The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia and/or low grade dysplasia) and malignant lesions.
Device Story
Cryosurgical system applying medical-grade liquid nitrogen (-196°C) to target tissue via small, low-pressure, open-tipped catheters. System comprises mobile console with touch panel computer, cryogen/suction/electronics modules, and disposable spray kits (active or passive venting). Physician operates via dual foot pedal and touch panel under direct visualization (endoscope/bronchoscope). Console manages cryogen levels, pressure, cooling, defrost, and suction. Safety features include RFID catheter verification, pressure transducers, redundant switches, mechanical relief valves, and emergency stop. Computerized monitoring aborts freezing upon system failure. Active defrost uses warm nitrogen gas. Benefits include controlled tissue destruction via freeze/thaw/freeze cycles for lesion ablation.
Clinical Evidence
No clinical data; bench testing only (not applicable as submission is for labeling changes only).
Indicated for patients requiring ablation of benign or malignant lesions in dermatology, gynecology, and general surgery, including Barrett's Esophagus with high or low grade dysplasia.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
K113021 — TRUFREEZE SYSTEM, TRUFREEZE SPRAY KIT · CSA Medical, Inc. · Feb 7, 2012
K160273 — truFreeze System · Csa Medial, Inc. · Mar 25, 2016
K143625 — truFreeze System · CSA Medical, Inc. · Jan 15, 2015
K150920 — truFreeze System · CSA Medical, Inc. · Jun 17, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 18, 2022
STERIS Corporation Mr. Carroll Martin Regulatory Affairs Director 5960 Heisley Road Mentor, Ohio 44060
Re: K222272
Trade/Device Name: truFreeze System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: July 29, 2022 Received: July 29, 2022
Dear Mr. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222272
Device Name truFreeze System
Indications for Use (Describe)
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia and/or low grade dysplasia) and malignant lesions.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a bold, sans-serif font, with the registered trademark symbol (®) to the right of the word. Below the word "STERIS" is a graphic of several horizontal, wavy lines in a light blue color.
# 510(k) Summary for the truFreeze System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060
| Contact: | Mr. Carroll Martin |
|----------|----------------------------------|
| | Regulatory Affairs Director |
| | Tel: 440-358-6259 |
| | Email: Carroll_Martin@steris.com |
Submission Date: July 29, 2022
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | truFreeze System |
|--------------------|-----------------------------------|
| Device Class: | Class II |
| Regulation Name: | Cryosurgical Unit and Accessories |
| Common/usual Name: | Cryosurgical Unit |
| Regulation Number: | 21 CFR 878.4350 |
| Product Code: | GEH |
#### 2. Predicate Device
truFreeze System, K171626
#### Device Description 3.
The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the ablation area via a small, low pressure, open tipped catheter. The truFreeze System consists of a console and a disposable spray kit.
The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction, and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the-shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction.
There are 2 types of spray kits available. One kit is available for active venting procedures and one is available for passive venting procedures. Both active and passive venting kits are provided in a carton of five (5) individually packaged sterile, single-use catheters with introducers (an introducer is a spring that is introduced into the accessory channel port of the endoscope to provide additional strain relief to the catheter during insertion and retraction from the endoscope) in individual pouches. These catheters are used to introduce the liquid nitrogen to the desired site.
Additionally, the active venting kit is provided with a carton of five (5) individually packaged sterile, single-use cryo decompression tubes (CDTs) with associated tubing in individual pouches. The CDTs are used for the removal of the gaseous nitrogen from the patient. Each carton within a spray kit contains the instructions for use.
#### Indications for Use 4.
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia and/or low grade dysplasia) and malignant lesions.
{5}------------------------------------------------
K222272
#### 5. Technological Characteristics Comparison Table
A comparison of technical characteristics between the proposed and predicate devices is summarized in Table 1.
| Features | truFreeze System<br>Predicate Device K171626 | Modified Device | Comparison |
|---------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use | The truFreeze System is<br>indicated for use as a<br>cryosurgical tool in the fields of<br>dermatology, gynecology, and<br>general surgery, to ablate benign<br>(e.g. Barrett's Esophagus with<br>high grade dysplasia and/or low<br>grade dysplasia) and malignant<br>lesions. | The truFreeze System is<br>indicated for use as a<br>cryosurgical tool in the fields of<br>dermatology, gynecology, and<br>general surgery, to ablate benign<br>(e.g. Barrett's Esophagus with<br>high grade dysplasia and/or low<br>grade dysplasia) and malignant<br>lesions. | Identical |
| Construction | The truFreeze System consists of<br>a console that is used to control<br>the application of the cryogen,<br>spray catheters, active and passive<br>venting sets and<br>cryodecompression tubing sets<br>and connectors | The truFreeze System consists of<br>a console that is used to control<br>the application of the cryogen,<br>spray catheters, active and passive<br>venting sets and<br>cryodecompression tubing sets<br>and connectors | Identical |
| Sterile/Non-<br>sterile | Non-sterile (console)<br>Sterile (spray catheters and<br>cryodecompression tube) | Non-sterile (console)<br>Sterile (spray catheters and<br>cryodecompression tube) | Identical |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Identical |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Identical |
| Usage | Single Use:<br>Spray catheters and<br>cryodecompression tubing | Single Use:<br>Spray catheters and<br>cryodecompression tubing | Identical |
| | Reusable:<br>Console | Reusable:<br>Console | |
| Principle of<br>Operation to<br>Achieve Cooling | Via pressurized liquid nitrogen<br>(LN2) | Via pressurized liquid nitrogen<br>(LN2) | Identical |
| Method to<br>Destroy Tissue | Uses freeze/thaw/freeze cycle | Uses freeze/thaw/freeze cycle | Identical |
| Ability to Reach<br>Equivalent Dose | Achieve a 4 cm ice ball within 12<br>min (total elapsed time) | Achieve a 4 cm ice ball within 12<br>min (total elapsed time) | Identical |
| Point Source for<br>Destruction of<br>Tissue | Catheter head | Catheter head | Identical |
| Liquid Nitrogen<br>Capability to<br>Freeze Tissue | -196°C | -196°C | Identical |
| Features | truFreeze System<br>Predicate Device K171626 | Modified Device | Comparison |
| Output<br>Temperature at<br>Catheter Tip | -196°C | -196°C | Identical |
| Equivalent Ice<br>Formation | Able to produce an average 23.4<br>mm (radius measurement) ice ball<br>within 5 minutes | Able to produce an average 23.4 mm<br>(radius measurement) ice ball within<br>5 minutes | Identical |
| Equivalent<br>Temperature<br>Distribution | @ freeze depth of 21.87mm at 0°C<br>-20°C isotherm at 15.5mm depth<br>-40°C isotherm at 11.8mm depth | @ freeze depth of 21.87mm at 0°C<br>-20°C isotherm at 15.5mm depth<br>-40°C isotherm at 11.8mm depth | Identical |
| Delivery of<br>Cryogen | A spray delivered through a 2.2mm<br>(7 F) sterile conduit; straight<br>catheter<br>Controlled by user | A spray delivered through a 2.2mm<br>(7 F) sterile conduit; straight catheter<br>Controlled by user | Identical |
| Consistent<br>delivery of<br>cryogen<br>controlled for<br>target site | Computer driven | Computer driven | Identical |
| Guidance<br>required for<br>procedure | Direct visualization via an<br>endoscope or bronchoscope | Direct visualization via an endoscope<br>or bronchoscope | Identical |
| Safety Features<br>/Mitigations | Users can stop dose at any time via<br>foot pedal activation or emergency<br>stop button | Users can stop dose at any time via<br>foot pedal activation or emergency<br>stop button | Identical |
| Features<br>/Mitigations<br>(catheter/probe) | Confirms use of a compatible<br>catheter using RFID | Confirms use of a compatible<br>catheter using RFID | Identical |
| Notifies physician<br>to stop spraying | Audible beeper to coincide with<br>visual display of timer. | Audible beeper to coincide with<br>visual display of timer. | Identical |
| Computerized test<br>of system prior to<br>use | Uses computer program to test that<br>system is properly operating before<br>exposure of cryotherapy | Uses computer program to test that<br>system is properly operating before<br>exposure of cryotherapy | Identical |
| Computerized<br>continuous<br>monitoring of<br>system during<br>procedure | Uses computer program to abort<br>freezing if a system failure is<br>detected | Uses computer program to abort<br>freezing if a system failure is<br>detected | Identical |
| Ensure patient is<br>not exposed to<br>high pressure<br>gases | Uses active suction pump and CDT,<br>or a natural orifice, as per<br>instructions for use.<br>Instructions for Use provide venting<br>guidance for proper gas egress | Uses active suction pump and CDT,<br>or a natural orifice, as per<br>instructions for use.<br>Instructions for Use provide venting<br>guidance for proper gas egress | Identical |
| Protect healthy<br>tissue from<br>excessive<br>temperatures | Uses endoscope or bronchoscope to<br>help insulate when using internally.<br>Contains insulation outside of scope<br>to protect user. | Uses endoscope or bronchoscope to<br>help insulate when using internally.<br>Contains insulation outside of scope<br>to protect user. | Identical |
| Pressure Controls | Valves and pressure transducer to<br>control pressure of liquid nitrogen<br>(LN2); Redundant pressure switch;<br>Mechanical Relief | Valves and pressure transducer to<br>control pressure of liquid nitrogen<br>(LN2); Redundant pressure switch;<br>Mechanical Relief | Identical |
| Features | truFreeze System<br>Predicate Device K171626 | Modified Device | Comparison |
| Thermal/Defrost | Active defrost capability to thaw<br>catheter using warm Nitrogen gas | Active defrost capability to thaw<br>catheter using warm Nitrogen gas | Identical |
| Safe storage of<br>cryogenic agent | Single Dewar tank | Single Dewar tank | Identical |
| Biocompatibility | Patient contact materials comply<br>with ISO -10993 | Patient contact materials comply<br>with ISO -10993 | Identical |
# Table 1. Technological Characteristics Comparison Table
{6}------------------------------------------------
{7}------------------------------------------------
#### 6. Summary of Changes
The following is a summary of the changes in this submission:
- To take the previous instructions for use that contained instructions for both passive and active venting and separate it into two instructions for use: one for active venting and one for passive venting.
- Update the previous owner's manual
These changes were initiated for the following reasons:
- Transfer of the truFreeze device from CSA to STERIS ●
- Clarification of some information in the original instructions for use
- Addition of information for user convenience ●
- . Standardization of information in accordance with STERIS instructions for use
- Formatting
None of the modified information is related to issues of device performance.
#### 7. Summary of Non-Clinical Performance Testing
No performance testing was conducted as the subject of this submission is changes to the labeling only.
#### Conclusion 8.
Based on the intended use and technological characteristics, the subject device is as safe and effective as the legally marketed predicate device (K171626).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.