K171626 · CSA Medical, Inc. · GEH · Aug 30, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171626
Device Name
truFreeze System
Applicant
CSA Medical, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Aug 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The truFreeze System is intended for cryogenic destruction of tissue using Liquid Nitrogen spray that has a boiling point of -196℃ requiring either active or passive venting during surgical procedures. The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (eg. Barrett's Esophagus with high grade and/or low grade dysplasia) and malignant lesions.
Device Story
The truFreeze System is a cryosurgical tool for tissue ablation using medical-grade liquid nitrogen. The system comprises a mobile console (touch panel computer, cryogen/suction/electronics modules) and a disposable spray kit (catheters with introducers). The physician operates the console via a dual foot pedal and touch panel. An off-the-shelf controller manages cryogen level sensing, filling, pressure, cooling, defrost, suction, and timing. Liquid nitrogen is delivered via a low-pressure, open-tipped catheter to the target tissue. The device is used in clinical settings to ablate lesions; the healthcare provider monitors the procedure via the console interface. By applying extreme cold, the device destroys unwanted tissue, providing a therapeutic benefit for conditions like Barrett's Esophagus.
Clinical Evidence
Clinical evidence includes a prospective post-market registry study (NCT01802203) and literature review. In the registry, 22 patients with Barrett's Esophagus with low-grade dysplasia (BE-LGD) were treated; 21 (95.5%) achieved complete eradication of dysplasia (CE-D). Safety data from 111 total patients showed a 2.7% stricture rate and 0.9% pancreatitis rate. Literature review of 40 BE-LGD patients showed a 93% CE-D success rate. Data supports the safety and effectiveness of the device for the expanded indication.
Technological Characteristics
Cryosurgical unit using liquid nitrogen (-196°C). System includes a mobile console with touch panel computer, cryogen/suction/electronics modules, and disposable sterile single-use catheters. Features include active/passive venting, cryogen level sensing, pressure relief valves, and isolated low-voltage power. Connectivity includes off-the-shelf controller for data management. Sterilization is provided for disposable components.
Indications for Use
Indicated for patients requiring ablation of benign or malignant lesions in dermatology, gynecology, and general surgery, specifically including Barrett's Esophagus with high-grade or low-grade dysplasia.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Coldplay CryoBalloon™ FocalAblation System (K161202)
Related Devices
K163244 — truFreeze System · CSA Medical, Inc. · May 24, 2017
K133258 — TRUFREEZE SYSTEM · CSA Medical, Inc. · Jun 19, 2014
K152668 — truFreeze System · CSA Medical, Inc. · Oct 14, 2015
K150920 — truFreeze System · CSA Medical, Inc. · Jun 17, 2015
K160273 — truFreeze System · Csa Medial, Inc. · Mar 25, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three faces in profile, one behind the other. The seal is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 30, 2017
CSA Medical, Inc. Ms. Sherrie Coval-Goldsmith, M.S. VP of Regulatory & Quality 91 Hartwell Avenue Lexington, Massachusetts 02421
Re: K171626
Trade/Device Name: truFreeze® System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: June 2, 2017 Received: June 2, 2017
Dear Ms. Coval-Goldsmith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely. Jennifer R. Stevenson -53
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171626
Device Name truFreeze ® System
Indications for Use (Describe)
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia and/or low grade dysplasia) and malignant lesions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510K SUMMARY
| Applicant | CSA Medical, Inc. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number | 3010140265 |
| Contact Person | SherrieCoval-Goldsmith, M.S. Vice<br>President RA/QA<br>CSA Medical<br>91 Hartwell Ave<br>Lexington, Ma 02421<br>Phone:781-538-7447<br>Fax:781-538-4730<br>sgoldsmith@csamedical.com |
| Summary Date | June 2, 2017 |
| Proprietary Name | truFreeze® System |
| Classification | Class II |
| Classification Name | Cryosurgical Unit, Cryogenic Surgical Device |
| Regulation Number | 21 CFR878.4350 |
| Classification Product Code | GEH |
| Predicate Device | K163244(truFreeze® System) |
| Reference Devices | K152329 (Coldplay CryoBalloon™ FocalAblation<br>System), K083737 (Barrx System), and<br>K 161202 ((Coldplay CryoBalloon™ FocalAblation<br>System) |
#### DEVICE DESCRIPTION
The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the ablation area via a small, low pressure, open tipped catheter. The truFreeze System consists of a console and a disposable spray kit.
#### CONSOLE:
The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction, and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the-shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction. The console is the same as
{4}------------------------------------------------
previously cleared for the predicate truFreeze System (K163244). Therefore, there are no new issues of safety or effectiveness raised.
#### DISPOSABLE SPRAY KIT:
There are 2 types of spray kits available for active venting procedures and one is available for passive venting procedures. Both active and passive venting kits are provided in a carton of five (5) individually packaged sterile, single-use catheters with introducers in individual pouches.
Additionally, the active venting kit is provided with includes a carton of five (5) individually packaged sterile, single-use CDTs with associated tubing in individual pouches. Each carton within a spray kit contains the instructions for use. The disposable kit is the same as previously cleared for the predicate truFreeze System (163244). Therefore, there are no new issues of safety or effectiveness raised.
### LABELING (INTENDED USE/INDICATIONS FOR USE AND INSTRUCTIONS FOR USE DOCUMENT)
The Intended Use/Indications for Use statementis identical in its Intended Use as the predicate device with the exception of adding a specific indication for use as an example of benign tissue that can be ablated with truFreeze. Clinicaldata have been generated to substantiate a change in label claim. The proposed modification adds Barrett's Esophagus with low grade dysplasia into the Indications for Use statement as an example of benign tissue to be ablated.
#### PROPOSED INTENDED USE/INDICATIONS FOR USE
The truFreeze System is intended for cryogenic destruction of tissue using Liquid Nitrogen spray that has a boiling point of -196℃ requiring either active or passive venting during surgical procedures.
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (eg. Barrett's Esophagus with high grade and/or low grade dysplasia) and malignant lesions.
### TECHNICAL AND OPERATIONAL CHARACTERISTICS
The truFreeze System is identical in design, operational and technological characteristics as the predicate device and supports that no new safety concerns are being raised by changes in the intended use/indications for use statement and thus raises no new issues of safety or effectiveness.
#### SUMMARY OF TESTING
The truFreeze System was previously subjected to a comprehensive test program that included electrical safety and electromagnetic compatibility testing, software testing, biocompatibility and sterilization testing. Data was previously submitted to support that the use of liquid nitrogen is cold enough to ablate benign or malignant lesions (K133258) and reference is being made to that data. Clinical data are summarized below.
Previous, clinical data was generated to support the addition of BE-HGD (K163244) to the indications for use. In a post-market registry study, 46 patients with Barrett's Esophagus high grade dysplasia received ablation with the truFreeze device. Patients received an average of 1.89 ablations sessions consisting of an average of 2.23 cycles of 21.17seconds to achieve CE-D in 87% of the patients. Details are presented in the tables 1-2 below.
{5}------------------------------------------------
#### Table 1 BE-HGD Effectiveness Data
| | Response CE-D |
|------------------------------------|---------------|
| Efficacy Population | 46 |
| N Responders | 40 (87.0%) |
| Response rate by BE segment length | |
| Unknown | 2 (4.3%) |
| ≤ 3 cm | 27 (58.7%) |
| 3-6 cm | 7 (15.2%) |
#### Table 2 BE-HGD Safety Data
| Safety | truFreeze Per Patient rate |
|---------------------|----------------------------|
| Stricture | 2.7% |
| Abdominal Pain | 0% |
| Pancreatitis | 0.9% |
| Chest pain | 0% |
| GI Hemorrhage | 0% |
| Mucosal Lacerations | 0% |
#### RATIONALE FOR SUBSTANTIAL EQUIVALENCE
The Intended Use/Indications for Use statement and technological characteristics of the truFreeze System and the predicate device were compared. The Intended Use/Indications for Use statement of the two devices had equivalent general claims except that BE-HGD (as an example of a specific benign lesion that can be ablated) was added to the indications for use statement.
{6}------------------------------------------------
Clinical data are presented to support the safety and effectiveness of adding Barrett's Esophagus with low grade dysplasia as an example of specific benign tissue that can be ablated.
A clinical summary of data collected from a post market registry study (ClinicalTrials.gov Identifier: NCT01802203) supports this submission. This post-marketing registry prospectively collected data on patients treated with the FDA-cleared cryotherapy device (truFreeze® System). The instructions for use document guided the dosimetry used for ablation procedures.
Similar to any clinical trial, physicians offered "all comers" with BE to participate in the registry if they met all inclusion criteria and none of the exclusion criteria. Patients were sequentially and consecutively enrolled into the registry at participating trial.
The registry had pre-defined inclusion / exclusion criteria, which are consistent with the instructions for use of the cleared device. CSA Medical applied 3 additional exclusion criteria that were sequentially applied to the population of BE patients. These exclusion criteria included items 3-5 in Table 3 below.
This yielded patients for the effectiveness population.
Two medical societies cite endoscopic radiofrequency ablation (RFA) as a recognized ablation method for BE LGD. 2/2 A rate of 90.5% CE-D is cited in the literature. 4
The clinical data for the patients with Barrett's Esophagus low grade dysplasia are provided in Tables 3- 9 below.
| Inclusion Criteria | 1. Able to read, comprehend, and complete the informed consent form.<br>2. Aged 18 or older.<br>3. Receiving spray cryotherapy procedure with the truFreeze device for the first<br>time for removal of unwanted tissues including malignant or pre- malignant<br>conditions of the aerodigestive system. |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Exclusion Criteria | 1. Contraindication to spray cryotherapy.<br>2. Prior treatment with spray cryotherapy. Previous or concurrent<br>treatment using other mucosal therapies such as endoscopic mucosal<br>resection or radiofrequency ablation is acceptable.<br>3. Patients not having Barrett's Esophagus with low grade dysplasia.<br>4. Patients without a pre-therapy and post-therapy histology.<br>5. Patients who received other therapies (such as endoscopic mucosal<br>resection) during the ablation therapyperiod. |
| Safety | 111 Patients |
| Population/Effectiveness | 22 patients |
| Device Used | truFreeze System |
### Table 3 Summary of Populations analyzed
{7}------------------------------------------------
### Table 4 Demographics
| | BE Patients<br>(Safety Population)<br>(N=111) | BE LGD Patients<br>(Effectiveness Population)<br>(N=22) |
|----------------------------------------|-----------------------------------------------|---------------------------------------------------------|
| Age (years) | | |
| n | 111 | 22 |
| Mean (SD) | 67.8 (10.8) | 69.2 (7.84) |
| Range | 24 - 89 | 51 - 83 |
| Gender, no. (%) | | |
| Female | 29 (26.1%) | 5 (22.7%) |
| Male | 82 (73.9%) | 17 (77.3%) |
| Segment length (cm) | | |
| <= 3 | 65 (29%) | 9 (41%) |
| 3-6 | 17 (15%) | 3 (14%) |
| >=6 | 21 (19%) | 8 (36%) |
| Unknown | 8 (7%) | 2 (9%) |
| Endoscopic mucosal resection | | |
| Previous EMR, no. (%) | 30 (28%)<br>(N=109) | 2 (10%)<br>(N=21) |
| Mean time since EMR(months), mean (SD) | 6.9 (11.8) | 4.5 (5.1) |
### Table 5. Procedure Information to Achieve Best Response
| | 22 BE-LGD Patients in Effectiveness<br>Population |
|-------------------------------------------------------------|---------------------------------------------------|
| Patients receiving 1 procedure session to best<br>response | 10 (45%) |
| Patients receiving 2 procedure sessions to best<br>response | 5 (23%) |
| Patients receiving 3 procedure sessions to best<br>response | 5 (23%) |
| Patients receiving >3 procedure sessions to best response | 2 (9%) |
| Mean Procedure Sessions (SD) | 2 (1.1) |
| Mean Cycles per procedure (SD) | 2.19 (0.72) |
| Mean Time (sec) per cycle (SD) (seconds) | 21.76 (3.67) |
{8}------------------------------------------------
### Table 6 Procedure or Device Related Adverse Event Rate per Procedure Session
| | BE Patients<br>Safety Population Procedures<br>(258 procedures) | BE LGD Patients Effectiveness<br>Population Procedures<br>(52 procedures) |
|---------------|-----------------------------------------------------------------|---------------------------------------------------------------------------|
| Stricture | 3 (1.2%) | 0 (0%) |
| Pancreatitis* | 1 (0.4%) | 0 (0%) |
* This is the only serious adverse event reported. The physician stated the relationship of truFreeze treatment and development of pancreatitis was highly unlikely. However, a possible causal relation with the device could not be ruled out because of the temporal association.
### Table 7 Barrett's segment length for CE-D Results
| | Response<br>CE-D |
|------------------------------------|------------------|
| Effectiveness Population | 22 |
| N Responders | 21 (95.5%) |
| Response rate by BE segment length | |
| Unknown | 2 (10%) |
| <= 3cm | 9 (43%) |
| 3-6cm | 3 (14%) |
| >=6cm | 7 (33%) |
### Table 8 Safety Data
| | truFreeze Per Patient Rate Safety Population (N=111) |
|---------------------|------------------------------------------------------|
| Stricture | 3 (2.7%) |
| Abdominal Pain | 0 (0%) |
| Pancreatitis | 1 (0.9%) |
| Chest Pain | 0 (0%) |
| GI Hemorrhage | 0 (0%) |
| Mucosal Lacerations | 0 (0%) |
### Table 9 Effectiveness Data
| | CE-D Patients (N=22) |
|-----------|----------------------|
| truFreeze | 21 (95.5%) |
Additionally, 3 peer reviewed articles support the effectiveness of the truFreeze System to ablate BE-LGD and are presented in table 10.
{9}------------------------------------------------
| Reference<br>No. | No. of BE LGD<br>patients treated | No Achieving CE-<br>D | No Achieving<br>CEIM | Comments |
|------------------|-----------------------------------|-----------------------|----------------------|-----------------------------------------------------------------------------|
| 3 | 23 | 21 | | |
| 4 | 4 | 3 | | |
| 5 | 13 | Not reported | 13 | Note: achieving CEIM is a more<br>difficult endpoint to achieve vs CE-<br>D |
| Total | 40 | 24 | 13 | 93 % of patients achieved CE-D |
#### Table 10 Summary of articles/presentations
#### DISCUSSION
The data show that the majority of BE-LGD patients were successfully treated with the truFreeze device. Among the 22 BE-LGD patients in the effectiveness population, 21 (95.5 %) achieved a response of CE-D. Additionally, among 40 BE-LGD patients presented from peer reviewed literature, 37 (93%) were successfully treated with the truFreeze device.
This success rates are comparable to the success rate among the LGD phenotype of BE patients for the Barrx System and the Coldplay CryoBalloon™ Focal Ablation System.
The safety rate of the trufreeze device is also substantially equivalent to these reference devices.
#### CONCLUSION
In summary, endoscopic ablation of Barrett's Esophagus with low grade dysplasia with the truFreeze device does not raise new questions of safety or effectiveness as compared to the reference devices or the predicate device and warrant adding Barrett's Esophagus with low grade dysplasia as an example of benign tissues that can be ablated, to the truFreeze label.
#### References
- 1. ASGE Standards of Practice Committee. The role of endoscopy in Barrett's esophagus and other premalignant conditions of the esophagus. Gastrointestinal Endoscopy 2012; 76 (6): 1087 (available at https://www.asge.org/docs/default-source/education/practice_guidelines/doc-therole-of-endoscopyin-barretts- esophagus-and-other-premalignant-conditions-of-the- esophagus.pdf?sfvrsn=8).
- 2. Nicholas J. Shaheen , MD, MPH, FACG , Gary W. Falk , MD, MS, FACG, Prasad G. Iyer , MD, MSc, FACG andLauren Gerson , MD, MSc, FACG. ACG Clinical Guideline: Diagnosis and Management of Barrett's Esophagus. Am J Gastroenterol advance online publication, 3 November 2015; doi: 10.1038/aig.2015.322(availableathttps://gi.org/guideline/diagnosis-and-management-of-barrettsesophagus/)
- 3. Ghorbani S, et al. "Safety and efficacy of endoscopic spray cryotherapy for Barrett's dysplasia: results of the National Cryospray Registry." Diseases of the Esophagus. 2015: 29: p241-247. doi: 10.11111/dote.12330. http://onlinelibrary.wiley.com/doi/10.1111/dote.12330/abstract;jsessionid=6FECC3D65017B2C6003584B066D 31F5E.f01t04
- 4. Barthel JS, et al. "Cryoablation of persistent Barrett's epithelium after definitive chemoradiation therapy for esophageal adenocarcinoma." Gastrointestinal Endoscopy. 2011: Volume 74, Issue 1, p.51-57. http://www.giejournal.org/article/S0016-5107%2811%2901343-5/abstract
{10}------------------------------------------------
- 5. Eluri S, et al. "Liquid Nitrogen Spray Cryotherapy Effectively Eradicates Barrett's Esophagus irrespective of Severity of Baseline Histology: Results of a US multicenter Registry. Gastrointestinal Endoscopy. 2017 : Volume 85 No 5S.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.