-Freeze H Plus | | |
| Please provide your Indications for Use below. | | ? |
| The Verruca-Freeze H Plus cryosurgical system is intended for use by trained medical professionals to treat
the following benign skin lesions:
- Actinic Keratosis
- Genital Warts
- Lentigo
- Molluscum Contagiosum
- Seborrheic Keratosis
- Skin Tags
- Verruca Plantaris
- Verruca Vulgaris
- Verruca Plana | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
CRYO SURGERY, INC.
510(k) Summary
Verruca-Freeze H Plus
K252903
1. Submission Sponsor
Ronald McDow
CryoSurgery, Inc.
5829 Old Harding Pike
Nashville, TN 37205
(615) 354-0414
info@cryosurgeryinc.com
2. Primary Correspondent
Kaitlyn Clements
5829 Old Harding Pike
Nashville, TN 37205
3. Date Prepared
08/27/2025
4. Device Identification
Trade / Proprietary Name
Verruca-Freeze® H Plus
Common / Used Name
Prescription Cryosurgical Device
510(k) Number
K252903
Regulation
878.4350 - Cryosurgical unit and accessories
Product Code
GEH
Class
2
5. Primary Predicate Device
Device Name
Verruca-Freeze® H
510(k) Number
K243454
Regulation Number
878.4350 - Cryosurgical unit and accessories
Product Code
GEH
Class
2
6. Predicate Device
Device Name
Nuance Freeze Spray System
510(k) Number
K130995
Regulation Number
878.4350 - Cryosurgical unit and accessories
Product Code
GEH
Class
2
7. Reference Device
Device Name
Verruca-Freeze® H
510(k) Number
K233347
Version: 1
{5}
CRYO SURGERY, INC.
510(k) Summary
Verruca-Freeze H Plus
K252903
Regulation Number 878.4350 - Cryosurgical unit and accessories
Product Code GEH
Class 2
# 8. Description of the Subject Device
The main component of the Verruca-Freeze® H Plus cryosurgical system is the pressurized aerosol canister containing cryogen spray. The canister is used with foam swabs, which are non-sterile, single-use applicators that are disposed of after each use. The foam swabs allow the cryogen to be localized to a target lesion.
The procedure begins by saturating a foam swab with cryogen and then placing it directly onto the lesion for 15–40 seconds. Upon removal of the swab from the skin, the frozen tissue appears white and begins thawing. The typical thaw time is 20–40 seconds. This treatment methodology is simple and has been an accepted practice used by physicians for decades. It utilizes a cryogen formulation to freeze lesions and destroy the underlying tissue through evaporative cooling. On the cellular level, both extracellular and intracellular ice formation occur. This submission includes a modification to the cryogen formulation. No other changes were made to the device design, intended use, or operating principle.
# 9. Indications for Use
The Verruca-Freeze® H Plus cryosurgical system is indicated for use in the treatment of the following skin lesions:
- Actinic Keratosis
- Genital Warts
- Lentigo
- Molluscum Contagiosum
- Seborrheic Keratosis
- Skin Tags
- Verruca Plantaris
- Verruca Vulgaris
- Verruca Plana
# 10. Technological Comparison
A summary comparison of technological characteristics for the subject device vs predicate devices is provided in the following table:
| Characteristic | Subject Device Verruca-Freeze H Plus | Predicate Device Verruca-Freeze H | Predicate Device Nuance Freeze Spray |
| --- | --- | --- | --- |
| Device Name | Verruca-Freeze® H Plus | Verruca-Freeze® H | Nuance Freeze Spray System |
| 510(k) Number | K252903 | K243454 | K130995 |
| Type of Use | Rx | Rx | Rx |
Version: 1
Page 2 of 4
{6}
C BYOSURGERY, INC.
510(k) Summary
Verruca-Freeze H Plus
K252903
| Indications for Use | Treatment of benign skin lesions, including:
• Actinic Keratosis
• Genital Warts
• Lentigo
• Molluscum Contagiosum
• Seborrheic Keratosis
• Skin Tags
• Verruca Plantaris
• Verruca Vulgaris
• Verruca Plana | Treatment of benign skin lesions, including:
• Actinic Keratosis
• Genital Warts
• Lentigo
• Molluscum Contagiosum
• Seborrheic Keratosis
• Skin Tags
• Verruca Plantaris
• Verruca Vulgaris
• Verruca Plana | Treatment of skin lesions, including:
• Verruca (warts), including plantar warts
• Seborrheic keratoses
• Actinic keratosis
• Acrochordon (skin tags)
• Molluscum contagiosum
• Age spots
• Dermatofibroma
• Small keloids
• Granuloma annulare
• Porokeratosis plantaris
• Angiomas
• Keratoacanthoma
• Chondrodermatitis
• Epithelial nevus
• Leukoplakia
• Granuloma pyogenicum
• Pyogenic granuloma |
| --- | --- | --- | --- |
| Product Code | GEH | GEH | GEH |
| Regulation Number | 21 CFR 878.4350 | 21 CFR 878.4350 | 21 CFR 878.4350 |
| Classification Name | Unit, Cryosurgical, Accessories | Unit, Cryosurgical, Accessories | Unit, Cryosurgical, Accessories |
| Device Design | Pressurized aerosol canister with valve, actuator, and foam swabs | Pressurized aerosol canister with valve, actuator, and foam swabs | Pressurized aerosol canister with valve, actuator, foam swabs, and isolation funnels |
| Mechanism of Action | Application of extreme cold causing tissue destruction | Application of extreme cold causing tissue destruction | Application of extreme cold causing tissue destruction |
| Swab Types | Arrow (2 mm), Round (5 mm) | Arrow (2 mm), Round (5 mm) | Arrow (2 mm), Round (5 mm) |
| Sterility | Non-sterile | Non-sterile | Non-sterile |
## 11. Non-Clinical Performance Data
Non-clinical bench testing was conducted to evaluate the performance of the modified formulation. Testing methods were consistent with those used in the reference device's submission and included:
- Temperature of Cryogen
- Freeze Charge Hold Time
- Ice-Ball Formation
All acceptance criteria were met. The modified cryogen formulation performed as intended and is substantially equivalent to the predicate.
## 12. Statement of Substantial Equivalence
Version: 1
Page 3 of 4
{7}
CRYO SURGERY, INC.
510(k) Summary
Verruca-Freeze H Plus
K252903
The Verruca-Freeze® H Plus cryosurgical system is substantially equivalent to the predicate device (K243454). The modification does not change the device's intended use, indications, or technological characteristics other than the cryogen formulation. The change to the formulation does not introduce new risks or affect safety or effectiveness. Verification and validation activities demonstrate that the modified device meets all acceptance criteria and continues to perform as intended.
Version: 1
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.